METHA SHORT STEM HIP SYSTEM; MODULAR HIP SYSTEM; EXCIA HIP SYSTEM
K081973 · Aesculap Implant Systems, Inc. · LZO · Aug 7, 2008 · Orthopedic
Device Facts
Record ID
K081973
Device Name
METHA SHORT STEM HIP SYSTEM; MODULAR HIP SYSTEM; EXCIA HIP SYSTEM
Applicant
Aesculap Implant Systems, Inc.
Product Code
LZO · Orthopedic
Decision Date
Aug 7, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Excia Hip System is intended to replace a hip joint. The device is intended for: Patients suffering from severe hip and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis Patients suffering from disability due previous fusion Patients with acute femoral neck fractures. The Excia Hip System is available with two femoral stems. One is manufactured from CoCMo and is intended for cemented fixation. The other femoral stem is for uncemented fixation and manufactured from Ti with Plasmapore with or without u CaP®. The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip joint. The device is intended for: skeletally mature individuals undergoing primary surgery for total hip replacement patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur, patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis patients suffering from disability due to previous fusion patients with acute femoral neck fractures. Indications for use of the CONSENSUS® HIP SYSTEM-PRIMARY HIP: A) Significantly impaired joints resulting from rheumatoid, osteo, and posttraumatic arthritis. B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. C) Proximal femoral fractures. D) Avascular necrosis of the femoral head. E) Non-union of proximal femoral neck fractures. F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteoporosis, pseudarthrosis conversion, and structural abnormalities. Acetabular components are indicated for cemented and cementless use.
Device Story
Consensus Acetabular Cup is a titanium alloy shell with highly cross-linked polyethylene (UHMWPE) insert; designed for use with Aesculap Excia and Metha hip systems. Device replaces hip joint in total hip arthroplasty; used in orthopedic surgery by surgeons. Shells available in various sizes with/without screws; inserts available in 28mm, 32mm, 36mm IDs. Device provides stable acetabular component for hip joint reconstruction; facilitates joint mobility; reduces pain; restores function in patients with severe hip disability. No software or algorithm involved.
Clinical Evidence
Bench testing only. Engineering evaluation of component dimensions was performed to establish substantial equivalence; no clinical data required.
Technological Characteristics
Titanium alloy shell; highly cross-linked polyethylene (UHMWPE) insert. Available in 28mm, 32mm, 36mm IDs. Standard or hooded versions. Cemented or cementless fixation. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring total hip replacement due to rheumatoid, osteoarthritis, posttraumatic arthritis, avascular necrosis, femoral neck fractures, congenital dysplasia, or failed previous hip procedures. Contraindications include active infection or insufficient bone stock.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
#### 510(k) SUMMARY (as required by 21 CFR 807.92) B.
### Consensus Acetabular Cups in Aesculap Hip Systems AUG - 7 2008 8 July 2008
| COMPANY: | Aesculap® Implant Systems, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Matthew M. Hull<br>800-258-1946 (phone)<br>610-791-6882 (fax)<br>matt.hull@Aesculap.com (e-mail) |
| TRADE NAME: | Excia and Metha Hip Systems (Consensus Acetabular Cup |
| COMMON NAME: | Total Hip System |
| CLASSIFICATION NAME: | Prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous uncemented prosthesis<br><br>Prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented |
- 888.3353/ 888.3358 REGULATION NUMBER:
- LZO/ LPH PRODUCT CODE:
# SUBSTANTIAL EQUIVALENCE
Aesculap®, Inc. believes that the Consensus Acetabular Cup is a line extension of Aesculap's Excia (K042344 - K062684) and Metha (K071916 & K080584) Hip Systems that were previously cleared for use with acetabular cups. The Consensus Acetabular Cup components have been cleared in K020153 – K070061 for use with the Consensus and Unisyn hip systems from Hayes Medical.
## DEVICE DESCRIPTION
The Consensus Acetabular Cup from Hayes Medical is a Titanium alloy shell with a highly cross linked polyethylene insert. The Consensus acetabular cups are available in 28mm, 32mm, and 36mm ID's, a variety of sizes, and in either standard or hooded versions. The shells are available with or without screws. The Aesculap Excia hip systems also comes in 28mm - 36 mm ID's, it is available in both cemented and uncemented variants, and a wide range of sizes. Aesculap's Metha hip system also comes in 28 - 36 mm ID's but is for uncemented use, and also comes in a variety of sizes. The PlasmaCup acetabular cup from Aesculap is cleared for use with both the Excia and Metha systems.
{1}------------------------------------------------
K081978 (pg 2/2)
## INDICATIONS FOR USE
The Excia Hip System is intended to replace a hip joint. The device is intended for:
· Patients suffering from severe hip and disability due to rheumatord arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur.
- · Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis
· Patients suffering from disability due previous fusion
- · Patients with acute femoral neck fractures
The Excia Hip System is available with two femoral stems. One is manufactured from CoCMo and is intended for cemented fixation. The other femoral stem is for uncemented fixation and manufactured from TT with Plasmapore with or without u CaP®.
### The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip joint.
The device is intended for:
- · skeletally mature individuals undergoing primary surgery for total hip replacement
· patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis,
- polvarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur.
- · patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis
- · patients suffering from disability due to previous fusion
• patients with acute femoral neck fractures
#### Indications for use of the CONSENSUS® HIP SYSTEM-PRIMARY HIP:
A) Significantly impaired joints resulting from rheumatoid, osteo, and positraumatic arthritis.
B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C) Proximal femoral fractures.
- D) Avascular necrosis of the femoral head.
- E) Non-union of proximal femoral neck fractures.
F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa valga, developmental conditions, metabolic and tumorous conditions, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
Acetabular components are indicated for cemented and cementless use.
## TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s))
There are no changes to the Hayes Consensus acetabular cup nor to the Aesculap Excia and Metha hip systems. The inner dimensions of the Aesculap PlasmaCup acetabular cup inserts are identical to those of the Hayes Consensus acetabular cups. The difference is that the Consensus inserts are highly crosslinked polyethylene (UHMWPE) and the Aesculap PlasmaCup inserts are regular polyethylene (UHMWPE).
### PERFORMANCE DATA
Based upon the engineering evaluation of the dimensions of these components no additional performance data was required.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three overlapping, curved lines representing its wings. The words "DEPARTMENT OF HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
800 - 1000 - 100
# AUG - 7 2008
Aesculap Implant Systems, Inc. % Mr. Matthew M. Hull, RAC Regulatory Affairs Manager 3773 Corporate Parkway Center Valley, Pennsylvania 18034
K081973 Re:
Trade/Device Name: Consensus Acetabular Cups for use with the Aesculap Excia and Metha Hip Systems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO, LPH Dated: July 9, 2008 Received: July 10, 2008
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assessment to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmetier fore, market the device, subject to the general controls provisions of the Act. The Tota may, ateres provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of actived that I B r mination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any i vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Matthew M. Hull, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Millison
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT A.
Image /page/4/Picture/3 description: The image shows the 510(k) number, which is K081973. The number is followed by the page number, which is 1/1. The text is written in black ink on a white background. The text is slightly blurry, but still legible.
Device Name: Consensus Acetabular Cups for use with the Aesculap Excia and Metha Hip Systems
## Indications for Use:
The Excia Hip System is intended to replace a hip joint.
The device is intended for:
Patients suffering from severe hip and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur.
- Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis
- Patients suffering from disability due previous fusion ◆
- Patients with acute femoral neck fractures .
The Excia Hip System is available with two femoral stems. One is manufactured from CoCMo and is intended for cemented fixation. The other temoral stor uncemented fixation and manufactured from Ti with Plasmapore with of without u CaP®.
## The Metha® Hip System (uncemented, press-fit fixation) is Intended to replace a hip joint.
The device is intended for:
- skeletally mature individuals undergoing primary surgery for total hip replacement
patients suffering from severe hip pain and disability due to meumatoid arthritis, traumatic arthritis,
- polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur, patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis
- patients suffering from disability due to previous fusion
- patients with acute femoral neck fractures
#### Indications for use of the CONSENSUS® HIP SYSTEM-PRIMARY HIP:
A) Significantly impaired joints resulting from rheumatoid, osteo, and posttraumatic arthritis.
- B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C) Proximal femoral fractures.
- D) Avascular necrosis of the femoral head.
- E) Non-union of proximal femoral neck fractures.
F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
Acetabular components are indicated for cemented and cementless use.
x and/or Over-the-Counter Use Prescription Use
(per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device E
Barbara Breen
Division Sign-Off
and Neurological Devices
**510(k) Number** K081977³ 002
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.