VACUETTE BLOOD COLLECTION TUBE WITH CLOT ACTIVATOR AND GEL SEPARATOR

K081929 · Greiner Bio-One North America, Inc. · JKA · Dec 16, 2008 · Clinical Chemistry

Device Facts

Record IDK081929
Device NameVACUETTE BLOOD COLLECTION TUBE WITH CLOT ACTIVATOR AND GEL SEPARATOR
ApplicantGreiner Bio-One North America, Inc.
Product CodeJKA · Clinical Chemistry
Decision DateDec 16, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

VACUETTE® Blood Collection Tubes with Clot Activator, Gel Separator, and Standard Stopper (16 x 100 mm, 8 mL) are used to collect, transport and process blood for testing serum in the clinical laboratory. In addition, the tubes are suitable for therapeutic drug monitoring (TDM).

Device Story

VACUETTE® blood collection tube; 16 x 100 mm, 8 mL capacity; contains clot activator and gel separator; standard stopper. Used by clinical laboratory personnel for venous blood collection. Tube facilitates blood clotting, separation of serum from cells via gel barrier during centrifugation, and subsequent transport/processing for serum-based diagnostic testing and therapeutic drug monitoring. Benefits include standardized specimen handling and improved serum quality for laboratory analysis.

Clinical Evidence

Bench testing only. Aerosol comparison studies performed between predicate and candidate devices to validate cap modification performance.

Technological Characteristics

Blood collection tube; modified cap design; no change in raw materials; fundamental scientific technology remains consistent with predicate devices.

Indications for Use

Indicated for use in clinical laboratories to collect, transport, and process blood specimens for serum testing and therapeutic drug monitoring (TDM).

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER: k081929 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) Greiner VACUETTE blood collection tube, k960858, k983952 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: Modifying the design of the cap of the blood collection tube without any raw materials change. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, analytes and performance characteristics. Studies include aerosol comparisons between predicate and candidate devices. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. FEMA b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...