K081845 · Versamed Medical Systems, Inc. · CBK · Mar 16, 2009 · Anesthesiology
Device Facts
Record ID
K081845
Device Name
IVENT101
Applicant
Versamed Medical Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Mar 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The intended use of the iVent101 ventilator is to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The intended patient population includes infants from 5 kg through adult patients. The iVent101ventilator is suitable for use in institution, home and portable settings
Device Story
iVent101 is a compact, turbine-powered, microprocessor-controlled ventilator providing continuous/intermittent ventilatory support. Inputs include internal flow and pressure sensor data. Device transforms inputs via software algorithms to control air delivery; supports single or dual limb circuits. Outputs include real-time clinical data (Tidal Volume, Rate, PIP, FiO2, Peak Flow, I:E, Minute Volume) displayed on an LCD touch-screen. Used in institutions, homes, and portable settings by trained clinicians. Optional software displays waveforms, loops, and trends. Device operates on AC/DC or internal battery. Output allows clinicians to monitor patient status and adjust ventilation parameters, facilitating respiratory support for patients from infants (5kg) to adults.
Clinical Evidence
No clinical data. Safety and efficacy established via literature and bench testing per ASTM F 1100-90, ASTM F 1246-91, and relevant IEC/ISO standards.
Technological Characteristics
Turbine-powered, microprocessor-controlled ventilator. Internal compressor for air generation. LCD touch-screen interface. Supports single/dual limb circuits. Powered by AC/DC or internal battery. Complies with ASTM F 1100-90, ASTM F 1246-91, IEC 60601-1, IEC 60601-2-12, IEC 60601-1-2, IEC 60601-1-8, and ISO 10651-2/6.
Indications for Use
Indicated for infants (>=5 kg) through adults requiring invasive or non-invasive mechanical ventilation. Used in institutional, home, and portable settings by trained personnel under physician direction. Modalities include Assist/Control, SIMV (Volume, Pressure, or PRVC), CPAP with Pressure Support, and Adaptive Bi-Level.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K081845
page 1 of 4
## VersaMed Medical Systems Inc. 2 Blue Hill Plaza Bldg. 2 Pearl River, NY 10965 USA
# MAR 1 6 2019
#### Non-Confidential Summary of Safety and Effectiveness
The Summary of Safety and Effectiveness
# Submitter's Name:
VersaMed Medical System Inc.
### Contact Person:
Mr. Jerry Korten Tel: 845 770 8240 Fax: 845 770 8250
### Owner / Operator:
General Electric Co. 3135 Easton Turnpike Fairfield, CT 06828
### Owner / Operator Number:
9912004
### Trade Name:
The iVent101
### Classification:
Name: Product code: Classification panel: Class: 2 Panel:
Continuous ventilator 73 CBK, 73 NOU 868-5895 Anesthesiology
{1}------------------------------------------------
### Device Description:
The iVent101 is a compact, portable, microprocessor-controlled ventilator intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation .A turbine-powered air source and a rechargeable internal battery provide freedom from wall air and power outlets . Internal flow and pressure are read through flow/pressure sensors. Clinical data [Tidal Volume, Rate, PIP, FiO2, Peak Flow, Inspiratory Time, I:E Minute Volume] are presented on machine screen.
All the operator actions are performed on the LCD touch-screen on the front panel, allows, rapid control and continuous real-time monitoring of patient ventilation. · Alarm settings are fully adjustable. Optional Waveform and Diagnostic Software package displays pressure and flow waveform data, loops, trends, and logged totals in a full array of time slices and presentation modes.
The iVent101 can use external AC or DC power supply and contains an integrated battery.
The iVention is equipped with two configurations, which differs in the patient circuit type that connects to the machine. One configuration is to be used with a standard one limb patient circuit, and the second configuration is to be used with a standard dual limb patient circuit. The two configurations are incorporating the same infrastructure and can be easily replaced by the user.
The iVent101 is capable of providing the following types of ventilatory support:
- Assist/Control and SIMV with either Volume, Pressure, or Pressure Regulated . Volume Control (PRVC)
- � CPAP with Pressure Support
- Adaptive Bi-Level for either NIV or invasive ventilation. .
#### Intended use:
·
The intended use of the iVent101 ventilator is to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The intended patient population includes infants from 5 kg through adult patients.
The iVent101ventilator is suitable for use in institution, home and portable settings
{2}------------------------------------------------
#### Predicate Devices:
The iVent101 is substantially equivalent to:
- · LTV1200 cleared under K060647 ·
- · iVent201 cleared under K061627
- · PLV Continuum II cleared under K034032
- · Legendair XL2 cleared under K070899
### Substantial Equivalence and Technological Characteristics:
The iVent101 is viewed as substantially equivalent to the following predicate devices;
- · LTV1200 cleared under K060647
- · iVent201 cleared under K061627
- · PLV Continuum II cleared under K034032
- · Legendair XL2 cleared under K070899
The iVent101 is viewed as substantially equivalent to the predicate device since them:
1. Have the same intended use:
2. Have the same environment for use:
3. Have the same patient populution :
4. Employ a same technology [utilized an internal compressor to generate compressed air and breathe delivery is controlled by software algorithms as the predicates]
## Performance Standards:
No performance standards have been established for such devices under Sections 514 of the Federal Food, Drug, and Cosmetic Act. However, the iVent101 complies with the following voluntary standards:
ASTM F 1100-90, "Standard specification for ventilators intended for use in critical care ventilators".
ASTM F 1246-91," Standard specification for ventilators intended for use in home care ventilators".
Draft Reviewer guidance for ventilators , July 1995
{3}------------------------------------------------
ISO 10651-2," Lung Ventilators for medical use - Particular requirements for Home Care Ventilators", Second Edition (2004)
ISO 10651-6," Medical Electrical Equipment" part 1-6: General requirements for safety-collateral standard: Usability, First Edition (2004)
IEC 60601-1," Medical Electrical Equipment - General requirements for safety", Second edition (1990), #2 (1995), 13(1996)
IEC 60601-2-12," Medical electrical equipment. Part 2: Particular requirements for the safety of lung ventilators for medical use", Second Edition(2001).
IEC 60601-1-2, "Electromagnetic Compatibility (EMC)", 2005
IEC 60601-1-8, "General requirements for safety - collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems", First Edition (2003).
UL 60601-1," Medical Electrical Equipment, Part1: General Requirements for safety " , first edition (2003) .
### Summary of clinical performance:
The safety and efficacy of mechanical ventilation is well established in scientific research and literature. Due to comprehensive scientific literatures, performance tests per ASTM 1246-91, ASTM 1100-90, IEC and ISO standards, VersaMed believes that clinical studies are not required to determine the safety and efficacy of the device.
#### Conclusions from Non-clinical Tests:
The iVent101 performs as intended according to its performance specification. The iVent101 is substantially equivalent to the predicate devices.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure.
MAR 1 6 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Jerry Korten CEO VersaMed Medical Systems, Incorporated 2 Blue Hill Plaza Building 2, 300 Floor Pearl River. New York 10965
Re: K081845
> Trade/Device Name: The iVent101 Regulation Number: 21 CFR 868,5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK , NOU Dated: March 9, 2009 Received: March 12, 2009
#### Dear Mr. Korten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Korten
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Saiste y. Michien Dnis
Ginette Y. Michaud. M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known):
Device Name:
..
The iVent101
1681843
Indications for Use:
The iVent101 ventilator [with single or dual limb configuration] is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for infants from 5 kg through adult patients who require invasive or non-invasive support via the following ventilatory modalities:
- Assist/Control and SIMV with either Volume, Pressure, or Pressure Regulated . Volume Control (PRVC)
- CPAP with Pressure Support .
- Adaptive Bi-Level for either NIV or invasive ventilation. ●
The iVent101 ventilator is suitable for use in institution, home and portable settings.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Ruma
(Division Sign-Off)___ Division of Anesthesiology, General Hospital, Infection Control, Dental Devices-
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.