← Product Code [NBW](/productcode/NBW) · K081796

# EASYPLUS MINI R2N SELF MONITORING GLUCOSE TEST SYSTEM (K081796)

_Eps Bio Technology Corp. · NBW · Jan 23, 2009 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081796

## Device Facts

- **Applicant:** Eps Bio Technology Corp.
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Jan 23, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The EasyPlus mini R2N Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The EasyPlus mini R2N Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or forearm. EasyPlus mini R2N Blood Glucose Test Strips must be used with the EasyPlus mini R2N Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The EasyPlus mini R2N Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or forearm. EasyPlus mini R2N Blood Glucose Test Strips must be used with the EasyPlus mini R2N Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The EasyPlus mini R2N Glucose Normal/High Control Solution are for use with the EasyPlus mini R2N meter and EasyPlus mini R2N Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

## Device Story

System measures glucose in capillary whole blood via electrochemical biosensor. User applies blood to disposable test strip; glucose oxidase catalyzes oxidation of glucose; mediator transfers electrons to carbon electrode, generating current proportional to glucose concentration. Meter processes current to display result in 5 seconds. Used at home by patients or in clinics by professionals. Output aids diabetes management. System includes meter, test strips, auto-lancet, check strip, and control solutions. Benefits include rapid, quantitative monitoring of blood glucose levels.

## Clinical Evidence

Bench testing only. Precision/reproducibility studies (N=200-400) showed CVs ranging from 2.10% to 4.44%. Linearity confirmed (R2=0.9950). Method comparison (N=165) against YSI 2300 analyzer met ISO 15197 accuracy requirements: 95% of results within ±15 mg/dL (for <75 mg/dL) or ±20% (for ≥75 mg/dL). Hematocrit interference tested (20-60% range).

## Technological Characteristics

Amperometric glucose oxidase assay; carbon electrodes; potassium ferricyanide mediator. Dimensions: 3.5" x 2.1" x 0.97". Power: 3V (2x CR2032 batteries). No-code calibration. Standalone device; no wireless/data transmission. Standards: CLSI EP5-A, EP6-A, EP7-A; ISO 15197:2003; ISO 14971:2000; EN 13640:2002.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- EasyPlus Self-Monitoring Blood Glucose System (k061992)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k081796

B. Purpose for Submission:
Clearance of a new device

C. Measurand:
Whole blood glucose

D. Type of Test:
Whole blood glucose concentration through a quantitative amperometric assay (Glucose Oxidase)

E. Applicant:
EPS Bio Technology Corp.

F. Proprietary and Established Names:
EasyPlus mini R2N Self-Monitoring Blood Glucose Test System

G. Regulatory Information:

1. Regulation section:
21 CFR § 862.1345, Glucose Test System
21 CFR § 862.1660, Quality control material (assayed and unassayed)

2. Classification:
Class II
Class I (reserved)

3. Product code:
NBW, CGA
JJX

4. Panel:
75 (Clinical Chemistry)

H. Intended Use:

1. Intended use(s):
See indications for use below.

2. Indication(s) for use:
The EasyPlus mini R2N Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from

{1}

the fingertips or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The EasyPlus mini R2N Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or forearm. EasyPlus mini R2N Blood Glucose Test Strips must be used with the EasyPlus mini R2N Meter. Testing is done outside the body (In Vitro diagnostic use).

It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The EasyPlus mini R2N Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or forearm. EasyPlus mini R2N Blood Glucose Test Strips must be used with the EasyPlus mini R2N Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The EasyPlus mini R2N Glucose Normal/High Control Solution are for use with the EasyPlus mini R2N meter and EasyPlus mini R2N Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

3. Special conditions for use statement(s):
- Not for use on critically ill patients, dehydrated patients or hyperosmolar patients
- Not for neonatal use
- Not for screening or diagnosis of diabetes mellitus
- Alternative site testing is for use at times of steady state only

4. Special instrument requirements:
EasyPlus mini R2N Self-Monitoring Blood Glucose Test System

I. Device Description:
The EasyPlus mini R2N Self-Monitoring Blood Glucose System consists of the EasyPlus mini R2N Blood Glucose Meter, EasyPlus mini R2N Glucose Test Strips, Auto-Lancet Device, Check Strip and Control Solution. The control solutions available are used to test the performance of the device.

J. Substantial Equivalence Information:
1. Predicate device name(s):
EPS Bio Technology Corp., EasyPlus Self-Monitoring Blood Glucose System
2. Predicate 510(k) number(s):
k061992

{2}

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Detection method | Amperometry | Amperometry  |
|  Enzyme | Glucose oxidase (Aspergillus niger) | Glucose oxidase (Aspergillus niger)  |
|  Mediator | Potassium ferricyanide | Potassium ferricyanide  |
|  Electrode | Carbon electrode | Carbon electrode  |
|  Test time | 5 seconds | 5 seconds  |
|  Temperature range | 10-40°C | 10-40°C  |
|  Humidity range | R.H. ≤ 90% | R.H. ≤ 90%  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Sample volume | ≥ 0.6 uL | ≥ 2.0 uL  |
|  Test range | 20-600 mg/dL | 30-550 mg/dL  |
|  Coding | One Code | Code card  |
|  Hematocrit range | 20-60% | 30-55%  |
|  Memory capability | 480 tests with date and time | 100 tests with date and time  |
|  Power | 3V 2X CR2032 batteries | 1.5V (AAA) batteries  |
|  Battery life | Approx. 2000 tests | Approx. 1000 tests  |
|  Size L x W x H (inch) | 3.5”x 2.1”x 0.97” | 3.2”x 2”x 0.7”  |
|  Weight | 2.05 oz (without batteries) | 1.6 oz (without batteries)  |

K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A, Precision Performance of Clinical Chemistry Devices
- CLSI EP6-A, Evaluation of the Linearity of Quantitative Analytical Methods; Proposed Guideline
- CLSI EP7-A, Interference Testing in Clinical Chemistry; Proposed Guideline
- ISO 15197:2003, In Vitro Diagnostic Test Systems – Requirements for Blood Glucose Monitoring Test Systems for Self Managing Diabetes Mellitus
- ISO 14971:2000, Medical Devices – Application of Risk Management to Medical Devices
- EN 13640:2002 Stability Testing of In Vitro Diagnostic Reagents

L. Test Principle:
Glucose measurement is based on electrical potential caused by the reaction of glucose with the reagents contained on the strip's electrodes. The glucose in the sample is oxidized by the enzyme glucose oxidase, and the current resulting from this enzymatic reaction is measured and converted to glucose concentration by the meter. The magnitude of the current is proportional to the concentration of glucose in the sample.

{3}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Within-Run

The testing was conducted using venous blood, collected in a heparin blood collection tube. Glucose was added to the blood to prepare 6 different levels of glucose for the testing. The glucose concentration ranges were: 20-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL, and 401~600 mg/dL. For each testing range, 10 Meters, 10 test strips for each meter, and 2 lots of test strips were used. (N=10 Meter x10 tests x 2 lots =200)

|  Range (mg/dL) | N | Lot: 021074801  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Mean (mg/dL) | SD (mg/dL) | CV (%)  |
|  20~50 | 100 | 53.8 | 1.57 | 2.92%  |
|  51~110 | 100 | 76.0 | 1.98 | 2.61%  |
|  111~150 | 100 | 126.8 | 3.28 | 2.59%  |
|  151~250 | 100 | 214.6 | 7.05 | 3.29%  |
|  251~400 | 100 | 363.0 | 8.17 | 2.25%  |
|  401~600 | 100 | 561.6 | 12.39 | 2.21%  |
|  Normal control solution | 100 | 111.6 | 3.92 | 3.51%  |
|  Range (mg/dL) | N | Lot: 021075201  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Mean (mg/dL) | SD (mg/dL) | CV (%)  |
|  20~50 | 100 | 55.9 | 2.48 | 4.44%  |
|  51~110 | 100 | 77.5 | 2.80 | 3.62%  |
|  111~150 | 100 | 132.3 | 3.27 | 2.47%  |
|  151~250 | 100 | 221.2 | 5.92 | 2.67%  |
|  251~400 | 100 | 368.9 | 7.74 | 2.10%  |
|  401~600 | 100 | 563.4 | 13.96 | 2.48%  |
|  Normal control | 100 | 119.7 | 4.38 | 3.66%  |

{4}

5

|  solution |  |  |  |   |
| --- | --- | --- | --- | --- |

# Day-to-Day Precision

10 Meters, 2 lots of test strip, and 3 control solutions of Low, Normal and High were prepared. Each control was tested twice a day, once in the morning and once in the afternoon for 10 days. (N=10 Meter x 2 Lots x 2 tests x10 days =400)

|  Control solution | N | mean (mg/dL) | SD (mg/dL) | CV (%)  |
| --- | --- | --- | --- | --- |
|  Low | 40
0 | 47.1 | 1.67 | 3.54%  |
|  Normal | 40
0 | 122.2 | 4.62 | 3.78%  |
|  High | 40
0 | 398.5 | 10.57 | 2.65%  |

## b. Linearity/assay reportable range:

Testing was performed using whole blood supplemented with $\beta$-D-glucose to provide 630 samples at blood glucose levels between 14 and $613\,\mathrm{mg/dL}$. The range of samples tested on the meter that gave a numeric result was between 22 and $600\,\mathrm{mg/dL}$.

The linear regression was as follows:

|  N= | 630  |
| --- | --- |
|  Slope | 1.0263  |
|  Y-intercept | -1.6952  |
|  R² | 0.9949  |

The sponsor claims $20\,\mathrm{mg/dL}$ as the lowest detectable limit.

## c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The sponsor claims that the device is traceable to a laboratory analyzer which is calibrated to a glucose standard (NIST SRM 965a).

The two control solutions (normal and high levels) consist of buffered aqueous solutions of $\beta$-D-glucose and non-reactive ingredients. Lot-specific ranges are printed on the test strip bottle label.

Closed Vial Stability of Control Solution: Three lots of each glucose level of control solution were tested using an accelerated stability study. The results showed that during the simulated time frame of 19 months at room temperature, the controls were within the specified ranges (96 mg/dL – 144 mg/dL for normal level and 240 mg/dL – 420 mg/dL for high level).

{5}

Opened Vial Stability of Control Solution: The measured glucose levels of 2 lots of each level of control solution were tested during the use lifetime period (90 days) and were found to be within the specified ranges (96 mg/dL – 144 mg/dL for normal level and 240 mg/dL – 420 mg/dL for high level), and the average bias with respect to the mean value throughout the lifetime is less than ≤20% and a CV of 6%.

d. Detection limit:
The measuring range of the system is 20 - 600 mg/dL. This range was verified by the linearity study (above section M.1.b.).

e. Analytical specificity:
Spiked whole blood samples containing three levels of glucose (50, 250, and 500 mg/dL), with and without interfering substances, were prepared to test common endogenous and exogenous substances for interference. Levels tested for each interferant (in mg/dL) are summarized below:

|  Interferent (mg/dL) | Level 1 | Level 2 | Level 3  |
| --- | --- | --- | --- |
|  Acetaminophen | 2 | 8 | 15  |
|  Ascorbic acid | 2.5 | 3.0 |   |
|  Dopamine | 2 | 13 |   |
|  Gentisic acid | 6 | 8 |   |
|  Ibuprofen | 4.2 | 40 |   |
|  L-Dopa | 2 | 10 |   |
|  Methyldopa | 1 | 2 | 2.5  |
|  Sodium Salicylate | 30 | 40 | 50  |
|  Tetracycline | 0.4 |  |   |
|  Tolbutamide | 10 | 64 | 100  |
|  Bilirubin – unconjugated | 1.2 |  |   |
|  Cholesterol | 500 |  |   |
|  Creatinine | 1.5 | 30 |   |
|  Triglycerides | 190 | 3000 |   |
|  Uric acid | 10 | 13 | 14  |
|  Galactose | 10 | 20 |   |
|  Maltose | 5 | 10 |   |
|  Xylose | 10 | 20 |   |

Interference testing showed that the system results in less than ±10% bias in the presence of high concentrations of interferants even beyond clinically significant ranges.

Positive errors greater than 10% were found in the presence of the following and are presented in the labeling under limitations for healthcare professionals:
- Acetaminophen (15 mg/dL at glucose levels ≤250 mg/dL)
- Ascorbic acid (3 mg/dL at glucose levels of 50 mg/dL)
- Dopamine (13 mg/dL at glucose levels ≥50 mg/dL)

{6}

- Gentisic acid (8 mg/dL at glucose levels of 50 mg/dL)
- L-Dopa (10 mg/dL at glucose levels ≥50 mg/dL)
- Methlydopa (1 mg/dL at glucose levels of 50 mg/dL and 2.5 mg/dL at glucose levels of 50 mg/dL)
- Triglycerides (3000 mg/dL at glucose levels of 250 mg/dL)
- Uric acid (14 mg/dL at glucose levels of 50 and 500 mg/dL)

An altitude study was performed with 9 blood samples ranging from 29 to 595 mg/dL and spiked whole blood samples of 60 and 310 mg/dL. Testing was performed at altitudes of 160 feet and 10,360 feet. The test data using capillary blood samples showed that the bias versus YSI at 10,360 feet is the same bias versus YSI as that observed at 160 feet. These data indicate no additional effect due to altitude up to the claimed altitude of 10,000 feet.

Hematocrit interference was evaluated by adjusting the glucose concentrations and hematocrit levels of venous blood samples from several donors. The venous blood samples were spiked to 4 glucose concentrations (50, 120, 260, and 500 mg/dL). The hematocrit levels were adjusted to 0%, 20%, 40%, 50%, 60% and 70%. Each sample was assayed n=10 for each of two strip lots and the maximum percent bias was calculated compared to the 40% hematocrit samples. The sponsor demonstrated that the bias did not exceed ±15 mg/dL when glucose concentration is ≤75 mg/dL and bias did not exceed ±15% when glucose concentration is &gt;75 mg/dL for hematocrit concentrations within the claimed range of 20% to 60%.

f. Assay cut-off:
Not Applicable.

2. Comparison studies:

a. Method comparison with predicate device:

To demonstrate the accuracy performance of the EasyPlus mini R2N Self-Monitoring Blood Glucose Test System a total of 156 diabetes patients performed a finger stick and collected blood from the forearm using the EasyPlus mini R2N system at 3 sites. A healthcare professional then performed the test on the EasyPlus mini R2N and the YSI 2300 analyzer. The range of tested values for these samples was 74.2-486 mg/dL. In order to fully cover the measuring range, 9 of the samples were spiked with glucose or allowed to glycolyze. The total 165 (156+9) samples ranged in concentration from 32 to 591 mg/dL. Each sampling site met the ISO 15197 requirement of ninety-five percent (95%) of the individual glucose results falling within ±15 mg/dL of the results of the manufacturer's measurement procedure at glucose concentrations for samples &lt;75 mg/dL and within ±20% at glucose concentrations ≥75 mg/dL. Results are summarized below:

|  Linear Regression | Y=1.0263x – 1.6952  |
| --- | --- |
|  R² | 0.9949  |

{7}

|  Lay User | Fingerstick vs. YSI | Forearm vs. Fingerstick  |
| --- | --- | --- |
|  Total | 150/156 (96%) | 148/156 (95%)  |
|  Professional | Fingerstick vs. YSI | Forearm vs. Fingerstick  |
| --- | --- | --- |
|  Total | 160/165 (97%) | 150/156 (96%)  |

b. Matrix comparison:
Not applicable. Capillary whole blood is the only indicated matrix.

3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided a readability study that indicated that the user manual, test strip labeling, and control solution labeling is at an 8th grade reading level or below.

4. Clinical cut-off:
Not Applicable.

5. Expected values/Reference range:
The sponsor included the following expected values for normal glucose levels in their strip labeling:

70-110 mg/dL

Two-hours after meals: less than 120 mg/dL

American Diabetes Associations: Standards of Medical Care for Patients with Diabetes Mellitus, Diabetes Care, 25(2002), p.S37.

N. Instrument Name:
EasyPlus mini R2N Self-Monitoring Blood Glucose Test System

O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?

{8}

Yes ______ or No ☐ X

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ______ or No ☐ X

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X or No ☐

9

{9}

10

3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.

4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger, palm, and forearm. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues.

5. Calibration:
This meter is a no code meter. A single calibration code has been selected to be the code for all test strips for the meter which removes the need to manually code the meter.

6. Quality Control:
The sponsor has two levels of controls available for this meter with both levels coming with the kit and also being available through the distributor. When a test strip is inserted into the meter, a control can be run. An acceptable range for each control level is printed on the test strip vial label. The user is referred to a troubleshooting section at the end of the control test instructions of the owner's manual to identify possible reasons control results fall outside these ranges.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None.

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K081796](https://fda-staging.innolitics.com/device/K081796)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
