ECHELON RAPID PV COIL, MODEL MR-PVC-150, ECHELON RAPID FOOT COIL, MODEL MR-RFC-150 AND ECHELON RAPID KNEE COIL
K081745 · Hitachi Medical Systems America, Inc. · MOS · Aug 12, 2008 · Radiology
Device Facts
Record ID
K081745
Device Name
ECHELON RAPID PV COIL, MODEL MR-PVC-150, ECHELON RAPID FOOT COIL, MODEL MR-RFC-150 AND ECHELON RAPID KNEE COIL
Applicant
Hitachi Medical Systems America, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The ECHELON Specialty Coils are receive-only devices that detect the MR signal used to produce transverse, coronal, sagittal, oblique, and/or curved cross-sectional images that display the internal structure of the body. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The indications for use for the ECHELON Specialty Coils are as follows: - The MR-PVC-150, ECHELON RAPID PV Coil is a receive-only multiple array device . used for MRI imaging of the lower body region. - The MR-RFC-150, ECHELON RAPID Foot Coil is a receive-only multiple array device . used for MRI imaging of the feet. - The MR-RKC-150N, ECHELON RAPID Knee Coil is a receive-only multiple array device used for MRI imaging of the knee.
Device Story
ECHELON RAPID Specialty Coils (PV, Foot, Knee) are receive-only multiple array MRI coils. Devices detect MR signals from hydrogen nuclei in static magnetic fields; signals are spatially encoded via magnetic field gradients. Output consists of raw MR signal data processed by the host MRI system to reconstruct cross-sectional images. Used in clinical settings by trained physicians to provide diagnostic information. Benefits include non-invasive visualization of internal body structures without ionizing radiation. PV Coil features 16 elements; Foot Coil features 8 channels; Knee Coil features 12 channels. Signals from each element are independently processed by the MRI system to enhance image performance.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Receive-only multiple array MRI coils. PV Coil: 16 elements. Foot Coil: 8 channels. Knee Coil: 12 channels. Independent signal processing per element. Compatible with existing MRI system reconstruction techniques.
Indications for Use
Indicated for MRI imaging of the lower body region (PV Coil), feet (Foot Coil), and knee (Knee Coil) in patients requiring non-invasive diagnostic imaging.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MR-RBC-150N
ECHELON RAPID Torso Coil
Submission Summary (Full Text)
{0}------------------------------------------------
K081745
### Submitter Information
| Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371<br>ph: (330) 425-1313<br>fax: (330) 963-0749 | | AUG 12 2008 |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--|-------------|
| Contact: | Douglas J. Thistlethwaite | | |
| Date: | June 12, 2008 | | |
| Device Name | | | |
| Classification Name: | Coil, magnetic resonance, specialty | | |
| Classification Number: | 90MOS | | |
| Trade/Proprietary Name: | ECHELON RAPID PV Coil | | |
| Predicate Device(s): | MR-RBC-150N, ECHELON RAPID Torso Coil | | |
| Classification Name: | Coil, magnetic resonance, specialty | | |
| Classification Number: | 90MOS | | |
| Trade/Proprietary Name: | ECHELON RAPID Foot Coil | | |
| Predicate Device(s): | MR-RBC-150N, ECHELON RAPID Torso Coil | | |
| Classification Name: | Coil, magnetic resonance, specialty | | |
| Classification Number: | 90MOS | | |
| Trade/Proprietary Name: | ECHELON RAPID Foot Coil | | |
| Predicate Device(s): | MR-RBC-150N, ECHELON RAPID Torso Coil | | |
## Device Intended Use
The MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The ECHELON Specialty Coils are receive-only devices that detect the MR signal used to produce transverse, coronal, sagittal, oblique, and/or curved cross-sectional images that display the internal structure of the body. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
The indications for use for the ECHELON Specialty Coils are as follows:
- The MR-PVC-150, ECHELON RAPID PV Coil is a receive-only multiple array device . used for MRI imaging of the lower body region.
- The MR-RFC-150, ECHELON RAPID Foot Coil is a receive-only multiple array device . used for MRI imaging of the feet.
- The MR-RKC-150N, ECHELON RAPID Knee Coil is a receive-only multiple array � device used for MRI imaging of the knee.
{1}------------------------------------------------
## Device Description
#### Function
The MR-PVC-150, ECHELON RAPID PV Coil is a receive-only multiple array device used for MRI imaging of the lower body region.
The MR-RFC-150. ECHELON RAPID Foot Coil is a receive-only multiple array device used for MRI imaging of the feet.
The MR-RKC-150N. ECHELON RAPID Knee Coil is a receive-only multiple array device used for MRI imaging of the knee.
#### Scientific Concepts
Magnetic Resonance Imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field. referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2.
A RF emission or echo that can be measured accompanies these relaxation events. The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.
#### Physical and Performance Characteristics
The MR-RHC-150N, ECHELON RAPID PV Coil is a receive-only device suitable for imaging the lower body region. The coil consists of 16 elements. The signal output of each element is independently processed by the MRI system to enhance performance.
The MR-RBC-150N. ECHELON RAPID Foot Coil is a receive-only device suitable for imaging feet. The coil consists of 8 channels. The signal output of each element is independently processed by the MRI system to enhance performance.
The MR-RBC-150N, ECHELON RAPID Knee Coil is a receive-only device suitable for imaging the knee. The coil consists of 12 channels. The signal output of each element is independently processed by the MRI system to enhance performance.
{2}------------------------------------------------
# Device Technological Characteristics
The technological characteristics of the ECHELON RAPID PV Coil, ECHELON RAPID Foot Coil, and ECHELON RAPID Knee Coil are similar to the predicate devices as listed in Section 10 - Substantial Equivalence Discussion.
# Conclusions
It is the opinion of Hitachi Medical Systems America, Inc. that the ECHELON RAPID PV Coil, ECHELON RAPID Foot Coil, and ECHELON RAPID Knee Coil are substantially equivalent to the listed predicate devices.
. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 1 2 2008
Mr. Doug Thistlewaite Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087-1313
Re: K081745
Trade/Device Name: ECHELON PV Coil, Foot Coil, and Knee Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: June 13, 2008 Received: June 19, 2008
Dear Mr. Thistlewaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. 'The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and liasing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to I ogally marketed predicate device results in a classification for your device and thus, verse to volu device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance a one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, release contact CDRH s Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 24/0276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Division of One Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.btml
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): ECHELON RAPID PV Coil Device Name:
#### Indications for Use:
The MR-PVC-150, ECHELON RAPID PV Coil is a receive-only multiple array device used for MRI imaging of the lower body region.
Prescription Use × AND/OR
Over-the-Counter Use
(21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Hubert Lunen
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices ് ' ()(k) Number __
{6}------------------------------------------------
510(k) Number (if known): ECHELON RAPID Foot Coil Device Name:
#### Indications for Use:
The MR-RFC-150, ECHELON RAPID Foot Coil is a receive-only multiple array device used for MRI imaging of the feet.
Prescription Use × AND/OR
Over-the-Counter Use
(21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Huko Leun
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{7}------------------------------------------------
510(k) Number (if known): Device Name: ECHELON RAPID Knee Coil
#### Indications for Use:
The MR-RKC-150N, ECHELON RAPID Knee Coil is a receive-only multiple array device used for MRI imaging of the knee.
Prescription Use × AND/OR
Over-the-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helwteens
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.