INNOVANCE™ D-Dimer: For the quantitative determination of cross-linked fibrin degradation products (Ddimers) in human plasma on the Siemens and Sysmex Coagulation Systems. The INNOVANCE™ D-Dimer assay is intended for use as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. INNOVANCE™ D-Dimer Controls: INNOVANCE™ D-Dimer Control 1 and INNOVANCE™ D-Dimer Control 2 are assayed, normal and pathological level, intralaboratory quality controls for assessment of precision and analytical bias in the quantitative determination of D-dimer on the Siemens and Sysmex Coagulation Systems.
Device Story
In vitro diagnostic assay for quantitative D-dimer measurement in human plasma; used on Dade Behring and Sysmex Coagulation Systems (BCS/BCS XP). Principle: Polystyrene particles coated with monoclonal antibody (8D3) aggregate in presence of D-dimer; aggregation detected via turbidimetry (increased turbidity). Input: Human plasma samples. Output: Quantitative D-dimer concentration (mg/L FEU). Used in clinical laboratory settings by trained personnel. Results aid clinicians in diagnosing VTE (DVT/PE) by ruling out or confirming suspected cases in conjunction with Wells' pre-test probability models and objective diagnostic imaging. Benefits include rapid, automated assessment of fibrin degradation products to support clinical decision-making for VTE.
Clinical Evidence
Prospective clinical study of 902 out-patients (ages 18-95) with suspected VTE. Patients evaluated via Wells' PTP models and objective diagnostic tests (spiral CT/compression ultrasonography) with 4-month follow-up. Results at 0.5 mg/L cutoff: 97% sensitivity, 42% specificity, 98% NPV. Additional derivation study (n=359) showed 98% sensitivity, 38% specificity, 98% NPV. Bench testing included precision (CLSI EP5-A2), linearity (CLSI EP06-A), LoB/LoD (CLSI EP17-A), and interference (CLSI EP7-A2).
Technological Characteristics
Turbidimetric immunoassay using polystyrene particles coated with monoclonal antibody (8D3). Operates on automated Siemens and Sysmex Coagulation Systems. Quantitative measurement of D-dimer in human plasma.
Indications for Use
Indicated for the quantitative determination of D-dimer in human plasma as an aid in the diagnosis of venous thromboembolism (VTE), including deep vein thrombosis (DVT) or pulmonary embolism (PE), in patients aged 18 to 95 years presenting to the emergency room with clinically suspected VTE.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k081732
B. Purpose for Submission:
Clearance of a new assay
C. Measurand:
D-Dimer
D. Type of Test:
Quantitative Turbidometry
E. Applicant:
SIEMENS HEALTHCARE DIAGNOSTICS
F. Proprietary and Established Names:
Innovance D-Dimer
Innovance D-dimer Controls
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DAP | Class II | 21 CFR 864.7320 | 81 Hematology |
H. Intended Use:
1. Intended use(s):
INNOVANCE™ D-Dimer:
{1}
For the quantitative determination of cross-linked fibrin degradation products (D-dimers) in human plasma on the Dade Behring and Sysmex Coagulation Systems. The INNOVANCE™ D-Dimer assay is intended for use as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)].
## INNOVANCE™ D-Dimer Controls:
INNOVANCE™ D-dimer Control 1 and INNOVANCE D-Dimer Control 2 are assayed, normal and pathological level, intralaboratory quality controls for assessment of precision and analytical bias in the quantitative determination of D-dimer on the Dade Behring and Sysmex Coagulation Systems.
2. Indication(s) for use:
The Indications for Use for the INNOVANCE™ D-Dimer and INNOVANCE™ D-Dimer Controls are attached.
3. Special conditions for use statement(s): Prescription
4. Special instrument requirements:
INNOVANCE™ D-Dimer and INNOVANCE™ D-Dimer Controls are for use on the Dade Behring Coagulation Systems - K970431 and its family member BCS® XP
## I. Device Description:
## INNOVANCE™ D-Dimer:
INNOVANCE™ D-Dimer Reagent - lyophilized mouse monoclonal antibodies to D-dimer (3 or 6 vials, 4.0 mL per vial)
INNOVANCE™ D-Dimer Buffer - liquid saline buffer (3 or 6 vials, 5.0 mL per vial)
INNOVANCE™ D-Dimer Supplement - liquid saline buffer with heterophilic blocking reagent (3 or 6 vials, 2.6 mL per vial)
{2}
INNOVANCE™ D-Dimer Diluent - liquid saline buffer to dilute samples (3 or 6 vials, 5.0 mL per vial)
INNOVANCE™ D-Dimer Calibrator - lyophilized, single analyte, human plasma based product containing D-dimer preparation (2 vial, 1.0 mL per vial)
Empty Vials - 3 vials per INNOVANCE™ D-Dimer Reagent, INNOVANCE™ D-Dimer Buffer, INNOVANCE™ D-dimer Supplement and INNOVANCE™ D-Dimer Diluent
Polystyrene particles covalently coated with a monoclonal antibody (8D3)² are aggregated when mixed with samples containing D-dimer. The D-dimer cross-linkage region has a stereo symmetrical structure, i.e. the epitope for the monoclonal antibody occurs twice. Consequently, one antibody suffices in order to trigger an aggregation reaction, which is then detected turbidimetrically via the increase in turbidity
## INNOVANCE™ D-Dimer Controls
Content: 5 vials each level, two levels, 1 mL per vial
INNOVANCE™ D-Dimer Controls 1 and 2 are lyophilized, single analyte, human plasma based products containing D-dimer.
J. Substantial Equivalence Information:
| Predicate | Stratus CS DDMR TestPak, CALPak and DILPak- K063356 |
| --- | --- |
| Describe the item being compared | |
| Proposed device versus predicate. | |
| Similarities | |
| Similarities between the Proposed Device and Predicate Device include: | |
| Intended Use: Both are in vitro diagnostic tests for the quantitative measurement of D-dimer in human plasma, as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE). | |
| Antibody: The antibody for both assays is monoclonal from mice. | |
| Differences | |
| Differences between the Proposed Device and the Predicate Device include: | |
| Technology: | |
{3}
4
Proposed Device - immunochemical reactions that are measured using turbidimetric technology
Predicate Device- the enzymatic rate of the bound fraction increases directly with the concentration of D-dimer in the sample. The reaction rate can then be measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
Instrument:
Proposed Device-
BCS®/BCS® XP System
Predicate Device - Stratus® CS Analyzer
| Predicate | Advanced D-Dimer Control Plasma 1 and 2 - K992957 |
| --- | --- |
| Describe the item being compared | |
| Propose Device versus Predicate Device | |
| Similarities | |
| Similarities between the Proposed Device and Predicate Device include: | |
| Intended Use: Both are assayed, normal and pathological level, intralaboratory quality controls for assessment of precision and analytical bias in the quantitative determination of D-dimer. Constituents: D-dimer Traceability: In house reference standard Levels: Two | |
| Form: | |
| Proposed Device- Lyophilized | |
| Predicate Device- Liquid | |
| Differences | |
| Differences between the Proposed Device and the Predicate Device include: | |
| Proposed Device: | |
| For use with the INNOVANCE D-Dimer assay | |
| Predicate Device | |
| For use with the Advanced D-Dimer assay | |
{4}
5
K. Standard/Guidance Document Referenced (if applicable):
STANDARDS
Title and Reference Number
CLSI - EP7A2, Interference testing in Clinical Chemistry; Approved Guideline
CLSI EP-5A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.
CLSI EP9-A2; Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline
CLSI EP6-P2 Evaluation of the Linearity of Quantitative Analytical Methods: A Statistical Approach
NCCLS/CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation
Other Standards
GUIDANCE
| Document Title | Office | Division | Web Page |
| --- | --- | --- | --- |
| Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material | | | http://www.fda.gov/cdrh/oivd/guidance/2231.html |
L. Test Principle:
Polystyrene particles covalently coated with a monoclonal antibody (8D3) are aggregated when mixed with samples containing D-dimer. The D-dimer cross-linkage region has a stereosymmetrical structure, i.e., the epitope for the monoclonal antibody occurs twice. Consequently, one antibody suffices in order to trigger an aggregation reaction, which is then detected turbidimetrically via the increase in turbidity.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{5}
Precision testing was conducted by following CLSI EP5-A2.
Testing was performed over 20 days, two separate runs with two test samples for each test material.
One operator, one reagent, and one instrument were used.
n 80
Mean [mg/L] 0.279
Median [mg/L] 0.280
Repeatability [%] 4.1 (SD: 0.011)
Within-day-CV [%] 4.1 (SD: 0.011)
Between-day-CV [%] 1.2 (SD: 0.003)
Within-device/lab-CV [%] 4.3 (SD: 0.012)
b. Linearity/assay reportable range:
The linearity study covered a range of 0.12 to 4.46 mg/L FEU. Linearity testing was performed using a citrated plasma sample spiked with D-Dimer, concentration of 4.46 mg/L FEU that was diluted with INNOVANCE™ D-Dimer Diluent.
The linear range was determined according to CLSI EP06-A. Based on the results of this testing and that from the Limit of Blank and Limit of Detection Study, the measuring range was established.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The calibrator is traceable to an internal reference standard. A new calibration is required for each new lot of INNOVANCE™ D-Dimer, after major maintenance or service, as indicated by quality control results, as indicated in laboratory quality control procedures, when required by government regulations.
The INNOVANCE™ D-Dimer Controls 1 and 2 are single analyte, lyophilized, human plasma based products containing D-dimer. The control is traceable to an in house reference standard. INNOVANCE™ D-Dimer Controls must be tested at least every eight hours on each testing day and for each vial of reagent for the respective measurement range to ensure that the system is functioning correctly.
Stability Protocol
{6}
7
INNOVANCE™ D-Dimer Calibrator and Controls 1 and 2
## Closed Vial
Study duration: 25 months
Replicates: 3 vials, 5 replicates per vial.
Testing Frequency: Product is stored at +2 to +8°C throughout testing cycle and tested on day 0 and after 3, 6, 12, 18 and 24 and 25 months.
Acceptance Criteria: Results obtained must be within a range of 80 to 120% of the control assigned value.
## Stability after reconstitution
Replicates: 1 vial, 2 replicates per vial
Testing Frequency (Calibrator):
0, 2, 4, 6 hours at +15 to +25°C
Testing Frequency (Controls):
0, 4, 6, 8, 10 hours at +15 to +25°C
0, 3, 5, 7, 10, 14, 15 days at +2 to +8°C
0, 2, 3, 4, 5 weeks at < or = 18°C
Acceptance Criteria: Results obtained must be within a range of 80 to 120% of the control assigned value.
## Onboard Stability
### Calibrators
Replicates: 1 vial, 1 replicate per vial
Testing Frequency:
BCS®/BCS® XP System - 0, 1, 2, 4, 6, 8 hours
Acceptance Criteria: Results obtained must be within a range of 90 to 110% of
{7}
the initial value of the calibrator.
## Controls
Replicates: 1 vial, 1 replicate per vial
Testing Frequency:
BCS®/BCS® XP System- 0, 4, 6, 8, 8.5, 9 hours
Acceptance Criteria: Results obtained must be within a range of 90 to 110% of the initial value of Control 1 and within a range of 85% to 115% of the initial value of Control 2.
Refer to the attached INNOVANCE™ D-Dimer Calibrator and Control value assignment flow chart, Section 5.1.4.
## d. Detection limit:
The Limit of Blank (LoB) and Limit of Detection (LoD) for INNOVANCE™ D-Dimer was determined to be 0.02 mg/L and 0.05 mg/L, respectively.
CLSI EP17-A was followed to establish the Limit of Blank and Limit of Detection.
Four blank samples, four low analyte citrated plasma samples spiked with D-dimer (diluted with INNOVANCE D-Dimer Diluent) and two reagent lots were tested over 4 days.
The Limit of Blank (LoB) for INNOVANCE™ D-Dimer was determined to be 0.02 mg/L. The Limit of Detection (LoD) L was determined to be 0.05 mg/L.
## e. Analytical specificity:
Test samples were prepared by spiking the potential interferent into plasma. D-Dimer concentrations ranged from 0.45 mg/L to 0.55 mg/L, except for hemoglobin, bilirubin and triglycerides, where D-Dimer concentrations ranged from 0.28 mg/L to 0.29 mg/L and 2.32 mg/L to 2.43 mg/L.
Interference testing was performed according to CLSI EP7-A2 to determine the effect of various endogenous and exogenous substances on the INNOVANCE™ D-Dimer assay. For all exogenous interferents, hemoglobin,
8
{8}
bilirubin and triglycerides the percent bias was determined by testing a control sample without the interferent and comparing to the value obtained from a test sample to which the potential interferent had been added.
To evaluate cholesterol, the endogenous concentration of the base pool was determined and additional cholesterol was spiked to the test concentration. The percent bias was determined by testing a normal control sample and comparing the value obtained from the test pool to which the cholesterol had been added.
To evaluate creatine, albumin, fibrinogen, Immunoglobulin G, urea, uric acid the endogenous concentration of the base pool and test pool were not determined. The test concentration was spiked into the test pool without taking the endogenous concentration into account. The test pool percent bias was determined by testing a normal control sample and comparing it to the value obtained from a test sample to which the potential interferent had been added.
To evaluate interference from rheumatoid factors interference, samples which had RF concentrations 3640 IU/mL and samples with no detectable RF concentration were used to prepare samples for the study. Different mixtures, e.g., 1+1, 1+3, 3+1, of samples with high concentrations of RF were prepared and the INNOVANCE D-Dimer assay concentrations were determined in replicates of five. The recovery of the D-dimer value ranged from 92% to 101%.
f. Assay cut-off:
2. Comparison studies:
a. Method comparison with predicate device:
2 clinical sites, 318 samples. Passing-Babcock regression yielded a slope of 0.951, intercept of 0.059 mg/L FEU and r of 0.97.
b. Matrix comparison:
3. Clinical studies: Frozen samples were collected from 902 out-patients presenting to the emergency room with clinically suspected venous thromboembolism (VTE). Patients were evaluated using the Wells' pre-test probability (PTP) models to estimate the probability of DVT or PE. Patients were diagnosed as DVT or PE positive by standard objective tests as appropriate.
9
{9}
Patients initially diagnosed as negative were followed for four months.
Samples were collected from 2 clinical sites, frozen, and tested at 2 additional sites. Of the 902 patients, 533 were female and 369 were male; all patients were between the ages of 18 and 95.
| Instrument | VTE Patients | Cutoff (mg/L) | Sensitivity (%) | Specificity (%) | NPV (%) | Frequency of VTE (%) |
| --- | --- | --- | --- | --- | --- | --- |
| BCS® Systems | 902 | 0.5 | 97 | 42 | 98 | 22.0 |
a. Clinical Sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
In a prospective single center study, plasma specimens from 359 out-patients suspected of VTE were tested with the INNOVANCE D-Dimer assay using a cutoff of .5 mg/L FEU. PE was ruled out or confirmed by spiral CT and a 3-month follow-up. DVT was ruled out or confirmed by compression ultrasonography and a 3-month follow-up. 89 of the 359 patients received a positive diagnosis for VTE resulting with a VTE frequency of 24% in this market.
INNOVANCE D-Dimer Derivation Study Summary
| Instrument | VTE Patients | Cutoff mg/L FEU | Sensitivity (%) | Specificity (%) | NPV (%) |
| --- | --- | --- | --- | --- | --- |
| BCS® System | 359 | 0.5 | 98 | 38 | 98 |
{10}
5. Expected values/Reference range:
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
{11}
12
{12}
13
{13}
14
{14}
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.