K081726 · Tyson Bioresearch, Inc. · NBW · Aug 26, 2008 · Clinical Chemistry
Device Facts
Record ID
K081726
Device Name
TYSONBIO LINK HEALTH MANAGEMENT SYSTEM
Applicant
Tyson Bioresearch, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Aug 26, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The TysonBio Link Health Management System is an optional software accessory for use with the following models with data management capabilities: a) EZ Smart-608 series meter, b) DIACHEX series Blood Glucose Meter, and c) DIACHEX+ series Blood Glucose Meter. When used with one of these meters, TysonBio Link Health Management System transfers data from the device’s memory into a computer for enhanced data management. TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level.
Device Story
TysonBio Link Health Management System is a Windows-based software accessory for glucose meters (EZ Smart-608, DIACHEX, DIACHEX+ series). It functions by connecting the meter to an Intel-compatible PC via a USB cable. The software extracts stored blood glucose data from the meter's memory; it does not modify or change the data. The system provides data management tools, including statistics, pie charts, line graphs, and histograms for review by patients and healthcare professionals. It is used in home and clinical settings to support diabetes management. The software does not provide medical advice or dosage recommendations. Benefits include improved data visualization and record-keeping for diabetes monitoring.
Clinical Evidence
No clinical data. The device underwent software verification and validation testing in accordance with FDA guidance for software contained in medical devices.
Technological Characteristics
Windows-based software application; requires Intel-compatible PC, Windows 2000/XP or later, 128 MB RAM, 100 MB free hard disk space, and USB port. Connectivity via USB cable. Data management functionality includes export and memory clearing. Does not modify meter measurements.
Indications for Use
Indicated for people with diabetes and their healthcare professionals in home and clinical settings to review, analyze, and evaluate blood glucose test results.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Lifescan, In Touch Diabetes Management Software (k984527)
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
k081726
B. Purpose for Submission:
New submission for an accessory data management software application for glucose meters
C. Manufacturer and Instrument Name:
Tyson Bioresearch, Tysonbio Link Health Management System
D. Type of Test or Tests performed:
Tysonbio Link Health Management System is a software-only medical device which serves as an internet-based accessory which interfaces between the software in personal glucose monitoring devices and the Tysonbio Link Health Management System.
E. System Descriptions:
1. Device Description:
The TysonBio Link Health Management System software provides users the ability to export data from compatible TysonBio meters to a computer via a cable connecting to the computer's required USB port. The TysonBio Link HMS is a Microsoft Windows based software application for diabetes data management. TysonBio Link HMS software is designed to operate an Intel compatible PC with Microsoft Windows 2000/XP or later operating system. The export of the data allows the data to be viewed only in the TysonBio Link HMS software. Aside from the ability to export the data stored in the meter's memory, the software also allows the ability to clear the meter's memory. Data transferred from the device cannot be changed or modified.
2. Principles of Operation:
The TysonBio Link Health Management System is an accessory to compatible meters: EZ Smart-608 series meter (k062829), DIACHEX series Blood Glucose Meter (k072854), and DIACHEX+ series Blood Glucose Meter (k073492)
3. Modes of Operation:
TysonBio Link HMS is compatible with Microsoft® Windows® 2000/XP version 2
Hardware specifications:
☐ Intel® Pentium® III 850 MHz processor or better
☐ Minimum 128MB of RAM or better
☐ Minimum 100 MB of free hard disk space or more
☐ USB port
☐ TysonBio Link PC link Cable
☐ Mouse and keyboard
☐ Printer for printing reports
4. Specimen Identification:
{1}
Page 2 of 7
Meter controlled
5. Specimen Sampling and Handling:
Not Applicable
6. Calibration:
Not Applicable
7. Quality Control:
Not Applicable
8. Software:
FDA has reviewed the applicant’s Hazard Analysis and software
Documentation: Yes ☐ X or No ☐
F. Regulatory Information:
1. Regulation Section:
21CFR §862.1345 -Glucose test system.
21CFR Sec.-862.2100 - Calculator/data processing module for clinical use.
2. Classification:
Class 2 and 1 respectively
3. Product Code:
NBW - System, Test, Blood Glucose, Over The Counter
JQP-Calculator/Data Processing Module, For Clinical Use
4. Panel:
Chemistry (75)
G. Intended Use:
1. Indication(s) for Use:
The TysonBio Link Health Management System is an optional software accessory for use with the following models with data management capabilities: a) EZ Smart-608 series meter, b) DIACHEX series Blood Glucose Meter, and c) DIACHEX+ series Blood Glucose Meter. When used with one of these meters, TysonBio Link Health Management System transfers data from the device’s memory into a computer for enhanced data management.
TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level.
2. Special Condition for use Statement(s):
Not Applicable
H. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) numbers:
Lifescan, In Touch Diabetes Management Software (k984527)
{2}
Page 3 of 7
2. Comparison with Predicate Device:
Similarities
| Items | IN TOUCH DIABETES Management Software | TysonBio Link Health Management System |
| --- | --- | --- |
| Intended Use | The IN TOUTCH Diabetes Management Software is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. | TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level |
| Cable availability | Serial cable, cable available separately | Serial cable, cable available separately |
| Capable of uploading data from various devices | Software driver must be uploaded on the device or install PC | Same |
| Computer system requirements | Minimum of 128 MB of RAM | Minimum of 128 MB of RAM or better |
| | 100-200 MB minimum free hard disk space during, 100 MB after installation | Minimum 100 MB of free hard disk space or more |
| | 9-pin or 25 pin COM or USB port | USB port |
| | Video monitor and adapter with at least 800x600 pixels screen resolution and 256 colors | Same |
| Software use indications | Home or clinic settings | Single or multiple user settings |
| Deleting patients and all accompanying records | Yes | Yes |
{3}
| Items | IN TOUCH DIABETES Management Software | TysonBio Link Health Management System |
| --- | --- | --- |
| Ability to uninstall program | Yes | Yes |
| Ability to personalize target ranges | Yes | Yes |
| Unit of measure display | Choice of mmol/L or mg/dL | mmol/L or mg/dL |
| Ability to enter hypoglycemic range | Yes | Yes |
| Report Types | Data list, data statistics, average reading, pie chart, Line graph, histogram, | Data list, data statistics, average reading, pie chart, Line graph, histogram, |
| Auto-detect COM port | Yes | Yes |
| Downloaded results cannot be edited or deleted | Yes | Yes |
| Ability to modify meter average results | Yes,7,14,30,60,90 days | Yes |
| Ability to clear meter results in memory | Yes | Yes |
| Required information on use entry | Yes | Yes, user ID, name, photo |
Differences
| Items | IN TOUCH DIABETES Management Software | TysonBio Link Health Management System |
| --- | --- | --- |
| Computer system requirements | Windows operating system 98 SE, Windows 2000, Windows XP home and Professional | only Windows 2000, Windows XP home and Professional |
| Manual Entry | Ability to add records manually | No |
| Copy database to separate file | Yes | No |
| Deleting results | Yes | No |
{4}
| Items | IN TOUCH DIABETES Management Software | TysonBio Link Health Management System |
| --- | --- | --- |
| Language capabilities | English, Spanish | English |
| Viewing the Owner's manual | Link provided via icom | No |
| Default glucose target ranges available | Yes | No |
| Search patient capability | Yes | No |
| Result type display | Choice of whole blood or plasma result types | No |
| Ability to set default favorite report | Yes | No |
| Ability to synchronize meter clock to PC upon download | Yes | No |
| Ability to display 12 or 24 hour clock format | Yes | No |
| Ability to change date format | Yes | No |
| Ability to link to older database versions | Yes | No old versions available at this time |
| Diabetes Educator information | One diabetes educator may be entered | No |
| Diabetes control | Insulin list, medication list, diet/exercise options | No |
| Specifying/Entering medication/insulin | Yes, up to 3 different medications/insulin types | No |
| Ability to enter insulin regiment | Yes | No |
| Report types | Standard Day, Average Reading, Insulin, Exception, Data List Histogram, Health checks | No Insulin, Exception, standard day report. |
| Ability to set meter | Yes | No |
{5}
Page 6 of 7
| Items | IN TOUCH DIABETES Management Software | TysonBio Link Health Management System |
| --- | --- | --- |
| clock to a specific date and time | | |
| Ability to email report from PC directly from program | Yes | No |
| Ability to input additional information on manual result | Yes | No |
I. Standard/Guidance Document Referenced (if applicable):
None
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
see above associated devices k062829, k072854 and k073492
The accuracy of PC download feature was investigated in k073492. There is a 300 test memory capacity in DIACHEX+ INFINITY Blood Glucose Meter. The PC download feature was validated by installing the TysonBio Link Health Management System software, setting meter in PC transfer data mode and connecting to computer through RS 232 port. The accuracy of transfer data was validated by comparison to the original data in DICHEX+ INFINITY device.
b. Precision/Reproducibility:
see above associated devices k062829, k072854 and k073492
c. Linearity:
see above associated devices k062829, k072854 and k073492
d. Carryover:
Not Applicable
e. Interfering Substances:
see above associated devices k062829, k072854 and k073492
2. Other Supportive Instrument Performance Data Not Covered Above:
125 typical diabetic users ranging in age, education and about equally divided between males and females were included in the user study. Participants were first provided with a DIACHEX Basic or DIACHEX Superb glucose meter (with previously stored test data), cable and both the meter and software User's manuals. They were asked to read the user's
{6}
manual and follow the instructions to install the software to transfer the data from meter to PC and use all the software application functions. The statistical analysis of study participants' response to consumer questionnaire indicated the software is easy to use and met the expectation of end users.
Page 7 of 7
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.