K081678 · Kls Martin GmbH + Co. KG · GEI · Jul 29, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081678
Device Name
CONMED BEAMER SYSTEM C600
Applicant
Kls Martin GmbH + Co. KG
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 29, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ConMed Beamer System CE600 with Accessories is intended to deliver electrosurgical current and Argon gas for the cutting, coagulation and argon beam assisted coagulation on tissue when used in conjunction with compatible applicators or probes.
Device Story
The ConMed Beamer System CE600 is an electrosurgical unit (ESU) comprising the Beamer Mate CE200 generator and Beamer Plus CB200 argon gas supply unit. The system delivers high-frequency (HF) electrical current for monopolar and bipolar cutting and coagulation. The CB200 unit regulates argon gas flow for argon beam-assisted coagulation. The system is operated by surgeons in clinical settings using a color TFT display, rotary switch, and channel selection buttons to manage power, current modes, and activation sources. HF output is triggered via finger switches or footswitches. The system includes a pressure reducer, cart, and various cables. The ESU monitors return electrodes for safety. By providing precise control over electrical waveforms and gas flow, the device assists surgeons in achieving hemostasis and tissue cutting, potentially reducing blood loss and improving surgical outcomes.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison of technological characteristics and performance specifications to predicate devices.
Technological Characteristics
Electrosurgical generator (monopolar/bipolar) and argon gas supply unit. Features color TFT display, rotary switch navigation, and return electrode monitoring. Outputs: 2 monopolar, 2 bipolar. Connectivity: standard 3-pin and Bovie-plug monopolar, 2-pin bipolar. Gas supply: argon cylinder with pressure reducer (CGA580 connector). System is non-sterile and reusable. Software is program-based with user-definable settings.
Indications for Use
Indicated for cutting, coagulation, and argon beam-assisted coagulation of tissue during surgical procedures using compatible applicators or probes. Intended for prescription use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
# KO41678
# 510(k) Summary
Page 1 of 5
# 5 510(k) Summary
Submitted by: KLS Martin GmbH + Co. KG Am Gansacker 1b 79224 Umkirch Germany
JUL 2 9 2008
Contact Person: Dr. Bernhard Hug
Date Prepared: April 23 2008
Common Name: Electrosurgical Generator (ESU) System and Argon Beamer System
# Trade/Proprietary Name:
Conmed Beamer System CE600
# Classification Name:
Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400)
Product Code: 79GEI
### Legally Marketed Predicate Devices:
Erbe VIO 300D ( K023886 & K060484) and Erbe VIO APC2 (K024047) and Erbe APC 300 (K963189) and Erbe APC Connector Hose (K013348) and Erbe ICC 200 (K933157) and Sutter Electrosurgical Cables (K073450 )
# Device Description
The Conmed Beamer System CE600 with Accessories is an electrosurgical system that uses high frequency (HF) electrical current waveforms to cut and/or coagulate tissue whereby coagulation can be assisted by Argon gas. The system comprises of the following components:
# Electrosurgical unit (ESU) Beamer Mate CE200
The Beamer Mate CE200 is an electrosurgery generator for monopolar and bipolar application. It features two monopolar and two bipolar outputs that can be set to output power, current mode and activation source independently. For each output, these settings can be provided for cutting mode and coagulation mode in separate arrays, called channels, on a color TFT display. This TFT display provides the user with actual settings and operational information of all four outputs and enables operation of the device by navigation in menus on the display using channel selection buttons, a MENU button and a rotary switch with axial pushbutton function. The device is intended as a program-based HF device. The manufacturer delivers the unit with exemplary programs for a whole range
{1}------------------------------------------------
# K081678
Page 2 of 5
# 510(k) Summarv
of different applications. However, the surgeon is free to modify these programs and store them under a new name in accordance with his/her own preferences and individual requirements. The unit has various cutting and coagulation current forms with defined grades or flexible wattage ranges to provide flexibility in interventional applications. Further features are monitoring of the return electrode, activation of HF output power by finger switches or by footswitch and service functions, accessible by Setup menu. Also, the unit can be used in association with an Argon Unit, the Beamer Plus CB200. The ESU is supplied nonsterile and is reusable.
#### Argon gas supply unit Beamer Plus CB200
The Beamer Plus CB200 enables and controls the gas flow required for the surgical process of argon beam coagulation. Terminals include argon gas inlet from the gas cylinder's pressure reducer, gas outlet for connection to the argon probe and a socket for connection with the Beamer Mate CE200 by means of CE/CB interface cable CB200-A03. Power supply and control is performed by the ESU, setting of flow rates and other beamer parameters as well as gas flow activation is done by the ESU. The only operation element at the CB200 is a "Purge" key that starts a three second's gas flow for filling a freshly connected probe with argon gas. The housings of CE200 and CB200 are matched for stacking both units with the CE200 upside. The housings are equipped with means to prevent horizontal movement of the units which enables placing on inclined surfaces without the hazard of slipping off or being torn off by connected cables or accessories. The CB200 is supplied nonsterile and is reusable.
#### Pressure reducer
The pressure reducer (CB200-A02) is an essential accessory delivered with the CB200 Beamer Unit. It drops the gas pressure of the argon gas cylinder which may be as high as 200 bar to a pre-regulated pressure in the range of about 4 bar which is input to the Argon Unit. The pressure reducer is equipped with a CGA580 connector at the high pressure side and a fast-lock connector at the beamer's side.
#### Cart CC200
The cart is intended as the system's carrier with a top plate to pick up the stacked ESU and Beamer Unit as well as the ESU alone. Inside the cart's body, an argon gas cylinder with pressure reducer can be placed and fixed. The cart is equipped with four swivel castors on stems protruding from the lower corners which provide mobility combined with high tilt stability. Two of four castors feature foot-tip operable brake levers to prevent random movement of the cart. A non-detachable just by the help of a tool) mains cord feeds the mains power through the cart to two cables with mains plug which allow mains supply of two units assembled at the cart, i.e. the ESU and a further unit that requires mains power. Further properties are three terminals for common ground connection, a basket for picking up accessories, a standard rail for infusion holders or else, a pickup for the foot switch and a handle for moving the cart.
#### Footswitches
A double footswitch (CE200-A01) is an essential accessory delivered with the ESU. It enables activation of cutting mode HF output power by pressing the yellow pedal or activation of coagulation mode HF output power by pressing the blue pedal. Allocation of the footswitch to one of the HF outputs is done in the main menu of the output channels. A single footswitch (CE200-A02) is optional delivered with the ESU.
{2}------------------------------------------------
K081678
# 510(k) Summary
Page 3 of 5
# Reusable Cable for Neutral Electrodes (CE200-A04)
The Reusable Cable for Neutral Electrodes (CE200-A04) connects disposable return electrodes (neutral electrode, patient plate) attached to the patient's skin to the ESU. This cable is intended for use with disposable neutral electrodes designed for cable clip connection. This cable belongs to the Applied Part of the system, is reusable and non sterile.
# Connecting Cable for flexible argon probes
The Connecting Cable for flexible Argon Probes (CB200-A01) is an essential accessory in combination with the Argon Unit to connect argon assisted instruments with the ESU and with the Argon Unit which means that it features an electric conductor for HF power and a hose as guidance for the argon gas. The connecting cable is provided non sterile and is reusable. The cleaning and sterilization processes have been validated and are provided in the Instructions for Use to the user.
### Intended use
The ConMed Beamer System CE600 with Accessories is intended to deliver electrosurgical current and Argon gas for the cutting, coagulation and argon beam assisted coagulation on tissue when used in conjunction with compatible applicators or probes.
# Similarities and Differences of the Device to the Predicate Device
# Electrosurgical unit (ESU) Beamer Mate CE200:
#### Similarities to Predicate Device
Both devices are intended for the same use. The handling of both devices is based on programs, both units use a color TFT display to show actual settings of the output channels and to manage the "background" settings in menus. Both units feature channel selection buttons to select a channel for being set. Both units use finger switches or foot switches for HF power activation. In both systems, the foot switch pedals can be allocated to any output. Both devices feature monopolar cutting currents with low, medium and to any currostasis during cutting and feature special current modes for gastrointestinal procedures where cutting phases alternate with coagulation phases. Both devices feature monopolar coagulation current modes for soft coagulation, for preparation and for noncontact coagulation with high voltage what is a prerequisite to generate a consistent argon beam. The output characteristics (Power, Impedance Matching, Voltages and Frequencies) are equivalent and same effective. Both devices feature bipolar cutting and coaqulation currents. Return electrode monitoring of single-plate and dual-plate electrodes is possible with both units.
#### Differences to Predicate Device
The CE200 features two fixed combo sockets M1 and M2 for monopolar current which accept the three-pin connector which is most common in the US market as well as the "Bovie-Plug", and two fixed sockets for bipolar current for connection of two-pin connectors with 28mm separation. The upper bipolar socket B1 is a combo socket which accepts a coaxial 12mm plug also. The lower bipolar socket B2 accepts the two-pin connector and
{3}------------------------------------------------
# KO81678
Page 4 of ⑤
# 510(k) Summary
features two additional 2mm sockets for bipolar finger switch activation and bipolar instrument recognition. The Predicate Device uses three modular sockets in interchangeable drawers which may be monopolar, bipolar or both as well. Channel selection at the CE200 is done by pressing the yellow key of an output to set or change the cutting properties of this output or by pressing the blue key of this output to set or change its coagulation properties. During this procedure, the settings of all other channels are displayed simultaneously which gives an overall overview of the system's setting. Display of settings of a channel which is not used may be blanked by pressing the channel selection button and keeping it pressed for two seconds. With the Predicate Device, output selection is done by pressing the selection button at the output. The screen then displays the properties of this selected channel only, selection of a certain property is done by touch pads which are embedded in the frame of the display unit. With the CE200, changing of values and log-in is done by means of a rotary switch with axial push function while the Predicate Device uses three pushbuttons for the same purpose.
#### Argon gas supply unit Beamer Plus CB200:
#### Similarities to Predicate Device
Both devices generate a controlled argon gas flow which occurs on activation of an argon coagulation current at the ESU. Both beamer systems are designed to operate with their dedicated ESU only since power supply, flow activation and flow control is done by the ESU. Argon beamer system state is displayed on the ESU in both systems. Both systems feature a "Purge" key to remove residual air out of an argon probe and flush the probe with argon qas.
# Differences to Predicate Device
ESU CE200 and Argon Unit CB200 are connected by means of a connection cable that supplies auxiliary power and control signals from the ESU to the Argon System. HF-power is not fed through the Argon Unit. With the Predicate Device, the connection between both system components is done by case-to case connectors which are incorporated into the bottom of the ESU and the top of the argon unit. This connection serves also as a HF power feedthrough from the ESU to the argon unit. With the system CE200 and CB200, the connection of the argon probe is done at one of the monopolar outputs M1 or M2 of the CE200 using the standard three-pin connector, and at the male luer-lock gas outlet of the CB200. With the predicate Device, electric and gas connection is done at the argon unit using a special multifunctional plug and a female luer-lock gas outlet.
### Pressure reducer:
### Similarities to Predicate Device
The pressure reducer CB200-A02 is the same model (same technology, same materials, same characteristics) from the same manufacturer (Linde company) as the Predicate Device.
# Cart CC200:
#### Similarities to Predicate Device
Functionality of both carts is the same.
#### Differences to Predicate Device
The cart CC200 has a mains cord which is detachable only by use of a tool. The Predicate Device has a detachable mains cord and outlet sockets for mains power instead of cables so that additional connection cables are required unless the mains cord is connected to the ESU directly.
{4}------------------------------------------------
# 510(k) Summary
Page 5 of 5
# Connecting Cable for flexible argon probes Similarities to Predicate Device
Both - the new and the predicate Connecting Cables have the same intended use, principles of operation and technological characteristics. They are non-sterile and reusable. The connector receptable for the flexible argon probes is compatible to the Conmed CET Argon Probes. The products have no patient contact.
The conecting cable has the same dimensions in length and diameter.
### Differences to Predicate Device
The Gas Connector of both devices is Luer-Lock, whereas the Predicate is a male connector and the New Device has a female connector. Within the New Device the HF-Plug is a Standard 3 pin-plug, whereas the Predicate has a special designed plug. The minor differences do not affect the safety or effectiveness of the product.
# Conclusion
The Conmed Beamer System with its component CE200 shows same performance of the current modes in comparison with the Predicate Device and thus represents good substantial equivalence to the Predicate Device.
The fact that the HF power is not fed through the CB200 Argon Unit as it is the case with the Predicate Device does not affect the clinical properties at all. Since all other technical details are the same, the CB200 with accessories represents also substantial equivalence to its predicate Device.
The CC200 cart shows same properties as its Predicate Device. Minor differences do not affect the clinical properties of the system.
Thus, the system CE600 with accessories raises no new issues of safety or effectiveness. e proventified in the status the first of the learning and the many of the many
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines representing its wings and tail. To the left of the eagle is a circular inscription that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA".
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 29 2008
KLS Martin GmbH + Co. KG % TÜV SÜD America, Inc. Mr. Norbert Stuiber Third Party Reviewer 1775 Old Highway 8 Northwest New Brighton, Minnesota 55112
Re: K081678
Trade/Device Name: ConMed Beamer System CE600 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 11, 2008 Received: July 14, 2008
Dear Mr. Stuiber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Norbert Stuiber
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for KLS martin, with the words "KLS martin" in a bold, sans-serif font. Below the logo, the text "510(k) submission Conmed BeamerSystem CE600" is written in a smaller, regular font. The logo is black and white, and the text below is also black and white.
#### Indications for Use Statement র্ব
510(k) Number (if known):
78
Device Name: Conmed Beamer System CE600.
Indications For Use:
The ConMed Beamer System CE600 with Accessories is intended to deliver electrosurgical current and Argon gas for the cutting, coagulation and argon beam assisted coagulation on tissue when used in conjunction with compatible applicators or probes.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Melleman
(Division Sign-C Division of General, Restorative, and Neurological Devi
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.