← Product Code [KPI](/productcode/KPI) · K081480

# KEGEL8 PELVIC EXERCISER, MODEL OPH 400 (K081480)

_Ne Services , Ltd. · KPI · Sep 12, 2008 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081480

## Device Facts

- **Applicant:** Ne Services , Ltd.
- **Product Code:** [KPI](/productcode/KPI.md)
- **Decision Date:** Sep 12, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5320
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The 'Kegel8' Pelvic Muscle Trainer is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvice floor muscles for the treatment of stress urge and mixed urinary incontinence in women Federal (USA) law restricts this device to sale by or on the order of a physician

## Device Story

Kegel8 is a battery-powered, dual-channel neuromuscular electrical stimulation (NMES) device. It utilizes a vaginal probe with two electrodes to deliver electrical stimulation to pelvic floor muscles. The device is designed for neuromuscular re-education to rehabilitate weak muscles and treat urinary incontinence. It is operated by the user or clinician. The device provides electrical pulses to induce muscle contraction, aiding in the strengthening of the pelvic floor. It is intended for use in clinical or home settings under the order of a physician.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Dual-channel neuromuscular electrical stimulation device; battery-powered; includes vaginal two-electrode probe; connects via cable and plug. Class II device (21 CFR 876.5320).

## Regulatory Identification

A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).

## Predicate Devices

- Athena Pelvic Muscle Trainer II ([K033256](/device/K033256.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## NE Services Ltd 14D iPark Innovation Drive HUS ISG UK

### 510(k) Summary

#### General Information 1.

| Trade Name of Device:        | 'Kegel8'                                                                                                               |
|------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:           | Pelvic muscle trainer                                                                                                  |
| Classification Name:         | Non-implanted electrical continence device                                                                             |
| Submitters Name and Address: | NE Services Ltd<br>14D iPark<br>Innovation Drive<br>HU5 1SG<br>UK<br>Tel +44 (0) 1482 873377<br>Fax+44 (0) 1482 873570 |
| Manufacturer:                | Mantra International (HK) Ltd<br>Registration number 3003741750                                                        |

#### 2. Device Description

The 'Kegel8' Pelvic Muscle Trainer is a small lightweight battery powered dual channel neuromuscular stimulation device supplied with a vaginal two electrode stimulation probe

The probe connects to the control unit by cable and plug The 'Kegel8' is intended to provide electrical stimulation and neuromuscular reeducation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women

#### 3. Indications for Use

The 'Kegel8' Pelvic Muscle Trainer is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress urge and mixed urinary incontinence in women

4. Substantial Equivalence The 'Kegel8' Pelvic Muscle Trainer is substantially equivalent to: Athena Pelvic Muscle Trainer II (K033256).

Kog1480 Page

{1}------------------------------------------------

### 5. Performance Studies

Performance testing was conducted on the Kegel8 Pelvic Muscle Trainer to demonstrate the integrity, suitability and substantial equivalence of the device

#### 6. Conclusion

Based upon the Indications for use and performance studies Kegel8 has been shown to be substantially equivalent for its intended use

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services (USA). The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 2 2008

NE Services Ltd. % Mr. Brent Reider President International Trade Group, Inc. 4663 Kate Lane OXFORD OHIO 45056

Re: K081480

Trade/Device Name: 'Kegel8' Pelvic Muscle Trainer Regulation Number: 21 CFR 876.5320 Regulation Name: Non-implanted electrical continence device Regulatory Class: II Product Code: KPI Dated: August 28, 2008 Received: September 5, 2008

Dear Mr. Reider:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-0100 |

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

loque M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K081480

Device Name: 'Kegel8'

Indications For Use:

The 'Kegel8' Pelvic Muscle Trainer is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvice floor muscles for the treatment of stress urge and mixed urinary incontinence in women

Federal (USA) law restricts this device to sale by or on the order of a physician

Prescription Use X (21 CFR 801 Subpart D) _ AND/OR Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Hubert Lemon

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number _

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K081480](https://fda-staging.innolitics.com/device/K081480)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).