OLYMPUS CHF TYPE V

K081456 · Olympus Medical Systems Corporation · FBN · Nov 6, 2008 · Gastroenterology, Urology

Device Facts

Record IDK081456
Device NameOLYMPUS CHF TYPE V
ApplicantOlympus Medical Systems Corporation
Product CodeFBN · Gastroenterology, Urology
Decision DateNov 6, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bile duct.

Device Story

The CHF-V is a flexible choledocho videoscope used for endoscopic diagnosis and treatment within the bile duct. It utilizes a charge-coupled device (CCD) to capture optical images, which are transmitted to an external Olympus video system center and displayed on a monitor for the physician. The device features a 120-degree field of view and 160/130-degree up/down angulation. It is designed for use with ancillary equipment, including light sources and EndoTherapy accessories. The physician operates the scope to visualize the bile duct, facilitating diagnostic or therapeutic interventions. The video output allows for real-time visualization, aiding clinical decision-making during endoscopic procedures.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Flexible video endoscope with color CCD optical system. Field of view: 120°. Direction of view: 0°. Angulation: 160° up / 130° down. Working length: 380mm. Instrument channel diameter: 2.0mm. Connectivity: Compatible with Olympus video system centers. Sterilization: Not specified.

Indications for Use

Indicated for endoscopic diagnosis and treatment within the bile duct in patients requiring such procedures. Intended for use by clinicians in a professional healthcare setting.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # OLYMPUS . # 510(k) SUMMARY NOV - 6 2008 ## 1. General Information | ■ | Applicant: | | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan<br>Establishment Registration No: 8010047 | |---|-----------------------------------|-------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ■ | Official Correspondent: | | Laura Storms-Tyler<br>Regulatory Affairs & Quality Assurance<br>Olympus America Inc.<br>3500 Corporate Parkway, PO Box 610<br>Center Valley PA 18034-0610<br>Phone: (484) 896-5688<br>Facsimile: (484) 896-7128<br>Email:Laura.storms-tyler@olympus.com<br>Establishment Registration No: 2429304 | | ■ | Manufacturer: | | | | | Light source/Video system center: | | SHIRAKAWA OLYMPUS CO., LTD.<br>3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura<br>Nishishirakawa-gun, Fukushima, Japan 961-8061<br>Establishment Registration No: 3002808148 | | | Choledocoscope: | | Aizu Olympus Co., Ltd.<br>500 Aza-Muranishi, Ooaza-Iidera, Monden-cho,<br>Aizuwakamatsu-shi, Fukushima, Japan 965-8520<br>Establishment Registration No.: 9610595 | | ■ | Date Prepared: | April | 2008 | ## 2. Device Identification : | • | Device Name: | CHF-V | |---|-------------------------|-------------------------------------------------------------------------------------------| | • | Common Name: | CHOLEDOCHO VIDEOSCOPE | | • | Class: | II | | • | Regulation Number/Name: | 876.1500 Endoscope and accessories | | • | Product Code: | NWB - Endoscope, accessories, narrow band spectrum<br>FBN - Choledochoscope, accessories, | | • | Classification Panel: | Gastroenterology/Urology | OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-42-642-2694, TELÉFAX +81-42-642-2307 {1}------------------------------------------------ # APUS # 3. Legally Marketed Device to which Substantial Equivalence is Claimed The following table shows the subject device and the predicate device to which we clalm substantial equivalence. #### Table 15-1: Primary Component & Predicate Device | Subject Device<br>(Part of this submission) | Predicate Device | PD's 610(k) No. | |---------------------------------------------|----------------------------------------------------|-----------------| | CHOLEDOCHO VIDEOSCOPE<br>CHF-V | OES Nephroscope / Cystoscope<br>Fiberscope CHF-P10 | K843084 | ## 4. Device Description The CHF-V choledochoscope is a flexible video endoscope used for endoscopic diagnosis and treatment within the bile duct . The CHF-V choledochoscope is basically identical to the predicate device, CHF Type P10 OES Nephroscope/Cystoscope, in intended use, spacifications, performance. The optical system of the CHF-V Is a charge coupled device (CCD) based system, allowing endoscopic image display on a video monitor. The new endoscope is basically identical to each predicate device shown in Table 15-1 In intended use, and similar in specifications, performance and materials. ### 5. Indications for Use #### CHOLEDOCHO VIDEOSCOPE CHF-V This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bile duct. > OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-42-642-2694, TELEFAX +81-42-642-2307 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are large and black, creating a strong visual impact. A thin line is present underneath the word, emphasizing the brand name. ## 6. Comparison of Technological Characteristics The CHF-V is similar to the predicate device CHF-P10 in specifications except for the material and optical system. Comparison between the subject and predicate devices is shown below. | Device: OES Nephroscope / Cystoscope Fiberscope CHF-P10 (K843084) | | | |-------------------------------------------------------------------|--------------------------|----------------------------------| | Specifications | Control Unit (VRC CHF-V) | Rigid Device (CHF-P10) (K843084) | | Field of View | 120° | 90° | | Direction of View | 0° | 0° | | Optical System | Color CCD | Image guide fiber bundle | | Angulation<br>Up / Down | 160° / 130° | 160° / 130° | | Working Length | 380mm | 330mm | | Inner Diameter of<br>Instrument Channel | φ2.0mm | φ2.0mm | #### Table 15-2. Comparison of Specifications Subject Device: CHOLEDOCHO VIDEOSCOPE CHF-V Predicate ### 6. Conclusion When compared to the predicate device, the CHF-V does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device. {3}------------------------------------------------ Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and three human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the symbol. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Olympus Medical Systems Corp. % Ms. Laura Storms-Tyler, Vice President Regulatory Affairs & Quality Assurance Olympus America Incorporated 3500 Corporate Parkway P.O. Box 610 CENTER VALLEY PA 18034-0610 Re: K081456 Trade Name: CHF-V Choledocho Videoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Codes: FBN Dated: October 6, 2008 Received: October 7, 2008 Dear Ms. Storms-Tyler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. NOV - 6 2008 {4}------------------------------------------------ Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html. Sincerely yours, loque M. Whang Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 08/456 510(k) Number (if known): Device Name: CHF-V CHOLEDOCHO VIDEOSCOPE Indications For Use: #### CHOLEDOCHO VIDEOSCOPE OLYMPUS CHF TYPE V This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bile duct. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Arsuththethy (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number. K081456
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...