RAYNER SINGLE USE SMALL INCISION DISPOSABLE INJECTOR, MODEL R-INJ-06
Applicant
Rayner Surgical, Inc.
Product Code
MSS · Ophthalmic
Decision Date
Sep 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Indications for Use
The single-use small incision disposable injector (model number R-INJ-06) is intended to compress and insert into the capsular bag only those IOL models that allow use of this injector in their approved labeling.
Device Story
Single-use, disposable plastic injector for small-incision cataract surgery. Device components (barrel, flap, nozzle, bush, sleeve) made of polypropylene; plunger made of polycarbonate. Surgeon manually loads IOL into injector loading bay in 'reverse-S' configuration; applies viscoelastic lubricant. Surgeon advances plunger to mechanically fold lens and inject it through nozzle into capsular bag. Device facilitates controlled delivery of IOL without damaging optic or haptics. Used in clinical settings by ophthalmologists. Benefits include consistent, safe IOL insertion through small incisions.
Clinical Evidence
Bench testing only. Included biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity per ISO 10993), visual/optical/mechanical testing of injected IOLs (MTF, dioptric power, spectral transmittance, haptic recovery, fatigue, pull strength), and packaging integrity (sterility, dye penetration, burst test). Testing demonstrated the injector delivers IOLs without compromising lens integrity or optical performance.
Technological Characteristics
Single-use, manual mechanical injector. Materials: Polypropylene (barrel, flap, bush, nozzle, sleeve) and Polycarbonate (plunger). Dimensions/form factor: Handheld syringe-type. Sterilization: Ethylene Oxide (EO). No electronic components or software.
Indications for Use
Indicated for use in patients undergoing small-incision cataract surgery requiring the insertion of an intraocular lens (IOL) into the capsular bag, specifically for IOL models labeled for use with this injector.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Rayner Single Use Soft Tipped Disposable Injector R-INJ-04 (K062512)
Reference Devices
Rayner Single Use Disposable Injector R-INJ-02 (K052651)
Submission Summary (Full Text)
{0}------------------------------------------------
K081455
6654 Church Street Los Angeles, California 90042-1555 USA
Telephone: 1-800-905-0143
### SEP - 5 2008
# 510(k) Summary
## Submitted by
Rayner Intraocular Lenses Ltd.
Andrew Wells Quality Assurance & Regulatory Affairs Manager
Sackville Road Hove, East Sussex, BN3 7AN England
Telephone # 0044-1273-205 401 Fax # 0044-1273-324 623
Summary prepared on May 12th 2008
## Device Name
- Trade/Proprietary Name: Rayner Single Use Small Incision Disposable Injector R-INJ-06 ●
- Common Name: Rayner Single Use Small Incision Injector ●
- Classification Name: Product Code is MSS. CRF section is TITLE 21, Part 886, Subpart E, Sec. 886.4300 . Intraocular lens guide. Device class is Class I. Classification Panel is Ophthalmic
## Information on devices to which substantial equivalence is claimed
- 510(k) Number: K062512 ●
- Trade or Proprietary or Model Name: Rayner Single Use Soft Tipped Disposable Injector R-INJ-04 �
- . Manufacturer: Rayner Intraocular Lenses Ltd.
## Intended Use
The single-use small incision disposable injector (model number R-INJ-06) is intended to compress and insert into the capsular bag only those IOL models that allow use of this injector in their approved labeling.
Page 1 of 5
Image /page/0/Picture/22 description: The image shows the number 000017. The number is written in a simple, sans-serif font. The digits are evenly spaced and clearly visible. The number is likely a serial number, code, or identifier.
{1}------------------------------------------------
Description of the device that is subject to of the application, including an explanation of how the device functions, basic scientific concepts, scientific physical and performance characteristics (design, material, physical properties)
The single-use small incision disposable injector (model number R-INJ-06) is intended to compress and insert into the capsular bag only those IOL models that allow use of this injector in their approved labeling. It is designed to mechanically fold the lens and insert it into the eye during normal, small-incision cataract surgery.
INJECTOR LOADING is described as follows: aseptically transfer the sterile field by tipping it from the tray. Fully retract the plunger ensuring that the tip does not protrude into the loading bay. Open the loading bay fialy to 90° and sparingly apply a commercially available viscoelastic inside the nozzle and to both grooves of the loading bay. Balanced salt solution alone should not be used as a lubricant. Carefully peel back the foil lid of the lens blister. Gently lift out the lens using parallel tipped, non-serrated forceps e.g. Kelman-McPherson. Rinse the IOL with sterile balanced salt solution. Position the lens in the loading bay in a "reverse - S" configuration. Ensure that the nearest edge of the optic is securely under the edge (lip) of the barrel. Hold open the flap and press down on lens with closed forceps to ensure the furthest edge of the optic is securely under the edge (lip) of the flap. Ensure the haptics are completely within the loading bay. While keeping the lens in position with open forceps, gently close the flaps of the injector ensuring that no part of the optic or haptics is trapped, before locking the flaps firmly together. Visually observe that the lens is symmetrically folded within the loading bay. Advance the plunger in a slow controlled manner. Anticipate an initial slight resistance, Excessive resistance could indicate a trapped lens. Observe that the lens stays symmetrically folded within the nozzle. When the lens exits the nozzle, stop depressing the plunger is only to be depressed once.
The small incision injector (model number R-INJ-06) is a plastic, single-use disposable device.
The injector components barrel, flap, nozzle, bush and sleeve are made of polypropylene. The plunger is made of polycarbonate. The injector is transparent and the plunger is white.
{2}------------------------------------------------
## Summary of how the technological characteristics of the device compare with the predicate device identified - Device comparison table:
| Characteristics | Rayner Single Use Disposable<br>Soft Tipped Injector R-INJ-04 | Rayner Single Use Small<br>Incision Disposable Injector<br>R-INJ-06 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Intended Use | Folds and delivers IOL into eye<br>during normal small incision<br>cataract surgery | Same |
| Operating Principle | - Load IOL into the inserter<br>mechanically and insert IOL<br>into the eye<br>- IOL delivered by direct<br>forward motion applied to a<br>syringe type plunger | - Same |
| Folding Operation | IOL is loaded into cartridge and<br>closed. Opposing contact edges<br>are folded towards each other | Same |
| Folding Direction of the Lens | Lens decompresses in a<br>horizontal plane | Same |
| Cartridge design | None | None |
| Sterilization Method | EO for entire device | Same |
| Materials | Polypropylene barrel, flap, bush<br>and nozzle. The plunger is<br>polypropylene with a<br>thermoplastic elastomer soft tip. | Polypropylene barrel, flap, bush,<br>nozzle and sleeve. The plunger is<br>polycarbonate. |
| Surface Treatment | None | None |
| Patient contact portion of the<br>device | Nozzle and plunger tip | Same |
{3}------------------------------------------------
## Non-clinical performance data - discussion and conclusions
Substantial equivalence is based on the assessment of non-clinical performance data
More specifically this contains the following information:
- Biocompatibility testing on the injector .
- Visual, optical and mechanical testing on injected IOL. t
- Visual and mechanical testing on injector.
- Packaging performance testing .
The performance data indicates that the Rayner Single Use Disposable Soft Tipped Injector R-INJ-06 delivers those IOL models, that allow use of this injector in their approved labeling, without significantly impacting the optical performance, the dimensions or the cosmetic appearance of the lens.
The following series of tests were conducted with the injection media Balanced Salt Solution (BSS) and a viscoelastic currently approved and used on the US market.
#### a) Biocompatibility testing on the injector
Biocompatibility testing on the injector was undertaken on the nozzle and plunger tip as these components incorporate the materials in contact with the tissues of the injector components barrel, flap, bush and sleeve are made from the same polypropylene material as the nozzle was tested in a previous version of the single use disposable injector (model R-INJ-02 / K052651). The plunger is a polycarbonate material which was tested in a previous version of the single use disposable injector (model R-INJ-02 / K052651. The injector is transparent and the plunger is white. Using the scheme as outlined in ISO 10993-1 and the US Blue Book Memorandum G95-1 the following tests have been undertaken Cytotoxicity (Quantitative Growth Inhibition Test (ISO 10993-5), Maximization Test according to Magnusson and Kligman (ISO 10993-10) Intracutaneous Reactivity (ISO 10993-10) and Acute Systemic Toxicity (ISO 10993-11). Testing on the final packaged and terminally sterilized Single Use Disposable Injector shows the materials to be biocompatible/toxicologically safe for the intended clinical application (limited exposure duration with the device in contact with a breached/compromised surface).
#### b) Visual, optical and mechanical testing on injected IOL
#### VISUAL TESTING
Observation at magnification under optimal lighting conditions for the following:
- . No optic lens tears for properly loaded lenses.
- . No haptic damage.
- t Absence of 'Fold lines' and/or deposits/debris on the lens surface.
- Evaluation of haptic fixation recovery time, to 11 mm diameter dimension. OPTICAL TESTING
- . Modulation Transfer Function
- . Dioptric power
- . Spectral transmittance
- MECHANICAL TESTING
- Dimensions .
- . Lens sagittal dimension
- . Lens overall diameter
- . Haptic compression force
Page 4 of 5
000020
{4}------------------------------------------------
- . Dynamic fatigue durability
- Limb/loop pull strength .
- Optic decentration .
- . Optic tilt
- Axial displacement .
The aforementioned mechanical testing is an assessment of the haptic function. Dioptric power does not affect the property tested. Therefore testing as per FDA IOL Guidance document Oct 10th 1997, was performed on 10 lenses each of both the highest and lowest powers.
Testing was carried out as per FDA guidelines in that the lens is folded for a minimum of 3 minutes. The IOL was allowed to return to its original and designed configuration. Compliance with applicable mechanical and optical requirements was demonstrated at 24+/-2 hours post folding/injection. Reference ISO 11979-3, section 4.1 & CDRH IOL Guidance Document Oct 10th 1997).
#### c) Visual and mechanical testing on the single use disposable injector
#### VISUAL/PRODUCTION
- Surface finish & dimensional check .
#### MECHANICAL
- Nozzle tip detachment from barrel/main body of injector .
#### d) Packaging performance testing
The following tests were performed:
- Sterility test .
- Dye penetration .
- Burst Test .
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a person with outstretched arms.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 5 2008
Rayner Surgical, Inc. c/o Mr. Andrew Wells Rayner Intraocular Lenses 6654 Church Street Los Angeles, CA 90042-1555
Re: K081455
Trade/Device Name: Rayner Single Use Small Incision Disposable Injector R-INJ-06 Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I Product Code: MSS Dated: August 8, 2008 Received: August 13, 2008
Dear Mr. Wells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 ) CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electromic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egleston, m.d.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Rayner Surgical Inc.
### 510k submission
# Indications for Use
510(k) Number (if known): K 08 14 5 5
Device Name Rayner Single Use Small Incision Disposable Injector R-INJ-06
-
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Indications For Use:
## Statement of Indications for use
"The single-use small incision disposable injector (model number R-INJ-06) is intended to be used to compress and insert into the capsular bag only those IOL models that allow use of this injector in their approved labeling "
Prescription Use \ / Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No. No. -(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Widnets Virran 9/3/08 Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic Ear, Division Throat Devises 510(k) Number_KO&THE 5
Page 1 of 1
May 12th 2008
000016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.