ADVANCED MEDICAL INNOVATIONS DISPOSABLE TROCARS AND LAPAROSCOPIC ACCESSORIES
K081398 · Advanced Medical Innovations, Inc. · GEI · Feb 6, 2009 · General, Plastic Surgery
Device Facts
Record ID
K081398
Device Name
ADVANCED MEDICAL INNOVATIONS DISPOSABLE TROCARS AND LAPAROSCOPIC ACCESSORIES
Applicant
Advanced Medical Innovations, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 6, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Advanced Medical Monopolar Scissors, Maryland Dissecting Forceps, Lap Clinch, Rat Tooth LapClinch, Babcock, Straight Grasping Forceps with Ratchet are indicated for use in gynecological and general endoscopic procedures for mobilization and transection of tissues. The device with a 5mm diameter insulated shaft has a male cautery connection on top of the handle and may be used for monopolar cautery when attached to standard cautery cables and their generators. The Advanced Medical Reducer has application in gynecologic laparoscopy and other abdominal procedures and is designed to be used with the Advanced Medical Disposable Surgical Trocar to allow the surgeon to insert undersized instruments through the Trocar without losing insufflation. The Advanced Medical Laparoscopic Smoke Filter is intended for the filtration of contaminants from smoke, generated during laparoscopic surgery, which can contain volatile organic products of combustion (including those which are perceived as odor) and particulates (including cellular debris, bacteria and viruses). The Advanced Medical Hasson Trocar and Dilating Tip Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments. The Advanced Medical Shielded Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments. Instruments provided with a stopcock are intended for insufflation of the operative space when the trocar is in place. The Advanced Medical Secondary Trocar and Sleeve with Insufflation Port are manual surgical instruments used to support a cut down (lap approach). The Secondary Trocar is used in endoscopic surgery (abdominal and thoracic) for incision and peritoneal access for positioning of the hollow sleeve. Once the trocar is removed, the port of entry provided by the sleeve, through the cap, is used with manual surgical instruments, endoscopic instruments, laparoscopes and probes. There is a seal on the cap which closes the port of entry. The Advanced Medical Stainless Steel Veress Needle is for the initial induction of pneumoperitoneum. The Advanced Medical Suction/Irrigation Set is intended for use in endoscopic procedures such as laparoscopy, pelviscopy, and thorascopy, for suctioning of large clots, retrieving spilled stones, high flow irrigation and hydro-dissection.
Device Story
Disposable laparoscopic instruments and trocars; includes scissors, forceps, dissectors, suction/irrigation sets, Veress needles, and smoke filters. Used in OR by surgeons for tissue manipulation, cautery, and access. Instruments feature insulated shafts with male cautery connectors for monopolar electrosurgery. Trocars provide access ports for endoscopic tools; reducers maintain insufflation. Smoke filters remove combustion byproducts and particulates. Devices are single-use, sterile, and biocompatible. Surgeon operates instruments manually; output is physical tissue effect or access to body cavity. Benefits include reduced cross-contamination risk via single-use design and efficient surgical access.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety, EMC, and performance testing per IEC 60601-2-2:2006.
Technological Characteristics
Biocompatible materials; insulated shafts; male cautery connectors for monopolar electrosurgery; gamma sterilization per AAMI/ISO 11137 (SAL 10^-6); manual operation; disposable/single-use. Compliant with IEC 60601-2-2:2006.
Indications for Use
Indicated for patients undergoing gynecological, general, thoracic, or abdominal endoscopic procedures requiring tissue mobilization, transection, pneumoperitoneum induction, or access for endoscopic instruments.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
VECTEC Disposable Trocars and Laparoscopic Accessories (K071976)
Pall Medical Laparoscopic Smoke Filtration System (K992361)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB - 6 2009
Page 1 of 5
### 510(K) SUMMARY
This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________
#### Submitter's Identification: 1.
| Contact: | Mike Hoftman |
|---------------|-----------------------------------------------------|
| Company Name: | Advanced Medical Innovations |
| Address: | 9410 De Soto Avenue, Bldg J<br>Chatsworth, CA 91311 |
| Tel: | 818-701-7180 |
| Fax: | 818-701-9708 |
Date Prepared:
February 3, 2009
- 2. Name of the Device: Advanced Medical Innovations Disposable Trocars and Laparoscopic Accessories
These disposable devices include: Monopolar Scissors, Maryland Dissecting Forceps, Straight Grasping Forceps, Straight Grasping Forceps with Ratchet, Lapclinch with Ratchet, Rat Tooth Lapclinch with Ratchet, Babcock, Suction Irrigation Set, Trocars (Auto Shield, Hasson, Secondary and Dilating Tip), Veress Needles, Flip Top Reducer and Laparoscopic Smoke Filter.
- Electrosurgical, Cutting & Coagulation & 3. Common Name: Accessories 878.4400 Regulation: Product Code: GEI
Common Name: Laparoscopic, Gynecologic and accessories Regulation: 884.1720-Product Code: HET
{1}------------------------------------------------
KO 81 398
Page 2 of 5
.
1
:
.
.
i
.
#### 4. Predicate Device Information and Comparison:
:
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
. 1
.
.
| Device<br>Description | Predicate Device | 510(k)<br>Number | Comparison |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| American<br>Medical<br>Innovations<br>Disposable<br>Trocars and<br>Laparoscopic<br>Accessories | VECTEC Bioparc,<br>France. VECTEC<br>Disposable Trocars<br>and Laparoscopic<br>Accessories | K071976 | Both our subject device<br>and the predicate<br>device are similar in<br>their basic<br>characteristics. |
| Scissors,<br>Forceps,<br>Dissectors,<br>Hook,<br>Lapclinch,<br>Babcock | Applied Medical<br>Resources Corp. | K040295 &<br>K062169 | The predicate device is<br>basically identical to our<br>subject device with the<br>exception of the tip<br>design which is based<br>on physician's<br>preference |
| Suction<br>Irrigation Set | Conmed . Gleeson<br>FloVac Hand<br>controlled Suction<br>Irrigation<br>Instrument. | K940439 | The subject device(s)<br>are basically identical to<br>the predicate device<br>with respect to<br>technological<br>characteristics and<br>function. |
| Veress<br>Needles | Northgate<br>Technologies, Inc<br>Veress Needle | K971837 | The predicate device is<br>basically identical to the<br>subject device. |
| Dilating Tip<br>Trocars,<br>Autoshield<br>Trocars and<br>Hasson<br>Trocars | Ethicon Endo-<br>Surgery Inc -<br>ENDOPATH Trocar | K020428 | Both our subject device<br>and the predicate<br>device are basically<br>identical. |
| Flip Top<br>Reducer | Ethicon, Inc.<br>Endopath<br>(Disposable<br>Surgical Trocar<br>Reducer | K904907 | Both our subject device<br>and the predicate<br>device are basically<br>identical. |
| Secondary<br>Trocar | Pilling Weck<br>Secondary Trocar<br>with a plastic<br>Sleeve and<br>Insufflation Port | K964450 | Both our subject device<br>and the predicate<br>device are basically<br>identical. |
| Smoke Filter | Pall Medical<br>Laparoscopic<br>Smoke Filtration<br>System | K992361 | Both our subject device<br>and the predicate<br>device are basically<br>identical. |
{2}------------------------------------------------
K081398
Page 3 of 5
#### 5. Device Description:
The Advanced Medical Innovations Disposable Trocars and Laparoscopic Accessories can be used in a variety of procedures (general, vascular, gynecological and surgical procedures) to cut, dissect, manipulate and/or cauterize various tissues. Trocars are sterile, single-use devices that allow visualization of body cavities and use of the laparoscopic accessories.
The accessory devices are disposable, single use, individually packaged devices that are composed of biocompatible materials. Scissors, forceps, and dissectors have a handle attached to an insulated shaft with different tips, which allows the shaft and tip to rotate. They include a male cautery connector when attached to standard monopolar cautery cables and their generators. All devices are sterilized using a traditional, validated gamma procedure per AAMI/ISO International standard 11137 (EN 552) to a SAL of 10-6.
#### 6. Intended Use/Indications for Use
The Advanced Medical Monopolar Scissors, Maryland Dissecting Forceps, Lap Clinch, Rat Tooth LapClinch, Babcock, Straight Grasping Forceps with Ratchet are indicated for use in gynecological and general endoscopic procedures for mobilization and transection of tissues. The device with a 5mm diameter insulated shaft has a male cautery connection on top of the handle and may be used for monopolar cautery when attached to standard cautery cables and their generators
The Advanced Medical Reducer has application in gynecologic laparoscopy and other abdominal procedures and is designed to be used with the Advanced
{3}------------------------------------------------
K081398
Page 4 of 5
Medical Disposable Surgical Trocar to allow the surgeon to insert undersized instruments through the Trocar without losing insufflation.
The Advanced Medical Laparoscopic Smoke Filter is intended for the filtration of contaminants from smoke, generated during laparoscopic surgery, which can contain volatile organic products of combustion (including those which are perceived as odor) and particulates (including cellular debris, bacteria and viruses).
The Advanced Medical Hasson Trocar and Dilating Tip Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments.
The Advanced Medical Shielded Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments. Instruments provided with a stopcock are intended for insufflation of the operative space when the trocar is in place.
The Advanced Medical Secondary Trocar and Sleeve with Insufflation Port are manual surgical instruments used to support a cut down (lap approach).
The Secondary Trocar is used in endoscopic surgery (abdominal and thoracic) for incision and peritoneal access for positioning of the hollow sleeve. Once the trocar is removed, the port of entry provided by the sleeve, through the cap, is used with manual surgical instruments, endoscopic instruments, laparoscopes and probes. There is a seal on the cap which closes the port of entry.
The Advanced Medical Stainless Steel Veress Needle is for the initial induction of pneumoperitoneum.
The Advanced Medical Suction/Irrigation Set is intended for use in endoscopic procedures such as laparoscopy, pelviscopy, and thorascopy, for suctioning of large clots, retrieving spilled stones, high flow irrigation and hydro-dissection.
{4}------------------------------------------------
K 0 81398
Page 5-g(5)
#### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
Advanced Medical disposable trocars and laparoscopic accessories met the electrical, safety, EMC, and applicable testing requirements of IEC 60601-2-2:2006 4th Edition .
None of the testing demonstrated any design characteristics that violated the requirements of the standards or resulted in any safety hazards.
#### 8. Discussion of Clinical Tests Performed:
Clinical testing was not conducted.
#### 9. Conclusions:
Based on the information provided in this submission we conclude that the Advanced Medical Disposable Trocars and Laparoscopic Accessories are substantially equivalent to the predicates and is safe and effective for its intended use.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Public Health Service
FEB = 6 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Medical Innovations, Inc. % MDI Consultants, Inc. Ms. Maria F. Griffin 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K081398
Trade/Device Name: Advanced Medical Disposable Trocar and Laparoscopic Accessories Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II
Product Code: GEI, HET, GCJ
Dated: January 28, 2009
Received: January 29, 2009
### Dear Ms. Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Ms. Maria F. Griffin
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N. Wilkinson
- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
FEB - 6 2009
Attachment A
### Indications for Use
Page of 2
#### K081398 510(k) Number (if known):
Device Name: Advanced Medical Disposable Trocar and Laparoscopic Accessories
#### Indications For Use:
The Advanced Medical Monopolar Scissors, Maryland Dissecting Forceps, Lap Clinch, Rat Tooth LapClinch, Babcock, Straight Grasping Forceps with Ratchet are indicated for use in gynecological and general endoscopic procedures for mobilization and transection of tissues. The device with a 5mm diameter insulated shaft has a male cautery connection on top of the handle and may be used for monopolar cautery when attached to standard cautery cables and their generators
The Advanced Medical Reducer has application in gynecologic laparoscopy and other abdominal procedures and is designed to be used with the Advanced Medical Disposable Surgical Trocar to allow the surgeon to insert undersized instruments through the Trocar without losing insufflation.
The Advanced Medical Laparoscopic Smoke Filter is intended for the filtration of contaminants from smoke, generated during laparoscopic surgery, which can contain volatile organic products of combustion (including those which are perceived as odor) and particulates (including cellular debris, bacteria and viruses).
The Advanced Medical Hasson Trocar and Dilating Tip Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments.
**Prescription Use** *X*
(Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number
{8}------------------------------------------------
FEB = 6 2009
## Attachment A
### Indications for Use
Page 2 of of 22
# 510(k) Number (if known): _ KO8 | 398
Device Name: The Advanced Medical Disposable Trocar and Laparoscopic Accessories
## Indications For Use:
The Advanced Medical Shielded Trocar has application in thoracic, gynecologic laparoscopy, and other abdominal procedures to establish a path of entry for endoscopic instruments. Instruments provided with a stopcock are intended for insufflation of the operative space when the trocar is in place.
The Advanced Medical Secondary Trocar and Sleeve with Insufflation Port are manual surgical instruments used to support a cut down (lap approach).
The Secondary Trocar is used in endoscopic surgery (abdominal and thoracic) for incision and peritoneal access for positioning of the hollow sleeve. Once the trocar is removed, the port of entry provided by the sleeve, through the cap, is used with manual surgical instruments, endoscopic instruments, laparoscopes and probes. There is a seal on the cap which closes the port of entry.
The Advanced Medical Stainless Steel Veress Needle is for the initial induction of pneumoperitoneum.
The Advanced Medical Suction/Irrigation Set is intended for use in endoscopic procedures such as laparoscopy, pelviscopy, and thorascopy, for suctioning of large clots, retrieving spilled stones, high flow irrigation and hydro-dissection.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off Of Chine of General, Restorative, Ind Neurological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.