← Product Code [DRN](/productcode/DRN) · K081359

# GENESIS ECP (K081359)

_Scottcare Corporation · DRN · May 29, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081359

## Device Facts

- **Applicant:** Scottcare Corporation
- **Product Code:** [DRN](/productcode/DRN.md)
- **Decision Date:** May 29, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5225
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the ScottCare Genesis ECP device is to provide external counterpulsation (ECP) therapy, and is indicated for use in the treatment of stable or unstable angina pectoris, congestive heart failure, cardiogenic shock, and acute myocardial infarction.

## Device Story

Genesis ECP is non-invasive microprocessor-controlled system for external, sequential counterpulsation therapy. Device uses three pairs of air cuffs wrapped around patient calves, thighs, and buttocks. Microprocessor detects R-wave from ECG input, calculates heart rate, and triggers pneumatic valves to inflate cuffs sequentially during diastole (approx. 200ms) and evacuate simultaneously before systole (approx. 120ms). Inflation increases coronary perfusion pressure, coronary blood flow, and venous return; improves stroke volume and cardiac output. System comprises patient bed with pneumatic controls/valves/pump and operator console with all-in-one computer/monitor. Trained operator uses display to fine-tune inflation/deflation signals. Treatment typically 35 1-hour sessions, 5 days/week. Device automatically stops compression upon timer completion. Used in clinical settings.

## Clinical Evidence

Bench testing only. Device performance verified through design control procedures, confirming functionality as designed and compliance with IEC60601 medical device safety and EMC standards.

## Technological Characteristics

Microprocessor-controlled pneumatic system. Materials: aluminum bed base. Energy: 220VAC/50Hz or 120VAC/60Hz. Sensing: R-wave detection (ECG). Actuation: 6 magnetic bobbin valves per bed (24V DC). Connectivity: serial port for display data. Form factor: patient bed with floor-standing pole console. Software: firmware-based control.

## Regulatory Identification

An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.

## Special Controls

*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Model NCP-1 Counter Pulsation Device ([K023016](/device/K023016.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "ScottCare" in a stylized font. There is a line that runs underneath the word, and it resembles an EKG readout. The line has a peak in the middle, which is directly under the space between the two words. The image is in black and white.

800 243.9412 phone 216.267.6129 fax www.scottcare.com web

### 510(K) SUMMARY

## (SPECIAL CONSIDERATION)

MAY 2 9 2008

This summary of 510(k) safety and effectiveness information is being supplied in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92

| The assigned 510(k) number is         | K081359                                                                                                        |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Date:                                 | April 18, 2008                                                                                                 |
| Submitted by:                         | ScottCare Corporation<br>Registration No: 1527715<br>4791 West 150 <sup>th</sup> Street<br>Cleveland, OH 44135 |
| Contact Person:                       | Mr. Ronald J. Clines<br>216-362-0550 # 113<br>216-264-6129 Fax<br>rclines@scottcare.com                        |
| Manufacturing Site:                   | ScottCare Corporation<br>Registration No: 1063268<br>4897 W. Waters Ave, Suite J<br>Tampa, Florida 33634       |
| Trade Name:                           | ScottCare Genesis ECP Device                                                                                   |
| Common Name:                          | External Counter Pulsation (ECP) device                                                                        |
| Classification:                       | 870.5225<br>External counter-pulsating device,                                                                 |
| Product Code:                         | DRN                                                                                                            |
| Legally Marketed Predicate Device(s): | Model NCP-1 Counter Pulsation Device<br>K023016, December 12, 2002                                             |
|                                       | 2-0001                                                                                                         |

## દર

TeleRehab®

ROLINN

Page 1 of 4

NICORE

Telemetry

Diagnostic Cardiology

External Counterpulsation

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "ScottCare" in a stylized font. Above the "t" in ScottCare is a line that resembles an EKG reading. Below the word ScottCare is another line that resembles an EKG reading.

800.243.9412 phone 216.267.6129 fax www.scottcare.com web

#### Device Description:

The "Genesis" is a non-invasive medical device for performing external, sequential counterpulsation. It is a microprocessor-controlled system that inflates and deflates three pairs of air cuffs, which compress vascular beds in the muscles of the calves, thighs, and buttocks to achieve the desired therapy. The microprocessor's primary function is to serve as an R-wave detector, calculate the R-wave rate (heart rate), and deliver triggering signals to actuate valves that allow the cuffs to be filled during the diastolic period and then emptied prior to systole. The secondary function of the microprocessor is to output, through a serial port, information necessary for displaying patient treatment data on a video monitor. This data allows a trained operator to finetune the inflation/deflation signals to optimize a patient's diastolic augmentation. The cuffs are snugly wrapped around the patient's calves, thighs, and buttocks to allow compression of vascular beds in these body areas. To prevent skin irritation the patient is typically dressed in tight fitting, stretchable treatment pants prior to being fitted with the cuffs. As diastole begins, the cuffs inflate sequentially proceeding from the calves, to the thighs, to the buttocks, with the inflation sequence taking a total of approximately 200ms. The inflation sequence generates and drives an arterial counter pulsation wave creating an increase in coronary perfusion pressure and coronary blood flow. The compression sequence also increases venous return, which increases stroke volume and cardiac output. At the end of diastole, and just before the next QRS complex, the cuffs are evacuated simultaneously over a period of approximately 120ms. The course of treatment is typically 35 1-hour treatments administered 5 days a week. The device automatically stops compression of the cuffs and releases all pressure in the cuffs upon completion of the pre-set treatment timer.

The system consists of two major assemblies: 1) the patient bed that houses the electronic pneumatic controls / valves, compressor / vacuum pump and associated power supplies and 2) the operator's console that consists of an all-in-one computer/monitor, computer peripherals (keyboard and pointing device) and a control panel.

#### Indications for Use:

The ScottCare Genesis ECP is a non-invasive external counterpulsation (ECP) device indicated for use in the treatment of stable or unstable angina pectoris, congestive heart failure, cardioqenic shock and acute myocardial infarction.

2-0002

Page 2 of 4

TeleRehab® Telemetry

#### ROTANN

NICORE™

Diagnostic Cardiology

External Counterpulsation

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "ScottCare" in a stylized font. Above the "i" in ScottCare is a line that resembles a heartbeat. Below the word ScottCare is a line that also resembles a heartbeat. 

800.243.9412 phone 216.267.6129 fax www.scottcare.com web

## Substantial Equivalence:

| Features                 | The Modified Genesis Model                                                                                                                                                                                                                                                            | Predicate Device<br>NCP-1                                                                                                                                                                                                                                                 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number            | TBD                                                                                                                                                                                                                                                                                   | K023016                                                                                                                                                                                                                                                                   |
| Date Cleared             | TBD                                                                                                                                                                                                                                                                                   | 12/12/2002                                                                                                                                                                                                                                                                |
| Intended Use             | The intended use of the ScottCare Genesis<br>device is to provide external counterpulsation<br>(ECP) therapy, and is indicated for use in the<br>treatment of stable or unstable angina pectoris,<br>congestive heart failure, cardiogenic shock,<br>and acute myocardial infarction. | The intended use of the NCP-1 device is to<br>provide external counterpulsation (ECP)<br>therapy, and is indicated for use in the<br>treatment of stable or unstable angina pectoris,<br>congestive heart failure, cardiogenic shock, and<br>acute myocardial infarction. |
| Bed Enclosure            | Aluminum panels / welded                                                                                                                                                                                                                                                              | Furniture grade compressed wood / fasteners &<br>glue                                                                                                                                                                                                                     |
| Pump                     | Integral compressor / vacuum<br>220VAC 15A 60Hz (UL listed)<br>50Hz Option (UL Listed)                                                                                                                                                                                                | Integral compressor / vacuum<br>120VAC, 30A, 60Hz                                                                                                                                                                                                                         |
| nflate / Deflate Valve   | 6 per bed, Voltage controlled<br>24VDC                                                                                                                                                                                                                                                | 6 per bed, Voltage controlled<br>24VDC                                                                                                                                                                                                                                    |
| Emergency relief valve   | 3 per bed, Normally open, voltage controlled<br>24VDC                                                                                                                                                                                                                                 | 3 per bed, Normally open, voltage controlled<br>110VAC                                                                                                                                                                                                                    |
| Vacuum Tank Relief Valve | 1 per bed, Voltage controlled<br>24VDC                                                                                                                                                                                                                                                | 1 per bed, Voltage controlled<br>24VDC                                                                                                                                                                                                                                    |
| Air Valve Type           | Magnetic bobbin, Normally Closed, spring                                                                                                                                                                                                                                              | Magnetic bobbin, Normally Closed, spring                                                                                                                                                                                                                                  |
| Pressure Tank Relief     | Mechanical Type Calibrated at 8PSI                                                                                                                                                                                                                                                    | Mechanical Type Calibrated at 8PSI                                                                                                                                                                                                                                        |
| Power Supply             | (2) Switching Medical Grade<br>Located in bed base                                                                                                                                                                                                                                    | (2) Step down transformer<br>Located in the Operator's Console cabinet                                                                                                                                                                                                    |
| Bed Function Control     | Xilog Z-80 CPU integrated on motherboard                                                                                                                                                                                                                                              | Xilog Z-80 CPU integrated on motherboard                                                                                                                                                                                                                                  |
| Data Display Device      | Flat Panel LCD Computer Display                                                                                                                                                                                                                                                       | CRT Display                                                                                                                                                                                                                                                               |
| Panel Control Type       | Membrane style, momentary, panel type                                                                                                                                                                                                                                                 | Non-sealed, latching pushbutton, panel type                                                                                                                                                                                                                               |
| Operator's Console       | Floor standing pole type W/casters                                                                                                                                                                                                                                                    | Metal enclosure, floor type W/casters                                                                                                                                                                                                                                     |

Page 3 of 4

# 2-0003

TeleRehab®

Image /page/2/Picture/8 description: The image shows the word "ROZINN" in a bold, sans-serif font. The "Z" in the word is stylized with a diagonal line that extends beyond the top and bottom of the letter. There is a trademark symbol to the right of the word.

NICORE External Counterpulsation

Telemetry

.

Diagnostic Cardiology

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "ScottCare" in a stylized font. A graphic resembling an electrocardiogram (ECG) tracing is incorporated into the design, with a sharp peak positioned above the "tt" in "ScottCare." The ECG line extends horizontally beneath the word, adding a medical or healthcare-related visual element to the logo.

800.243.9412 phone 216.267.6129 fax www.scottcare.com web

As the above chart demonstrates, the Genesis model ECP bed is substantially equivalent to the NCP-1 ECP bed that was cleared on December 12, 2002. Basic control functions remain unchanged and there are no significant technological changes to the device. The changes are predominantly cosmetic in nature and include:

- Changing from a CRT display screen to a flat panel LCD computer type display. .
- . Elimination of the large roll-around control console (that housed the CRT display) and utilize a pole style, roll around to mount the all-in-one computer /LCD display and control panel.
- Changing the construction of the bed base from compressed wood to aluminum ● and present a smaller overall footprint.
- Changing the control panel switches from latching, push-button style to a . momentary, membrane style. To support this change a minor software (firmware) change was required.
- Provision for 50Hz operation where required. .
- Simplification of the internal power supply to use switching, medical grade power . supplies instead of discrete step down / rectification.
- Emergency relief valves (normally open) were changed to operate at 24V DC . instead of at 120V AC. The control mechanism, construction and overall action of the valves remain unchanged. Lower voltage valve standardizes the control voltage within the device to a safer, 24V DC level.

The intended use of the modified device, as described in its labeling, has not changed as a result of these modifications. No technologically significant changes have occurred.

#### Conclusion:

Appropriate testing was conducted in accordance with established design control procedures. Testing demonstrated that the device functioned as designed and that the device was in compliance with IEC60601 Medical Device Safety and EMC testing requirements. The modified device, which is to be marketed as the "ScottCare Genesis" ECP Treatment device, is substantially equivalent to the predicate device.

2-0004

Page 4 of 4

NICORE External Counterpulsation

TeleRehab

Image /page/3/Picture/18 description: The image contains the logo for "ROZINN" in a stylized font with a trademark symbol. Below the logo, the text "Diagnostic Cardiology" is present. The text is in a simple, sans-serif font and is smaller than the logo.

Telemetry

{4}------------------------------------------------

Food and Drug Administration

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

9200 Corporate Boulevard Rockville MD 20850

MAY 2 9 2008

ScottCare Corporation Ronald J. Clines Manager Regulatory/Quality 4791 West 150th Street Cleveland, OH 44135

Re: K081359

Trade/Device Name: ScottCare Genesis External Counter Pulsation (ECP) Device Regulation Number: 21 CFR 870.5225 Regulation Name: External counter-pulsation device Regulatory Class: Class III (three) Product Code: DRN Dated: April 30, 2008 Received: May 15, 2008

Dear Mr. Clines:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Ronald J. Clincs

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

R. Vahmer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

| 510(k) Number (if known): | K081359 |
|---------------------------|---------|
|---------------------------|---------|

Device Name: ScottCare Genesis External Counterpulsation Device

### Indications for Use:

The intended use of the ScottCare Genesis ECP device is to provide external counterpulsation (ECP) therapy, and is indicated for use in the treatment of stable or unstable angina pectoris, congestive heart failure, cardiogenic shock, and acute myocardial infarction.

Prescription Use -(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. Holmes

2-0005

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_K081359

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K081359](https://fda-staging.innolitics.com/device/K081359)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
