FIXXSURE CROSS LINK

K081331 · Spineworks, LLC · MNH · Jul 23, 2008 · Orthopedic

Device Facts

Record IDK081331
Device NameFIXXSURE CROSS LINK
ApplicantSpineworks, LLC
Product CodeMNH · Orthopedic
Decision DateJul 23, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SpineWorks FixxSure Cross Link is intended work with the Talon Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The SpineWorks FixxSure Cross Link must be used with the Talon Pedicle Screw system.

Device Story

Spinal implant system; cross link connects dual-rod constructs to increase rigidity and construct strength. Multi-span and fixed designs available; features proprietary dual locking mechanism for rod connection and multi-plane angulation. Used by spinal surgeons in clinical/OR settings to stabilize spinal segments during fusion procedures. Implanted as adjunct to Talon Pedicle Screw system. Benefits patient by providing enhanced mechanical stability for spinal fusion.

Clinical Evidence

Bench testing only. Testing conducted per ASTM F1717 and F1798 standards to verify mechanical strength.

Technological Characteristics

Fabricated from Ti6Al4V alloy (ASTM F136). Multi-span and fixed design configurations. Proprietary dual locking mechanism. Mechanical device; no software or energy source.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization and stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including severe spondylolisthesis (L5-S1, grades 3-4), degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and pseudarthrosis. Must be used with Talon Pedicle Screw system.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081331 (pg. 1 of 2) JUL 23 2008 # for the Spineworks FixxSure Cross Link In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the SpineWorks FixxSure Cross Link. Date Prepared: May 7, 2008 - 1. Submitter: SpineWorks LLC 16742 Gothard St. Suite 101 Huntington Beach, CA 92647 Contact Person: J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, TX 78681 Telephone: 512-388-0199 - 2. Trade name: FixxSure Cross Link Common Name: pedicle screw system crosslink Classification Name: Pedicle screw spinal system Class II 21 CFR 888.3070 MNI/MNH - Predicate or legally marketed devices which are substantially equivalent: 3. Sea Spine Crossbar (K032739) Acme/Talon Pedicle Screw system (K071824). #### 4. Description of the device: The SpineWorks FixxSure Cross Link allows spinal surgeons to convert a dual-rod construct into a frame and increase overall construct strength. The FixxSure Cross Link comes in a multi-span and fixed design, both of which can be rigidly locked onto a dual-rod construct and has the capability of being manipulated into various planes of angulation. The SpineWorks implant has a proprietary dual locking mechanism allowing maximum Cross Link/rod connection while offering superior ease of insertion/use. ## Materials: The components in this submission are fabricated from Ti6Al4V alloy, conforming to ASTM F136, which is known to have good biocompatibility. ## Function: Cross Link systems were introduced to increase the rigidity of the pedicle screw constructs. #### ပ် Intended Use: The SpineWorks FixxSure Cross Link is intended work with the Talon Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The SpineWorks FixxSure Cross Link must be used with the Talon Pedicle Screw system. {1}------------------------------------------------ $$\begin{pmatrix} \mathfrak{s} \ \mathfrak{s} \ \mathfrak{s} \end{pmatrix} \begin{pmatrix} \mathfrak{s} \ \mathfrak{s} \end{pmatrix} \begin{pmatrix} \mathfrak{s} \ \mathfrak{s} \end{pmatrix} \begin{pmatrix} \mathfrak{s} \end{pmatrix}$$ - 6. Comparison of the technological characteristics of the device to predicate and legally marketed devices: The FixxSure Cross Link is similar to the predicate device in terms of material, design and indications. ## 7. Summary of Nonclincal Tests Testing was conducted according to ASTM F1717 and F1798 with adequate strength. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circle around the eagle. The eagle is facing to the right and the stripes are curved. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SpineWorks, LLC. % OrthoMedix Group, Inc. Mr. J.D. Webb 1001 Oakwood Boulevard Round Rock, TX 78681 JUL 23 2008 Dear Mr. Webb: Rc: K081331 > Trade/Device Name: FixxSure Cross Link Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: May 7, 2008 Received: May 12, 2008 Dear Mr. Webb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. J.D. Webb This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark H. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): _长ひ8) 331 Device Name: Indications for Use: The SpineWorks FixxSure Cross Link is intended work with the Talon Pedicle Screw system to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The SpineWorks FixxSure Cross Link must be used with the Talon Pedicle Screw system. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...