ETEST DORIPENEM - ANTIMICROBIAL SUSCEPTIBILITY TEST - MIC AT 0.002-32 UG/ML

K081325 · Ab Biodisk · JWY · Jul 24, 2008 · Microbiology

Device Facts

Record IDK081325
Device NameETEST DORIPENEM - ANTIMICROBIAL SUSCEPTIBILITY TEST - MIC AT 0.002-32 UG/ML
ApplicantAb Biodisk
Product CodeJWY · Microbiology
Decision DateJul 24, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

This submission is for Etest Doripenem for MIC determinations across 0.002-32 µg/mL with Gram negative aerobic bacteria such as Enterchacteriaceae, Accretobacer baumanii, Pseudomonas aeraginesa and Gram negative anaerobic bacteria such as Bacteroides aacae, B. fragilis, B.thetaiotaorracron, B. uniformis and B. ulgatus. Etest is a quantitative technique for determination of antimicrobial susceptibility of both nonfastidious Gram negative and Gram positive aerobic bacteria such as Enterchaceriane, Pseudomnas, Staphylococus and Enteroxous species and fastidious bacteria, such as anaerobes, N. gonombeae, S. preumoniae, Streptoxous and Haemophilus species. The system comprises a predelined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC) in ug/ mL of different antimicrobial agents against microorganisms as tested on agar using overnight incubation.

Device Story

Etest® is a manual, quantitative antimicrobial susceptibility test. Device consists of a 5mm x 60mm inert plastic strip containing a predefined exponential gradient of dried, stabilized antibiotic. Strip is applied to inoculated agar plates; antibiotic diffuses into agar, creating a concentration gradient. Following overnight incubation, a symmetrical inhibition ellipse forms. User visually identifies the MIC value where the ellipse edge intersects the strip. Results provide clinicians with quantitative susceptibility data to guide antibiotic therapy. Intended for use in clinical microbiology laboratories.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing of the Etest system's performance in determining MIC values for Doripenem against specified bacterial isolates compared to reference methods.

Technological Characteristics

Inert, non-porous plastic strip (5mm x 60mm) with dried, stabilized antibiotic gradient. Manual reading of MIC values. No electronic components, software, or connectivity. Sterilization method not specified.

Indications for Use

Indicated for testing Acinetobacter baumannii, E. coli, K. pneumoniae, P. mirabilis, P. aeruginosa, Bacteroides caccae, Bacteroides fragilis, Bacteroides thetaiotaomicron, Bacteroides uniformis, and Bacteroides vulgatus using doripenem at concentrations of 0.002 – 32 µg/mL.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k081325 B. Purpose for Submission: Addition of doripenem to the Etest® strip C. Measurand: Doripenem 0.002 – 32 µg/mL D. Type of Test: Quantitative Antimicrobial Susceptibility Test (AST) growth based detection E. Applicant: AB BIODISK F. Proprietary and Established Names: Etest® for Antimicrobial Susceptibility Testing G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: JWY - Manual Antimicrobial Test Systems 4. Panel: Microbiology {1} H. Intended Use: 1. Intended use(s): Etest® is a quantitative technique for determining antimicrobial susceptibility testing of both non-fastidious Gram-negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient to determine the Minimum Inhibitory Concentration (MIC) in ug/mL of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. 2. Indication(s) for use: This submission is for the addition of the antibiotic doripenem at concentrations of 0.002 – 32 µg/mL to the Etest® strip for testing Acinetobacter baumannii, E. coli, K. pneumoniae, P. mirabilis, P. aeruginosa, Bacteroides caccae, Bacteroides fragilis, Bacteroides thetaiotaomicron, Bacteroides uniformis, and Bacteroides vulgatus. 3. Special conditions for use statement(s): Prescription use 4. Special instrument requirements: Manual readings only There was not sufficient data to evaluate the performance of doripenem with Peptostreptococcus. An alternate method should be performed. I. Device Description: Etest® consists of a thin, inert and non-porous plastic strip, 5mm wide and 60 mm long. One side of the strip carries a two-letter code designating the identity of the antibiotic and is calibrated with MIC values in terms of ug/mL. A predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method. J. Substantial Equivalence Information: 1. Predicate device name(s): The Etest® {2} 2. Predicate K number(s): k913459 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Etest® is a quantitative technique for determining antimicrobial susceptibility testing of both non-fastidious Gram-negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient to determine the Minimum Inhibitory Concentration (MIC) in ug/mL of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. | Same | | Inoculation | Isolated colonies from culture used | Same | | Result | MIC | Same | | Type of Test | Manual read only | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antibiotic | Doripenem | Various antibiotics | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S18) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard; CLSI M11-A6, "Methods for Antimicrobial Susceptibility Testing of Anaerobic Bacteria; Approved Standard - Sixth Edition." # L. Test Principle: The Etest® gradient technology is based on a combination of the concepts of dilution and {3} diffusion test methods for susceptibility testing. Etest® directly quantifies antimicrobial susceptibility in terms of discrete MIC values. When the Etest® strip is applied to an inoculated agar plate, the antibiotic is immediately released from the plastic surface into the agar. A predefined, continuous gradient of antibiotic concentrations is created and maintained directly underneath the strip. After incubation whereby bacterial growth becomes visible, a symmetrical inhibition ellipse centered along the strip will be seen. The MIC value in $\mathrm{ug/mL}$ is read where the ellipse edge intersects the strip. Since Etest® generates MIC values which fall between two-fold dilutions for interpretation, the MIC value read must be recorded to the next two-fold dilution. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Reproducibility testing was performed on 26 gram negative isolates, and 23 Bacteroides spp. All organisms were tested once at all three sites which demonstrated $>95\%$ reproducibility. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended Quality Control (QC) isolates were tested at the concentrations listed (see tables below). The results demonstrated that the device system could produce QC results in the recommended range. Quality Control results demonstrated the ability of the device to produce acceptable results $>95\%$ of the time. Quality Control Table for Gram Negative Isolates | ORGANISM | Conc ug/mL | Reference | Etest® | | --- | --- | --- | --- | | | | | | | E. coli ATCC 25922 Expected Range : 0.015 – 0.064 μg/mL | 0.016 | 47 | 35 | | | 0.032 | 23 | 35 | | | | | | | | | | | | | | | | | P. aeruginosa ATCC 27853 Expected Range : 0.12 – 0.5 μg/mL | 0.125 | 25 | 27 | | | 0.25 | 45 | 43 | | | 0.5 | 1 | 1 | | | | | | | | | | | {4} Quality Control Table for Anaerobes | ORGANISM | Conc ug/mL | Reference | Etest® | | --- | --- | --- | --- | | | | | | | B. fragilis ATCC 25285 Expected Range: 0.12 – 0.5 μg/mL | 0.125 | 52 | 61 | | | 0.25 | 13 | 4 | | | | | | | | | | | | B. thetaiotaomicron | 0.125 | 6 | 21 | | ATCC 29741 | 0.25 | 39 | 38 | | Expected Range: | 0.5 | 20 | 6 | | 0.125 – 1 μg/mL | | | | | | | | | A 0.5 McFarland is used to determine the correct inoculum. Colony counts for the clinical/stock isolates were performed periodically at each site to demonstrate that the inoculum procedure results were in the expected CFU/mL range. All colony count inoculum check means results were within the acceptable ranges. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: CLSI recommended broth microdilution and agar dilution media were utilized as the reference methods, and were used to compare with the Etest® results. Clinical testing was performed at three sites. The testing included fresh/stock, clinical aerobic Gram negative isolates; and fresh/stock clinical isolates of Bacteroides species. Approximately 60% of the Clinical strains were fresh isolates. The Challenge isolates selected for testing each organism group was appropriate for this antibiotic. Challenge testing was performed at one site. Low numbers from each organism group were tested, although the {5} overall numbers were sufficient. The performance of the study is demonstrated below. Summary Table | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA Tot | CA N | CA % | #NS | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | A. baumanii | 189 | 189 | 100 | 133 | 133 | 100 | 133 | 131 | 98.5 | 104 | | P. aeruginosa | 171 | 169 | 98.8 | 123 | 121 | 98.4 | 123 | 119 | 96.7 | 80 | | Enterobacteriaceae | 358 | 357 | 99.7 | 354 | 353 | 99.7 | 354 | 353 | 99.7 | 19 | | Bacteroides spp. | 266 | 266 | 100 | 258 | 258 | 100 | 258 | 256 | 99.2 | 21 | EA - Essential Agreement CA - Category Agreement NS - Not Susceptible EA is when there is agreement between the reference method and the Etest® panel within plus or minus one serial two-fold dilution of antibiotic. Category agreement (CA) is when the Etest® panel result interpretation agrees exactly with the reference panel result interpretation. Evaluable EA is when the MIC result is on scale for both the Etest® and the reference and have on-scale EA. The no growth rate is <10% for all organisms tested. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: {6} Enterobacteriaceae ≤0.5 P. aeruginosa ≤2 Acinetobacter baumannii ≤1 Anaerobes ≤1 * Limitation statement: The current absence of resistant isolates precludes defining any category other than “Susceptible”. Isolates yielding MIC results suggestive of “Nonsusceptible” category should be subjected to additional testing. However, in the in vitro diagnostic studies, 104/189 Acinetobacter baumanii isolates, and 80/171 P. aeruginosa isolates were identified as Nonsusceptible (NS). N. Proposed Labeling: The expected value range, interpretive criteria and QC are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The information submitted in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...