K081248 · Inova Diagnostics, Inc. · OKM · Oct 27, 2008 · Immunology
Device Facts
Record ID
K081248
Device Name
QUANTA LITE OMP PLUS ELISA
Applicant
Inova Diagnostics, Inc.
Product Code
OKM · Immunology
Decision Date
Oct 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5785
Device Class
Class 2
Indications for Use
The QUANTA Lite™ OMP Plus kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of anti-OMP antibodies of the IgA class in human serum. It is intended to be used in conjunction with anti-Saccharomyces cerevisiae (S. cerevisiae) antibody (ASCA) IgG and/or IgA test systems. The presence of OMP (outer membrane proteins) IgA antibodies, used in conjunction with clinical findings and other laboratory tests, may aid in the diagnosis of patients with Crohn's disease.
Device Story
QUANTA Lite™ OMP Plus is an ELISA-based in vitro diagnostic test. It detects anti-outer membrane protein (OMP) IgA antibodies in human serum samples. The assay is performed in a clinical laboratory setting by trained personnel. The device transforms the presence of specific antibodies into a semi-quantitative optical density measurement via an enzyme-linked reaction. Results are interpreted by clinicians alongside ASCA IgG/IgA test results and clinical findings to support the diagnosis of Crohn's disease. The test provides diagnostic information to assist in patient management.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) for semi-quantitative detection of IgA antibodies. Utilizes outer membrane protein (OMP) antigens. Standard laboratory plate-based format.
Indications for Use
Indicated for use as an aid in the diagnosis of Crohn's disease in patients, to be used in conjunction with ASCA IgG and/or IgA test systems. Prescription use only.
Regulatory Classification
Identification
The Anti-Saccharomyces cerevisiae (S. cerevisiae ) antibody (ASCA) test system is an in vitro diagnostic device that consists of the reagents used to measure, by immunochemical techniques, antibodies to S. cerevisiae (baker's or brewer's yeast) in human serum or plasma. Detection of S. cerevisiae antibodies may aid in the diagnosis of Crohn's disease.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Guidance for Industry and FDA Reviewers: Class II Special Control Guidance Document for Anti-*Saccharomyces cerevisiae* (*S. cerevisiae* ) Antibody (ASCA) Premarket Notifications.”
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k081248
B. Purpose for Submission:
New Device
C. Measurand:
Anti-OMP IgA Antibodies
D. Type of Test:
Semi-quantitative ELISA
E. Applicant:
INOVA Diagnostics, Inc.
F. Proprietary and Established Names:
QUANTA Lite™ OMP Plus IgA ELISA
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5785 - Anti-Saccharomyces cerevisiae (S. Cerevisiae) Antibody (ASCA) Test Systems.
2. Classification:
Class II
3. Product code:
OKM, Antibodies, outer-membrane proteins
4. Panel:
Immunology 82
H. Intended Use:
1. Intended use(s):
The QUANTA LiteTM OMP Plus kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of anti-OMP antibodies of the IgA class in human serum. It is intended to be used in conjunction with anti-Saccharomyces cerevisiae (S. cerevisiae) antibody (ASCA) IgG and/or IgA test systems. The presence of OMP (outer membrane proteins) IgA antibodies, used in conjunction with clinical findings and other laboratory tests, may aid in the diagnosis of patients with Crohn's disease.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
Prescription use only.
4. Special instrument requirements:
Microplate reader capable of measuring OD at 450 nm (or 620 for dual wavelength readings)
I. Device Description:
QUANTA Lite™ OMP Plus ELISA consists of a polystyrene microwell ELISA plate coated with a proprietary lysate of outer membrane and associated proteins derived from 2 strains of anaerobic bacteria; high positive, low positive, and negative controls; sample diluent; wash concentrate; goat anti-human IgG horseradish peroxidase
{1}
conjugate; TMB chromogen; and stop solution.
# J. Substantial Equivalence Information:
1. Predicate device name(s): QUANTA Lite™ ASCA IgG ELISA
2. Predicate K number(s): k000732
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indication for Use | To aid in the diagnosis of patients with Crohn's disease. | Same |
| Technology | ELISA | Same |
| Assay Format | Semi-quantitative | Same |
| Platform | 96 well microtiter plates | Same |
| ELISA Negative control | Pre-diluted human serum, Ready to use | Same |
| Sample type and dilution | Serum, 1:101 | Same |
| Sample volume | 5 uL | Same |
| Enzyme conjugate | Horseradish peroxidase, goat anti-human IgG | Same |
| Substrate | TMB Chromogen | Same |
| Incubation times | 30-30-30 minutes | Same |
| O.D. Reading | 450 nm (or 620 for dual wavelength readings) | Same |
| Result Interpretation | <20 units = negative 20-24.9 units = equivocal ≥ 25 units = positive | Same |
| Cut-off | 25 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Semi-quantitative detection of anti-OMP (outer membrane proteins) antibodies of the IgA class in human serum. To be used in conjunction with ASCA (IgG and IgA). | Semi-quantitative detection of anti-Saccharomyces cerevisiae antibodies (ASCA) of the IgG class in human serum. |
| Antigen | Partially purified lysate of outer membrane and associated proteins from two different species of intestinal bacteria | Partially purified and disrupted S.cerevisiae antigen |
| High Positive and Low Positive Controls | OMP Plus IgA ELISA High and Low positive controls. Pre-diluted. Ready to use | ASCA IgG ELISA High and Low positive controls. Pre-diluted. Ready to use |
# K. Standard/Guidance Document Referenced (if applicable):
Not referenced.
{2}
L. Test Principle:
A proprietary lysate of partially purified outer membrane and associated proteins derived from 2 species of anaerobic bacteria are bound to the wells of a polystyrene microwell plate under conditions that will preserve the antigen in its native state. Pre-diluted controls and diluted patient sera are added to separate wells, allowing any OMP antibodies present to bind to the immobilized antigen. Unbound sample is washed away and an enzyme labeled anti-human IgA conjugate is added to each well. A second incubation allows the enzyme labeled anti-human IgA to bind to any patient antibodies, which have become attached to the microwells. After washing away any unbound enzyme labeled anti-human IgA, the remaining enzyme activity is measured by adding a chromogenic substrate and measuring the intensity of the color that develops. The assay can be evaluated spectrophotometrically by measuring and comparing the color intensity that develops in the patient wells with the color in the control wells.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Intra-assay studies: Eight samples (3 specimens with moderate OMP antibody levels (46.2, 45.5, 41.8 units, respectively), 2 specimens just over the cut-off (26.9, 26.3 units, respectively), and 3 equivocal samples (21.8, 23.9, 24.5 units, respectively)) were assayed in 10 replicates each. Results showed acceptable %CV ranging from 2.1 to 6.4.
Intra-assay Performance of QUANTA Lite™ OMP Plus ELISA
| | Specimen | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
| Mean units | 46.2 | 26.9 | 45.5 | 26.3 | 41.8 | 24.5 | 23.9 | 21.8 |
| SD | 1.17 | 0.78 | 2.91 | 0.72 | 0.93 | 0.86 | 0.83 | 0.46 |
| CV % | 2.5 | 2.9 | 6.4 | 2.8 | 2.2 | 3.5 | 3.5 | 2.1 |
Inter-assay: Nine specimens (2 equivocal, 3 low positive, and 3 moderate positive) plus the high positive control were tested in duplicate twice daily for 3 days. The results showed acceptable %CV ranging from 1.1 to 4.3.
Inter-assay Performance for QUANTA Lite™ OMP Plus ELISA
| | Specimen | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | HPC | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
| Mean units | 69.8 | 50.0 | 26.6 | 54.1 | 29.2 | 46.2 | 27.3 | 24.7 | 23.5 |
| SD | 2.35 | 0.74 | 0.89 | 2.23 | 1.07 | 1.63 | 1.14 | 1.07 | 0.89 |
| CV % | 3.4 | 1.1 | 3.4 | 4.1 | 3.7 | 3.5 | 4.2 | 4.3 | 3.8 |
b. Linearity/assay reportable range:
No claims were made regarding linearity for the assay. Reactivity is related to the quantity of antibody present in a non-linear fashion.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
There is no reference standard for OMP Antibodies. The results are reported in arbitrary units
{3}
d. Detection limit:
Not applicable.
e. Analytical specificity:
Interference:
Interference by endogenous substances: No data provided. The package insert states that grossly hemolyzed, icteric, microbially contaminated, or heat-treated samples should be avoided in this assay.
Cross-reactivity with other autoantibodies:
Sera from 36 patients with various autoimmune or infectious disease and positive for associated antibodies including 8 rheumatoid arthritis, 2 soluble liver antigen (SLA), 2 liver kidney microsome (LKM), 2 chromatin, 2 Jo-1, 2 centromere, 1 SS-A, 2 SS-B, 2 anti-mitochondrial antibody (AMA), 2 tissue transglutaminase (tTG), 1 Scl-10, 1 RNP, 3 glomerular basement membrane (GBM), 2 cytomegalovirus (CMV), 2 rubella virus, 2 herpes simplex virus (HSV), were tested with the QUANTA Lite™ OMP Plus ELISA to assess the assay's specificity. No specimens were interpreted as positive and only one sample (an autoimmune hepatitis patient positive for SLA) was interpreted as equivocal at 22.2 units on the OMP Plus ELISA. Additional testing showed the specimen was moderately positive for ASCA IgG (76.1 units) and had an equivocal level of ASCA IgA antibodies. Sera from 226 patients with celiac disease were tested with the OMP Plus IgA ELISA and $27\%$ were found positive. On the same cohort, $27\%$ were positive for ASCA IgA and $31\%$ for ASCA IgG antibodies. 14 $(6.2\%)$ of the celiac patients were triple positive (OMP Plus IgA, ASCA IgG, and ASCA IgA).
f. Assay cut-off:
To validate the cut-off, a panel of 500 apparently healthy individuals from the US (250 male and 250 female) with a median age of 41 years (range 18-80 years old) were tested. The median value obtained was 12.5 units. The cutoff was set at 25 units. This resulted in a specificity of $92\%$ (468/500).
2. Comparison studies:
a. Method comparison with predicate device:
A panel of 1186 sera consisting of 365 specimens from patients with Crohn's disease (CD) and 821 patients from patients without Crohn's disease including specimens from 234 patients with ulcerative colitis (UC) and 587 healthy controls were tested with both the QUANTA Lite™ OMP Plus ELISA and the predicate QUANTA Lite™ ASCA IgG ELISA. Demographic information for the patient population was not available. The CD, UC and normal samples for the clinical method comparison were of European and Canadian origin.
| | QUANTA Lite™ ASCA IgG ELISA | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Pos | Eq* | Neg | Total |
| QUANTA Lite™ OMP Plus ELISA | Pos | 117 | 13 | 70 | 200 |
| | Eq* | 34 | 14 | 52 | 100 |
| | Neg | 167 | 45 | 674 | 886 |
| | Total | 318 | 72 | 796 | 1186 |
*Equivocal results counted as negative (not-positive).
{4}
Positive percent agreement: 36.8% (117/318)
Negative percent agreement: 90.4% (674+14+45+52/868 = 785/868)
Overall Agreement 76.1% (117+785/1186)
b. Matrix comparison:
Both assays use serum as the matrix.
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
The study consisted of the following: CD (365), UC (234), and Healthy Normal (587) for a total of 1186 samples.
Differentiation of Crohn's Disease from non-Crohn's Disease (ulcerative colitis and normal controls) Specimens
| N=1186 | | Crohn's Disease | Non-Crohn's Disease | Total |
| --- | --- | --- | --- | --- |
| Quanta Lite OMP Plus ELISA | Pos | 137 | 63 | 200 |
| | Neg | 228 | 758 | 986 |
| | Total | 365 | 821 | 1186 |
Equivocal results counted as negative (i.e. not positive)
Clinical Sensitivity: 37.5% (137/365) (95% CI, 0.325 to 0.427)
Clinical Specificity: 92.3% (758/821) (95% CI, 0.903 to 0.941)
Differentiation of Crohn's Disease from Ulcerative Colitis Specimens
| N =599 | | Crohn's Disease | Ulcerative Colitis | Total |
| --- | --- | --- | --- | --- |
| Quanta Lite OMP Plus ELISA | Pos | 137 | 26 | 163 |
| | Neg | 228 | 208 | 436 |
| | Total | 365 | 234 | 599 |
Equivocal results counted as negative (i.e. not positive)
Clinical Sensitivity: 37.5% (137/365) (95% CI, 0.325 to 0.427)
Clinical Specificity: 88.9% (208/234) (95% CI, 0.841 to 0.926)
The OMP Plus assay is intended as a supplemental assay to ASCA IgG and ASCA IgA. The clinical sensitivities of the new device and the predicate device were tested with both the Quanta Lite™ OMP Plus ELISA and both the Quanta Lite™ ASCA IgG and ASCA IgA ELISA kits.
The simultaneous presence of ASCA IgG, ASCA IgA, and OMP Plus antibodies (triple positive) may prove to be of value for clinical assessment. As shown in the table below, when the whole 510(k) cohort was examined, the presence of at least one of the three markers (ASCA IgG, ASCA IgA, or OMP Plus) resulted in a specificity of 79.4% (652/821), the presence of at least 2 markers resulted in a specificity of 96.3% (791/821), and the presence of all three markers had a
{5}
specificity of $99.5\%$ (817/821). The increasing specificity of the marker combination assessment results in decreasing sensitivity. It may however, aid in ruling-in or ruling-out some specimens. For example of the total of 97 triple positive specimens, 93/97 $(95.9\%)$ were CD patients. The other 4 consisted of 3 UC and 1 "normal" healthy donor.
| At least 1 marker Positive | | At least 2 markers positive | | 3 markers positive | |
| --- | --- | --- | --- | --- | --- |
| Sensitivity | Specificity | Sensitivity | Specificity | Sensitivity | Specificity |
| 72.1% (263/365) | 79.2% (652/821) | 46% (168/365) | 96.3% (791/365) | 25.5% (93/365) | 99.5% (817/821) |
A comparison between the OMP Plus, ASCA IgG, and ASCA IgA demonstrated the added value of the OMP assay results to the results obtained from CD and the ASCA IgG and ASCA IgA is shown in the venn diagrams below.
Fig. 1: Sensitivity of ASCA IgG, ASCA IgA, and OMP for CD, non-CD (and healthy normal), and UC patients.
Crohn's Patients - Whole Cohort

Note: Equivocal results counted as neg (i.e not positive)
{6}
7
# Non-Crohn's Patients

Non-Crohn's Patients
- ASCA G and/or ASCA IgA pos= 123; Specificity =85%
- Omp Plus additional 46+ pos (5.6%)
- 1 or more ASCA G,ASCA A, Omp Plus = 20.5% (169/821) Specificity = 79.4% (652/821)
- 2 or more pos Specificity = 96.3% (30/821)
- Triple Pos = 4/821 Specificity = 99.5% (4/821)
Note: Equivocal results counted as neg (i.e not positive)
# Ulcerative Colitis Patients (510K)

UC Patients
- ASCA G and/or ASCA IgA pos= 40; specificity = 82.9%
- Omp Plus additional 26+ pos (11.1%)
- 1 or more ASCA G,ASCA A, Omp Plus = 57 (20.5%) Specificity = 75.6% (177/234)
- 2 or more pos =11 Specificity = 95.3% (223/234)
- Triple Pos = 3 Specificity = 98.7% (231/234)
Note: Equivocal results counted as neg (i.e not positive)
4. Clinical cut-off:
See assay cut-off
5. Expected values/Reference range:
The expected value in the normal population is negative, however the performance of the assay has not been assessed in patients with Irritable Bowel Syndrome (IBS). The median value obtained from the 500 healthy normal individuals (see assay cutoff) was 12.5 units.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.