TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES

K081232 · Praxim · HAW · Aug 6, 2008 · Neurology

Device Facts

Record IDK081232
Device NameTOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES
ApplicantPraxim
Product CodeHAW · Neurology
Decision DateAug 6, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. It is specifically indicated for: Total Knee Arthroplasty

Device Story

Image-guided surgical navigation system used in OR by surgeons. Inputs: optical tracking data from reflective markers on patient anatomy and surgical instruments. System components: mobile computer station (Surgetics/NanoStation), optical localizer, and motorized cutting block (Praxiteles). Operation: system tracks anatomical structures and guides alignment of endoprostheses; motorized cutting block automatically or manually positions cutting slot to desired cut. Output: visual guidance and mechanical positioning of cutting block. Benefit: assists surgeon in precise anatomical localization and implant alignment during Total Knee Arthroplasty.

Clinical Evidence

Bench testing only. Non-clinical assessment using specimens confirmed system accuracy and performance were adequate and comparable to the predicate device.

Technological Characteristics

Optical navigation system; mobile computer station; motorized cutting block; reflective markers for registration. Connectivity: standalone navigation station. Software-controlled positioning of cutting slot. Energy source: electrical.

Indications for Use

Indicated for patients undergoing Total Knee Arthroplasty requiring stereotaxic surgical navigation for anatomical localization and endoprosthesis alignment.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K08/232 Page 1 of 3 # 510(k) Summary AUG - 6 2008 for TOTAL KNEE SURGETICS Navigation System with Praxiteles # 1. Submitter Name and Address: Praxim "Le Grand Sablon" 4, Avenue de l'Obiou 38 700 La Tronche France Contact Name: Annie Pollier Telephone: (0033) 4 76 54 95 03 Date Prepared: April 25th 2008 ## 2. Device Name: | Proprietary Name: | TOTAL KNEE SURGETICS Navigation System with<br>Praxiteles | |----------------------|-----------------------------------------------------------| | Common/Usual Name: | Image guided surgical navigation system | | Classification Name: | Computed tomography x-ray system (accessory) | # 2. Equivalent to: | 510(k) number | K060282 | K061362 | K063408 | |------------------------|----------------------------------------------|--------------------------------------------------------|-------------------------------| | Proprietary Name | PRAXIM | PLUS<br>ORTHOPEDICS AG | PRAXIM | | Common/Usual<br>Name | TOTAL KNEE<br>SURGETICS<br>Navigation System | PIGALILEO<br>TOTAL KNEE<br>REPLACEMENT<br>(TKR) SYSTEM | CTLOGICS<br>Navigation System | | Classification<br>Name | instrument,<br>stereotaxic | instrument,<br>stereotaxic | instrument,<br>stereotaxic | {1}------------------------------------------------ Ke 8/232 Page 83 #### 4. Intended Use: The TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning the endoprostheses with the anatomical structures. It is specifically indicated for: - . Total Knee Arthroplasty #### న. Device Description: As the equivalent TOTAL KNEE SURGETICS NAVIGATION SYSTEM, the TOTAL KNEE SURGETICS NAVIGATION SYSTEM WITH PRAXITELES consists of the following major components and subsystems: - The Station (Surgetics or NanoStation), consisting of a mobile computer . system and an optical localizer - Ancillary instruments, specific motorized cutting block and reflective . markers used for reference and registration - TOTAL KNEE SURGETICS with Praxiteles software ● The main modification to the predicate device K060282 concerns the use of a new motorized cutting block PRAXITELES®. ### Technological Characteristics and Substantial Equivalence 6. The underlying technology of the TOTAL KNEE SURGETICS Navigation System with PRAXITELES is the same as for the predicate device K060282. The system is based on the same operating principle and control mechanism to provide the user with the same kind of information and guidance for the same surgery. The main changes with respect to the predicate device concern the addition of a motorized cutting block Praxiteles. Praxiteles is a motorized cutting block controlled by the navigation system that automatically positions the cutting slot to the desired cut. Then the position of the motorized cutting block is controlled and the surgeon can proceed to the cut as it is done in the conventional way. The system enables the cutting slot to be positioned either automatically or manually. {2}------------------------------------------------ Ko81232 Page 3 of 3 # 7. Performance Testing The TOTAL KNEE SURGETICS Navigation System with Praxiteles was tested in a non clinical setting (bench testing, specimen) to assess that no new safety and efficiency issues were raised in the device. Analyses show that the accuracy and performance of the system was adequate for its intended use and not reduced in comparison to the predicate device. In conclusion the modified device TOTAL KNEE SURGETICS Navigation System with Praxiteles is substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol. The logo is simple, yet recognizable, and represents the department's role in protecting the health of all Americans. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG - 6 2008 Praxim % Ms. Annie Pollier VP, Operations, Quality & Regulatory Le Grand Sablon 4. Avenue de l'Obiou 38 700 La Tronche France Re: K081232 Trade/Device Name: TOTAL KNEE SURGETICS Navigation System with Praxiteles Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: July 10, 2008 Received: July 17, 2008 Dear Ms. Pollier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Ms. Annie Pollier forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for PRAXIM Computer Integrated Surgery. The logo is in all caps and has a unique font. The words "COMPUTER INTEGRATED SURGERY" are written in a smaller font below the main logo. PRAXIM S.A. le Grand Sablon 4, Avenue DE L'Oblou 38700 LA TRONCHE FRANCE PHONE : 33 4 76 54 95 03 FAX +33 4 76 51 98 29 S.A. AU CAPITAL DE 2'769'610 € NO SIRET 400 491 304 00020 CODE APE 331 B www.praxim.com # STATEMENT OF INDICATIONS FOR USE TOTAL KNEE SURGETICS Navigation System with Praxiteles 510(k) Number (if known): K08/232 Device Name: TOTAL KNEE SURGETICS Navigation System with Praxiteles The TOTAL KNEE SURGETICS Navigation System with Praxiteles is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures and aligning they endoprostheses with the anatomical structures. It is specifically indicated for : * Total Knee Arthroplasty Mark X Millhouse Division of General, Restorative, and Neurological I 510(k) Number K0812 S2 Prescription Use X AND / OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Praxim 510(k) [04/25/2008] TOTA KNEE SURGETICS Navigation System with Praxiteles Confidential Page 1 / 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...