# MODIFICATION TO ANGIOSCULPT PTA SCORING BALLOON CATHETER (K081220)

_Angioscore, Inc. · PNO · May 28, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081220

## Device Facts

- **Applicant:** Angioscore, Inc.
- **Product Code:** [PNO](/productcode/PNO.md)
- **Decision Date:** May 28, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

## Device Story

AngioSculpt PTA Scoring Balloon Catheter is a two-lumen catheter featuring a distal nylon-blend balloon wrapped with a nitinol scoring element. Operated by physicians in clinical settings for peripheral angioplasty. During procedure, nitinol element creates focal concentrations of dilating force upon balloon inflation; this mechanism minimizes balloon slippage and assists luminal expansion of stenotic arteries. Radiopaque markers facilitate precise positioning under fluoroscopy. Device provides controlled expansion at specific pressures to treat arterial and dialysis fistula lesions. Benefits include improved procedural control and effective lesion dilation compared to standard balloon angioplasty.

## Clinical Evidence

Bench testing only. Design verification and validation testing demonstrated adequate device performance.

## Technological Characteristics

Two-lumen catheter; nylon-blend balloon; nitinol scoring element; radiopaque markers for positioning. Mechanical dilation principle. Non-powered device.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- AngioSculpt® Scoring Balloon Catheter ([K072225](/device/K072225.md))
- AngioSculpt® Scoring Balloon Catheter ([K080151](/device/K080151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 31, 2016

AngioScore, Inc. c/o Ms. Kimberley Kline Senior Regulatory Specialist 5055 Brandin Court Fremont, CA 94538

Re: K081220

Trade/Device Name: AngioSculpt PTA Scoring Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: PNO Dated: April 28, 2008 Received: April 30, 2008

Dear Ms. Kline:

This letter corrects our substantially equivalent letter of May 28, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Ms. Kimberley Kline

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Misti L. Malone -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for AngioScore. The logo consists of a stylized letter 'A' inside a square on the left, followed by the word 'AngioScore' in a distinct font. There is a small circle with a line through it, which is a registered trademark symbol, to the upper right of the word 'AngioScore'.

# INDICATIONS FOR USE STATEMENT

K 081220

510(k) Number (if known):

Device Name: AngioSculpt® Scoring Balloon Catheter

Indications for Use:

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mama R. Vclunes
(Division Sign-Off)

Division of Cardiovascular Devices

Page 1 of 1

510(k) Number_K68 | 220

5055 Brandin Court Fremont CA 94538 www.angioscore.com Toll Free: 1.510.93.7900 [9]6]619.9339993

{3}------------------------------------------------

K081220 page 1082.

Image /page/3/Picture/1 description: The image shows the word "AngioScore" with a registered trademark symbol. The word "AngioScore" is written in a bold, sans-serif font. To the left of the word is a logo that appears to be the letter "A" inside of a square. The logo and the word are both in black.

MAY 2 8 2008

# 510(k) Summary for the AngioSculpt Scoring Balloon Catheter

# 1. Submitter's Name / Contact Person

Submitter:

AngioScore, Inc. 5055 Brandin Court Fremont, CA 94538

Contact Person:

Kimberley Kline Senior Regulatory Specialist Phone: 510-933-7989 Fax: 510-933-7994

Summary Preparation Date: May 27, 2008

## 2. General Information

| Trade Name:          | AngioSculpt® PTA Scoring Balloon Catheter                      |
|----------------------|----------------------------------------------------------------|
| Common / Usual Name: | Angioplasty catheter                                           |
| Classification Name: | Percutaneous catheter                                          |
| Product Codes:       | DQY and LIT                                                    |
| Predicate Devices:   | AngioSculpt® Scoring Balloon Catheter<br>(K072225 and K080151) |

#### 3. Intended Use / Indications

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

#### 4. Device Description

The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

:

5055 Brandin Court Fremont CA 94538 www.angioscore.com Toll Free: 1.510.933.7900 Fax: 1.510.933.7901

{4}------------------------------------------------

### 5. Substantial Equivalence Comparison

The AngioSculpt catheter shares the same indications for use, fundamental design, scientific technology (operating principle), functional performance, and materials as the currently marketed AngioSculpt catheter. Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised. The changes to the subject AngioSculpt PTA Scoring Balloon Catheter do not affect the intended use of the device, alter the fundamental scientific technology of the device, or raisc new issues of safety and effectiveness. The AngioSculpt PTA Scoring Balloon Catheter is therefore, substantially equivalent to the predicate AngioSculpt catheters.

---

**Source:** [https://fda-staging.innolitics.com/device/K081220](https://fda-staging.innolitics.com/device/K081220)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).