BIOMERS TRANSLUCENT ORTHODONTIC WIRE

K081143 · Biomers Product, LLC · DYW · May 2, 2008 · Dental

Device Facts

Record IDK081143
Device NameBIOMERS TRANSLUCENT ORTHODONTIC WIRE
ApplicantBiomers Product, LLC
Product CodeDYW · Dental
Decision DateMay 2, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The BioMers Translucent Orthodontic Wire is indicated for use as an orthodontic arch wire to provide force to the teeth to effect movement in the early (leveling and aligning) stage of orthodontic treatment.

Device Story

BioMers Translucent Orthodontic Wire is a translucent arch wire used in orthodontic treatment. Device consists of glass fibers for force reinforcement, a polymer composite resin matrix, and a USP Class VI polycrystalline/amorphous linear polymer coating for abrasion resistance. Used by orthodontists in clinical settings to provide force to teeth for leveling and aligning. Mechanical properties facilitate tooth movement. Device is non-electronic and non-software based.

Clinical Evidence

Bench testing only. Device performance evaluated using ASTM D 790-03 (flexural properties), ASTM D 3916-02 (tensile properties), and ASTM D 570-98 (water absorption).

Technological Characteristics

Orthodontic arch wire composed of glass fibers, polymer composite resin, and USP Class VI polymer coating. Tested per ASTM D 790-03, ASTM D 3916-02, and ASTM D 570-98. Mechanical force application.

Indications for Use

Indicated for patients requiring orthodontic treatment during the early leveling and aligning stage to effect tooth movement.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K08//43 ## SioMers We Make You Smile MAY - 2 2008 ### 510(k) Summary (per 21 CFR 807.92) - I. Applicant BioMers Products, LLC. 2316 Pine Ridge Road, Suite 459 Naples, FL, 34109 Contact Person: George Aliphtiras, Vice President of Operations Tel: (239) 776-6344 Fax: (239) 213-0062 Email: george(a)biomersbraces.com Date Prepared: December 19, 2007 ### II. Device Name | Proprietary Name: | BioMers Translucent Orthodontic Wire | |-----------------------|---------------------------------------| | Common/ Usual Name: | Orthodontic wire | | Classification Name: | Orthodontic appliance and accessories | | Regulation Number: | 872.5410 | | Product Codes: | DZC | | Classification: | I | | Classification Panel: | Dental | #### III. Predicate Device The BioMers Translucent Orthodontic Wire is substantially equivalent to the Optiflex from Ormco Corp. and the Super Elastic Nickel Titanium Arch Wire from Acme Monaco Corp. The Optiflex was cleared by the FDA on September 15, 1989 under 510(k) K894781. The Super Elastic Nickel Titanium Arch Wire is exempt from 510(k) filings. ### IV. Intended Use of the Device The BioMers Translucent Orthodontic Wire is indicated for use as an orthodontic arch wire to provide force to the teeth to effect movement in the early (leveling and aligning) stage of orthodontic treatment. ### V. Description of the Device The BioMers Translucent Orthodontic Wire is a translucent arch wire comprised of glass fibers, a polymer composite resin, and a polymer coating. The embedded glass fibers function as the reinforcement, providing the necessary force to straighten teeth. The translucent polymer composite resin serves as the matrix, binding together the individual glass fibers. The outer coating, made of a USP Class VI polycrystalline and amorphous linear polymer, increases theabrasion resistance properties of the wire. {1}------------------------------------------------ # SioMers Me Make You Smile ### VI. Summary of the Technical Characteristics The BioMers Translucent Orthodontic Wire is an orthodontic arch wire used to provide force to the teeth to effect movement in the early (leveling and aligning) stage of orthodontic treatment. The BioMers Translucent Orthodontic Wire was designed and tested using the following standards: - ASTM D 790 03 Standard test methods for flexural properties of . unreinforced and reinforced plastics and electrical insulating materials - . ASTM D 3916 - 02 Standard test method for tensile properties of pultruded glass fiber reinforced plastic rods - . ASTM D 570 - 98 (Reapproved 2005) Standard test method for water absorption of plastics #### VII. Safety & Effectiveness There are no known substantial differences between the BioMers Translucent Orthodontic Wire defined in this 510(k) submission and the predicate devices. They have the same intended use and any differences in technological characteristics do not raise issues of safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, representing the department's role in protecting the health of all Americans. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## MAY - 2 2008 BioMers Products, LLC C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087 Re: K081143 Trade/Device Name: BioMers Translucent Orthodontic Wire Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: DYW, DZC Dated: April 21, 2008 Received: April 22, 2008 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Lehtonen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Quitte Y. Michau Ous Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## BioMers Be Malz Von Smile ### 4. Indication for Use Statement 510(k) Number (if known): Device Name: BioMers Translucent Orthodontic Wire Indications for Use: - An orthodontic arch wire used to provide force to the teeth to effect movement in . the early (leveling and aligning) stage of orthodontic treatment. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | Susan Ruppe | |--|-------------| |--|-------------| (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices | 510(k) Number: | K081143 | |----------------|---------| |----------------|---------| Page 1 of ____
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...