← Product Code [GZJ](/productcode/GZJ) · K081141

# COMBOCARE 2000 (K081141)

_Xanacare Technologies, LLC · GZJ · Aug 29, 2008 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081141

## Device Facts

- **Applicant:** Xanacare Technologies, LLC
- **Product Code:** [GZJ](/productcode/GZJ.md)
- **Decision Date:** Aug 29, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain.

## Device Story

ComboCare 2000 combines multiple therapeutic modalities into a single portable device: pulsed infrared light (870 nm), pulsed red light (640 nm), vibratory massage, and electrical stimulation (E-Stim TENS and Micro-current TENS). Device operates on rechargeable NiMH batteries or wall-mount power. Intended for pain management; used by patients or clinicians. Output includes light therapy, vibration, and electrical pulses delivered via pads. Healthcare providers use device to provide adjunctive pain relief; patient benefits from temporary reduction in muscle/joint pain and muscle relaxation. Device integrates modalities previously found in separate predicate systems.

## Clinical Evidence

No clinical data. Substantial equivalence determined via bench testing and design validation against voluntary standards: EN 60601-1 (General safety), EN 60601-1-2 (EMC), and EN 60601-2-10 (Nerve/muscle stimulator safety).

## Technological Characteristics

Materials: NiMH batteries. Energy: 100-240 VAC or 9 VDC battery. Modalities: Pulsed IR LED (870 nm), Red LED (640 nm), vibratory massage (3 levels), E-Stim/Micro-current TENS. Waveform: 8 Hz asymmetrical biphasic square wave. Pulse width: 200 μs. Connectivity: None. Sterilization: N/A. Standards: EN 60601-1, EN 60601-1-2, EN 60601-2-10, ISO 14971.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- TerraQuant MQ2000 v.5 with TQ-ITENS ([K071445](/device/K071445.md))
- Apex Medical Corp., Apex TS1211 ([K021755](/device/K021755.md))
- EMPI, EMPI Select TENS ([K061650](/device/K061650.md))
- Excalibur Light Therapy System ([K041530](/device/K041530.md))
- Lazr Pulsr 4x ([K041565](/device/K041565.md))
- GRT LITE Model 8-A Light Therapy System ([K050668](/device/K050668.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Number: _ _ K081141

Date: _

Page: 1 of 3

# 510(k) Summary

#### Introduction

This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.

## 510(k) Applicant

Xanacare Technologies LLC 9185 East Kenyon Ave., Ste. 270 Denver, CO 80237 USA Tel: 720-554-9262 / Fax: 720-554-9264

## 510(k) Correspondent

Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave., Golden, CO 80401 USA Tel: 303-463-0900 / Fax: 303-558-3833

## Date Prepared

August 19, 2008

## Trade Name of Device

ComboCare 2000

#### Common Name of Device

Lamp, non-heating, for adjunctive use in pain therapy

#### Classification Name

Infrared Lamp

## 510(k) Classification

Class II

#### Device Description and Intended Use

The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain.

{1}------------------------------------------------

| 510(k) Number: | K081141 |
|----------------|---------|
| Date:          |         |
| Page: 2 of 3   |         |

## Predicate Devices

The ComboCare 2000 does not differ significantly from predicate devices in its technological characteristics, except that it combines modalities using Red LED light, infrared LED light, therapeutic vibratory massage and E-Stim or Micro-current TENS into a single package. Modalities are similarly combined in the TerraQuant device, except that the TerraQuant uses higher power laser IR diodes and does not include a vibrator. The ComboCare 2000 has technological characteristics similar to the following predicate device(s):

- K071445 Escada International, Inc. TerraQuant MQ2000 v.5 with TQ-ITENS .
- K021755 Apex Medical Corp., Apex TS1211 .
- K061650 EMPI, EMPI Select TENS ●
- K041530 Excalibur Light Therapy System .
- . K041565 Lazr Pulsr 4x
- K050668 GRT LITE Model 8-A Light Therapy System .

#### Summary of Technological Characteristics Compared to Predicate Devices

|                         | ComboCare 2000                                                                                                             | Predicate Device                                                                                                |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Electrical Power Source | Class II Wall-Mount Battery Charger Input<br>100-240 VAC 50-60 Hz, 0.6A Output: 9 VDC<br>at 2A                             | Equivalent to other predicate devices rated for<br>international electrical use (i.e. 100-240 VAC<br>50-60 Hz). |
| Battery Type            | 4 rechargeable Size AA NiMH batteries                                                                                      | Equivalent to other predicate devices that use<br>rechargeable batteries for portability.                       |
| Infrared Light          | Pulsed infrared light emitting diodes (IR<br>LED). 24 IR LEDs @ 870 nm (standard pad),<br>48 IR LEDs @ 870 nm (large pad). | Equivalent to other predicate devices cleared<br>as non-heating lamps for adjunctive use in<br>pain therapy.    |
| Red Light               | Pulsed red light emitting diodes (RED LED).<br>32 RED LEDs @ 640 nm (standard pad), 64<br>RED LEDs @ 640 nm (large pad).   | Equivalent to other predicate devices cleared<br>as non-heating lamps for adjunctive use in<br>pain therapy     |
| Vibratory massage       | Off + 3 levels                                                                                                             | Equivalent to other predicate therapeutic<br>electrical massagers and vibrators.                                |
|                         | E-Stim TENS mode                                                                                                           |                                                                                                                 |
| Maximum Output Current  | Maximum peak output current (1 KΩ load) =<br>30 mA / Maximum peak output current (500 Ω<br>load) = 40 mA                   | K021755: 0-80 mA (500 Ω load)<br>K061650: 0-60 mA, adjustable (500 Ω load)                                      |
| Maximum Output Voltage  | Maximum peak output voltage (1 KΩ load) =<br>30 V / Maximum peak output voltage (500 Ω<br>load) = 20 V                     | K021755: 40 V (500 Ω load)<br>K061650: 0-60V (1K Ω load)                                                        |
| Pulse Frequency         | 72 pps/Hz nominal                                                                                                          | K021755: 2-150 pps/Hz<br>K061650: 2-150 pps/Hz                                                                  |
| Pulse Width             | 200 μs                                                                                                                     | K021755: 50-300 μs<br>K061650: 0-400 μs                                                                         |
|                         | Micro-Current TENS Mode                                                                                                    |                                                                                                                 |
| Maximum Output Current  | Maximum peak output current (1 KΩ load) –<br>9.3 mA / Maximum peak output current (500<br>Ω load) = 14.4 mA                | K021755: 0-80 mA (500 Ω load)<br>K061650: 0-60 mA, adjustable (500 Ω load                                       |
| Maximum Output Voltage  | Maximum peak output voltage (1 KΩ load) =<br>9.3v V / Maximum peak output voltage (500<br>Ω load) = 7.2 V                  | K021755: 40 V (500 Ω load)<br>K061650: 0-60V (1K Ω load)                                                        |
| Waveform                | 8 Hz Asymmetrical biphasic square wave                                                                                     | K021755: Asymmetrical biphasic square wave<br>K061650: Asymmetrical biphasic square wave                        |

{2}------------------------------------------------

| 510(k) Number: | K081141 |
|----------------|---------|
| Date:          |         |
| Page: 3 of 3   |         |

## Clinical & Non-Clinical Testing

Xanacare did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design against the company's specified design requirements, and to assure conformance with the following voluntary design standards:

- EN 60601-1 "Medical electrical equipment Part 1: General requirements for safety". .
- EN 60601-1-2 "Medical electrical equipment -- Part 1-2: General requirements for safety Collateral ● Standard"
- EN 60601-2-10 "Medical electrical equipment Part 2: Particular requirements for the safety of nerve . and muscle stimulators"

#### Risk Management

This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program according to standard ISO 14971.

## Substantial Equivalence

Based on the above Xanacare Technologies LLC believes that the ComboCare 2000 is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate device(s).

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

#### AUG 2 9 2008

Xanacare Technologies, LLC. % Medical Device Regulatory Advisors, Inc. Mr. Robert N. Clark President and Senior Consultant 13605 West 7th Avenue Golden, Colorado 80401

Re: K081141

Trade/Device Name: ComboCare 2000 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief. Regulatory Class: Class II Product Code: GZJ, NHN, ISA Dated: August 19, 2008 Received: August 21, 2008

Dear Mr. Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Mulkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K081141 ComboCare 2000 Device Name: _ Indications for Use: The ComboCare 2000 is intended for the temporary relief of minor muscle and joint pain, promoting relaxation of muscle tissue, symptomatic relief and management of chronic intractable pain, and adjunctive relief of post-surgical or post-traumatic acute pain. Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .............................................................................................................................................................................. waluation (ODE) Concurrence of CDRH, Office of Device E (Division Sign-C Division and Neur 510(k) Number

Page 1 of 1

ComboCare 2000 K081141/S002 Page 8 of 29

---

**Source:** [https://fda-staging.innolitics.com/device/K081141](https://fda-staging.innolitics.com/device/K081141)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).