K081095 · Quantel S.A. · GEX · Dec 12, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081095
Device Name
PROLITE III TOWER AND PR AND HR HANDPIECE
Applicant
Quantel S.A.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProLite III TM Pulsed Light System with the HR handpiece (containing a 630-1100 nm, 630-990 nm, or 750 - 1100 nm filter) is intended for the removal of unwanted hair in all Fitzpatrick skin types. The ProLite III TM Pulsed Light System with the PR handpiece (containing a 530-1100 nm, 550-1100 nm, or 580-1100 nm filter) is intended for the treatment of benign pigmented lesions and the removal of tattoos, the treatment of vascular lesions, and the treatment of shallow veins, telangiectasia, facial hemangiomas, and rosacea vascular lesions.
Device Story
ProLite III is an Intense Pulsed Light (IPL) system for dermatological use. Device consists of a tower (power supply, cooling system) and a handpiece (flash lamp, bandpass filter, waveguide). System emits light pulses (450-2000 nm) tailored by user-inaccessible filters to specific wavelengths (HR or PR handpieces). Waveguide cooled by Peltier devices (setpoint 5°C to room temp). Operated by clinicians via touchscreen GUI. Device delivers selectable pulse durations (5-120 ms) and fluence (5-30 J/cm²). Output targets hair follicles or vascular/pigmented lesions to achieve therapeutic dermatological effects.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance specifications compared to predicate devices.
Indicated for removal of unwanted hair in all Fitzpatrick skin types; treatment of benign pigmented lesions; removal of tattoos; treatment of vascular lesions including shallow veins, telangiectasia, facial hemangiomas, and rosacea.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
KO81095
# Section 5
## 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and
807.92. Summary preparation date 04-10-08 [21 CFR 807.92(a)(1)].
Contact Information [21 CFR 807.92(a)(1)] A.
Quantel USA
P O Box 8100
Bozeman, MT 59715
Tel: 406-586-0131
Fax: 406-586-2924
Contact person: Michael Johnson M.D.
#### B. Device Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: ProLite III Intense Pulsed Light System
Device Common Name: Intense Pulsed Light (IPL)
Classification Name: Laser Instrument, Surgical Powered (per 21 CFR 878.4810)
Product Code: GEX
Panel: Dermatology and Plastic Surgery
Device Classification: Class II
#### C. Predicate Devices [21 CFR 807.92(a)(3)]
The ProLite III IPL device uses similar technology and has equivalent physical output characteristics as the following predicate devices:
ProLite, manufactured by Big Sky Laser Technologies, now Quantel USA (K021304)
Dermalux, manufactured by Palomar (K010618)
4
{1}------------------------------------------------
#### D. Device Description [21 CFR 807.92(a)(4)]
The ProLite III IPL is a device designed for dermatological use. It produces an intense light pulse within the wavelengths of 450 to 2000 nanometers, which is tailored to specific indications by a bandpass filter. The system is composed of a Tower which encloses the power supply and cooling system and a Handpiece which contains the flash lamp, filter and waveguide. The device is controlled by a touchscreen graphic user interface.
#### E. Device Specifications [21 CFR 807.92(a)(6)]
The ProLite III IPL outputs a flash with selectable pulse durations of 5 to 120 milliseconds with a shot fluence adjustable in the range of 5-30 J/cm². The maximal repetition rate is 1 shot ver second. The photorejuvenation "PR" handpieces output the spectrum from 530 to 1100 nm, 550-1100 nm, or 580-1100 nm using a userinaccessible bandpass filter. The hair removal "HR" handpieces output the spectrum from 630 to 990 nm, 630 to 1100 nm, or 650 to 1100 nm using a user inaccessible bandpass filter. The waveguide area is 6 cm (1.5 x 4 cm). The waveguide is cooled by thermo-electric "Peltier" devices with the temperature setpoint adjustable from room temperature to 5℃.
#### F. Indications for Use [21 CFR 807.92(a)(5)]
The ProLite III TM Pulsed Light System with the HR handpiece (containing a 630-1100 nm, 630-990 nm, or 750 - 1100 nm filter) is intended for the removal of unwanted hair in all Fitzpatrick skin types.
The ProLite III TM Pulsed Light System with the PR handpiece (containing a 530-1100 nm, 550-1100 nm, or 580-1100 nm filter) is intended for the treatment of benign pigmented lesions and the removal of tattoos, the treatment of vascular lesions, and the treatment of shallow veins, telangiectasia, facial hemangiomas, and rosacea vascular lesions.
#### G. Conclusion [21 CFR 807.92(b)(3)]
Technologically, the ProLite III was found to be substantially equivalent to the currently cleared ProLite (K021303) and Dermalux (K021304). Thus, the risks and benefits for the ProLite III are comparable to the predicate devices.
The indications for use are exactly the same as the previously cleared pulsed light system, the ProLite (K021303).
We believe that there are no new questions of safety or efficacy raised by the introduction of the ProLite III.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Quantel USA % Michael Johnson, M.D. 601 Haggerty Lane Bozeman, Montana 59715
## DEC 1 2 2008
Re: K081095
Trade/Device Name: ProLite III Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: October 17, 2008 Received: October 20, 2008
Dear Dr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Michael Johnson, M.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Section 4
Indications for Use
510(k) Number (if known): _ _ OR 109 S
Device Name: ProLite III
Indications for Use:
The ProLite III M Pulsed Light System with the HR handpiece (containing a 630-1100 nm, or 750-1100 nm filter) is intended for the removal of unwanted hair in all Fitzpatrick skin types.
The ProLite III TM Pulsed Light System with the PR handpiece (containing a 530-1100 nm, 550-1100 mm, or 580-1100 nm filter) is intended for the treatment of benign pigmented lesions and the removal of tattoos, the treatment of vascular lesions, and the treatment of shallow veins, telangiectasia, facial hemangiomas, and rosacea vascular lesions.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
3
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Reost farmer
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.