← Product Code [DRA](/productcode/DRA) · K081049

# FLEXCATH STEERABLE SHEATH & DILATOR, MODELS 3FC10, 3FC12 (K081049)

_Cryocath Technologies, Inc. · DRA · Jun 27, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K081049

## Device Facts

- **Applicant:** Cryocath Technologies, Inc.
- **Product Code:** [DRA](/productcode/DRA.md)
- **Decision Date:** Jun 27, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1280
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The FlexCath® Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The Sheath deflection facilitates catheter positioning.

## Device Story

FlexCath® Steerable Sheath and Dilator is a deflectable catheter introducer; facilitates percutaneous catheter placement into vasculature and heart chambers. Device consists of shaft and handle; includes dilator. Used by physicians in clinical settings to assist in positioning catheters. Sheath deflection mechanism allows for controlled navigation. Benefits patient by enabling precise catheter placement during cardiac procedures. Does not perform transseptal puncture.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Deflectable catheter introducer comprising shaft and handle; includes dilator. Materials and design substantially equivalent to predicate. Class II device (21 CFR 870.1280).

## Regulatory Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

## Predicate Devices

- FlexCath® Steerable Sheath & Dilator ([K070357](/device/K070357.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081049

# 510(k) Summary CryoCath Technologies Inc. FlexCath® Steerable Sheath and Dilator

JUN 2 7 2008

#### 1. SPONSOR

CryoCath Technologies Inc. 16771 Chemin Ste-Marie Kirkland, Quebec H9H 5H3, CANADA

| Official Correspondant: | Jean-Pierre Desmarais<br>Chief Scientific Officer |
|-------------------------|---------------------------------------------------|
| Telephone:              | 514-694-1212 ext 226                              |
| Fax:                    | 514-694-7075                                      |
| Date Prepared:          | April 09, 2008                                    |

#### 2. DEVICE NAME

| Device Trade Name:     | FlexCath® Steerable Sheath & Dilator |
|------------------------|--------------------------------------|
| Common/Usual Name:     | FlexCath® Steerable Sheath & Dilator |
| Classification Name:   | Steerable Catheter, 21CFR 870.1280   |
| Device Classification: | Class II                             |

#### 3. PREDICATE DEVICE

FlexCath® Steerable Sheath & Dilator (K070357)

#### DEVICE DESCRIPTION 4.

The FlexCath® Steerable Sheath is a deflectable catheter introducer used to facilitate placing a catheter through the skin into the artery or vein.

It is comprised of the following two (2) main sections: the shaft and the handle. A dilator is also part of the package.

#### 5. Intended Use

The FlexCath® Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The Sheath deflection facilitates catheter positioning.

#### Technological Characteristics and Substantial Equivalence 6.

Technological characteristics such as product design and materials of the FlexCath® Steerable Sheath are substantially equivalent to the predicate devices. Where dimensional and material differences exist between the proposed device and the predicate devices, bench testing and biocompatibility testing demonstrated that these differences do not adversely affect safety and effectiveness.

{1}------------------------------------------------

#### 7. Performance Testing

Information submitted in this Special 510k for the FlexCath® Steerable Sheath and Dilator is substantially equivalent to the predicate devices listed.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 7 2008

CryoCath Technologies Inc. c/o Fred Milder Applied Physics 52 West Basin Ridge Galisteo, NM 87540

Re: K081049

Trade Name: FlexCath® Steerable Sheath & Dilator Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: June 5, 2008 Received: June 6, 2008

Dear Mr. Milder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-240-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

R. Lachner

ABram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

SPECIAL 510(K) - FLEXCATH® STEERABLE SHEATH & DILATOR - DEVICE MODIFICATIONS

## PREMARKET NOTIFICATION

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name: FlexCath® Steerable Sheath & Dilator

### Indications For Use:

The FlexCath® Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The Sheath deflection facilitates catheter positioning.

### Contraindications:

The FlexCath® Steerable Sheath is contraindicated for placement in the left atrium or ventricle if:

- The patient has an intra-atrial septal patch or has had other surgical intervention in or adjacent to the intra-atrial septum.
- The patient has had a previous embolic event from the left side of the heart within 2 months of the procedure.
- The patient has known or suspected atrial myxoma.

FlexCath® Steerable Sheath should not be used to perform the transseptal puncture.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|

|                                    | Duana R. Vanes |
|------------------------------------|----------------|
| (Division Sign-Off)                |                |
| Division of Cardiovascular Devices |                |

| 510(k) Number | K681049 |
|---------------|---------|
|---------------|---------|

| Page 1 of | I |
|-----------|---|
|-----------|---|

Confidential

Page viii of ix

---

**Source:** [https://fda-staging.innolitics.com/device/K081049](https://fda-staging.innolitics.com/device/K081049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
