K081001 · Sutures India Pvt., Ltd. · NEW · Jun 9, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081001
Device Name
PD SYNTH, ABSORBABLE (POLY DIOXANONE) SUTURE
Applicant
Sutures India Pvt., Ltd.
Product Code
NEW · General, Plastic Surgery
Decision Date
Jun 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PD SYNTH IS SYNTHETIC ABSORBABLE (MONOFILAMENT POLYDIOXANONE) SURGICAL SUTURE, STERILE AND FLEXIBLE STRAND, PREPARED FROM THE POLYMER, POLYDIOXANONE, (HOMO POLYMER OF DI OXANONE 100%) AND IS INDICATED FOR USE IN SOFT TISSUE APPROXIMATION, INCLUDING USE IN OPHTHALMIC SURGERY, BUT NOT FOR USE IN CARDIO VASCULAR AND NEUROLOGICAL TISSUES
Device Story
PD Synth is a sterile, flexible, monofilament absorbable surgical suture composed of polydioxanone polymer. Device is used by surgeons for soft tissue approximation. Suture is dyed with D&C Violet #2 and is uncoated. Device is supplied for single use only. Clinical benefit derived from providing temporary wound support during healing process before being absorbed by body. Healthcare providers use suture during surgical procedures to close incisions or wounds; output is physical closure of tissue. Device is equivalent to existing polydioxanone sutures in material, design, and manufacturing process.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included USP 29 performance requirements (diameter, length, knot pull tensile strength, needle attachment strength, extractable color, sterility) and ISO 10993 biocompatibility/toxicity testing.
Indicated for soft tissue approximation in patients requiring surgical sutures, including ophthalmic procedures. Contraindicated for use in cardiovascular and neurological tissues.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
PDS II Absorbable Monofilament Polydioxanone Suture (N 18331)
Unicryl M Absorbable Monofilament Polydioxanone Suture (K 042285)
Submission Summary (Full Text)
{0}------------------------------------------------
JUN - 9 2008
# SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE POLYDIOXANONE SUTURE
#### PAGE No: 17.0-1 SECTION No: 17.0
### 510k SUMMARY as required by: 21CFR 807.92
### A. APPLICANT INFORMATION
Name
#### : SUTURÈS INDIA PVT. LTD
Address
PHONE NO.
FAX NÓ.
E mail
Web Address
B. Contact Person
C. Date Prepared
#### D. DEVICE NAME
Trade Name
- Common name
SUTURES INDUSTRY P. LTD
: 472 D 13 th Cross, 4 th Phase, Peenya Industrial Area, Bangalore-560058. India
: 91-80-41868000 (30 lines)
: 91-80-41171056
: sutures(@vsnl.com
: www.suturesin com
: L.G.Chandrasekhar MANAGING DIRECTOR
: 1.4.2008
#### : PD SYNTH
: Absorbable Surgical Suture, Synthetic (Monofilament Polydioxanone)
Classification Name
: Suture, absorbable, synthetic, Polydioxanone.
### E. PREDICATE DEVICES
A. PDS II Absorbable Monofilament Polydioxanone Suture, P.M.A. Number : N 18331
- B. Unicryl M Absorbable Monofilament Polydioxanone Suture, 510(k) Number K 042285, United Medical Industries Co.Ltd (UNIMED), Riyadh, SA 11553
{1}------------------------------------------------
# SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE POLYDIOXANONE SUTURE
081001
SECTION No: 17.0 PAGE No: 17.0-2
## F. DESCRIPTION OF THE DEVICE
PD Synth is a synthetic absorbable surgical suture, (Monofilament Polydioxanone). PD Synth is a sterile flexible monofilament thread, composed of the polymer, Polydioxanone. The sutures are inert, non collageneous and non antigenic.
PD Synth is dyed with D&C Violet #2 and being monofilament it is uncoated
#### G. INTENDED USE OF THE DEVICE
PD Synth, Absorbable (Polydioxanone) suture, is indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures.
{2}------------------------------------------------
K.081001
page 3 of 5
# SUTURES INDIA PVT.LTD
# SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE POLYDIOXANONE SUTURE
#### SECTION No: 17.0 PAGE No: 17.0-3 COMPARISON TABLE OF SUTURES INDIA'S "PD SYNTH" ABSORBABLE POLYDIOXANONE SUTURE TO PREDICATE DEVICES
| Comparison parameters | PD Synth | PDS II | Unicryl M |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|--------|-----------|
| Absorbable (Polydioxanone) suture is a synthetic absorbable surgical suture. (Monofilament). It is a sterile flexible multifilament thread, composed of the homo polymer, Polydioxanone. | Same | Same | Same |
| The sutures are inert, non collageneous and non antigenic. | Same | Same | Same |
| Absorbable Polydioxanone suture is dyed with D&C Violet #2 and being monofilament it is uncoated | Same | Same | Same |
| Absorbable (Polydioxanone) suture is indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures. | Same | Same | Same |
| Absorbable Polydioxanone suture is supplied for single use only. | Same | Same | Same |
| Absorbable Polydioxanone suture is sterilized by E.O. gas method | Same | Same | Same |
| Absorbable Polydioxanone suture is packaged in the same or equivalent manner, and has the same or equivalent labeling claims as the predicate devices including indications, warnings, cautions and precautions | Same | Same | Same |
| Absorbable Polydioxanone suture meets the official monograph of the United States Pharmacopeia current edition U.S.P. 29 for extractable color. | Same | Same | Same |
| Finished suture material does not meet the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for Diameter<861> | Same | Same | Same |
| Finished suture material meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for Tensile strength<881> | Same | Same | Same |
| Comparison parameters | PD Synth | PDS II | Unicryl M |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>U.S.P. 29 for Needle attachment<871> | Same | Same | Same |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>U.S.P.29 for length requirement (95% of<br>length stated on the label) | Same | Same | Same |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>U.S.P. 29 for sterility | Same | Same | Same |
| Finished suture material packaged in a same<br>or equivalent manner with sterile single or<br>double packing having labeling conforming<br>to 21CFR and USP 29 | Same | Same | Same |
| Absorbable Polydioxanone suture is<br>biologically compatible when tested as per<br>ISO-10993 | Same | Same | Same |
| Absorbable Polydioxanone suture is tested<br>and proved to be non toxic, when tested as<br>per ISO-10993 for toxicity | Same | Same | Same |
{3}------------------------------------------------
1 :
: : 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上一篇: 上
:
:
.
:
-
:
1
のお気になる。
:
{4}------------------------------------------------
K08100
page 5 of 5
# SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE POLYDIOXANONE SUTURE
#### PAGE No: 17.0-4 SECTION No: 17.0
#### CONCLUSION
Sutures India's PD Synth Absorbable (Polydioxanone) suture is composed of the same material, as of the predicate devices and has the same design, as that of the predicate devices. The suture is manufactured in a manner typical of the industry and equivalent to that used to produce predicate devices. Further the subject device is offered with the same colorant D&C Violet No.2 at a concentration that conforms to the requirements of Title 21 CFR § 74.3602, as are of the predicate devices.
Testing of suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterility to methods outlined in U.S.P. 29 demonstrates Sutures India's PD Synth (Absorbable Polydioxanone suture) meets or exceeds U.S.P. specifications and are equivalent in terms of the above mentioned predicate devices.
# For SUTURES INDIA PVT. LTD.
L. G. CHANDRASEKHAR
MANAGING DIRECTOR
L.G.Chandrasekhar
Managing Director
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 9 2008
Sutures India Private Limited % L.G. Chandrasekhar 472 D 13th Cross, 4th Phase Peenya Industrial Area Bangalore-560058 India
Re: K081001 Trade/Device Name: PD SYNTH Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable polydioxanone surgical suture Regulatory Class: II Product Code: NEW Dated: May 22, 2008 Received: May 27, 2008
Dear L.G. Chandrasekhar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - L.G. Chandrasekhar
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mardi N. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number K081001
PD SYNTH Device Name:
> ABSORBABLE SURGICAL SUTURE (SYNTHETIC) (MONOFIL AMENT POLYDIOXANONE)
Indications For Use:
PD SYNTH IS SYNTHETIC ABSORBABLE (MONOFILAMENT POLYDIOXANONE) SURGICAL SUTURE, STERILE AND FLEXIBLE STRAND, PREPARED FROM THE POLYMER, POLYDIOXANONE, (HOMO POLYMER OF DI OXANONE 100%) AND IS INDICATED FOR USE IN SOFT TISSUE APPROXIMATION, INCLUDING USE IN OPHTHALMIC SURGERY, BUT NOT FOR USE IN CARDIO VASCULAR AND NEUROLOGICAL TISSUES
Neil R. Oslen for nxn
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_KOS 100
Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.