K080991 · DePuy Orthopaedics, Inc. · KWA · Jul 2, 2008 · Orthopedic
Device Facts
Record ID
K080991
Device Name
DEPUY ASR XL MODULAR ACETABULAR CUP SYSTEM
Applicant
DePuy Orthopaedics, Inc.
Product Code
KWA · Orthopedic
Decision Date
Jul 2, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The device is part of a modular system for use in total hip replacement in which the acetabular component articulates with a femoral component. The DePuy ASR™ XL Modular Acetabular Cup System is compatible with DePuy ASR™ femoral components. The DePuy ASR™ XL Modular Acetabular Cup System is indicated for use in the following conditions, where there is evidence of sufficient sound bone to seat and support the components: 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia. 2. Avascular necrosis of the femoral head. 3. Acute traumatic fracture of the femoral head and neck. 4. Failed previous hip surgery, including internal fixation, arthrodesis and hemiarthroplasty. 5. Certain cases of ankylosis. Porous-coated DePuy ASR™ XL Modular Acetabular Cups are indicated for cementless application.
Device Story
Modular acetabular cup system for total hip replacement; replaces natural articular surface of hip joint. Components: cobalt-chrome-molybdenum (CoCrMo) alloy one-piece cup with porous and hydroxyapatite (HA) coating; available in 64mm-70mm diameters. Optional "spiked" configuration for adjunct fixation. System includes CoCrMo femoral heads (57mm-63mm) and taper sleeve adapters for mating to femoral stems. Used in surgical setting by orthopedic surgeons. Device provides articulating surface for hip joint; restores joint function; reduces pain. Line extension of previously cleared ASR systems.
Clinical Evidence
No clinical data provided. Substantial equivalence based on design and material similarities to predicate devices.
Technological Characteristics
Materials: Cobalt-chrome-molybdenum (CoCrMo) alloy. Surface: Porous coating with hydroxyapatite (HA). Form factor: One-piece acetabular cup (64-70mm) and femoral heads (57-63mm) with taper sleeve adapters. Fixation: Cementless. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients with severely painful/disabled hip joints due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, congenital hip dysplasia, avascular necrosis of femoral head, acute traumatic fracture of femoral head/neck, failed previous hip surgery (internal fixation, arthrodesis, hemiarthroplasty), or ankylosis, provided sufficient sound bone exists for component seating/support.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Predicate Devices
DePuy ASR™ Modular Acetabular Cup System (K040627)
{0}------------------------------------------------
:
K080991 # '1/3
## 510(k) Summary (As required by 21 CFR 807.92 and 21 CFR 807.93)
JUL - 2 *2008
| NAME OF SPONSOR: | DePuy Orthopaedics, Inc.<br>700 Orthopacdic Drive<br>Warsaw, Indiana 46582<br>Establishment Registration Number: 1818910 |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER: | DePuy International Limited<br>St. Anthony Road<br>Leeds, United Kingdom LS11 8DT<br>Establishment Registration Number: 8010379 |
| 510(K) CONTACT: | Dawn Sinclair<br>Regulatory Affairs Associate<br>Telephone: (574) 372-5023<br>Facsimile: (574) 371-4987<br>Electronic Mail: Dsincla3@dpyus.jnj.com |
| DATE PREPARED: | February 25, 2008 |
| PROPRIETARY NAME: | DePuy ASR™ XL Modular Acetabular Cup<br>System |
| COMMON NAME: | Acetabular Cup Prosthesis |
| CLASSIFICATION: | Class III per 21 CFR 888.3330, Hip joint<br>metal/metal semi-constrained, with an<br>uncemented acetabular component, prosthesis |
| DEVICE PRODUCT CODE(S): | 87 KWA |
| SUBSTANTIALLY EQUIVALENT<br>DEVICE(S): | DePuy ASR™ Modular Acetabular Cup System<br>(K040627)<br>DePuy ASR™ 300 Acetabular Cup System<br>(K073413)<br>Wright Medical Metal TRANSCEND®<br>Articulation System (Larger Sizes)<br>(K021349)<br>DePuy Ultima® Unipolar Head and Adapter<br>Sleeves (K965156)<br>DePuy ASR™ Taper Sleeve Adapter (K070359)<br>Corail AMT™ Hip Prosthesis (K042992)<br>DePuy Tri-Lock® Bone Preservation Stem<br>(K073570) |
:
・
.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a handwritten text string that appears to be a combination of letters and numbers. The string reads 'K080991 #2/3'. The text is written in black ink on a white background and has a casual, handwritten style.
## DEVICE DESCRIPTION:
The subject DePuy ASR™ XL Modular Acetabular Cup components are part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The acetabular cup is designed as a cobalt-chrome-molybdenum (CoCrMo) alloy one-piece cup with a porous coating and is available in outer diameter sizes 64mm through 70mm in two-millimeter increments. The outer surface of the cup has a porous coating with the addition of a hydroxyapatite (HA) coating. There are no separate liner components to this system, as the liners are integral to the one-piece acetabular cups.
Two cup configurations will be offered: a "spiked" cup with three fixation spikes on the outer surface of the cup for adjunct fixation, and an acetabular cup with no spikes. Both configurations are specific to the DePuy ASRTM Modular Cup System cleared in K040627 and K073413. This submission is a line extension of the acetabular cup components. These acctabular cups will be compatible with DePuy ASR™ femoral components.
The uni femoral head is manufactured from cobalt-chrome-molybdenum (CoCrMo) alloy and is available in a range of diameters from 57mm to 63mm in two-millimeter increments. The uni femoral heads have an internal taper which mates with a taper sleeve adapter specific to DePuy 12/14 or 11/13 tapers. The femoral heads articulate with corresponding one-piece metal acetabular cups. The ASR Uni femoral heads (sizes
39mm through 55mm) were cleared in the DePuy ASR™ Modular Acetabular Cup System, K040627.
The subject heads use taper sleeve adapters to mate the DePuy femoral heads to DePuy femoral stems and are manufactured from cobalt-chrome-molybdenum (CoCrMo) alloy.
#### INDICATIONS AND INTENDED USE:
#### Indications for Use:
The DePuy ASR™ XL Modular Acetabular Cup System is indicated for use in the following conditions, where there is evidence of sufficient sound bone to seat and support the components:
- 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head and neck.
- 4. Failed previous hip surgery, including internal fixation, arthrodesis and hemiarthroplasty.
- 5. Certain cases of ankylosis.
{2}------------------------------------------------
080991 # 3/3
Porous-coated DePuy ASR™XL Modular Acetabular Cups are indicated for cementless application.
#### Intended Use:
The device is part of a modular system for use in total hip replacement in which the acetabular component articulates with a femoral component.
The DePuy ASR™ XL Modular Acetabular Cup System is compatible with DePuy ASR™ femoral components.
# BASIS OF SUBSTANTIAL EQUIVALENCE:
The DePuy ASR" XL Modular Acetabular Cup components described in this submission are, in our opinion, substantially equivalent to those included in the previously cleared DePuy ASR" Modular Acetabular Cup System (K040627); the DePuy ASR" 300 Acetabular Cup System (K073413); the Wright Medical Metal TRANSCEND® Articulation System (K021349); the DePuy ASR™ Adapter Sleeve (K070359); the DePuy Ultima "Adapter Sleeves (K965156); the Corail AMT" Hip Prosthesis (K042992); and the DePuy Tri-Lock® Bone Preservation Stem (K073570), based upon the similarities in design, material composition and intended use/indications for use. A modification is simply being made to add additional cup sizes. In addition, minor revisions are being made to the Indications for Use and the Instructions for Use (IFU) to mininimize the necessity for multiple IFUs and to update the contents to reflect current practice unc subject device does not raise any new issues of safety or effectiventess.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 2 2008
DePuy Orthopaedics, Inc. % Ms. Dawn Sinclair Regulatory Affairs Associate 700 Orthopaedic Drive Warsaw, Indiana 46582
Re: K080991
Trade/Device Name: DePuy ASRIM XL Modular Acetabular Cup System Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA Dated: March 17, 2008 Received: April 7, 2008
Dear Ms. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Dawn Sinclair
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Miller
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
| 510 (k) Number (if known): | K080991 |
|----------------------------|---------|
|----------------------------|---------|
Device Name:
## Indications for Use:
The DePuy ASR™ XL Modular Acetabular Cup System is indicated for use in the following conditions, where there is evidence of sufficient sound bone to seat and support the components:
- 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
- 2. Avascular necrosis of the femoral head.
- 3. Acute traumatic fracture of the femoral head and neck.
- 4. Failed previous hip surgery, including internal fixation, arthrodesis and hemiarthroplasty.
- 5. Certain cases of ankylosis.
Porous-coated DePuy ASR™ XL Modular Acetabular Cups are indicated for cementless application.
| Prescription Use | X | AND/OR | Over-The-Counter Use |
|-----------------------------|---|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 807 Subpart C) |
(Please do not write below this line. Continue on another page if needed.)
| Concurrence of CDRH, Office (Division Sign-Off) (ODE) | |
|-----------------------------------------------------------|-------------|
| (Posted November 13, 2003) | |
| Division of General, Restorative and Neurological Devices | Page 1 of 1 |
| 510(k) Number | 1080997 |
Page 14 of 152
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.