FRAXEL RE:PAIR (FRAXEL III SR) LASER SYSTEM AND ACCESSORIES
K080915 · Reliant Technologies, Inc. · GEX · Jul 1, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080915
Device Name
FRAXEL RE:PAIR (FRAXEL III SR) LASER SYSTEM AND ACCESSORIES
Applicant
Reliant Technologies, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 1, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fraxel re:pair (Fraxel III SR) Laser System, including its resurfacing handpiece delivery system, is intended for use in: -Dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. -Treatment of wrinkles, rhytides, furrows, fine lines; -Pigmented lesions; -Textural irregularities; -Vascular dyschromia. The Fraxel re:pair (Fraxel III SR) Laser System, including its incisional and debulking delivery systems, are intended for ablation (removal), for the reduction, treatment and/or removal of: - Actinic and seborrheic keratoses; -Chelitis; -Cutaneous horns; -Cutaneous horns; -Hemangiomas; -Keloids; -Nevi, including spider, epidermal and protruding; -Rhinophyma; -Syringomas; -Warts; -Laser incision and/or excision for the performance of upper and lower eyelid blepharoplasty and vermilionectomy of the lip.
Device Story
Fraxel re:pair (Fraxel III SR) is a CO2 laser surgical system used for dermatological ablation, resurfacing, and coagulation. System comprises CO2 laser source, articulating arm, and three interchangeable handpieces (resurfacing, incisional, debulking) with associated treatment tips and spatulas. Operated by physicians in clinical settings. Device delivers laser energy to target soft tissue; enables precise incision, excision, or resurfacing. Output allows clinicians to treat various skin conditions and perform minor surgical procedures like blepharoplasty. Benefits include controlled tissue removal and coagulation, facilitating skin rejuvenation and lesion management.
Clinical Evidence
Clinical performance supported by non-significant risk and investigational device exemption studies for the resurfacing handpiece. Substantial equivalence for incisional and debulking handpieces was established via comparison to predicate performance specifications.
Technological Characteristics
CO2 laser surgical system; energy type: laser; delivery via articulating arm with interchangeable handpieces (resurfacing, incisional, debulking) and spatulas. Class II device. Technical specifications and operating principles are equivalent to predicate CO2 laser systems.
Indications for Use
Indicated for patients requiring dermatological soft tissue ablation, resurfacing, and coagulation, including treatment of wrinkles, rhytides, furrows, fine lines, pigmented lesions, textural irregularities, vascular dyschromia, actinic/seborrheic keratoses, chelitis, cutaneous horns, hemangiomas, keloids, nevi, rhinophyma, syringomas, warts, and eyelid blepharoplasty or vermilionectomy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fraxel re:pair (Fraxel III SR) Laser System and Accessories (K063038, K071051)
Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device Accessories (K022060, K030147)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### REGULATORY AUTHORITY
Safe Medical Devices Act of 1990, 21 CFR 807.92
#### COMPANY NAME/CONTACT
Heather MacFalls Reliant Technologies, Inc. 464 Ellis St. Mountain View, CA 94043 650 605-2257 650 605-2057 fax hmacfalls@fraxel.com
## NAME OF DEVICE
Trade Name:
Common Name: Regulation Number Product code: Device Panel: Device Classification: Fraxel re:pair™ (Fraxel III SR) Laser System and Accessories Laser Surgical Instrument 878.4810 GEX General Surgery/Restorative Devices Class II
## LEGALLY MARKETED PREDICATE DEVICES
Name: Fraxel re:pair (Fraxel III SR) Laser System and Accessories 510(k) #: K063038, K071051
Name: Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device Accessories 510(k) #: K022060, K030147
## DEVICE DESCRIPTION
The Fraxel re:pair Laser System consists of a CO2 laser source and three delivery handpieces for resurfacing, incision and debulking procedures. Each handpiece attaches to an articulating arm. Additional device accessories include interchangeable resurfacing treatment tips, incisional handpiece spatulas, and debulking handpiece spatulas.
{1}------------------------------------------------
Koyogll'
#### INDICATION FOR USE STATEMENT
The Fraxel re:pair (Fraxel 111 SR) Laser System, including its resurfacing handpiece delivery system_is intended for use in:
-Dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue.
-Treatment of wrinkles, rhytides, furrows, fine lines;
-Pigmented lesions;
-Textural irregularities:
- Vascular dyschromia.
The Fraxel re:pair (Fraxel III SR) Laser System, including its incisional and debulking delivery systems, are intended for ablation (removal), for the reduction, treatment and/or removal of:
- Actinic and seborrheic keratoses;
-Chelitis;
-Cutaneous horns;
-Hemangiomas;
-Keloids;
-Nevi, including spider, epidermal and protruding;
-Rhinophyma:
-Syringomas;
-Warts:
-Laser incision and/or excision for the performance of upper and lower evelid blepharoplasty and vermilionectomy of the lip.
## SUBSTANTIAL EQUIVALENCE COMPARISON
#### Indications for Use
Substantial equivalence for the Fraxel re:pair Laser System and Accessories is supported by the predicate devices listed in this submission, which have identical or similar indications for use statements and equivalent performance specifications.
#### Clinical Performance Data
Non-Significant Risk and Investigational Device Exemption studies to support the clinical performance of the Fraxel re:pair Laser System and its resurfacing handpiece. The substantial equivalence of the incisional and debulking handpieces supported the determination of safety and effectiveness for the Fraxel re:pair Laser System and Accessories. Unique device features do not introduce new issues of safety and effectiveness.
{2}------------------------------------------------
\$\kcirc\$ \$80\$ \$91/\$
## Technological Characteristics
Key technological characteristics for the resurfacing, incisional and debulking delivery systems for the Fraxel re:pair Laser System, such as energy type and operating principle, are equivalent to the Fraxel re:pair Laser System as described in submissions K063038 and K071051 and to the Lumenis Ultrapulse Encore Laser System as described in K020660 and K030147.
## CONCLUSION
Based on the design, materials, function, specifications and intended use, the Fraxel re:pair Laser System and Accessories is substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. Safety and effectiveness are reasonably assured, justifying 510(k) clearance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 1 2008
Reliant Technologies, Inc. % Ms. Heather MacFalls Clinical and Regulatory Affairs 464 Ellis Street Mountain View, California 94043
Re: K080915
Trade/Device Name: Fraxel re:pair (Fraxel III SR) Laser System and accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 31, 2008 Received: April 02, 2008
Dear Ms. MacFalls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Heather MacFalls
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Wilkinson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): k080915
Fraxel re:pair (Fraxel !!! SR) Laser System and accessories Device Name: Indications For Use:
The Fraxel re:pair (Fraxel III SR) Laser System, including its resurfacing handpiece delivery system, is intended for use in:
-Dermatological procedures requiring ablation (removal), resurfacing and coagulation of soft tissue. -Treatment of wrinkles, rhytides, furrows, fine lines;
-Pigmented lesions;
-Textural irregularities;
-Vascular dyschromia.
The Fraxel re:pair (Fraxel II) SR) Laser System, including its incisional and debulking delivery systems, are intended for ablation (removal), for the reduction, treatment and/or removal of:
- Actinic and seborrheic keratoses:
-Chelitis:
-Cutaneous horns:
-Hemangiomas;
-Keloids;
-Nevi, including spider, epidermal and protruding;
-Rhinophyma;
-Syringomas:
-Warts;
-Laser incision and/or excision for the performance of upper and lower eyelid blepharoplasty and vermilionectomy of the lip.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)Concurrence of CDRH, Office of
Neuro:ion. Iu sea
11. of D918
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.