EASYRA ALB REAGENT AND EASYRA ALP REAGENT, MODEL 10218/10214, EASYRA AST REAGENT, MODEL 10206, EASYRA CO2 REAGENT
Applicant
Medica Corp.
Product Code
CIX · Clinical Chemistry
Decision Date
Feb 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1035
Device Class
Class 2
Indications for Use
The EasyRA Albumin (ALB) reagent is intended for the quantitative determination of Albumin in human serum, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. The EasyRA Alkaline Phosphatase (ALP) reagent is intended for the quantitative determination of Alkaline Phosphatase in human serum, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Measurement of Alkaline Phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases. The EasyRA Aspartate Aminotranferase (AST) reagent is intended for the quantitative determination of the enzyme Aspartate Aminotranferase in human serum, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Measurement of Aspartate Aminotranferase measurements are used in the diagnosis and treatment of certain types of liver and heart diseases. The EasyRA Carbon Dioxide (CO2) reagent is intended for the quantitative measurement of Carbon Dioxide (CO2) in human serum, Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. The Carbon Dioxide calibrator establishes points of reference that are used in the determination of values in the measurement of Bicarbonate/Carbon dioxide on the EasyRA clinical chemistry analyzer when used in conjunction with Medica's CO2 reagent. The EasyRA CREA reagent is a device intended to measure creatinine levels in serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis and as a calculation basis for measuring other urine analytes. The EasyRA CK reagent is a device intended to measure creatinine kinase activity in serum. Measurements of CK are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Device Story
System consists of liquid, ready-to-use reagents for quantitative colorimetric chemistry analysis on Medica EasyRA Chemistry Analyzer. Input: human serum samples. Operation: reagents react with analytes (enzymatic or non-enzymatic); analyzer measures absorbance changes via spectrophotometry (Beer's Law). Output: quantitative concentration values for ALB, ALP, AST, CO2, CREA, and CK. Used in clinical laboratories by trained personnel. Results assist clinicians in diagnosing/monitoring liver, kidney, heart, and muscle diseases and acid-base status.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing of analytical performance characteristics.
Technological Characteristics
Quantitative colorimetric chemistry reagents. Dual or single reagent systems. Sensing principle: spectrophotometric absorbance measurement (Beer's Law). Energy source: integrated into EasyRA analyzer. Storage: 2–8 °C. Connectivity: proprietary to Medica EasyRA Analyzer. Standards: CLSI EP5-A2, EP6-A, EP9-A2, EP17-A.
Indications for Use
Indicated for quantitative determination of Albumin, Alkaline Phosphatase, Aspartate Aminotransferase, Carbon Dioxide, Creatinine, and Creatine Kinase in human serum for diagnosis and treatment of liver, kidney, bone, parathyroid, intestinal, heart, and muscle diseases, and acid-base balance disorders. Prescription use only.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k080874
B. Purpose for Submission:
New device
C. Measurand:
Albumin (ALB), Alkaline Phosphatase (ALP), Aspartate Aminotransferase (AST), Total Carbon Dioxide (CO2), Creatinine (CREA) and Creatine Kinase (CK)
D. Type of Test:
Quantitative colorimetric chemistry tests
E. Applicant:
Medica Corporation
F. Proprietary and Established Names:
EasyRA Albumin Reagent
EasyRA Alkaline Phosphatase Reagent,
EasyRA Aspartate Aminotransferase Reagent,
EasyRA Carbon Dioxide (CO2) Reagent,
EasyRA Carbon Dioxide (CO2) Calibrator,
EasyRA Creatinine Reagent
EasyRA Creatine Kinase Reagent
{1}
G. Regulatory Information:
| Device Classification Name | Regulation section | Classification | Product Code | Panel |
| --- | --- | --- | --- | --- |
| Albumin | 21 CFR § 862.1035 | II | CIX | Chemistry (75) |
| Alkaline Phosphatase | 21 CFR § 862.1050 | II | CJE | Chemistry (75) |
| Aspartate Aminotransferase | 21 CFR § 862.1100 | II | CIT | Chemistry (75) |
| Bicarbonate/carbon dioxide | 21 CFR § 862.1160 | II | KHS | Chemistry (75) |
| Calibrator | 21 CFR § 862.1150 | II | JIT | Chemistry (75) |
| Creatinine | 21 CFR § 862.1225 | II | JFY | Chemistry (75) |
| Creatine Kinase | 21 CFR § 862.1215 | II | CGS | Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
**EasyRA Albumin**
The EasyRA Albumin (ALB) reagent is intended for the quantitative determination of Albumin in human serum, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
**EasyRA Alkaline Phosphatase**
The EasyRA Alkaline Phosphatase (ALP) reagent is intended for the quantitative determination of Alkaline Phosphatase in human serum, using the MEDICA "EasyRA Chemistry Analyzer" in clinical laboratories. Measurement of Alkaline Phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases.
**EasyRA Aspartate Aminotransferase**
The EasyRA Aspartate Aminotransferase (AST) reagent is intended for the quantitative
{2}
determination of the enzyme Aspartate Aminotransferase in human serum, using the MEDICA “EasyRA Chemistry Analyzer” in clinical laboratories. Measurement of Aspartate Aminotransferase measurements are used in the diagnosis and treatment of certain types of liver and heart diseases.
## EasyRA Carbon Dioxide
The EasyRA Carbon Dioxide (CO2) reagent is intended for the quantitative measurement of Carbon Dioxide (CO2) in human serum. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
## EasyRA Carbon Dioxide Calibrator
The Carbon Dioxide calibrator establishes points of reference that are used in the determination of values in the measurement of Bicarbonate/Carbon dioxide on the EasyRA clinical chemistry analyzer when used in conjunction with Medica’s CO2 reagent.
## EasyRA Creatinine
The EasyRA CREA reagent is a device intended to measure creatinine levels in serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis and as a calculation basis for measuring other urine analytes.
## EasyRA Creatine Kinase
The EasyRA CK reagent is a device intended to measure creatinine kinase activity in serum. Measurements of CK are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
Medica EasyRA Analyzer
I. Device Description:
The EasyRA Alkaline Phosphatase, EasyRA Aspartate Aminotransferase, EasyRA Creatinine, and EasyRA Creatine Kinase reagents are dual reagent systems containing reagents, stabilizers and/or diluents. The EasyRA Albumin and EasyRA CO2 are single reagents containing reagent, stabilizers and/or diluents. All of the above listed reagents are for use on the Easy RA analyzer.
3
{3}
The EasyRA CO2 calibrator is an aqueous calibrator with pre-assigned value for CO2 used for a one point calibration. It is provided in a ready to use dripper bottle with end-user instructions for use on the EasyRA clinical chemistry analyzer.
J. Substantial Equivalence Information:
| New Device | Predicate Device Name | Predicate K Number(s) |
| --- | --- | --- |
| EasyRA Albumin Reagent | Stanbio LiquiColor Albumin | k771771 |
| EasyRA Alkaline Phosphatase Reagent | Roche Alkaline Phosphatase Reagent | k902560 |
| EasyRA Aspartate Aminotransferase Reagent | Roche AST Reagent | k924244 |
| EasyRA CO2 Reagent | Roche CO2 Reagent | k844987 |
| EasyRA CO2 Calibrator | Raichem Carbon Dioxide Calibrator | k864819 |
| EasyRA Creatinine Reagent | Stanbio Creatinine LiquiColor Liquicolor Reagent | k050283 |
| EasyRA Creatine Kinase | Roche Creatine Kinase Reagent | k891462 |
Comparison with predicate:
| | Albumin | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica ALB Reagent | Stanbio Albumin Reagent |
| Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of albumin in human serum, using the EasyRA chemistry analyzer. | Clinical chemistry reagent used to provide a quantitative measurement of albumin in human serum, using an automated chemical analyzer |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 0.4 – 7.0 g/dL | Up to 7 g/dL |
| Wavelength | 600/700 nm | 550 nm |
| Reaction type | End point | End point |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | Non-enzymatic reaction based on the binding of the analyte to an organic compound to produce a chromogen at endpoint. The chromogen absorbs light of specific wavelength, where the EasyRA measures absorbance according to Beer's law. | Non-enzymatic reaction based on the binding of the analyte to an organic compound to produce a chromogen at endpoint. The chromogen absorbs light of specific wavelength, where the COBAS-Mira measures absorbance according to Beer's law. |
{4}
| | Alkaline Phosphatase (ALP) | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica ALP Reagent | Roche Alkaline phosphatase Reagent |
| Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of alkaline phosphatase in human serum, using the EasyRA chemistry analyzer. | Clinical chemistry reagent used to provide a quantitative measurement of alkaline phosphatase in human serum, using an automated chemical analyzer |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 8 – 800 U/L | Up to 1500 U/L |
| Wavelength | 405 nm | 405 nm |
| Reaction type | Rate | Rate |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the EasyRA measures absorbance according to Beer's law. | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the COBAS-Mira measures absorbance according to Beer's law. |
| | Aspartate Aminotransferase(AST) | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica AST Reagent | Roche AST Reagent |
| Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of AST in human serum, using the EasyRA chemistry analyzer. | Clinical chemistry reagent used to provide a quantitative measurement of AST in human serum, using an automated chemical analyzer |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 5.5 – 400 U/L | Up to 440 U/L |
| Wavelength | 550 nm | 550 nm |
| Reaction type | Rate | Rate |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the EasyRA | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the COBAS-Mira |
{5}
| Carbon Dioxide CO2 | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | EasyRA CO2 Reagent | Roche Carbon Dioxide Reagent |
| Intended Use | Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. | Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 2.3 – 45.0 mmol/L | Up to 40 mmol/L |
| Wavelength | 405 nm | 340 nm |
| Reaction type | End point | End point |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | An enzymatic reaction based on the carboxylation of Phosphoenolpyruvate (PEP). The product of this reaction is oxidized and a chromogen is produced. The rate of chromogen loss is monitored by the EasyRA over time. | An enzymatic reaction based on the carboxylation of Phosphoenolpyruvate (PEP). The product of this reaction is oxidized and a chromogen is produced. The rate of chromogen loss is monitored by the Cobas-Mira over time. |
| CO2 Calibrator | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica CO2 Calibrator | RAICHEM CO2 Calibrator |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | The calibrator is used to establish the calibration factor in Beer's equation using a know concentration reagent and the measured absorbance. The calibration factor is used to determine the CO2 concentration in the patient sample. | The calibrator is used to establish the calibration factor in Beer's equation using a know concentration reagent and the measured absorbance. The calibration factor is used to determine the CO2 concentration in the patient sample. |
{6}
| | Creatinine | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica CREA Reagent | Stanbio Creatinine Reagent |
| Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of Creatinine in human serum, using the EasyRA chemistry analyzer. | Clinical chemistry reagent used to provide a quantitative measurement of Creatinine in human serum, using an automated chemical analyzer |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 0.2 – 15.0 mg/dL | Up to 30 mg/dL |
| Wavelength | 550/700 nm | 550 nm |
| Reaction type | End point | End point |
| Reagent storage | 2 – 8 °C | 2 – 8 °C |
| Test Methodology | An enzymatic reaction based on the conversion of creatinine to glycine and hydrogen peroxide. The latter, is reduced in the presence of peroxidase to form a chromogen. The chromogen absorption is monitored by the EasyRA and is directly related to the amount of Creatinine present in the sample. | An enzymatic reaction based on the conversion of creatinine to glycine and hydrogen peroxide. The latter, is reduced in the presence of peroxidase to form a chromogen. The chromogen absorption is monitored by the Cobas-Mira and is directly related to the amount of Creatinine present in the sample. |
| | Creatine Kinase (CK) | |
| --- | --- | --- |
| Item | Device | Predicate |
| Product Attribute | Medica CK Reagent | Roche CK Reagent |
| Intended Use | Clinical chemistry reagent used to provide a quantitative measurement of CK in human serum, using the EasyRA chemistry analyzer. | Clinical chemistry reagent used to provide a quantitative measurement of CK in human serum, using an automated chemical analyzer |
| Sample | Serum | Serum |
| Reagent type | Liquid ready-for-use | Liquid ready-for-use |
| Linearity range | 3 – 1200 U/L | Up to 2000 U/L |
| Wavelength | 340 nm | 340 nm |
| Reaction type | Rate | Rate |
| Reagent | 2 – 8 °C | 2 – 8 °C |
{7}
8
| storage | | |
| --- | --- | --- |
| Test Methodology | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the EasyRA measures absorbance according to Beer’s law. | An enzymatic reaction based on the catalytic activity of a serum based enzyme, which is quantified by monitoring the reaction rate. The chromogen absorbs light of specific wavelength, where the COBAS-Mira measures absorbance according to Beer’s law. |
## K. Standard/Guidance Document Referenced (if applicable):
Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline- Second Edition (CLSI EP5-A2)
Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (CLSI EP6-A)
Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition (CLSI EP9-A2)
Protocols for Determination of Limits f Detection and Limits of Quantitation; Approved Guideline (CLSI EP17-A)
## L. Test Principle:
### EasyRA Albumin
Albumin binds with bromocresol green to produce a blue-green color. This color is measured at 600 nm with a blanking wavelength of 700 nm. The color intensity is proportional to the concentration of Albumin in the sample.
### EasyRA Alkaline Phosphatase
ALP hydrolyses 4-nitrophenyl phosphate substrate to form 4-nitrophenol and phosphates. The 4-nitrophenol component is yellow in color at pH 10.4, with an absorbance peak at 405 nm. The rate of the formation of 4-nitrophenol is directly proportional to the Alkaline Phosphatase activity in the sample.
### EasyRA Aspartate Aminotransferase
The AST enzyme catalyzes the transfer of the Aspartate amino group to α-Ketoglutarate (α-KG) with the formation of L-glutamate and oxaloacetate (OAA). The OAA is then reduced to L-malate by reacting with NADH in a reaction catalyzed by malate dehydrogenase (MDH). In this second reaction the amount of NADH that is oxidized to NAD results in a decrease in absorbance at 340 nm. This decrease is
{8}
followed spectrophotometrically and is directly proportional to the activity of AST in serum.
## EasyRA Carbon Dioxide
Carbon dioxide (in the form of bicarbonate ion) reacts with phosphoenolpyruvate (PEP), catalyzed by phosphoenol pyruvate carboxylase (PEPC), to form oxaloacetate and phosphate. The oxaloacetate, in turn, is oxidized by malate dehydrogenase (MDH) in the presence of NADH or an analog. The decrease in NADH analog concentration is monitored at 405 nm and is proportional to the total carbon dioxide in the sample.
## EasyRA Creatinine
The method uses multi-step enzymatic reactions. The creatinine is hydrolyzed to sarcosine with Creatinine Amidohydrolase. The sarcosine is then oxidized via sarcosine oxidase to produce glycine and hydrogen peroxide. The hydrogen peroxide is reacted with 4-aminoantipyrene and N-ethyl-N-sulfopropyl-m-toluidine (ESPMT) to produce a quinoneimine dye. The reaction is monitored at 550 nm. The increase in absorbance is proportional to the level of creatinine in the sample.
## EasyRA Creatine Kinase
CK catalyzes transfer of phosphate groups from creatine phosphate to ADP to form ATP. The rate of production of ATP is measured by two coupled reactions. Hexokinase (HK) catalyzes the reaction of Glucose and ATP to form glucose-6-phosphate. The glucose-6-phosphate with NAD in the presence of glucose-6-phosphate dehydrogenase (G6PDH) forms 6-phosphogluconate and NADPH. The increase in NADPH absorbance at 340 nm is directly related to CK activity in the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Within-Run and Total precision evaluations were determined following CLSI EP5-A2. Three levels of commercial serum-based Quality Control material were tested on one EasyRA analyzer twice a day over a twenty-day period. In addition, a within-run precision study was performed in the extended measuring range to verify precision.
## EasyRA Albumin
9
{9}
10
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 2.66 | 4.58 | 3.06 |
| Within Run Precision: | | | |
| Std. Dev. | 0.03 | 0.03 | 0.03 |
| CV % | 1.21 | 0.62 | 0.91 |
| Total Precision: | | | |
| Std. Dev. | 0.03 | 0.05 | 0.03 |
| CV % | 1.27 | 1.18 | 0.97 |
## EasyRA ALP
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 170 | 85 | 353 |
| Within Run Precision: | | | |
| Std. Dev. | 2.4 | 3.0 | 4.4 |
| CV % | 1.4 | 3.5 | 1.3 |
| Total Precision: | | | |
| Std. Dev. | 3.6 | 3.5 | 7.4 |
| CV % | 2.1 | 4.1 | 2.1 |
## EasyRA AST
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 35.5 | 31.2 | 132.6 |
| Within Run Precision: | | | |
| Std. Dev. | 1.05 | 1.34 | 1.60 |
| CV | 2.96 | 4.28 | 1.21 |
| Total Precision: | | | |
| Std. Dev. | 1.05 | 1.46 | 1.86 |
| CV | 2.97 | 4.69 | 1.40 |
## EasyRA CO2
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mmol/L) | 27.8 | 32.2 | 18.3 |
| Within Run Precision: | | | |
| Std. Dev. | 0.48 | 0.35 | 0.30 |
| CV % | 1.73 | 1.09 | 1.63 |
| Total Precision: | | | |
| Std. Dev. | 1.49 | 1.71 | 0.93 |
| CV % | 5.34 | 5.30 | 5.05 |
{10}
11
# EasyRA CREA
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 0.72 | 1.24 | 6.73 |
| Within Run Precision: | | | |
| Std. Dev. | 0.01 | 0.01 | 0.04 |
| CV % | 1.25 | 1.07 | 0.66 |
| Total Precision: | | | |
| Std. Dev. | 0.02 | 0.02 | 0.09 |
| CV % | 2.50 | 1.50 | 1.30 |
# EasyRA CK
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (U/L) | 96 | 191 | 500 |
| Within Run Precision: | | | |
| Std. Dev. | 1.1 | 1.6 | 4.3 |
| CV % | 1.1 | 0.9 | 0.9 |
| Total Precision: | | | |
| Std. Dev. | 1.7 | 4.7 | 10.8 |
| CV % | 1.7 | 2.5 | 2.2 |
b. Linearity/assay reportable range:
# EasyRA Albumin
Linearity was determined using the guidelines provided by CLSI EP6-A. Commercially available albumin standards were used. The sponsor obtained values from 0.3 mg/dL to 8.3 mg/dL. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the third order model demonstrating the best fit. The results are summarized below.
Fitted regression models are:
Linear: $y = 1.04624x - 0.2886$, $R^2 = 0.9957$, Std. Error = 0.178
Second order: $y = -0.0046x^2 + 1.0861x - 0.3497$, $R^2 = 0.9958$, Std. Error = 0.184
Third order: $y = -0.0116x^3 - 0.1569x^2 + 1.6336x - 0.8015$, $R^2 = 0.9979$, Std. Error = 0.138
{11}
The results of this study support the sponsor's claim that the device is linear from $0.4\mathrm{g / dL}$ to $7.0~\mathrm{g / dL}$.
## EasyRA ALP
Linearity was determined using the guidelines provided by CLSI EP6-A. Commercially available alkaline phosphatase standards were used. These seven standards had target values from $8.0\mathrm{U} / \mathrm{L}$ to $817\mathrm{U} / \mathrm{L}$. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the first order regression demonstrating the best fit. The results are summarized below.
Fitted regression models are:
Linear: $y = 0.9255x + 3.9244, R^2 = 0.9992$, Std. Error = 8.503
Second order: $y = -9E - 06x^{2} + 0.9327x + 3.2684, R^{2} = 0.9992$, Std. Error = 8.856
Third order: $y = 1E - 08x^{3} - 3E - 05x^{2} + 0.9386x + 0.3.0036,$
$R^2 = 0.9992$, Std. Error = 9.285
The results of this study support the sponsor's claim that the device is linear from $8\mathrm{U} / \mathrm{L}$ to $800\mathrm{U} / \mathrm{L}$.
## EasyRA AST
Commercially available standards ranging in value from $6.4\mathrm{U} / \mathrm{L}$ to $450.8\mathrm{U} / \mathrm{L}$ were used for the study. The sponsor obtained values from 3.3-463.3 U/L. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the third order regression demonstrating the best fit. The results are summarized below.
Fitted regression models are:
Linear: $y = 1.0215x - 4.2094, R^2 = 0.9991$, Std. Error = 2.646
Second order: $y = 0.0002x^{2} + 0.9502x + 0.0242, R^{2} = 0.9995$, Std. Error = 4.099
Third order: $y = 1E - 06x^{3} - 0.0007x^{2} + 1.1028x - 4.217,$
$R^2 = 0.9998$, Std. Error = 2.646
The results of this study support the sponsor's claim that the device is linear from $5.5\mathrm{U} / \mathrm{L}$ to $400\mathrm{U} / \mathrm{L}$.
## EasyRA CO2
Seven linearity standards were gravimetrically prepared covering the target range of $2 - 52\mathrm{mmol / L}$. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the second order regression demonstrating the best fit. The results are summarized
{12}
below.
Fitted regression models are:
Linear: $y = 0.9911x + 17.661$, $R^2 = 0.9988$, Std. Error = 0.340
Second order: $y = -7E - 05x^2 + 1.0811x + 8.2543$, $R^2 = 0.9993$, Std. Error = 0.270
Third order: $y = 2E - 07x^3 + 0.0005x^2 + 1.2763x - 2.318$, $R^2 = 0.9999$, Std. Error = 0.273
The results of this study support the sponsor's claim that the device is linear from $2.3\mathrm{mmol / L}$ to $45\mathrm{mmol / L}$.
## EasyRA CREA
Six commercially available standards ranging in value from $0.20\mathrm{mg / dL}$ to $15.20\mathrm{mg / dL}$ were used for the study. The sponsor obtained values from 0.00 mg/dL to $15.89\mathrm{mg / dL}$. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the third order regression demonstrating the best fit. The results are summarized below.
Fitted regression models are:
Linear: $y = 1.0387x - 0.1646$, $R^2 = 0.9975$, Std. Error = 0.296
Second order: $y = 0.0036x^2 + 0.984x - 0.0541$, $R^2 = 0.9977$, Std. Error = 0.297
Third order: $y = 0.00134x^3 - 0.0269x^2 + 11608x - 0.1974$, $R^2 = 0.9982$, Std. Error = 0.285
The results of this study support the sponsor's claim that the device is linear from $0.2\mathrm{mg / dL}$ to $15.00\mathrm{mg / dL}$.
## EasyRA CK
Nine commercially available standards ranging in value from $2.5\mathrm{U} / \mathrm{L}$ to $1353.4\mathrm{U} / \mathrm{L}$ were used for the study. The data was analyzed using linear regression as well as second and third order non-linear fitted polynomial regression with the third order regression demonstrating the best fit. The results are summarized below.
Fitted regression models are:
Linear: $y = 1.0505x - 0.0033$, $R^2 = 0.9997$, Std. Error = 17.763
Second order: $y = -0.0002x^2 + 1.0517x - 0.0042$, $R^2 = 0.9997$, Std. Error = 14.135
Third order: $y = 0.0014x^3 - 0.0134x^2 + 1.0814x - 0.0127$, $R^2 = 0.9997$, Std. Error = 5.007
{13}
The results of this study support the sponsor's claim that the device is linear from 3 U/L to 1,200 U/L.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
1) Traceability of CO2 Calibrator: The CO2 calibrator is a ready-to-use liquid, single level calibrator. The sponsor provided a certificate of analysis showing that the calibrator was verified to NIST material.
2) Calibrator and Reagent Stability: The sponsor performed stability studies at 2°C to 8°C and found them to be adequate for the CO2 Calibrator, Albumin, ALP, AST, CO2, Creatinine, and CK reagents. Accelerated stability studies were performed by the sponsor to verify stability in the sponsor's containers.
The sponsor also performed on-board stability studies and calibration frequency studies for the various assays and the conclusions are below. Enzyme methods are not calibrated. This information is included in the labeling.
| Assay | On-board Stability (days) | Calibration Frequency (days) |
| --- | --- | --- |
| Albumin | 30 | 30 |
| Alkaline Phosphatase (ALP) | 14 | |
| Aspartate Aminotransferase (AST) | 30 | |
| Carbon Dioxide (CO2) | 10 | 10 |
| Creatinine (CREA) | 20 | 20 |
| Creatine Kinase (CK) | 30 | |
d. Detection limit:
To determine the Limits of Blank (LoB) for EasyRA Albumin, Alkaline Phosphatase (ALP), Aspartate Aminotransferase (AST), Carbon Dioxide (CO2), Creatinine (CREA), and Creatine Kinase (CK) the sponsor used deionized water assayed twenty (20) times on each of three Easy RA analyzers. Limits of Detection (LoD) were determined by preparing a serum sample with a value slightly above the LoBs for the assays listed above, also analyzed 20 times on three EasyRA analyzers. Statistical analyses following CLSI EP-17A were used for calculating the LoB and LoD results. Results are summarized below:
| Device | Reportable Range | LoB | LoD |
| --- | --- | --- | --- |
| Albumin | 0.4 -7.0 g/dL | 0.0 g/dL | 0.02 g/dL |
{14}
15
ALP 10-800 U/L 4.73 U/L 7.55 U/L
AST 5.5 – 400 U/L 4.0 U/L 5.5 U/L
CO2 2.3 – 45 mmoL/L 0.56 mmoL/L 1.0 mmoL/L
Creatinine 0.2 -15.0 mg/dL 0.02 mg/dL 0.03 mg/dL
CK 7 – 1,200 U/L 1.73 U/L 2.38 U/L
e. Analytical specificity:
**EasyRA Albumin**
**Endogenous Interferences:**
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of two concentrations for each measurand. Samples with increasing amounts of hemoglobin, bilirubin or triglycerides (Intralipid®) were tested on these samples and compared to the same sample without the interferent, unless otherwise noted.
The sponsor defined interference as the highest level tested that does not cause >10% change in analytical result.
The table below summarizes the lowest concentration of interferent determined to cause interference with the assays.
| | Analyte level tested | No interference up to: |
| --- | --- | --- |
| Albumin | | |
| | 3.3 g/dL, 4.0 g/dL | 500 mg/dL hemoglobin |
| | 3.5 g/dL, 4.4 g/dL | 1,350 mg/dL of triglyceride |
| | 3.7 g/dL, 4.8 g/dL | 25 mg/dL of bilirubin |
| ALP | | |
| | 145 U/L, 219 U/L | 500 mg/dL of hemoglobin |
| | 138 U/L, 380 U/L | 1,446 mg/dL of triglyceride |
| | 71 U/L, 1089 U/L | 47 mg/dL of bilirubin |
| AST | | |
| | Hemoglobin interference testing not performed | The sponsor did not test for hemoglobin interference since there are large amounts of AST in red blood cells. The sponsor is only recommending non-hemolyzed serum samples be used. 1 |
| | 31.2 U/L, 102 U/L | 31 mg/dL of triglyceride* |
| | 33 U/L, 109 U/L | 32.7 mg/dL of bilirubin |
| CO2 | | |
{15}
| | 19.3 mmoL/L, 25 mmoL/L | 125 mg/dL of hemoglobin* |
| --- | --- | --- |
| | 20.6 mmoL/L, 30.9 mmoL/L | 850 mg/dL of triglyceride |
| | 25.4 mmoL/L, 31.5 mmoL/L | 40 mg/dL of bilirubin |
| Creatinine | | |
| | 0.68 mg/dL, 2.2 mg/dL | 156 mg/dL of hemoglobin* |
| | 0.79 mg/dL, 2.59 mg/dL | 1,350 mg/dL of triglyceride |
| | 2.43, 4.82 | 20 mg/dL of bilirubin |
| CK | | |
| | 209 U/L, 386 U/L | 250 mg/dL of hemoglobin* |
| | 119 U/L, 218 U/L | 1,000 mg/dL of triglyceride |
| | 112 U/L, 189 U/L | 25 mg/dL of bilirubin |
$^{1}$ Demetriou JA et al. In Clinical Chemistry – Principles and Technics $2^{\text{nd}}$ ed. RJ Henry et al. Eds. Harper & Row, Hagerstown, MD 1974, p 873.
*The labeling recommends not using visibly hemolyzed specimens for: AST, CO2, Creatinine and CK. The labeling also recommends not using visibly lipemic specimens for AST.
## Exogenous Interference
The sponsor did not perform any studies to investigate the effect of exogenous substances such as pharmaceuticals; however a literature reference is included to alert users that these potential interferences should be checked.
The sponsor cites the following references for exogenous interference in the labeling.
Young DS. Effects of Drugs on Clinical Laboratory Tests 4th ed. Washington, DC: AACC Press; 1995.
Young DS. Effects of Preanalytical Variables on Clinical Laboratory Tests. $2^{\mathrm{nd}}$ ed. Washington, DC. AACC Press; 1997.
f. Assay cut-off:
Not Applicable
## 2. Comparison studies:
a. Method comparison with predicate device:
{16}
Serum samples for the candidate devices were tested in singlicate on the Medica system and compared to duplicate results of the predicate devices.
## EasyRA Albumin
A method comparison was performed with the Medica EasyRA Albumin test system and the Stanbio Albumin reagent on the COBAS MIRA test system. A total of 43 samples were tested across the range of the predicate assay (0.6 g/dL to 7.1 g/dL). Of these 43 samples, 38 were native samples and 5 were manipulated to create very low and very high level samples. The comparison resulted in a slope of 0.9651, an intercept of -0.0945, correlation coefficient of $R^2 = 0.9959$, and a std. error of 0.1325. The reportable range is 0.4-7.0 mg/dL.
## EasyRA Alkaline Phosphatase (ALP)
A method comparison was performed with the Medica EasyRA ALP test system and the Roche ALP reagent on the COBAS MIRA test system. A total of 62 samples were tested across the range of the predicate assay (10 U/L to 758 U/L). Of these 62 samples, 55 were native samples and 7 were manipulated to create very low or very high level samples. The comparison resulted in a slope of 0.977, an intercept of 9.743, correlation coefficient of $R^2 = 0.9988$ and a std. error of 6.308. The reportable range is 10-800 U/L.
## EasyRA Aspartate Aminotransaminase (AST):
A method comparison was performed with the Medica EasyRA AST test system and the Roche AST reagent on the COBAS MIRA test system. A total of 45 samples were tested across the range of the predicate assay (8.0-392.3 U/L). Of these 46 samples, 41 were native samples and 4 were spiked or diluted to span the measuring range. The comparison resulted in a slope of 1.000, an intercept of -1.6771, correlation coefficient of $R^2 = 0.9996$, and a std. error of 1.7410. The reportable range is 5.5 -400 U/L.
## EasyRA Carbon Dioxide (CO2):
A method comparison was performed with the Medica EasyRA CO2 test system and the Roche CO2 reagent on the COBAS MIRA test system. A total of 60 samples were tested across the range of the predicate assay (2.1-43.9 mmol/L). Of these 60 samples, 56 were native samples and 4 were spiked or diluted to span the measuring range. The comparison resulted in a slope of 0.9414, an intercept of 0.6015, correlation coefficient of $R^2 = 0.9921$, and a std. error of 0.7856. The reportable range is 2.3 - 45 mmol/L.
17
{17}
18
# EasyRA Creatinine:
A method comparison was performed with the Medica EasyRA Creatinine test system and the StanBio Creatinine reagent on the COBAS MIRA test system. A total of 62 samples were tested across the range of the predicate assay (0.37 mg/dL to 14.37 mg/dL). Of these 62 samples, 58 were native samples and 4 were spiked or diluted to span the measuring range. The comparison resulted in a slope of 1.0449, an intercept of –0.0819, correlation coefficient of R² = 0.9973, and a std. error of 0.1638. The reportable range is 0.2-15.0 mg/dL.
# EasyRA Creatine Kinase (CK):
A method comparison was performed with the Medica EasyRA CK test system and the Roche CK reagent on the COBAS MIRA test system. A total of 54 samples were tested across the range of the predicate assay (7-1182 U/L). Of these 42 samples, 49 were native samples and 5 were spiked or diluted to span the measuring range. The comparison resulted in a slope of 1.0185, an intercept of 5.5223, correlation coefficient of R² = 0.9987, and a std. error of 8.2879. The reportable range is 7-1200 U/L
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
EasyRA Albumin: Adult: 3.8-5.1 g/dL⁴
EasyRA Alkaline Phosphatase: Adult: 34-114 U/L²
{18}
EasyRA Aspartate Aminotransaminase: Adult: 8-40 U/L²
EasyRA Carbon Dioxide: 23-34 mmoL/L³
EasyRA Creatinine: 0.5-1.2 mg/dL¹
EasyRA Creatine Kinase: Male: 24-195 U/L, Female: 24-170 U/L²
The expected values for all the analytes were cited from the following references:
¹ Burtis, C.A. and Ashwood, E.R. (Eds), Tietz Textbook of Clinical Chemistry, 2nd edition, W.B. Saunders C0., Philadelphia (1994).
² Tietz NW. Editor, Clinical Guide to Laboratory Tests, 3rd ed., WB Saunders and Co., Philadelphia, PA, (1995).
³ Henry RJ, Carmon D Winkelman JW, editors, Clinical Chemistry: Principles and Techniques. Hagerstown, MD: Harper and Row Publishers; 1974.
⁴ Doumas BT, Watson WA, Biggs HG. Albumin standards and the measurement of serum albumin with bromocresol green. Clin Chem Acta. 1971;31:87-96.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.