K080857 · Berchtold Holding GmbH · FSY · Apr 10, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080857
Device Name
CHROMOPHARE E-SERIES
Applicant
Berchtold Holding GmbH
Product Code
FSY · General, Plastic Surgery
Decision Date
Apr 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The CHROMOPHARE® Exxx is intended to be used to provide of the device: visible illumination of the surgical field or the patient.
Device Story
Surgical lighting system (CHROMOPHARE® E805, E800, E655, E650, E550, E520) provides high-intensity, shadow-free, cold illumination for surgical fields. Uses gas-discharging lamps or Halogen bulbs with ThermoSorb® heat/UV filtering. Features include swivel arms, auto-switching to secondary lamps upon primary failure, and exchangeable lamp cartridges. Optional integrated CCD video camera and 'EndoLite' ambient lighting for minimally invasive surgery. Operated by surgical staff in clinical/OR settings. AUTOlux feature (E805 model) provides automatic intensity regulation. Output allows surgeons to visualize the surgical site clearly; benefits include reduced heat radiation and improved pattern consistency compared to previous generations.
Surgical lamp system; gas-discharging or Halogen bulb technology; ThermoSorb® heat/UV filter; single-piece polygon reflectors; swivel arm mounting; optional CCD camera and EndoLite ambient light; AUTOlux automatic intensity regulation (E805); reusable steam-sterilizable handles; Class I electrical safety; conforms to IEC 60601-2-41.
Indications for Use
Indicated for local illumination of the operating site on a patient's body with high intensity, shadow-free, cold light during surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
## 510 (k) Summary
Submitter:
BERCHTOLD Holding GmbH Ludwigstaler Straße 125 D- 78532 Tuttlingen Germany
Contact Person:
Silke Goral Requiatory Affairs Phone: 0049 7461 181-155 Fax: 0049 7461 181-8155 Silke.Goral@BERCHTOLD.biz
..............................................................................................................................................................................
January 16, 2008 Preparation Date:
Trade Name:
CHROMOPHARE® E 805 CHROMOPHARE® E 800 CHROMOPHARE® E 655 CHROMOPHARE® E 650
CHROMOPHARE® E 550 CHROMOPHARE® E 520
Common Name: Surgical lamp
Classification Name: Light, Surgical, Ceiling mounted
BERCHTOLD CHROMOPHARE® D650 (K965130) Predicate Device: BERCHTOLD CHROMOPHARE® X65 (K024132)
Device Description: The new BERCHTOLD CHROMOPHARE® E805, E800, E655, E650, E550, and E520 (collectively known and referred to in this document as CHROMOPHARE® Exxx) surgical lights are suitable for all types of surgical procedures. With the use of gas-discharging lamps or Halogen bulbs in these lights, BERCHTOLD realizes high illumination intensity with a lower heat radiation and less pattern variation compared to previous generations of BERCHTOLD products. The light features easy-to-operate swivel arms, auto switching to the secondary lamp in case of failure of the primary lamp, and an easy-to-exchange lamp cartridge. Each light functions with an optional integrated CCD video camera and/or with upward "EndoLite" for endoscopic procedures. The lights could be combined among each other.
Intended Use The CHROMOPHARE® Exxx is intended to be used to provide of the device: visible illumination of the surgical field or the patient.
The surgical lights BERCHTOLD CHROMOPHARE® E805, Indications for Use: E800, E 655, E 650, E 550 and E 520 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.
APR 1 0 2008
4080857
{1}------------------------------------------------
## Summary of technological: characteristics compared to the predicate device
. . .
The BERCHTOLD CHROMOPHARE® Exxx is substantially equivalent to the surgical lights BERCHTOLD CHROMOPHARE® D 650 (K96513 ) and CHROMOPHARE® X 65 (K024132) Similarities and differences are tabulated below. Any differences between the CHROMOPHARE Exxx and the predicate devices do not alter the safety or efficacy of the device.
| | Legally marketed device | New devices | |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------|-------------------|
| | CHROMOPHARE® X65 | CHROMOPHARE® E805 | CHROMOPHARE® E655 |
| Intended use | Illumination of the operating<br>site on a patient's body | same | same |
| Input power | 120V, 1- phase lines, 60Hz | same | same |
| Protection against<br>electrical shock | Class I | same | same |
| Diameter of light body | 650mm | 799mm | 649mm |
| Diameter of polygon<br>reflectors | 580mm | 680mm | 520mm |
| Number of mirrored<br>reflector elements | 800 | 720 | 720 |
| Lamp technology | Metal halide discharging | same | same |
| Polygon Reflector | Multi-piece | Single piece | Single piece |
| Power consumption of<br>bulb | 72W | 150W | same |
| Light / heat filter<br>technology incl. UV light<br>filter mechanism | ThermoSorb® | same | same |
| Color rendering index Ra | 94 | same | same |
| Color temperature | 4500°K | 4300°K | 4300°K |
| Central illuminance (at 1m) | 80000 - 160000lux | same | same |
| Light field diameter | 170 - 280mm | 200 - 330mm | same |
| Depth of Illumination | 1250mm | 1200mm | 1300mm |
| Total irradiance Ee | 550W/m² | 560W/m² | 560W/m² |
| UV- irradiance (≤400nm) | 2.0W/m² | same | same |
| Light focusing mechanism | Rotating of Handle | same | same |
| Life time of bulb | 5000h | same | same |
| Bulb life time indicator | Yes | same | same |
| Automatic switching to the<br>reserve bulb | Yes | same | same |
| Bulb replacement indicator | Yes | same | same |
| Reusable steam<br>sterilizable lamp handle | Yes | same | same |
| Additional light controls in<br>separate wall box | Standard | same | same |
| EndoLite in the light head<br>(ambient illumination for<br>minimally invasive<br>surgeries) | Optional feature | same | same |
| CCD video camera located<br>in sterilizable lamp handle | Optional feature | same | same |
| Automatic regulation of<br>intensity | No | AUTOlux | No |
| | Legally marketed<br>device<br>CHROMOPHARE®<br>D650 | New devices | | | |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|----------------------|----------------------|----------------------|----------------------|
| | | CHROMOPHARE®<br>E800 | CHROMOPHARE®<br>E650 | CHROMOPHARE®<br>E550 | CHROMOPHARE®<br>E520 |
| Intended use | Illumination of the<br>operating site on a<br>patient's body | same | same | same | Same |
| Input power | 120V, 1- phase lines,<br>60Hz | same | same | same | Same |
| Protection<br>against<br>electrical shock | Class I | same | same | same | Same |
| Diameter of<br>light body | 650mm | 799mm | 649mm | 579mm | 579mm |
| Diameter of<br>polygon<br>reflectors | 580mm | 680mm | 520mm | 460mm | 460mm |
| | | | | | |
| Number of<br>mirrored<br>reflector<br>elements | 800 | 720 | 720 | 720 | 720 |
| Lamp<br>technology | Halogen | same | same | same | Same |
| Polygon<br>Reflector | Multi-piece | Single piece | Single Piece | Single piece | Single piece |
| Power<br>consumption of<br>bulb | 150W | same | same | same | 150W |
| Light / heat<br>filter<br>technology incl.<br>UV light filter<br>mechanism | ThermoSorb® | same | same | same | Same |
| Color rendering<br>index Rₐ | 94 | 93 | 93 | 93 | 93 |
| Color<br>temperature | 4500°K | 4300°K | 4300°K | 4300°K | 4300°K |
| Central<br>illuminance (at<br>1m) | 70000 - 130000lux | 80000 - 160000lux | 80000 - 160000lux | 65000 - 130000lux | 80000lux |
| Light field<br>diameter | 150 - 280mm | 200-300mm | 170 - 280mm | 150 - 250mm | 200mm |
| Depth of<br>illumination | 1200mm | same | 1300mm | 1250mm | 1250mm |
| Total irradiance<br>Eₑ | 485W/m² | 560W/m² | 507W/m² | 455W/m² | 280W/m² |
| UV- irradiance<br>(≤400nm) | 2.6W/m² | same | Same | same | Same |
| Light focusing<br>mechanism | Rotating of handle | same | Same | same | No |
| Life time of<br>bulb | 1000h | same | Same | same | Same |
| Bulb life time<br>indicator | No | same | Same | same | Same |
| Automatic<br>switching to the<br>reserve bulb | Yes | same | Same | same | Same |
| Bulb<br>replacement<br>indicator | Yes | same | Same | same | Same |
| Reusable<br>steam<br>sterllizable<br>lamp handle | Yes | same | Same | same | Same |
| Additional light<br>controls in<br>separate wall<br>box | Optional feature | Standard | Standard | Standard | No |
| EndoLite in the<br>light head<br>(ambient<br>illumination for<br>minimally<br>invasive<br>surgeries) | Optional feature | same | Same | same | No |
| CCD video<br>camera located<br>In sterilizable<br>lamp handle | Optional feature | same | Same | same | No |
{2}------------------------------------------------
Performance Summary:
·
This device conforms to IEC 60601-2-41:2001 specifications for performance of surgical lamps. This device conforms to IEC 60601-1 and IEC 60601-1-2 for electrical safety.
.
.
·
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 0 2008
Berchtold Holding GmbH % Underwriters Laboratories, Inc. Mr. Jeff D. Rongero Senior Project Engineer 12 Laboratory Drive Research Triangle Park, North Carolina 27709
Re: K080857
I Trade/Device Name: CHROMOPHARE® E 805, CHROMOPHARE® E 800, CHROMOPHARE® E 655, CHROMOPHARE® E 650, CHROMOPHARE® E 550, CHROMOPHARE® E 520 Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: II Product Code: FSY Dated: March 25, 2008 Received: March 27, 2008
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 – Mr. Jeff D. Rongero
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, pcrmits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Milburn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K080857
Device Name:
CHROMOPHARE® E 805 CHROMOPHARE® E 800 CHROMOPHARE® E 655 CHROMOPHARE® E 650 CHROMOPHARE® E 550 CHROMOPHARE® E 520
Indications for Use:
The surgical lights BERCHTOLD CHROMOPHARE® E 805, E 800, E 655, E 650, E 550 and E 520 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free,
"cold" light.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division
Division of Gascal, Restorative,
and Neurological Dovices
Page 5 of 27
510(k) Number: K680857
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.