K080851 · Anticancer, Inc. · LPS · Oct 10, 2008 · Clinical Chemistry
Device Facts
Record ID
K080851
Device Name
A/C PORTABLE ENZYMATIC HOMOCYSTEINE ASSAY
Applicant
Anticancer, Inc.
Product Code
LPS · Clinical Chemistry
Decision Date
Oct 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1377
Device Class
Class 2
Indications for Use
The A/C Portable Enzymatic Homocysteine Assay on the A/C Diagnostics Reader (HyTek-205) is intended for the quantitative determination of total homocysteine (tHCY) in human plasma or serum. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia. The AiC Enzymatic Homocysteine Assay is for in vitro diagnostic use.
Device Story
Device performs quantitative determination of total homocysteine (tHCY) in human plasma or serum; intended for small laboratory settings. Input: plasma or serum samples. Principle: three-step enzymatic reaction; reduction followed by depletion via Homocysteinase (rHCYase) to produce hydrogen sulfide (H2S); H2S reacts with N,N-dibutyl phenylene diamine (DBPDA) to form a chromophore. Output: fluorescence measured by A/C Diagnostics Reader (HyTek-205). Operation: manual assay process; 80-minute duration at room temperature; reader provides quantitative results. Healthcare providers use output to assist in diagnosis and treatment of hyperhomocysteinemia. Benefits: provides portable, enzymatic tHCY testing capability for smaller clinical environments compared to high-throughput automated analyzers.
Clinical Evidence
Method comparison study performed on 50 plasma samples comparing the A/C Portable Enzymatic Homocysteine Assay to the predicate Hitachi 912 system. Results showed a correlation coefficient of R=0.95, regression y=1.01+0.91, and a mean difference of 1.05 umol/L. Precision data: within-assay CV 3.8%–4.8%; between-assay CV 5.0%–7.4%.
Technological Characteristics
Enzymatic assay; reagents include dithiothreitol, homocysteinase, DBPDA, potassium ferricyanide. Instrumentation: A/C Diagnostics Reader (fluorescence). Sample vessels: 0.5 mL polypropylene PCR tubes. Connectivity: RS-232 serial port. Software: Moderate level of concern. Calibration: Daily curve generation required. Stability: 24h at 25°C, 2mo at 4°C, 2yr at -20°C.
Indications for Use
Indicated for the quantitative determination of total homocysteine (tHCY) in human plasma or serum to assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia.
Regulatory Classification
Identification
A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.
Predicate Devices
A/C Enzymatic Homocysteine Assay on Hitachi 912 (K030754)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k080851
B. Purpose for Submission:
New device
C. Measurand:
Homocysteine
D. Type of Test:
Quantitative
E. Applicant:
AntiCancer, Inc.
F. Proprietary and Established Names:
A/C Portable Enzymatic Homocysteine Assay on the A/C Diagnostics Reader
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1377, Urinary Homocysteine Test System
2. Classification:
Class II
3. Product code:
LPS
{1}
4. Panel:
75, Chemistry
# H. Intended Use:
1. Intended use(s):
See Indications for Use below
2. Indication(s) for use:
The A/C Portable Enzymatic Homocysteine Assay on the A/C Diagnostics Reader (HyTek-205) is intended for the quantitative determination of total homocysteine (tHCY) in human plasma or serum. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia. The A/C Enzymatic Homocysteine Assay is for in vitro diagnostic use.
3. Special conditions for use statement(s):
Prescription Use only
This assay is for central laboratory use.
4. Special instrument requirements:
A/C Diagnostics Reader
# I. Device Description:
The kit consists of the following:
Kit Contents for 100 samples
(1) Reagents (Store at $2 - 8^{\circ}$ )
| Reagent | Compound | Components | Quantity |
| --- | --- | --- | --- |
| Reagent I | Reducing Reagent | Dithiothreitol (lyophilized powder) | 10 mg |
| Reagent II | Enzyme | Homocysteinase (lyophilized powder) | 0.5 mg |
| Reagent III | Chromophore | N,N-dibutyl phenylene diamine chloride) | 10 mL |
| Reagent IV | Oxidant | Potassium ferricyanide in buffer | 6 mL |
| Buffer I | Assay Buffer | Sodium phosphate buffer | 50 mL |
| Buffer II | Enzyme Buffer | Pyridoxyal 5-phosphate buffer | 1 mL |
| Surfactant | Surfactant | Triton X-100 | 1 mL |
{2}
(2) Calibrators and controls (Store at -20°C) supplied with the kit
Controls: Low and High Controls each 0.3 ml
Controls are prepared from a human plasma matrix spiked with L-homocysteine
Calibrators: Low, Medium and High Calibrators each 0.3 ml
Calibrators are prepared from a human plasma matrix spiked with L-homocysteine
(3) Thermowell TM tubes (120)
0.5 mL Costar® PCR Tubes - Thin-walled polypropylene PCR/centrifuge tubes. Tubes are plain polypropylene with no coating.
Human source material was tested by an FDA approved method for the presence of antibody to HIV-1/2, antibody to HCV, as well as for the Hepatitis B Surface Antigen (HbsAg), and found to be negative.
J. Substantial Equivalence Information:
1. Predicate device name(s):
A/C Enzymatic HCY Assay on Hitachi 912
2. Predicate 510(k) number(s):
k030754
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Principle | Enzymatic | Enzymatic |
| Indications for Use | Intended for the quantitative determination of total homocysteine (tHCY) in human plasma or serum. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia. | Intended for the quantitative determination of total homocysteine (tHCY) in human plasma or serum. The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia. |
| Calibrator levels | Low, medium and high | Low, medium and high |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Performance | Manual | Fully automatic |
| Instrumentation | A/C Diagnostics Reader | Hitachi 912 Analyzer |
| Throughput | 100 tests/hr | 360 tests/hr |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reportable ranges | 4.1 to 43.8 μmol/L | 2.5 - 80 μmol/L |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-T2, Evaluation of Precision Performance of Clinical Chemistry Devices;
# L. Test Principle:
The A/C Portable Enzymatic Homocysteine (HCY) Assay measures total homocysteine (tHCY) in plasma and serum. The principle of the assay is that tHCY is degraded by Homocysteinase (rHCYase) and produces hydrogen sulfide $(\mathrm{H}_2\mathrm{S})$ , which is measured using N,N-dibutyl phenylene diamine (DBPDA), the combination of which forms a chromophore. After oxidation, the A/C Diagnostics Reader, a fluorescence reader is used to obtain the tHCY values.
The A/C Portable Enzymatic Homocysteine Assay is a manual assay with the results read by the A/C Diagnostics Reader. Five $\mu$ L of EDTA plasma is used in a dithiothreitol reduction reaction for 60 minutes to release bound homocysteine. The rHCYase reaction is then run for 5 minutes. The DBPDA chromophore is then added and one minute later an oxidant, potassium ferricyanide, is added. Ten minutes later, the end-points are read at $\mathrm{Em}710\mathrm{nm}/\mathrm{Ex}660$ nm.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The precision of the assay on the A/C Diagnostics Reader was evaluated in a study based on CLSI EP5-T2 guidelines "Evaluation of Precision Performance of Clinical Chemistry Devices." In this study, precision was determined after analyzing three plasma samples containing low, medium and high tHCY in parallel assays. The within-assay precision was determined from results from 10 parallel analytical setups. The between-assay precision was calculated in 10 successive assays carried out over 10 days. The CVs, shown in the table below, for within-assay and between-assay were less than $5\%$ and $10\%$ , respectively.
{4}
5
Within-Assay Precision
| Assay | Sample 1 | Sample 2 | Sample 3 |
| --- | --- | --- | --- |
| Average tHCY (μmol/L) | 8.1 | 14.9 | 35.2 |
| SD | 0.31 | 0.71 | 1.64 |
| CV (%) | 3.8 | 4.8 | 4.7 |
Between-Assay Precision
| Assay | Sample 1 | Sample 2 | Sample 3 |
| --- | --- | --- | --- |
| Average tHCY (μmol/L) | 7.4 | 15.1 | 35.3 |
| SD | 0.5 | 0.9 | 1.8 |
| CV (%) | 7.4 | 5.9 | 5.0 |
b. Linearity/assay reportable range:
Five plasma samples containing different levels of homocysteine (3.2-44.6 μmol/L) were measured by the A/C Portable Enzymatic HCY Assay on the A/C Diagnostics Reader. The data demonstrated linearity of the assay up to 43.8 μmol/L with the regression equation of $y = 0.9536x + 0.5963$, $R^2 = 0.9969$. The studies support the stated measuring range of the assay of 4.1 to 43.8 μmol/L.
A recovery study was performed in which one sample with tHCY a concentration of 45.0 μmol/L was serially diluted 2, 4 and 8 times. The samples were analyzed by the Portable A/C Enzymatic HCY Assay in 10 replicates in a single run. The average % recoveries ranged from 91 to 104 %.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators and Controls were prepared by spiking homocysteine into a plasma matrix. The plasma matrix was manufactured from material such that each donor unit was tested by an FDA approved method for the presence of antibody to HIV-1/2, antibody to HCV, as well as for the Hepatitis B Surface Antigen (HbsAg), and found to be negative. An HPLC method was used for the value determination of Calibrators and Controls for the A/C Enzymatic Homocysteine Assay. Calibrators and controls are stated to be stable for 24 hours at 25 °C, 2 months at 4 °C and 2 years at -20 °C. The stability was determined by testing the vials stored at different temperatures for certain periods compared with the standard in the Portable A/C Enzymatic HCY Assay. Calibrators and controls are supplied with the kit.
c. Detection limit:
The limit of quantitation (LoQ) is used as the low detection limit of the assay. The limit of quantitation was defined as the lowest concentration having a CV < 20%.
{5}
The results demonstrate that the LoQ of the A/C Portable Enzymatic HCY Assay on the A/C Diagnostics Reader is 3.7 μmol/L. The sponsor has established the measuring range of the assay as from 4.1 to 43.8 μmol/L (see Section b. Linearity).
d. Analytical specificity:
To study the effect of bilirubin, plasma samples with tHCY levels of 5.7 and 14.8 μmol/L were spiked with bilirubin to achieve bilirubin levels of 12.5, 25 and 50 mg/dL. The samples were analyzed in quadruplicate by the assay. The results demonstrate that bilirubin concentrations of up to 50 mg/dL show < 10% interference and therefore, do not interfere with the measurement of tHCY by the assay. Interference from glycerol was studied using plasma samples with tHCY levels of 6.8 and 11.6 μmol/L which were spiked with glycerol to achieve levels of 1,250, 2,500 and 5,000 mg/dL. The results demonstrated that glycerol concentrations of up to 5000 mg/dL show < 10% interference and therefore, do not interfere with the measurement of tHCY by the assay. In addition, plasma samples with tHCY levels of 6.8 and 11.6 μmol/L spiked with a commercially available lipid mixture containing 450 mg/dL showed < 10% interference. To study the effect of hemoglobin plasma samples with 5.5 and 14.6 μmol/L tHCY were spiked with hemoglobin to achieve hemoglobin levels of 0.25 g/L, 0.5 g/L and 1.0 g/L. The results demonstrate that hemoglobin concentrations of up to 1.0 mg/mL (1.0 g/L) show < 10% interference and therefore, do not interfere with the measurement of tHCY by the assay.
Cross reactivity was tested and is < 10% for cystathione (up to 500 μmol/L), L-cysteine (up to 200 μmol/L), methionine (up to 0.2 mol/L) and glutathione (up to 0.5 μmol/L).
Other Limitations
Specimens from patients who are on drug therapy involving S-adenosyl methionine may show falsely elevated levels of homocysteine. Specimens from patients taking methotrexate, carbamazepine, phenytoin, nitrous oxide or 6-azauridine triacetate may have elevated levels due to their effect on the methionine-homocysteine metabolic pathway.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Fifty EDTA plasma samples with values ranging from 5.0 to 37.6 μmol/L were measured both by the A/C Portable Enzymatic HCY Assay on the A/C Diagnostics
{6}
Reader and the A/C Automatic Enzymatic HCY Assay on the Hitachi 912 Automatic Analyzer (predicate device). The comparison yielded a regression equation of $y = 1.0098x + 0.9124$ , $r^2 = 0.946$
# b. Matrix comparison:
Samples collected from 10 donors were immediately separated into three different vacutainers, one without additive, one containing $\mathrm{K}_2$ -EDTA and one containing sodium heparin, respectively. Plasma and serum were separated from whole blood within one hour by centrifugation. High HCY samples were prepared by spiking L-homocysteine into 6 samples to prepare a total of 16 samples. Total HCY values of the samples were measured by the A/C Portable Enzymatic HCY Assay on the A/C Diagnostics Reader. The results are shown in the table below.
| Sample type (x vs. y) | n | Sample range (μmol/L) | Regression equation | R2value |
| --- | --- | --- | --- | --- |
| EDTA Plasma vs. Heparin plasma | 16 | 5.6 – 34.8 | y = 0.990x + 0.11 | 0.996 |
| Serum vs. EDTA plasma | 16 | 5.4 – 36.1 | y = 0.970x + 0.33 | 0.995 |
| Serum vs. Heparin plasma | 16 | 6.2 – 35.0 | y = 0.964x + 0.4 | 0.997 |
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
Reference ranges are based on literature references provided in the labeling and referenced below. Total HCY values in plasma/serum from 5 to $15\mu \mathrm{mol} / \mathrm{L}$ are considered as normal. However, total HCY values in plasma/serum differ with age,
{7}
gender, stress and other physical conditions. Total HCY values in plasma/serum > 15 μmol/L are considered abnormal.
1. Ueland PM, Refsum H, Stabler SP, Malinow R, Anderson A, Allen RH. Total homocysteine in plasma or serum: Methods and clinical applications. Clin. Chem. 39, 1764-1779, 1993.
2. Ubbink JB, Vermaak WJH, van der Merwe A, Becker PJ. The effect of blood sample aging and food consumption on plasma total homocysteine levels. Clinica Chimica Acta 207, 119-128, 1992.
3. Moller J, Rasmussen K. Homocysteine in plasma: Stabilization of blood samples with fluoride. Clin. Chem. 41, 758-759, 1995.
N. Instrument Name:
A/C Diagnostics Reader
O. System Descriptions:
1. Modes of Operation:
The modes of operation are Setup, which includes setting up the number and value of the standards; Testing, which includes calculation of the standard curve and testing samples; and Test Reports, which includes looking up and/or printing the results.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
The software documentation for this device follows our software guidance for a moderate level of concern.
3. Specimen Identification:
Sample tubes (Thermowell TM tubes) are labeled with the ID number of the sample. During the test, the operator can choose the number of the sample using the keys on the instrument panel. After testing the sample, the instrument screen displays the result. Data can be sent out to a computer using the RS-232 serial port.
4. Specimen Sampling and Handling:
Manual assay – samples are put one by one into the test well and the lid closed to start testing.
{8}
5. Calibration:
Calibrators are supplied with the kit. Calibrator values are entered into the instrument in Setup mode. After the calibrators are tested, the instrument calculates the calibration curve. The labeling recommends to run a calibration curve each day.
6. Quality Control:
Controls are supplied with the kit. The labeling recommends to run a set of controls with patient samples in every batch.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.