HANDAN N95 PARTICULATE RESPIRATORS AND SURGICAL MASK, MODELS HY8510 AND HY9810
K080827 · Handan Hengyong Protective & Clean Products Co., L · MSH · Apr 4, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080827
Device Name
HANDAN N95 PARTICULATE RESPIRATORS AND SURGICAL MASK, MODELS HY8510 AND HY9810
Applicant
Handan Hengyong Protective & Clean Products Co., L
Product Code
MSH · General, Plastic Surgery
Decision Date
Apr 4, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Handan N95 Particulate Respirators and Surgical Masks HY8510, HY9810 are intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials. This includes use as a procedure mask, isolation mask or dental face mask. This device also meets CDC Guidelines for TB Exposure Control.
Device Story
Handan HY8510 and HY9810 are N95 particulate respirators and surgical masks; intended for single-use by healthcare personnel. HY8510 is a flat-pleated mask with single elastic head strap; HY9810 is a molded cup mask with dual elastic head straps. Both utilize multi-layer filtering media (polypropylene spunbond, melt-blown polypropylene, polyester) and include a PU foam nosepiece with plastic/steel wire. Devices function as physical barriers to protect patients and clinicians from microorganisms, body fluids, and airborne particulates. Used in clinical settings including operating rooms, isolation units, and dental offices. No sterilization; single-use only.
Clinical Evidence
No clinical data. Bench testing only, including NIOSH exhalation/inhalation resistance tests, NIOSH sodium chloride (NaCl) N95 filtration testing, 16 CFR 1610 flammability testing, and ISO 10993 biocompatibility testing.
Technological Characteristics
N95 particulate respirator; multi-layer construction (polypropylene spunbond, melt-blown polypropylene, polyester); PU foam nosepiece with plastic/steel wire; elastic head straps; non-sterile; single-use. NIOSH certified (TC-84A-4276, TC-84A-4521).
Indications for Use
Indicated for operating room personnel and health care workers for protection against microorganisms, blood, body fluids, and airborne particulates; includes use as procedure, isolation, or dental face mask; meets CDC guidelines for TB exposure control.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
### Handan Hengyong Protective &Clean Proc
1-1 1201, 455 Gongnong Road, Shijiazhuang, Hebei, P.R.China Tel: 86-311-83032925 Fax: 86-311-83995076
APR - 4 2008
Page 1 of 3
# 510(k) Summary
This summary of 510K safety and effectiveness information is being submitted in accrodance with the requirements of 21 CFR 807.92.
The assigned 510K number is K080827
1. Submitter's Identification:
Handan Hengyong Protective & Clean Products Co.,Ltd
1-1-1201,455 Gongnong Road, Shijiazhuang,
Hebei Province, P.R.China
Contact:
Maggie Zhong Tel:+86-21-33907930 Fax:+86-21-33907932 Mail: maggie-zhong2001@vip.sina.com Date of Summary: April 2, 2008
- 2. Device Name:
Handan N95 Particulate Respirators and Surgical Masks, HY8510 & HY9810
- 3. Classification Name: Surgical N95 Mask
- 4.. Device Description
Handan HY8510 N95 particulate respirators and surgical mask is constructed from a polypropylene spunbond used in the inner and outer cover. The polypropylene melt blown filter media and polyester filter fabric are layered between the inner and outer cover. The head strap is made of polyester elastic(for single head strap) which is circled to the mask. The inside nosepiece is a PU foam.
I landan HY9810 N95 particulate respirators and surgical mask is constructed from a polypropylene spunbond used in the outer cover, polyester filter fabric used in the inner cover. The polypropylene melt blown filter media is layered between the inner and out cover. The head straps are made of polyester elastic(for double headband) which is welded to the mask. The inside nosepiece is a PU foam.
Both items are no - sterilized and only for single use.
- 5. Intended Use:
Handan type N95 respirators are intended for single use by operating room personnel or general health care workers for protection against microscopic organisms, body fluids and particulates. These would include use as procedure mask, isolation mask or dental face mask.
- 6. Comparison to Predicate Devices
{1}------------------------------------------------
## Handan Hengyong Protective &Clean Products Co.,Ltd
1-1 1201, 455 Gongnong Road, Shijiazhuang, Hebei, P.R.China Fax: 86-311-83995076 Tel: 86-311-83032925
Handan N95 Particulate Respirators and Surgical Masks, HY8510 & HY9810 are substantially equivalent is safety and effectiveness to the predicate device. Aearo Company-- K041855 Pleated plus 1050 and 1050S Gerson Isolair Compay - k960778 APR Type N95 model 2735 for reference: FXX, 878, 4040. Class II
| Manufacturer | Handan Hengyong Protective & Clean<br>Products Co.,Ltd | Aearo Company<br>Predicate device-for reference |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | HY8510 N95 Surgical Mask<br>(New device) | Pleated plus 1050 and 1050S |
| 510K<br>Number | K080827 | K041855 |
| Product code | MSH, 878.4040 | SAME |
| Device<br>Description | 1. N95 Class Particuliar respirator<br>2. Multi-layer filteirng media<br>(white polypropylene spundbond,<br>polypropylene, meltblown, polyester,<br>polypropylene)<br>3. Plastic nose wire with steel<br>4. White elastic headband<br>5. Dimension<br>15.5" circumference<br>6. Flat pleated mask<br>7. Single elastic head strap | 1. N95 Class Particuliar respirator<br>2. Multi-layer filteirng media<br>(White spundbond polypropylene,<br>meltblown polypropylene)<br>3. Tic wire nose piece<br>4. White elastic headband<br>5. Dimension<br>Small (13.5" circumference)<br>Large(15.5" circumference)<br>6. Flat pleated mask<br>7. Dual elastic head strap |
| NIOSH<br>certification# | TC-84A-4276 | TC-84A-2630 |
| Manufacturer | Handan Hengyong Protective & Clean<br>Products Co.,Ltd | Gerson Isolair APR Company<br>Predicate device-for reference |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | HY9810 N95 Surgical Mask<br>(New device) | N95 model 2735 |
| 510K<br>Number | K080827 | K960778 |
| Product code | MSH, 878.4040 | SAME |
| Device<br>Description | 1. N95 Class Particuliar respirator<br>2. Multi-layer filteirng media<br>(white polyester, polypropylene<br>meltblown, polypropylene)<br>3. Plastic nose wire with steel<br>4. White elastic headband<br>5. Dimension<br>15.75" circumference<br>6. Molded Cup<br>7. Dual elastic head strap | 1. N95 Class Particuliar respirator<br>2. Multi-layer filteirng media<br>(White nonwoven polyester<br>meltblown polypropylene)<br>3. Plastic nose wire<br>4. Yellow elastic, latex free<br>5. Dimension<br>Small (13.75" circumference)<br>6. Molded Cup<br>7. Dual elastic head strap |
| NIOSH<br>certification# | TC-84A-4521 | TC-84A-160 |
Discussion of Non-clinical Test Performed for Determination of Substantial Equivalence
{2}------------------------------------------------
## Handan Hengyong Protective &Clean Products Co.,Ltd
1-1 1201, 455 Gongnong Road, Shijiazhuang, Hebei, P.R.China Fax: 86-311-83995076 Tel: 86-311-83032925
#### are as follows:
- NIOSH, Exhalation of Resistance Test, 84.180 I.
- NIOSH Inhalation of Resistance Test, 84.180 ][.
- NIOSH Sodium Chloride (Nacl) -N95 84.181 III.
- Flammibility, Complied with 16 CFR 1610 Class I, IV.
- V. Biocompatibility per ISO 10993
It is our conclusion that performance testing meet all relevant requirements of the aforementioned test standard.
#### Disscussion of Clinical Tests Performed
Not Applicable
- 7. Conclusions
Handan N95 Particulate Respirators and Surgical Masks, HY8510 & HY9810 has the same intended use and technology characterisitics as the predicate devices (K041855, K960778). Moreover, the bench testing contained in this submission supplied domonstrate that the technological characterisistics do not raise any new question of safety or effectiveness.
Handan N95 Particulate Respirators and Surgical Masks, HY8510 & HY9810 are substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 4 2008
Ms. Maggie Zhong Consultant Handan Hengyong Protective & Clean Products Company, Limited 1-1 1201, 455 Gongnong Road, Shijiazhuang Hebei Province P.R.CHINA
Re: K080827
Trade/Device Name: Handan N95 Particulate Respirators and Surgical Masks HY8510, HY9810 Surgical N95 Respirator Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: March 15, 2008 Received: March 25, 2008
Dear Ms. Zhong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Zhong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE
510(k) NUMBER (IF KNOWN): APPLICANT: DEVICE NAME:
080827 Handan Hengyong Protective & Clean Products Co.,Ltd Handan N95 Particulate Respirators and Surgical Masks HY8510, HY9810 Surgical N95 Respirator
INDICATION FOR USE:
The Handan N95 Particulate Respirators and Surgical Masks HY8510, HY9810 are intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials. This includes use as a procedure mask, isolation mask or dental face mask. This device also meets CDC Guidelines for TB Exposure Control.
Prescription Use __ (Part 21 CFR 801 Subpart D)
ア Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clue
Page 1 of _
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K080827
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.