K080812 · Beckman Coulter, Inc. · KHE · Jun 25, 2008 · Hematology
Device Facts
Record ID
K080812
Device Name
HEMOCCULT ICT
Applicant
Beckman Coulter, Inc.
Product Code
KHE · Hematology
Decision Date
Jun 25, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.6550
Device Class
Class 2
Indications for Use
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples. Fecal occult blood tests are useful screening aids for detecting primarily lower gastrointestinal (g.i.) disorders that may be related to iron deficiency anemia, diverticulitis, ulcerative colitis, polyps, adenomas, colorectal cancers or other g.i. lesions that can bleed. Hemoccult ICT is recommended for use by health professionals as part of routine physical examinations or when lower g.i. disorders are suspected.
Device Story
Hemoccult ICT is a rapid, qualitative, visually read immunochemical chromatographic test for human hemoglobin in fecal samples. Components include sample collection cards, applicator sticks, test devices, and buffer. Used by health professionals in clinical settings to screen for lower GI disorders. The test detects occult blood; results assist clinicians in identifying patients requiring further diagnostic evaluation for colorectal neoplasia or other bleeding lesions. Benefits include non-invasive screening for asymptomatic individuals or those with suspected GI pathology.
Clinical Evidence
Clinical evaluation included 1,734 average-risk individuals (ages 41-97) and 88 young volunteers (ages 17-33). In average-risk populations, multi-day screening positivity rate was 1.8% with a 15.6% positive predictive value for colorectal neoplasia. In high-risk patients, clinical sensitivity was 90% for colorectal cancer and 28% for large adenomas; low sensitivity observed for non-neoplastic lesions. Data supports screening utility for asymptomatic individuals age 50 or older.
Technological Characteristics
Immunochemical chromatographic test; qualitative visual readout. Components: sample collection cards, applicator sticks, test devices, buffer. No electronic components, software, or energy sources. Materials and formulation identical to predicate.
Indications for Use
Indicated for use by health professionals as a screening aid for detecting lower gastrointestinal disorders (e.g., iron deficiency anemia, diverticulitis, ulcerative colitis, polyps, adenomas, colorectal cancers) in patients suspected of having lower GI lesions or during routine physical exams.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
K080812
Summary of Safety & Effectiveness Hemoccult® ICT Immunochemical Fecal Occult Blood Test Kit
## 1.0 Submitted By:
Kim Walker, MS, RAC Global RA & QA Consultant C/O Beckman Coulter, Inc. 200 S. Kraemer Blvd., W-110 Brea, California 92822-8000 Telephone: (714) 273-9254 FAX: (714) 961-4234
JUN 2 5 2008
## 2.0 Date Submitted:
March 21, 2008
## 3.0 Device Name(s):
- 3.1 Proprietary Names Hemoccult® ICT
- 3.2 Classification Name Ocult blood test
## 4.0 Predicate Device:
| Candidate(s) | Predicate | Manufacturer | Docket<br>Number |
|---------------|---------------------------------------------|--------------------------|------------------|
| Hemoccult ICT | Hemoccult ICT<br>(formerly FlexSure<br>OBT) | Beckman<br>Coulter, Inc. | K961062 |
## 5.0 Description:
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples. It incorporates multiple components including separate sample collection card(s), applicator sticks, test devices, buffer, patient instructions, physician instructions and product instructions for use.
{1}------------------------------------------------
## 6.0 Intended Use:
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
## Clinical Significance:
Fecal occult blood tests are useful screening aids for detecting primarily lower gastrointestinal (g.i.) disorders that may be related to iron deficiency anemia, diverticulitis, ulcerative colitis, polyps, adenomas, colorectal cancers or other q.i. lesions that can bleed. Hemoccult ICT is recommended for use by health professionals as part of routine physical examinations or when lower g.i. disorders are suspected.
## 7.0 Comparison to Predicate(s):
The following tables show the similarities and differences between the predicates identified in Section 4.0 of this summary.
| Similarities to the Predicate | | |
|-------------------------------|---------------------------------------|----------------------------|
| Product | Aspect/Characteristic | Comments |
| Hemoccult ICT | Intended Use | Same as Predicate<br>HOICT |
| | Composition of Test Card<br>Materials | Same as Predicate<br>HOICT |
| | Buffer Solution | Same as Predicate<br>HOICT |
| | Testing Procedure Methodology | Same as Predicate<br>HOICT |
| | Sample Stability | Same as Predicate<br>HOICT |
| | Analytical Performance | Same as Predicate<br>HOICT |
| | Clinical Performance Source<br>Data | Same as Predicate<br>HOICT |
Similarities to the Predicate
{2}------------------------------------------------
| Product | Aspect/Characteristic | Comments |
|---------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hemoccult ICT | Clinical Performance Data<br>Presentation | Data provided in the new<br>product instructions includes<br>clinical performance for all<br>three days evaluated during<br>the original predicate clinical<br>study. |
| | 2-day Patient Screening<br>Kit Labeling | A 2-day kit option is being<br>created whereas the<br>predicate only offered a 3-<br>day kit. |
| | 3-day Patient Screening<br>Kit Labeling | The predicate Patient<br>Screening Kit Labeling is<br>now adding the<br>specifications of being a 3-<br>day kit. |
## 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through intended use, material composition, formulation, and clinical performance data analysis.
Immunochemical fecal occult blood test positivity rates of approximately 2% should be expected in a screening population of average risk, asymptomatic individuals age 50 or older. The Hemoccult ICT was evaluated using multiday fecal collections (all returned collections of the three dispensed slides regardless of number of days), one day fecal collections (day one results only), two day collections (day one and day two results) and three day collections (only those individuals who returned all three days of fecal collections). The positivity rate for multi-day collections was approximately 2% in a group of 88 young, presumed normal volunteers, ages 17-33, who did not follow a restricted diet. The Hemoccult ICT multi-day screening positivity rate and estimated positive predictive value for colorectal neoplasia were 1,8% and 15.6%, respectively, in a group of 1734 average risk individuals, ages 41-97, who followed a restricted diet. Among high risk patients, Hemoccult ICT multi-day screening had a clinical sensitivity of 90% for colorectal cancer and 28% for large adenomas; in this study, Hemoccult ICT had low sensitivity for non-neoplastic colorectal lesions.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wings, representing health, human services, and well-being. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUN 2 5 2008
Beckman Coulter, Inc. C/O Sylvia Zorich 200 South Kraemer Boulevard, W-110 PO Box 8000 Brea, California 92822
Re: k080812
Trade/Device Name: Hemoccult ICT Regulation Number: 21 CFR 864.6550 Regulation Name: Occult Blood Test Regulatory Class: Class II Product Code: KHE Dated: March 21, 2008 Received: March 24, 2008
Dear Ms. Zorich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter
{4}------------------------------------------------
Page 2 - Beckman Coulter, Inc.
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillancc at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
mckee
for Dr. Becker
Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510(k) Number (if known): 10808 | 2
Device Name: Hemoccult® ICT
Indications For Use:
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples. Fecal occult blood tests are useful screening aids for detecting primarily lower gastrointentinal (g.i.) disorders that may be related to iron deficiency anemia, diverticulitis, ulcerative colitis, polyps, adenomas, colorectal cancers or other g.i. lesions that can bleed. Hemoccult ICT is recommended for use by health professionals as part of routine physical examinations or when lower g.i. disorders are suspected.
Prescription Use X (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Division Sign-Off
Office/of In Witro Diagnostic Device Evaluation and Safety
510(k) K080812
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.