TINA-QUANT CYSTATIN C, CALIBRATOR F.A.S. CYSTATIN C AND CYSTATIN C ANDCYSTATIN C CONTROL SET
Device Facts
| Record ID | K080811 |
|---|---|
| Device Name | TINA-QUANT CYSTATIN C, CALIBRATOR F.A.S. CYSTATIN C AND CYSTATIN C ANDCYSTATIN C CONTROL SET |
| Applicant | Roche Diagnostics Corp. |
| Product Code | NDY · Clinical Chemistry |
| Decision Date | Jun 20, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1225 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K080811 · Jun 20, 2008 | TINA-QUANT CYSTATIN C, CALIBRATOR F.A.S. CYSTATIN C AND CYSTATIN C ANDCYSTATIN C CONTROL SET | Roche Diagnostics Corp. | Retrospective collection of banked frozen human plasma samples | Retrospective banked samples were used to perform a method comparison between the candidate device and the predicate device to demonstrate substantial equivalence. | Method comparison; Retrospective banked samples; Substantial equivalence |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective method comparison | Native human plasma samples from hospitalized and healthy blood donors; Sample Size: 94 | DAKO Cystatin C Immunoparticles (predicate device) | Correlation between candidate and predicate device (Passing/Bablock regression) |
Indications for Use
Reagent: Immunoturbidimetric assay for the quantitative in vitro determination of cystatin C in human serum and lithium-heparin plasma on Roche automated clinical chemistry analyzers. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Calibrator: Cfas (Calibrator for automated systems) Cystatin C is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. Control: Cystatin C Control Set is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.
Device Story
Tina-quant Cystatin C is a ready-to-use dual reagent assay; utilizes particle-enhanced immunoturbidimetric principle. Input: human serum or lithium-heparin plasma samples. Operation: latex particles coated with rabbit anti-cystatin C antibodies agglutinate with cystatin C in sample; precipitate measured turbidimetrically on Roche Hitachi 917 analyzer. Output: quantitative cystatin C concentration calculated via interpolation on standard curve. Used in clinical chemistry laboratories by trained personnel. Results aid clinicians in diagnosing and monitoring renal disease. System includes C.f.a.s. Cystatin C calibrator and bi-level control set for quality monitoring.
Clinical Evidence
Bench testing only. Precision studies (N=63 per material) showed total CVs 2.36-3.76%. Linearity confirmed 0.4-8.0 mg/L. Method comparison (N=94) against predicate yielded r²=0.9991. Interference testing showed no significant impact from common compounds or high-dose hook effect up to 20 mg/L. Reference range study conducted on 500 healthy donors.
Technological Characteristics
Particle-enhanced immunoturbidimetric assay. Reagents: latex particles coated with anti-cystatin C antibodies. Calibrator/Control: recombinant human cystatin C produced in E. coli in delipidated human serum matrix. Storage: 2-8°C. Analyzers: Hitachi 917, MODULAR P, cobas c 501. Connectivity: Automated clinical chemistry analyzer integration.
Indications for Use
Indicated for the quantitative determination of cystatin C in human serum and lithium-heparin plasma to aid in the diagnosis and treatment of renal diseases. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
- DakoCytomation Cystatin C Immunoparticles, Cystatin C Calibrator Kit and Cystatin C Control Set (K041627)