The Intel® Health Guide PHS6000 is a communication tool that allows caregivers to remotely access vital sign measurements of patients at home. It collects measurements captured on commercially available wireless or tethered medical devices designed for home use, stores and displays the collected information on the LCD screen, and transmits the information to a secure host server. Caregivers can review the transmitted information and select education and reminders for patients using a caregiver software interface to the host server. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers. Patients can also optionally engage in video conferences with caregivers and answer the caregivers' questions by participating in surveys. The Intel® Health Guide PHS6000 is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. It is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.
Device Story
Intel Health Guide PHS6000 is a remote patient monitoring system; collects vital sign data from external wireless/tethered home medical devices; stores/displays data on LCD screen; transmits data to secure host server via broadband/phone line. Caregivers access data via web-based Care Management Suite; system allows setting of upper/lower vital sign limits; triggers email/page alerts to caregivers if limits exceeded. Device includes touch screen, video camera, microphones, speakers, reminder light. Used in home setting by patients; operated by patients and remote caregivers. Facilitates patient education, surveys, and video conferencing. Does not provide real-time data; intended for non-time-critical care; supports clinical decision-making by providing remote access to patient measurements for caregiver review.
Clinical Evidence
No clinical data. Device relies on conformity to FDA-recognized consensus standards to demonstrate safety and efficacy.
Technological Characteristics
Electronic device in plastic enclosure; touch screen, video camera, microphones, speakers, reminder light. Interfaces with external wireless/tethered medical devices. Connectivity via broadband internet or standard phone line. Software includes patient-side application, Care Management Suite (server-side), and processor interface software.
Indications for Use
Indicated for patients at home willing and capable of managing device use. Contraindicated for patients requiring direct medical supervision or emergency intervention. Not for time-critical care.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
Philips Medical Systems' M3810A TeleMonitoring System with M3812B TeleStation (K023749)
Health Hero Network's Health Buddy Appliance (K063612)
WHBVMC, LLC's RemoteNurse™ Patient Monitoring system (K041308)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary As required by 21 CFR §807.92(c)
### Submitter
JUN 2 7 2008
| 510(k) Owner: | Intel Corporation |
|-----------------|---------------------------------------------------|
| Address: | 1900 Prairie City Road, FM7-197, Folsom, CA 95630 |
| Telephone: | (408) 765-2060 |
| Contact Person: | Tac-Woong Koo |
| Date Prepared: | March 19, 2008 |
#### Device Information
| Trade Name: | Intel® Health Guide PHS6000 |
|----------------------|---------------------------------------------------------------------------------------------------------|
| Common Name: | Remote Patient Monitoring System |
| Classification Name: | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency (21 CFR 870.2910, Product Code DRG) |
Substantial Equivalence is claimed to the following devices:
- 1. Philips Medical Systems' M3810A TeleMonitoring System with M3812B TeleStation (K023749)
- 2. Health Hero Network's Health Buddy" Appliance (K063612)
- 3. WHBVMC, LLC's RemoteNurse™ Patient Monitoring system (K041308)
### Device Description
The Intel* Health Guide PHS6000 is a communication tool that allows caregivers to remotely access vital sign measurements of patients at home. It collects measurements remotely access vitar ago available wireless or tethered medical devices designed for home use, stores and displays the collected information on the LCD screen, and transmits the information to a secure host scrver. Caregivers can review the transmitted information and select education and reminders for patients using a caregiver software interface to the host server. Patients can review the stored vital sign measurement information and receive Screef. Fattents can review the pregivers. Patients can also optionally engage in cultianonia and morractorers and answer the caregivers` questions by participating in surveys.
The Intel" Health Guide PHS6000 system consists of the:
- Intel* Health Guide PHS6000 hardware: (1)
The physical component of the Intel® Health Guide PHS6000 is an electronic device contained in a plastic enclosure with a touch screen, video camera with privacy screen, microphones, speakers and a reminder light which is mounted into the top of the case. On the back of the device is a power socket, a headphone socket. a
Section 5: 510(k) Summary
Page 1 of 3
{1}------------------------------------------------
#### 510(k) Notification Submission - Abbreviated Intel® Health Guide PHS6000
Broadband (high-speed) internet socket for connection to a broadband network. The device has medical device sockets for connection to specific physiological monitors, and may optionally have a phone socket for modem connection to a standard phone line.
- Intel " Health Guide software application: (2)
The software application captures, stores, and transmits information to a secure website via a standard telephone line or a LAN/WAN connection.
- Intel* Care Management Suite software application: (3)
The application allows caregivers to review patient vital signs on the secure website. The Intel® Care Management Suite allows for predetining upper and lower limits and, when either limit is exceeded, the system emails and/or pages the caregiver.
- Processor software application: (4)
The processor software application manages the interface between the Intel® Health Guide PHS6000 software application and the secure website.
The Intel® Health Guide PHS6000 is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. The device is contraindicated for patients requiring direct medical supervision or emergency intervention. The device is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.
## Indications for Use
The Intel* Health Guide PHS6000 is a communication tool that allows caregivers to remotely access vital sign measurements of patients at home. It collects measurements captured on commercially available wireless or tethered medical devices designed for home use, stores and displays the collected information on the LCD screen, and transmits the information to a secure host server. Caregivers can review the transmitted information and select education and reminders for patients using a caregiver software interface to the host server. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers. Patients can also optionally engage in video conferences with caregivers and answer the caregivers' questions by participating in surveys.
The Intel * Health Guide PHS6000 is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. It is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable
{2}------------------------------------------------
#### 510/k) Notification Submission - Abbreviated Intel® Health Guide PHS6000
of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.
## Technological Characteristics
The Intel® Health Guide PHS6000 is substantially equivalent to the predicate devices in terms of data collection software functionality, operating system for the patient device, communication method of patient device with central server, types of sensors which can be interfaced to the patient device, implementation method of collecting data from sensors, sensor software, connectivity, communication protocol, power source, and display method.
#### Safety and Efficacy
The Intel® Health Guide PHS6000 does not rely on an assessment of clinical performance data. The device conforms to FDA's recognized consensus standards and relies on its conformity to demonstrate the safety and efficacy. The device introduces no new questions concerning the safety or efficacy and is, therefore, substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The bird is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 7 2008
Digital Health Group c/o Ms. Maureen Glynn Director of Regulatory Affairs Intel Corporation 1900 Prairic City Road FM7-197 Folsom, CA 95630
Re: K080798
Trade/Device Name: Intel® Health Guide PHS6000 Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitters and Receivers Regulatory Class: Class II Product Code: DRG Dated: June 10, 2008 Received: June 12, 2008
Dear Ms. Glynn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Ms. Maureen Glynn
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Qomall
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use:
510(k) Number:
KOBO 798
Intel® Ilealth Guide PHS6000 Device Name:
Indications for Use:
The Intel® Health Guide PHS6000 is a communication tool that allows caregivers to remotely access vital sign measurements of patients at home. It collects measurements captured on commercially available wireless or tethered medical devices designed for home use, stores and displays the collected information on the LCD screen, and transmits the information to a secure host server. Caregivers can review the transmitted information and select education and reminders for patients using a caregiver software interface to the host server. Patients can review the stored vital sign measurement information and receive coucational and motivational content from caregivers. Patients can also optionally engage in video conferences with caregivers and answer the caregivers` questions by participating in surveys.
The Intel® Health Guide PHS6000 is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. It is contraindicated for patients requiring direct medical mine vision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.
Prescription Usc ਮ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sigh-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
| 510(k) Number | K080789 |
|---------------|---------|
|---------------|---------|
Pagege I of I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.