GILRAS SLIT LAMP, MODELS GR-SL7, GR-SL36, GR-SL54 AND GR-SL72
Applicant
Us Ophthalmic, LLC
Product Code
HJO · Ophthalmic
Decision Date
Sep 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Indications for Use
The GILRAS Slit Lamp is an AC-power slit lamp biomicroscope intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segments.
Device Story
The GILRAS Slit Lamp is an AC-powered biomicroscope used by eye care professionals in clinical settings to examine the anterior eye segment. The device uses a halogen light source and adjustable optics to illuminate and magnify the eye, allowing the clinician to visualize structures from the cornea to the posterior capsule. The clinician operates the device to perform visual inspections, aiding in the diagnosis of ocular diseases or trauma. The device features adjustable slit width, length, angle, and magnification levels, along with various filters to enhance visualization. It includes a chin-rest and forehead-rest for patient stabilization during the examination. The output is a direct magnified view of the patient's eye, which the clinician uses to inform clinical decision-making regarding ocular health and treatment.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and compliance with recognized consensus standards, specifically ISO 10939:2007 (Ophthalmic instruments - Slit-lamp microscopes) and IEC 60601-1 (Medical Electrical Equipment - General Requirements for Safety).
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties of the anterior eye segment. Prescription use only.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
# K080776
## Summary of Safety and Effectiveness for the GILRAS Slit Lamp
submitted by Al Weisenborn 19526 East Lake Drive Miami, Florida 33015 Phone: (305) 829-3437
SEP - 9 2008
on behalf of
USOphthalmic 7255 NW 68th Street Unit # 9 Miami, FL 33166 Phone: (305) 969-4545
Contact Person: Ezequiel D. Lukin Device Trade Name: GILRAS Slit Lamp Common Name: Slit Lamp Classification Name: Biomicroscope, Slit-lamp, AC-powered per 21 CFR § 886.1850
#### Identification of a Legally Marketed Predicate Device
The USOphthalmic GILRAS Slit Lamp is substantially equivalent to 66 Vision-Tech YZ Slit Lamp that is legally marketed and distributed by Suzhou 66 Vision-Tech Co., LTD pursuant to premarket notification K033190.
#### Device Description
The GILRAS Slit Lamp is an AC-power slit lamp biomicroscope intended for use in eye examination. There are 3 models GR-SL36, GR-SL54, and GR-SL72. These models differ only in the supplied accessories. All models have the same operating characteristics and intended use.
#### Intended Use
The GILRAS Slit Lamp is an AC-power slit lamp biomicroscope intended for use in eye examination of the anterior eve segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segments.
{1}------------------------------------------------
# Summary of Technological Characteristics
An 23-point comparison of the technological characteristics of the USOphthalmic GILRAS Slit Lamp and the predicate devices was performed. The devices were found to be substantially equivalent as shown in the table below.
| Feature | GILRAS Slit Lamp | 66 Vision-Tech YZ Slit Lamp |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Manufacturer: Shanghai<br>MediWorks Precision Instruments,<br>LTD<br>Relabeler: USOphthalmic | Suzhou 66 Vision-Tech Co., LTD |
| Trade name | GILRAS Slit Lamp | 66 Vision-Tech YZ Slit Lamp |
| 510(k) Number | K080776 | K033190 |
| Intended use | The GILRAS Slit Lamp is an AC-<br>power slit lamp biomicroscope<br>intended for use in eye examination<br>of the anterior eye segment, from the<br>cornea epithelium to the posterior<br>capsule. It is used to aid in the<br>diagnosis of diseases or trauma which<br>affect the structural properties of the<br>anterior eye segments. | The 66 Vision-Tech YZ Slit lamp is<br>an AC-power slit lamp biomicroscope<br>intended for use in eye examination<br>of the anterior eye segment, from the<br>cornea epithelium to the posterior<br>capsule. It is used to aid in the<br>diagnosis of diseases or trauma which<br>affects the structural properties of the<br>anterior eye segment. |
| Method of Operation | AC-powered | AC-powered |
| Exposure Parameters | 50000 Lux Maximum<br>Continuously adjustable from 0 to<br>50000 Lux | 75000 Lux at position ¼<br>150000 Lux at position ½<br>300000 Lux at position 1 |
| Data Collection and/or<br>Display Systems | None | None |
| Flammability of<br>Materials | Non-combustion supporting materials | Non-combustion supporting materials |
| Maximum temperature<br>of Parts of the Device<br>Held by the Operator or<br>Accessible to the | 30°C | 30°C |
| Eyepiece Power | 12.5X | 12.5X |
| Total Magnification | GR-SL36 - 10X, 16X, 25X<br>GR-SL54 - 6X, 10X, 16X, 25X, 40X<br>GR-SL72 - 10X, 16X, 25X | 6X, 10X, 16X, 25X, 40X |
| Stereo Angle | 13° | 13° |
| Diopter Adjustment | -6D to +6D | -5D to +3D |
| Pupillary Adjustment: | 55mm to 78.5mm | 55mm to 78.5mm |
| Slit Width | Continuous from 14mm to 0mm (at<br>14mm slit becomes a circle) | Continuous from 9mm to 0mm (at<br>14mm slit becomes a circle) |
| Feature | GILRAS Slit Lamp | 66 Vision-Tech YZ Slit Lamp |
| Slit Length | Continuous from 14mm to 0mm (at<br>14mm,slit becomes a circle) | Continuous from 9mm to 0mm (at<br>14mm, slit becomes a circle) |
| Slit Angle | 0 to 180 degree with horizontal<br>scanning capability | 0 to 180 degree with horizontal scan-<br>ning capability |
| Aperture Diameters | 14, 8, 5, 3, 0.5,and 0.2mm | 9mm, 8mm, 5mm, 3mm, 2mm, 1mm,<br>0.2mm |
| Filters | Heat-absorbing, Neutral Density,<br>Cobalt Blue, and Red-Free | Heat absorption, gray, red-free, and<br>blue |
| Slit Inclination | 5°, 10°, 15°, and 20° | 5°, 10°, 15°, and 20° |
| Illumination Lamp | 6V, 20W halogen lamp | 12V, 30W halogen lamp |
| Hruby Lens | Not available | Optionally Available |
| Fixation Target | Green LED | Red LED |
| Brightness Controls | Continuously adjustable from 0 to<br>50000 Lux | 3-mode: low, medium and high |
| Patient Contact<br>Materials | Chin-rest paper | Chin-rest paper |
| | Forehead-rest -<br>Polytetrafluoroethylene | Forehead-rest -<br>Polytetrafluoroethylene |
| Beam Splitter | GR-SL36 - No | |
| | GR-SL54 - No | Optional |
| | GR-SL72 - Yes | |
Error! Unknown document property name. Error! Unknown document property name., Error! Unknown document Revised August 21, 2008 property name.
{2}------------------------------------------------
## Summary of Performance Data
The USOphthalmic GILRAS Slit Lamp complies with the requirements of listed FDA Recognized Consensus Standards.
- ISO 10939:2007, Ophthalmic instruments -- Slit-lamp microscopes .
- IEC 60601-1, Medical Electrical Equipment Part 1: General . Requirements for Safety
The USOphthalmic GILRAS Slit Lamp is substantially equivalent to the 66 Vision-Tech YZ Slit Lamp that is legally marketed and distributed by Suzhou 66 Vision-Tech Co., LTD. This has been demonstrated through a 8-point technological comparison of features.
Because the USOphthalmic GILRAS Slit Lamp meets the requirements of the stated standards and embody technological characteristics essentially identical to the predicate device, we believe the device is safe and effective and performs as well as or better than the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 9 2008
USOphthalmic, LLC c/o Al Weisenborn 7255 NW 68th Street, Unit #9 Miami, FL 33166
Re: K080776
Trade/Device Name: GILRAS Slit Lamp, Models GR-SL36, GR-SL54 and GR-SL72 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slit-lamp Biomicroscope Regulatory Class: II Product Code: HJO Dated: August 14, 2008 Received: August 15, 2008
Dear Mr. Weisenborn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarlee, Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malvina B. Eggersta, und
Malvina B. Eydelman, M.L Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
Page 1 of 1
510(k) Number (if known):
Device Name: GILRAS Slit Lamp
Indications for Use:
The GILRAS Slit Lamp is an AC-power slit lamp biomicroscope intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segments .
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Danid L. Kaushn
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K080776
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.