SOFLENS DAILY DISPOSABLE TORIC (HILAFILCON B) VISIBILITY TINTED CONTACT LENS
K080755 · Bausch & Lomb, Inc. · MVN · Jun 10, 2008 · Ophthalmic
Device Facts
Record ID
K080755
Device Name
SOFLENS DAILY DISPOSABLE TORIC (HILAFILCON B) VISIBILITY TINTED CONTACT LENS
Applicant
Bausch & Lomb, Inc.
Product Code
MVN · Ophthalmic
Decision Date
Jun 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, and astigmatism) in not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 5.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D. The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
Device Story
Soft hydrophilic contact lens for correction of refractive ametropia and astigmatism. Material: hilafilcon B (copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone); 59% water content; tinted with Reactive Blue Dye 246. Cast-molded design. Prescribed by eye care professionals for daily wear; patient discards lens after each single use. Provides visual correction for patients with astigmatism up to 5.00 diopters. Benefits include convenience of daily disposable modality combined with toric design for astigmatism correction.
Clinical Evidence
No new clinical data provided. Safety and effectiveness of hilafilcon B material established in K994125; safety and effectiveness of daily disposable modality established in K061157; toric design performance established in K941370.
Technological Characteristics
Material: hilafilcon B (hydrophilic copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone). Water content: 59%. FDA Group II (high water/non-ionic). Manufacturing: Cast-molded. Design: Toric, aspheric optics. Diameter: 14.2mm. Base curve: 8.6mm. Tint: Reactive Blue Dye 246. Sterilization: Sterile saline solution.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, and astigmatism (≤5.00D) in non-aphakic, non-diseased eyes. Prescription range: +20.00D to -20.00D. Single-use disposable.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
080755
## JUN 1 0 2008
510(K) SUMMARY Bausch & Lomb, Inc. 510(k) Premarket Notification Bausch & Lomb® SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens
#### Applicant's Name and Address 1.
Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609
Susan Pate 2. Contact Person Associate Manager Global Regulatory Affairs Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609 (585) 338-6362 susan.m.pate@bausch.com
#### 3. Identification of Device
| Common Name: | soft contact lens (hydrophilic) |
|------------------------|---------------------------------------------------------------------|
| Trade Name: | SofLens® daily disposable toric (hilafilcon B)<br>Visibility Tinted |
| Classification: | Daily Wear Contact Lens |
| Device classification: | Class II (21 CFR 886.5925) |
| Product Code: | MVN |
Predicate Devices:
The predicate devices, SofLens daily disposable (hilafilcon B) and SofLens59 (hilafilcon B) Contact Lenses, were selected to demonstrate substantial equivalence of the lens material, hilafilcon B, and replacement modality, singleuse disposable lens. The predicate device SofLens toric (alphafilcon A) Contact Lenses was selected to demonstrate substantial equivalence to the Bausch & Lomb toric lens design.
{1}------------------------------------------------
The SofLens daily disposable toric (hilafilcon B) Visibility Contact Lens is substantially equivalent to the currently marketed SofLens daily disposable (hilafilcon B) Visibility Tinted Contact Lens cleared in 510(k) Premarket Notification No. K061158 on June 22, 2006, SofLens59 (hilafilcon B) Visibility tinted Contact Lens 510(k) Premarket Notification K994125 cleared on March 3, 2000 and SofLens Toric (alphafilcon A) Visibility Tinted Contact Lens 510(k) Premarket Notification No. K941370 cleared on April 2,1994.
The differences between the devices follow.
{2}------------------------------------------------
| Feature | SofLens daily disposable toric | SofLens daily disposable | SofLens59 | SofLens Toric |
|------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|
| FDA Group | Group II<br>(high water/non-ionic) | Group II<br>(high water/non-ionic)<br>Castmolded | Group II<br>(high water/non-ionic)<br>Castmolded | Group II<br>(high water/non ionic)<br>Castmolded |
| Manufacturing<br>Process | Castmolded | Castmolded | Castmolded | Castmolded |
| USAN Name | hilafilcon B | hilafilcon B | hilafilcon B | alphafilcon A |
| Indications | Toric<br>Daily Wear | Spherical<br>Daily Wear | Spherical<br>Daily Wear | Toric<br>Daily Wear |
| Modality | Single-use disposable | Single-use disposable | Frequent replacement | Single-use disposable or<br>Frequent replacement |
| Design Properties | | | | |
| Diameter | 14.2mm | 14.2mm | 14.2mm | 14.5mm |
| Center Thickness<br>(-3.00D) | 0.125mm | 0.090mm | 0.140mm | 0.190mm |
| Base Curve | 8.6mm | 8.6mm | 8.6mm | 8.5mm |
| Sphere powers | Plano to -6.00D in 0.25D<br>steps<br>-6.50D to -9.00D in 0.50D<br>steps | +6.50D to -6.50D in 0.25D<br>steps<br>-7.00D to -9.00D in 0.50D<br>steps | +6.50D to -6.50D in 0.25D<br>steps<br>-7.00D to -9.00D in 0.50D<br>steps | +6.00D to -6.00D in 0.25D<br>steps<br>-6.50D to -9.00D in 0.50D<br>steps |
| Cylinder Powers | -0.75D, -1.25D, -1.75D | n/a | n/a | -0.75D, -1.25D, -1.75D,<br>-2.25D and -2.75D |
| Axis | 0° to 180° in 10°<br>Increments | n/a | n/a | 0° to 180° in 10°<br>Increments |
| Aspheric Optics | Yes | Yes | No | No |
SIMILARITIES and DIFFERENCES
Section 5, Page 4
{3}------------------------------------------------
### 4. Description of device
The SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens is a soft hydrophilic contact lens which is available in a toric lens design for the correction of astigmatism. The lens is made from the hilafilcon B material, a hydrophilic copolymer of 2-hydroxyethyl methacrylate and N-vinyl pyrrolidone, and is 59% water by weight when immersed in a sterile saline solution. This lens is tinted blue with Reactive Blue Dye 246.
### 5. Intended use
The SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, and astigmatism) in not-aphakic persons with non-diseased eves, exhibiting astigmatism of 5.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D.
The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
### ે. Description of Safety and Substantial Equivalence
A series of preclinical and clinical studies were completed and previously submitted in Premarket Notifications K061157, K994125 and K941370.
### Non-Clinical Laboratory Testing:
A series of in vitro and in vivo preclinical toxicology and biocompatibility testing was performed to assess the safety and effectiveness of the contact lens material, hilafilcon B. Testing was performed in accordance with FDA quideline titled Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, May 1994. The non-clinical testing can be found in the currently marketed SofLens®59 (hilafilcon B) Visibility Tinted Contact Lens 510(k) Premarket Notification K994125 cleared on March 3, 2000.
Stability testing has demonstrated a five vear expiration date for the sterile lenses.
### Clinical Testing:
The clinical performance to confirm safety and effectiveness of the lens material, hilafilcon B, was conducted on the SofLens59 (hilafilcon B) Visibility Tinted Contact Lens and cleared in the Premarket Notification K994125 on March 3. 2000. The clinical performance to confirm the safety and effectiveness of the lens modality, daily disposable, was conducted on the SofLens daily disposable (hilafilcon B) Visibility Tinted Contact Lens cleared under Premarket
{4}------------------------------------------------
Notification K061157 on June 22, 2006. The Bausch & Lomb toric contact lens design was cleared in the SofLens Toric (alphafilcon A) Visibility Tinted Contact Lens Premarket Notification K941370 on April 2, 1994.
#### 7. Substantial Equivalence
The SofLens daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens is substantially equivalent to the currently marketed SofLens daily disposable (hilafilcon B) Visibility Tinted Contact Lens 510(k) Premarket Notification K061157 cleared on June 22, 2006; SofLens59 (hilafilcon B) Visibility Tinted Contact Lens 510(k) Premarket Notification K994125 cleared on March 3, 2000; and SofLens Toric (alphafilcon A) Visibility Tinted Contact Lens 510(k) Premarket Notification K941370 cleared on April 2, 1994. The lenses are similar in that all lenses fall into FDA Group II for soft (hydrophilic) contact lenses because the ionic content is less than 1% (nonionic polymer material), the water content is greater than 50% (hilafilcon B - 59% and alphafilcon A - 66%) and are manufactured with the same manufacturing process (cast molding).
{5}------------------------------------------------
Image /page/5/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 0 2008
Bausch & Lomb, Inc. c/o Ms. Debra Ketchum Manager, Global Regulatory Affairs 1400 North Goodman St. Rochester, NY 14609
Re: K080755
Trade/Device Name: Bausch & Lomb® SofLens® Daily Disposable Toric (hilafilcon B) Visibility Tinted Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: MVN Dated: March 14, 2008 Received: March 18, 2008
Dear Ms. Ketchum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 – Ms. Debra Ketchum
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
M. B. Egerlton, mD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K080755
Device Name:
SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens
Indication for Use
The SofLens® daily disposable toric (hilafilcon B) Visibility Tinted Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, and astigmatism) in not-aphakic persons with non-diseased eyes, exhibiting astigmatism of 5.00 diopters or less, that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +20.00D to -20.00D.
The lens is to be prescribed for single-use disposable wear, and is to be discarded after each removal.
Prescription Use X
Over-the -counter-use_ ાર
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Njudiet
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K080755
Section 4, Page 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.