MODIFICATION TO SPHYGMOCOR CARDIOVASCULAR MANAGEMENT SYSTEM
Applicant
Atcor Medical
Product Code
DSK · Cardiovascular
Decision Date
Apr 23, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1110
Device Class
Class 2
Indications for Use
The SphygmoCor® Cardiovascular Management System (CvMS) provides a derived ascending aortic blood pressure waveform and a range of central arterial indices. The CvMS is used with a tonometer over a radial artery calibrated with a standard cuff blood pressure measurement. It is to be used on those patients where information related to ascending aortic blood pressure is desired but in the opinion of the physician, the risks of cardiac catheterization procedure or other invasive monitoring may outweigh the benefits. The CvMS Heart Rate Variability (HRV) option is intended for use in obtaining HRV measurements in response to controlled exercises. The CvMS Pulse Wave Velocity (PWV) option is intended for use in obtaining PWV measurements. The PWV option is used on adult patients only.
Device Story
SphygmoCor CvMS is a computerized diagnostic system for non-invasive cardiovascular assessment. Inputs include radial artery pressure waveforms captured via applanation tonometry, standard cuff blood pressure measurements for calibration, and ECG signals for timing. The system processes these inputs to derive ascending aortic blood pressure waveforms, central arterial indices, HRV, and PWV. Used in clinical settings by healthcare professionals; output provides clinicians with hemodynamic data to inform cardiovascular management decisions, serving as an alternative to invasive cardiac catheterization. PWV calculation involves measuring distance between two arterial sites (e.g., carotid-femoral) and dividing by the pulse transit time (PTT) derived from ECG-referenced pressure waveforms. Benefits include reduced patient risk compared to invasive monitoring.
Clinical Evidence
Clinical comparison study conducted comparing the modified SphygmoCor CvMS side-by-side with the predicate device. Results demonstrated similar performance between the two devices. No specific statistical metrics (e.g., sensitivity, specificity) were provided in the summary.
Technological Characteristics
System consists of a computer-based platform with applanation tonometry sensors (Tecason and RTV Silicon) and ECG leads/electrodes. Connectivity is via standard physiological sensor interfaces. Software-based calculation of central arterial indices, HRV, and PWV. Operates as a non-invasive diagnostic tool.
Indications for Use
Indicated for adult patients requiring non-invasive assessment of central arterial indices, including ascending aortic blood pressure, heart rate variability (HRV) during controlled exercise, and pulse wave velocity (PWV).
Regulatory Classification
Identification
A blood pressure computer is a device that accepts the electrical signal from a blood pressure transducer amplifier and indicates the systolic, diastolic, or mean pressure based on the input signal.
Predicate Devices
SphygmoCor Cardiovascular Management System CvMS (K070795)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for AtCor Medical, along with the text "K080670/S1" and "Pi/5". The AtCor Medical logo is a stylized "AtCor" with the word "MEDICAL" underneath. To the right of the logo is a black square with white dots. The text "K080670/S1" and "Pi/5" are handwritten and appear to be some sort of identification or tracking number.
# 510(k) Summary - SpнygмoCor CvMS
| Date Prepared | 5th March 2008 | | APR 23 2008 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-------------------------------------------------------------|
| Submitter | AtCor Medical Pty Ltd<br>Suite 11, 1059-1063 Victoria Road<br>West Ryde, NSW 2114<br>Australia | | |
| Establishment Number | 9710654 | | |
| Official Contact | John Abram<br>Manager, Regulatory Affairs and Quality Assurance<br>Suite 11, 1059-1063 Victoria Road<br>West Ryde, NSW 2114<br>Australia<br>Ph: +61 2 9874 8761<br>Fax: +61 2 9874 9022<br>j.abram@atcormedical.com | | |
| Secondary Contact | Doug Kurschinski<br>AtCor Medical, Inc. (USA)<br>Sr. VP - US Commercial Operations<br>Ph: 630-799-8215 | | |
| Classification(s) | | | |
| | Product Code | Classification Reference | Common/Usual Name |
| | DSK | 21 C.F.R. § 870.1110<br>(Class II) | computer, blood-pressure |
| | DRT | 21 C.F.R. § 870.2300<br>(Class II) | monitor, cardiac (incl. cardio-<br>tachometer & rate alarm) |
| Proprietary Name | SphygmoCor® Cardiovascular Management System (CvMS) | | |
| Predicate Device(s) | Primary (un-modified device) - K070795 SphygmoCor Cardiovascular<br>Management System CvMS (AtCor Medical Pty Ltd)<br>Secondary - K013434 Vascular Profiling System VP-2000 (Colin Corp) | | |
| Reason for submission | Modification to existing product | | |
Intended Use
The SphygmoCor® Cardiovascular Management System (CvMS) provides a derived ascending aortic blood pressure waveform and a range of central arterial indices. The CvMS is used with a tonometer over a radial artery calibrated with a standard cuff blood pressure measurement. It is to be used on those patients where information related to ascending aortic blood pressure is desired but in the opinion of the physician, the risks of cardiac catheterization procedure or other invasive monitoring may outweigh the benefits.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the text "K080670/S1 P2/5" in the upper right corner. Below this text is the logo for AtCor Medical. The logo consists of the text "AtCor MEDICAL" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font size below "AtCor". To the right of the text is a stylized graphic of a square with rounded corners and a series of dots emanating from it.
The CvMS Heart Rate Variability (HRV) option is intended for use in obtaining HRV measurements in response to controlled exercises.
The CvMS Pulse Wave Velocity (PWV) option is intended for use in obtaining PWV measurements. The PWV option is used on adult patients only.
## Device Description
The SphygmoCor CvMS device (un-modified device) is a computerized tool for providing a derived ascending aortic blood pressure waveform and a range of central anterial indices. The CvMS is used with a tonometer over the radial artery to capture a pressure waveform which is used to derive central pressure and is calibrated with a standard blood pressure cuff measurement. The CvMS is intended for use on those patients where information related to ascending aortic blood pressure is desired but in the opinion of the physician, the risks of cardiac catheterization procedure or other invasive monitoring may outweigh the benefits. In addition, the CvMS incorporates an option to enable users to measure Heart Rate Variability (HRV) in response to controlled exercises.
The SphygmoCor CvMS with the PWV option (modified device) is a software addition to the SphygmoCor CvMS device (K070795). This software feature provides an additional central arterial indice - Pulse Wave Velocity (PWV).
#### Description of Modification
The PWV option provides a modular software addition to the SphygmoCor CvMS (K070795). This simple software addition allows Pulse Wave Velocity to be calculated between two sites in the arterial tree (typically carotid-femoral). This is achieved by capturing arterial pressure waveforms (using tonometry) simultaneously with ECG data. The tonometry and ECG capabilities are unchanged from the predicate device.
The PWV software then calculates the time difference (.st) between the two pressure waveforms using the ECG as a timing reference. The PWV is then simply calculated by measuring the distance between the two arteries and then dividing by At.
#### Device Comparison to Predicate(s)
A device comparison is provided in the table below which shows the differences between the Predicate devices and the modified device. In summary, the modified device has the following similarities to the previously cleared predicate devices:
- ం Similar intended use
- ಂ Same operating principle
- Similar technologies ்
- 0 Same manufacturing process
{2}------------------------------------------------
Atcor �
# Comparison Table
| | AtCor Medical<br>SphygmoCor CVMS<br>(Primary Predicate Un-modified Device<br>K070795) | Golm Corporation<br>Vascular Profiling System VP-2000<br>(including pulse wave unit 10-100)<br>Secondary Predicate K013434) | AtCor, Medical<br>SphygmoCor CVMS<br>(Modified Device) | Equivalence | Performance | | | | |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-------------------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Intended Use | The SphygmoCor® Cardiovascular<br>Management System (CVMS) provides a<br>derived ascending aortic blood pressure<br>waveform and a range of central arterial<br>indices. The CvMS is used with a tonometer<br>over a radial artery calibrated with a standard<br>cuff blood pressure measurement. It is to be<br>used on those patients where information<br>related to ascending aortic blood pressure is<br>desired but in the opinion of the physician, the<br>risks of cardiac catheterization procedure or<br>other invasive monitoring may outweigh the<br>benefits. | VP-2000/1000 is a non-invasive diagnostic<br>system designed to assist in the detection of<br>peripheral vascular diseases. It has a<br>capability of measuring; non-invasive blood<br>pressures, heart rate, pulse wave, and heart<br>sound. It also has a capability of calculating<br>ABI (Ankle Brachial Index), Pulse Wave<br>Velocity and Augmentation Index. The<br>instrument is used in a vascular laboratory,<br>clinic, hospital, doctor's office, and other<br>medical facilities where the non-invasive<br>peripheral vascular test is conducted. It is<br>used on adult patients only. | The SphygmoCor® Cardiovascular Management<br>System (CVMS) provides a derived ascending aortic<br>blood pressure waveform and a range of central<br>arterial indices. The CvMS is used with a tonometer<br>over a radial artery calibrated with a standard cuff<br>blood pressure measurement. It is to be used on<br>those patients where information related to ascending<br>aortic blood pressure is desired but in the opinion of<br>the physician, the risks of cardiac catheterization<br>procedure or other invasive monitoring may outweigh<br>the benefits. | Equivalent; The<br>scope of this<br>comparison is<br>regarding the PWV<br>function only. | PWV Calculation | Not Applicable | PWV is calculated by measuring the distance<br>between distal and proximal sites then<br>dividing by the pulse transit time (PTT)<br>between the same sites. | PWW is calculated by measuring the distance between<br>distal and proximal sites then dividing by the Mean<br>DeltaT time (ΔT) between the same sites. | Equivalent to<br>Secondary |
| Hardware | The CvMS Heart Rate Variability (HRV) option<br>is intended for use in obtaining HRV<br>measurements in response to controlled<br>exercises. | Modular Hardware component for use in an<br>existing system. Physiological Sensors:<br>- Applanation Tonometry pressure sensors<br>- BP Ankle & Brachial Cuff<br>- ECG | The CvMS Heart Rate Variability (HRV) option is<br>intended for use in obtaining HRV measurements in<br>response to controlled exercises.<br><br>The CvMS Pulse Wave Velocity (PWV) option is<br>intended for use in obtaining PWV measurements.<br>The PWW option is used on adult patients only. | Equivalent;<br>Primary - Same<br>hardware<br>Secondary - Same<br>technology | Distance<br>Measurement | Not Applicable | Performed manually by clinician. | Performed manually by clinician. | Equivalent to<br>Secondary |
| Software | | Modular Software component for use in an<br>existing system. | Modular Software component for use in an existing<br>system. | Equivalent | Time Measurement | Not Applicable | Single recording using two pulse waveform<br>measurements with the cuff-based Arterial<br>Applanation Tonometer. | Two recordings in sequence are made at the distal<br>and proximal sites using Applanation tonometry in<br>conjunction with the ECG. The ECG serves as a<br>reference to determine the Mean DeltaT time (ΔT)<br>between the two sites. | Equivalent to<br>Secondary |
| Materials | | Non-invasive physiological Sensors for PWV<br>& HRV function:<br>Tonometer: Tecason™ & RTV Silicon<br>ECG: ECG cables and leads (FDA Cleared<br>K945034)<br>ECG Electrodes (FDA Cleared K946273) | Existing Non-invasive physiological Sensors for PWV<br>function:<br>Tonometer: Tecason™ & RTV Silicon<br>ECG: ECG cables and leads (FDA Cleared K945034)<br>ECG Electrodes (FDA Cleared K946273) | Equivalent; | | | | | |
# K080670/si
中315
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "Atcores" in a vertical orientation. The letters are large and bold, with "Atc" stacked on top of "ores". To the right of the word, the word "Medical" is written vertically in smaller letters. There is also a logo at the top of the image.
...
Summary Comparison to Predicate Devices
K080670/si
P4/5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for AtCor Medical, which is in bold black font. To the right of the logo is a black square with a white arrow pointing up and to the right, surrounded by three dots. Above the logo is the text "K080670/si" and below that is "P 5/5".
# Testing Summary
Design and Verification activities were performed on the SphygmoCor CvMS as a result of the risk analysis and product requirements in accordance with 21 CFR 820.30.
These tests included:
- o Software verification
- Software validation (System Verification) ്
All test confirmed the product met the acceptance criteria.
Furthermore, a clinical comparison study was conducted which compared SphygmoCor CvMS side-byside with the predicate. The tests results demonstrated similar performance between the two devices.
### Conclusion
AtCor Medical has determined that the PWV software addition has not altered the safety and effectiveness of the device based on the company's design control procedures and test results from V&V and clinical comparison testing. The SphygmoCor CvMS System is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Public Health Service
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 2 8 2008
Atcor Medical c/o Mr. John Abram Manager, Regulatory Affairs & Quality Assurance 11/1059-1063 Victoria Road West Ryde NSW Australia, 2114
Re: K080670
Modification to Sphygmocor Cardiovascular Management System Regulation Number: 21 CFR 870.1110 Regulation Name: Blood Pressure Computer Regulatory Class: Class II (two) Product Code: DSK, DRT Dated: April 01, 2008 Received: April 04, 2008
Dear Mr. Abram:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
# Page 2 - Mr. John Abram
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, pcrmits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Bfimmerman fr
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for AtCor Medical. The logo features the company name in a bold, sans-serif font, with the word "MEDICAL" in smaller letters underneath. To the right of the name is a graphic of a square with dots around it. The logo is simple and professional, conveying a sense of reliability and expertise.
#### Indication for Use
510(k) Number (if known):
Device Name:
SPHYGMOCOR CVMS-PWV
Indication for Use
The SphygmoCor® Cardiovascular Management System (CvMS) provides a derived ascending aortic blood pressure waveform and a range of central arterial indices. The CvMS is used with a tonometer over a radial artery calibrated with a standard cuff blood pressure measurement. It is to be used on those patients where information related to ascending aortic blood pressure is desired but in the opinion of the physician, the risks of cardiac catheterization procedure or other invasive monitoring may outweigh the benefits.
The CvMS Heart Rate Variability (HRV) option is intended for use in obtaining HRV measurements in response to controlled exercises.
The CvMS Pulse Wave Velocity (PWV) option is intended for use in obtaining PWV measurements. The PWV option is used on adult patients only.
Prescription Use _ X AND/OR
Over-The-Counter Use_
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
Blummo
Division Sign Off
Page 1 of_____________________________________________________________________________________________________________________________________________________________________
5" March 2008
Division of Cardiovascular Devices
510(k) Number K080670
Page 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.