POWER DRIVE ART1 2304 WITH MOTOR HANDLES, TOOLS AND ACCESSORIES
K080617 · Richard Wolf Medical Instruments Corp. · HRX · Jun 27, 2008 · Orthopedic
Device Facts
Record ID
K080617
Device Name
POWER DRIVE ART1 2304 WITH MOTOR HANDLES, TOOLS AND ACCESSORIES
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
HRX · Orthopedic
Decision Date
Jun 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
In conjunction with a motorized handle (e.g. Power Stick M5/3), the POWER DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill MI, bones can be severed or machined. Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes: - In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications). - In thoracic surgery, e.g. for the removal of hematomas. - In sinus surgery, e.g. for the removal of polyps. - In spinal surgery (arthroscopic microdiskectomy (AMD) for removing degenerated tissue.
Device Story
POWER DRIVE ART1 2304 is a surgical motor control unit for endoscopic procedures. System comprises generator, motor handles, rotary blades/abraders, saws, and drills. Input: user control via touch screen or integrated keypads on handles. Operation: generator powers motor handles to drive surgical tools; RFID technology enables automatic tool recognition. Output: mechanical rotation/oscillation for tissue resection or bone machining; simultaneous suction for tissue evacuation. Used in OR by surgeons. Integration: CAN-BUS interface allows connection to R. Wolf RIWO NET SYSTEM for remote/speech control. Benefits: improved user comfort, reduced risk of incorrect settings, and efficient tissue removal/bone machining.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Generator-based motor control system. Materials: DLC (diamond-like carbon) coating on rotary blade/abrader sheaths; anodized aluminum for drill housing. Connectivity: CAN-BUS interface for RIWO NET SYSTEM integration. Features: RFID-based automatic tool recognition, touch screen interface, integrated keypads. Power: electrical motor drive. Sterilization: multiple-use components require manual sterilization; single-use components provided sterile.
Indications for Use
Indicated for patients undergoing endoscopic procedures requiring tissue removal or bone machining, including arthroscopy (meniscus resection, soft tissue removal, bone abrasion for ACL/shoulder), thoracic surgery (hematoma removal), sinus surgery (polyp removal), and spinal surgery (arthroscopic microdiskectomy for degenerated tissue).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Ko 80617
Page 1 of 3
# 510(k) Su
| Submitter: | Date of Preparation: | | |
|----------------------------------------|------------------------------------------------|----------|--------------------|
| | 25. February 2008 | | |
| | Revised: 05 June 2008 | | |
| Company / Institution name: | FDA establishment registration | | |
| RICHARD WOLF MEDICAL INSTRUMENTS CORP. | number: | | |
| | 14 184 79 | | |
| Division name (if applicable): | Phone number (include area code): | | |
| N.A. | (847) 913 1113 | | |
| Street address: | FAX number (include area code): | | |
| 353 Corporate Woods Parkway | (847) 913-0924 | | |
| City: | State/Province: | Country: | ZIP / Postal Code: |
| Vernon Hills | Illinois | USA | IL 60061 |
| Contact name: | Mr. Ron Haselhorst | | |
| Contact title: | Quality Assurance / Regulatory Affairs Manager | | |
Product Information:
| Trade name: | Model number: 2304.xxx, |
|--------------------------------------------------------------------------|----------------------------------------------------------------|
| POWER DRIVE ART1 2304 with motor handles, tools and Accessories | 8995X.xxx, 8564.xxx, 8996X.xxx,8997X.xxx, 8998X.xxx, 4997X.xxx |
| Common name: | Classification name: |
| Surgical instrument Motor Unit and accessories/attachments for endoscopy | Surgical instrument motors and accessories/attachments. |
Information on devices to which substantial equivalence is claimed:
| 510(k) Number | Trade or proprietary or model name | Manufacturer |
|---------------|-------------------------------------------------------------------------------------------------|-------------------|
| 1 K030082 | 1 Power Control 2303 with Power Stick M4 | 1 Richard Wolf |
| 2 K871250 | 2 Intra-articulated arthro power system 2161 | 2 Richard Wolf |
| 3 K970088 | 3 RIWO DRIVE Generator with Footswitch, Motor handles and Single Use Rotary Blades and Abraders | 3 Richard Wolf |
| 4 K984304 | 4 RIWO DRIVE Generator, single use or reusable Rotary Cutters and Burrs | 4 Richard Wolf |
| 5 K984521 | 5 RIWO DRIVE small handle | 5 Richard Wolf |
| 6. K030009 | 6. Powershaver SL | 6. Karl Storz |
| 7. K062849 | 7. DYONICS Power II Shaver System | 7. Smith & Nephew |
#### 1.0 Description
The POWER DRIVE ART1 2304 is an enhanced version of our POWER CONTROL 2303, cleared in K030082. The new device has new electronics, new software, additional motor handles, tools and tool holders such as rotary blades/ abraders, saws, sagittal and reciprocal saw attachments, drill chucks, drills and surgical wires.
{1}------------------------------------------------
Kos0617
2.0 Intended Use
Page 283
In conjunction with a motorized handle (e.g. Power Stick M5/3), the POWER Intended use: DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill MI, bones can be severed or machined.
Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes:
- . In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications).
- In thoracic surgery, e.g. for the removal of hematomas. .
- In sinus surgery, e.g. for the removal of polyps. ●
- In spinal surgery (arthroscopic microdiskectomy (AMD) for removing . degenerated tissue.
#### 3.0 Technological Characteristics
The POWER DRIVE ART1 2304, consisting of: generator, motor handles, tools (such as rotary blades/ abraders, sagittal and reciprocal saws, drills and surgical wires) and tool holders (such as sagittal and reciprocal saw attachments and drill chucks).
Additional features are new motor handles, some with integrated keypad as well as automatic tool recognition with the new rotary blades and abraders (via RFID technology) and a touch screen for user control.
Multiple use devices are not provided sterile; they must be startlized per their instruction before each use. Single use devices are provided sterife and designed for only one use on or in-a single patient.
Via the integrated CAN-BUS interface, the POWER CONTROL 2304 can be integrated into the R. Wolf RIWO NET SYSTEM with remote control, speech control, and touch-screen monitor.
#### 4.0 Substantial Equivalence
The submitted POWER Drive 2304 generator have same indications, equivalent design, functions and materials as the previous devices sold by Richard Wolf. The new device has new electronics, new software, additional motor handles, tools and tool holders.
Additional features are new motor handles, some with integrated keypad as well as automatic tool recognition for the new rotary blades and abraders. Together with the touch screen control, this increases the user comfort and avoids damage caused by incorrect settings.
The new inner sheaths of the rotary blades and abraders have a DLC (diamond-like carbon) coating to reduce friction. The housing of the new drill is made of anodized Aluminum. .
Submitted devices pose the same type of questions about safety or effectiveness as the compared devices and the new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing 510(k)-cleared devices sold by Richard Wolf, Karl Storz, and Smith & Nephew
{2}------------------------------------------------
Ko 80617
Page 383
### 5.0 Performance Data No performance standards are known.
- 6.0 Clinical Tests
No clinical tests performed.
### 7.0 Conclusions Drawn
These devices are designed and tested to guarantee the safety and effectiveness, when used according to the instructions manual.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, rendered in thick, black lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 7 2008
Richard Wolf Medical Instruments Corporation % Mr. Ron Haselhorst Quality Assurance/Regulatory Affairs Manager 353 Corporate Woods Parkway Vernon Hills, Illinois 60061
Re: K080617
Trade/Device Name: POWER DRIVE ART1 2304 with motor handles, tools and Accessories Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: II Product Code: HRX, HWE, ERL, GEY Dated: June 5, 2008 Received: June 13, 2008
Dear Mr. Haselhorst:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Ron Haselhorst
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N Mulhearn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kofobi7
## Indications for Use Statement
510(k) Number (if known):
Device Name: POWER DRIVE ARTI 2304 with motor handles, tools and Accessories
- Intended use: In conjunction with a motorized handle (c.g. Power Stick M5/3), the POWER DRIVE ART1 2304 serves as a drive unit for Richard Wolf rotary blades/abraders for the removal of tissue during endoscopic intervention. Simultaneous evacuation (suction) allows continuous removal of the ablated tissue. In conjunction with Power Drill M1, bones can be severed or machined.
Indications and field of use: The product is used in conjunction with endoscopic accessories for the following purposes:
- . In arthroscopy, e.g. resection of the meniscus, for removing soft tissue as well as intraarticular severing or abrasion of bone tissue (e.g. ACL or shoulder applications).
- In theracte surgery, e.g. for the removal of hematomas. .
- � In sinus surgery, e.g. for the removal of polyps.
- . In spinal surgery (arthroscopic microdiskectomy (AMD) for removing degenerated tissue.
Prescription Use > (Part 21CFR 801 Subpart D) AND/OR
Over-The Counter Use (Part 21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
**Division Sign-Off**
Division of General, Restorative, and Neurological Devi
**510(k) Number** K080617
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.