Invia's Corridor4DM application is intended to provide processing, quantification, and multidimensional review of reconstructed cardiac emission tomographic studies (SPECT and PET). Cardiac CT interpretation and calcium quantification are optional features that are integrated into Corridor4DM (SPECT-CT and PET-CT). The calcium-scoring package is a non-invasive diagnostic tool that can be used to evaluate the calcified plaques in the coronary arteries, a risk factor for coronary artery disease. Calcium scoring may be used to monitor the progression overtime of the amount or volume of calcium in the coronary arteries, which may be related to the prognosis of a cardiac attack. Coregistration or fusion of volumetric data (ECT and/or CT) is provided as a quality control for the identification of structures where correlative spatial information is necessary for a diagnostic interpretation.
Device Story
Corridor4DM is a cardiac software application for processing, quantifying, and reviewing SPECT, PET, and CT cardiac images. It accepts reconstructed cardiac emission tomographic data and CT data as input. The device performs quantitative analysis, including calculation of left ventricular chamber volumes, ejection fraction, and transient ischemic dilation (TID) values; it also provides calcium scoring for coronary artery disease risk assessment. It supports coregistration and fusion of ECT and CT volumetric data for spatial correlation. Used in clinical settings by healthcare professionals, the software provides multi-dimensional displays and slice-by-slice visualization. Outputs assist clinicians in diagnostic interpretation, monitoring disease progression, and assessing cardiac function. The device benefits patients by providing non-invasive diagnostic information regarding heart function and coronary artery calcification.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison and system performance characteristics relative to predicate devices.
Technological Characteristics
Software-based application for cardiac image processing and quantification. Features include multi-dimensional display, ECT/CT data fusion, and calcium scoring algorithms. Operates on standard computer hardware; supports local, network, or printed file storage. Connectivity includes digital image retrieval. No specific materials or sterilization methods applicable as it is a software-only device.
Indications for Use
Indicated for patients undergoing cardiac emission tomographic studies (SPECT/PET) or cardiac CT. Used for processing, quantification, and review of cardiac images, including calcium scoring to evaluate coronary artery calcified plaques as a risk factor for coronary artery disease and to monitor progression/regression of coronary calcium.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K080575
## 510k Summary
A) Submitted by: INVIA, LCC 3025 Boardwalk, Suite 200 Ann Arbor, MI 48108 Registration: 9069896
MAY 1 3 2008
Contact: MEDIcept 200 Homer Ave Ashland, MA 01721 F. David Rothkopf 508-231-8842 x20 508-231-8861 Fax
Corridor4DM B) Device Name: Common Name: System, Emission Computed Tomography & System, X-Ray Computed Tomography
| Device Class: | 21 CFR 892.1200 |
|---------------|-------------------------|
| | 21 CFR 892.1750 CLASS I |
KPS, JAK Product Code
C) Predicate:
Reagents of the University of Michigan 3D-MSPECT system (K001026) Siemens, Calcium Scoring Software Package (K990426)
D) Device Description:
Corridor4DM is a comprehensive cardiac emission computed tomographic (ECT) application designed to review and quantitatively analyze cardiac ECT (SPECT and PET) nuclear medicine patient studies. The application provides tools for processing and displaying standard ungated and ECG gated cardiac ECT images on both a slice-by-slice basis and as a 3-dimensional surface rendered images in many user selectable formats. Additionally, it provides quantitative assessments of heart function by computing and displaying left ventricular chamber volumes, ejection fraction and transient ischemic dilation (TID) values and provides an assessment of the data set(s) in comparison to a similar patient population with a low likelihood of cardiac disease. The application provides a number of different single or multi-dimensional display(s) to allow the users to select those that best fit their needs. Cardiac CT interpretation and cardiac calcium quantification are optional features that are integrated into Corridor4DM.
{1}------------------------------------------------
## E) Intended Use:
Invia's Corridor4DM application is intended to provide processing, quantification, and multidimensional review of reconstructed cardiac emission tomographic studies (SPECT and PET). Cardiac CT interpretation and calcium quantification are optional features that are integrated into Corridor4DM (SPECT-CT and PET-CT). The calcium-scoring package is a non-invasive diagnostic tool that can be used to evaluate the calcified plaques in the coronary arteries, a risk factor for coronary artery disease. Calcium scoring may be used to monitor the progression or regression overtime of the amount or volume of calcium in the coronary arteries, which may be related to the prognosis of a cardiac attack. Coregistration or fusion of volumetric data (ECT and/or CT) is provided as a quality control for the identification of structures where correlative spatial information is necessary for a diagnostic interpretation.
# F) Comparison to Predicate Device(s):
The Corridor 4DM has the same intended use, target population, and clinical setting as other SPECT software including the predicate devices. It uses the same technology as the predicate devices (3D-MSPECT system (K001026) and Calcium Scoring Software Package (K990426).
| Features | Corridor4DM | 3D-MSPECT | Siemens Calcium<br>Scoring Software<br>Package |
|-------------------------------------------------------------------------|--------------|-----------|------------------------------------------------|
| Computer based<br>templating system | Yes | Yes | NA |
| Digital image retrieval<br>of patient radiographs | Yes | Yes | NA |
| Patient Studies Type | ECT, PET, CT | ECT, PET | N/A |
| Measurement Tools | Yes | Yes | NA |
| Overlay of prosthetic<br>templates for proper<br>selection by physician | Yes | Yes | NA |
| Generation of Parts List | Yes | Yes | NA |
| Patient files can be<br>saved locally, on a<br>network, or printed. | Yes | Yes | NA |
| Allows for assessment<br>of wear. | Yes | Yes | NA |
| Identify and quantify<br>CT data via imaging | Yes | No | Yes |
{2}------------------------------------------------
## Corridor4DM 510(k) Premarket-Notification Submission
#### G) Guidance Document
Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems; Final, FDA December 3, 1998
H) Conclusion:
Invia, LLC believes that the Corridor4DM is substantially equivalent to the predicate devices based on intended usage, technology comparison and system performance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 2008
Invia Mr. F. David Rothkopf President MEDIcept, Inc. 200 Homer Ave ASHLAND MA 01721
Re: K080575
Trade/Device Name: Corridor4DM Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK, and LLZ Dated: February 20, 2008 Received: February 29, 2008
#### Dear Mr. Rothkopf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assession to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K080575
Device Name: Corridor4DM
Indications for Use:
Invia's Corridor4DM application is intended to provide processing, quantification, and multidimensional review of reconstructed cardiac emission tomographic studies (SPECT and PET). Cardiac CT interpretation and calcium quantification are optional features that are integrated into Corridor4DM (SPECT-CT and PET-CT). The calcium-scoring package is a non-invasive diagnostic tool that can be used to evaluate the calcified plaques in the coronary arteries, a risk factor for coronary artery disease. Calcium scoring may be used to monitor the progression overtime of the amount or volume of calcium in the coronary arteries, which may be related to the prognosis of a cardiac attack. Coregistration or fusion of volumetric data (ECT and/or CT) is provided as a quality control for the identification of structures where correlative spatial information is necessary for a diagnostic interpretation.
Prescription Use X OR Over-the-Counter Use _____21 CFR 801.109
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hory M. Khan
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
February 20, 2008
Corridor4DM
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.