# E-POLY TIBIAL BEARINGS (K080528)

_Biomet Orthopedics, Inc. · JWH · Jun 17, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080528

## Device Facts

- **Applicant:** Biomet Orthopedics, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Jun 17, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

For cemented and un-cemented use.

## Device Story

E-Poly™ Tibial Bearings are orthopedic implants designed for use with previously cleared Biomet knee systems. The device consists of highly-crosslinked ultra-high-molecular-weight polyethylene (UHMWPE) infused with vitamin E. The vitamin E infusion serves to stabilize free radicals and prevent oxidative degradation, thereby improving wear resistance compared to standard polyethylene. The bearings are provided in five sizes and five thicknesses to accommodate various patient anatomies. They are intended for implantation by orthopedic surgeons in clinical settings to restore joint function in patients suffering from degenerative or traumatic knee conditions. The device functions as a bearing surface within a knee prosthesis, reducing friction and wear to improve long-term implant performance and patient outcomes.

## Clinical Evidence

No clinical data provided as a basis for substantial equivalence. Non-clinical laboratory testing was performed to verify device functionality.

## Technological Characteristics

Material: Vitamin E-infused, highly-crosslinked UHMWPE. Form factor: Tibial bearings available in five sizes and five thicknesses. Intended for use with existing Biomet knee prosthesis systems. Sterilization: Not specified.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Maxim Accel Knee System ([K023546](/device/K023546.md))
- Maxim Accel (Vanguard™) PS & Bearings ([K041046](/device/K041046.md))
- Vanguard™ Anterior Stabilized Tibial Bearings ([K050222](/device/K050222.md))
- 100kGy E-Poly™ MaxRom Acetabular Liners ([K070364](/device/K070364.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for BIOMET INC. The logo is in black and white, and the text is in a bold, sans-serif font. The letters are all capitalized, and the word "INC" is smaller than the word "BIOMET". The logo is simple and modern.

086528

JUN 1 7 2008

# 510(k) Summary E-Poly™ Tibial Bearings

| Preparation Date:    | February 22, 2008                                                                  |
|----------------------|------------------------------------------------------------------------------------|
| Applicant/Sponsor:   | Biomet Manufacturing Corp.                                                         |
| Contact Person:      | Melissa Steinwedel<br>Regulatory Specialist                                        |
| Proprietary Name:    | E-Poly™ Tibial Bearings                                                            |
| Common Name:         | Vitamin E Polyethylene Tibial Bearings                                             |
| Classification Name: | Cemented semi-constrained polymer/metal/polymer knee<br>prosthesis (888.3560), JWH |

### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

| Maxim Accel Knee System                       | K023546 | Biomet Inc. |
|-----------------------------------------------|---------|-------------|
| Maxim Accel (Vanguard™) PS & Bearings         | K041046 | Biomet Inc. |
| Vanguard™ Anterior Stabilized Tibial Bearings | K050222 | Biomet Inc. |
| 100kGy E-Poly™ MaxRom Acetabular Liners       | K070364 | Biomet Inc. |

Device Description: The E-Poly™ Tibial Bearings are provided in five sizes. Each size comes in five thicknesses. (There are five basic styles of the E-Poly™ Bearings which are intended to be unsel with previously cleared Biomet knee products.)

#### Indications for Use:

- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

#### Intended Use:

For cemented and un-cemented use,

Summary of Technologies: The E-Poly™ UHMWPE tibial bearings are highly-crosslinked to improve wear resistance, and infused with vitamin E to stabilize free-radicals and prevent oxidative degradation.

Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use.

Clinical Testing: None provided as a basis for substantial equivalence.

All trademarks are property of Biomet

Pri Barces 11/2019 11/25/2017 11:51

5-1

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is a stylized image of an eagle.

UUL 2 1 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Biomet Orthopedics, Inc. % Mr. Gary Baker 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581

Re: K080528

Trade/Device Name: E-Poly Tibial Bearings Regulation Number: 21 CFR 888.3560 Regulation Name: Cemented semi-constrained polymer/metal/polymer knee prosthesis Regulatory Class: Class II Product Code: OIY, JWH, MBH, MBV Dated: June 4, 2008 Received: June 6, 2006

Dear Mr. Baker:

This letter corrects our substantially equivalent letter dated June 17, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, laheling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Gary Baker

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Millerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known): KO80528

E-Poly™ Tibial Liners Device Name:

Indications for Use:

- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

For cemented and un-cemented use.

Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil Rbe Ogle

Division of General, Restorative, and Neurological Devices

**510(k) Number** K080528

---

**Source:** [https://fda-staging.innolitics.com/device/K080528](https://fda-staging.innolitics.com/device/K080528)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).