K080420 · Sep 2, 2008
RESERT XL HLD HIGH LEVEL DISINFECTANT
STERIS Corporation
Patient-soiled endoscopes from clinical use
In-use testing was conducted on pre-cleaned, patient-soiled endoscopes to verify the high-level disinfection efficacy of the device under actual clinical conditions.
Resert™ XL HLD is a ready to use liquid chemical germicide. The product is a 2.0% nominal hydrogen peroxide solution. The minimum recommended concentration (MRC) is 1.5%. Resert XL HLD is intended to be used by health care practitioners in clinical settings as a ready to use liquid chemical disinfectant for the high level disinfection of semi-critical medical devices for which alternative methods of terminal reprocessing are not suitable or available. High Level Disinfectant: Resert™ XL HLD High Level Disinfectant is a high level disinfectant when used or reused undiluted for a maximum of 21 days at a minimum temperature of 20℃ (68°F) for a minimum immersion time of 8 minutes.
Device Story
Resert XL HLD is a ready-to-use 2.0% hydrogen peroxide solution buffered with furoic and phosphorous acids. It functions as a liquid chemical germicide via strong oxidation of cellular components. Used in clinical settings by healthcare practitioners for high-level disinfection of semi-critical medical devices (e.g., flexible endoscopes). The solution is applied via immersion for a minimum of 8 minutes at ≥20°C. It is reusable for up to 21 days. The device benefits patients by providing an effective disinfection method for heat-sensitive devices where other terminal reprocessing is not feasible. Output is a disinfected medical device ready for clinical use.
Clinical Evidence
Bench testing only. Efficacy validated via AOAC/EPA potency testing (bactericidal, fungicidal, virucidal, sporicidal, tuberculocidal) using stressed solutions (end of 21-day life, 5% organic challenge, 400 ppm hard water). Simulated use testing on endoscope sets achieved ≥6 log10 reduction of Mycobacterium terrae. Biocompatibility assessed per FDA guidance (acute dermal/oral toxicity, irritation, sensitization, mutagenicity). Material compatibility confirmed via 134-hour exposure (1005 cycles) on various plastics, metals, and elastomers.
Technological Characteristics
Liquid chemical disinfectant; 2.0% hydrogen peroxide solution buffered with furoic and phosphorous acids. Includes surfactants and corrosion inhibitors. Compatible with various plastics (Teflon, PVC, etc.) and metals (stainless steel, aluminum). Not compatible with brass, copper, or silver. Form factor: 4-liter polyethylene bottle. Shelf-life: 1 year unopened. No software or electronic components.
Indications for Use
Indicated for high-level disinfection of semi-critical medical devices in clinical settings by healthcare practitioners when alternative terminal reprocessing methods are unavailable or unsuitable.
Regulatory Classification
Identification
A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.
Special Controls
*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
Predicate Devices
SPOROX® Sterilizing and Disinfecting Solution (K970230)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP - 2 2008
Image /page/0/Picture/2 description: The image shows the logo for TERI, The Energy and Resources Institute. The logo consists of the letters "TERI" in a bold, sans-serif font, stacked above a stylized graphic. The graphic is composed of several horizontal lines that curve slightly, resembling waves or contours.
# 510(k) Summary For Resert™ XL HLD High Level Disinfectant
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459
| Contact: | John R. (Jack) Scoville, Jr.<br>STERIS Corporation<br>5960 Heisley Road<br>Mentor, Ohio 44060 |
|---------------|-----------------------------------------------------------------------------------------------|
| | Telephone: (440) 392-7330<br>Fax No: (440) 357-9198 |
| Summary Date: | June 05, 2008 |
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{1}------------------------------------------------
#### 1. Device Name
| Trade Name: | Resert™ XL HLD High Level Disinfectant |
|------------------------|------------------------------------------|
| Common/Usual Name: | Liquid Chemical Sterilants/Disinfectants |
| Classification Name: | Liquid Chemical Sterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED. |
#### 2. Predicate Device
SPOROX® Sterilizing and Disinfecting Solution (K970230) manufactured by Rickitt & Colman, Inc.
#### 3. Description of Device
Resert XL HLD High Level Disinfectant is a nominal 2.0% hydrogen peroxide solution, buffered with furoic acid and phosphorous acids. Hydrogen peroxide is the active ingredient in Resert XL HLD, which exerts its germicidal action via a strong oxidation reaction of cellular components.
#### 4. Intended Use
Resert™ XL HLD is a ready to use liquid chemical germicide. The product is a 2.0% nominal hydrogen peroxide solution. The minimum recommended concentration (MRC) is 1.5%. Resert XL HLD is intended to be used by health care practitioners in clinical settings as a ready to use liquid chemical disinfectant for the high level disinfection of semi-critical medical devices for which alternative methods of terminal reprocessing are not suitable or available.
High Level Disinfectant: Resert™ XL HLD High Level Disinfectant is a high level disinfectant when used or reused undiluted for a maximum of 21 days at a minimum temperature of 20℃ (68°F) for a minimum immersion time of 8 minutes.
#### ട. Description of Safety and Substantial Equivalence
Both Resert XL HLD High Level Disinfectant and SPOROX Sterilizing and Disinfecting Solution (SPOROX) formulations have common features and use indications. Their active ingredient is hydrogen peroxide. The major difference is
{2}------------------------------------------------
that Resert XL HLD has a nominal 2.0% hydrogen peroxide solution, buffered with furoic acid and phosphorous acids, while SPOROX has a 7.5% nominal hydrogen peroxide concentration, buffered with phosphoric acid, and they have high level disinfection contact times of 8 minutes and 30 minutes respectively.
The Resert XL HLD High Level Disinfectant has the same technological characteristics (i.e., design, material, chemical composition) as the predicate SPOROX device.
#### 6. Device Comparison Table
A summary of the technological characteristics of the Resert XL HLD High Level Disinfectant in comparison to those of the predicate device is shown in Table 5-1.
| Feature | Resert XL HLD | SPOROX |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Resert XL HLD is intended<br>to be used by health care<br>practioners in clinical<br>setting as a high level<br>disinfectant when used<br>according to the labeled<br>specified directions | SPOROX is intended to be<br>used by health care<br>practioners in clinical<br>setting as a Liquid chemical<br>sterilant and/or high level<br>disinfectant when used<br>according to the labeled<br>specified directions |
| Active Ingredient | 2.0% hydrogen peroxide | 7.5% hydrogen peroxide |
| Surfactants | Yes | Yes |
| Corrosion Inhibitor | Yes | Yes |
| Dilution required | No | No |
| Contact Time - High<br>Level Disinfection | 8 minutes @ 20°C | 30 minutes @20°C |
| Use-Reuse Period | 21 days | 21 days |
| Accessories | Resert Chemical Indicator | SPOROX Chemical<br>Indicator |
| Primary Packaging | 4 Liter Polyethylene bottle | 1 gal. Polyethylene bottle |
| Secondary Packaging | 4-4 Liter bottles per case | 4-1 gal bottles per case |
| Closure System | Vented | Vented |
| Storage Conditions | Cool dry place | Cool dry place |
| Shelf-life | 1 year un-opened | 2 years un-opened |
Table 5-1. Comparison Matrix for Resert XL HLD and SPOROX
#### 7. Performance Testing Summary - Bench
All AOAC/EPA Potency testing was conducted with Resert XL HLD solution, formulated to below the minimum manufacturing specification for hydrogen
{3}------------------------------------------------
peroxide, stressed to the end of its 21 day reuse life, and challenged with 5% organic challenge and 400 ppm hard water. The testing demonstrated the bactericidal, fungicidal, virucidal, sporicidal and tuberculocidal efficacy of Resert XL HLD, utilizing appropriate test methods. Also, Resert XL HLD was assessed for acute dermal toxicity, acute oral toxicity, primary skin irritation, skin sensitization, primary eye irritation, mutagenicity, and residue toxicity, according to the FDA Guidance on biocompatibility testing.
A stability program has been implemented to demonstrate a shelf-life for Resert XL HLD of at least 1 year, if unopened and stored according to label instructions, and reused for up to 21 days at a minimum temperature of 20℃ (68°F).
The test method criteria and a summary of results are presented in Tables 5-2 and 5-3.
| SPOROX Microbiological Efficacy Testing | | |
|--------------------------------------------------------------------------------------------------|-------------------------------------------|--------------------|
| Test Method | Resert XL HLD | SPOROX |
| AOAC Sporicidal Test<br>Bacillus subtilis<br>Clostridium sporogenes | Meets sporicidal efficacy<br>requirements | Same |
| AOAC Use-Dilution Test<br>Pseudomonas aeruginosa<br>Staphylococcus aureus<br>Salmonella enterica | Stressed solution is<br>bactericidal | Same |
| AOAC Fungicidal Test<br>Trichophyton<br>mentagrophytes | Stressed solution is<br>fungicidal | Same |
| EPA Virucidal Test<br>Poliovirus Type I<br>Herpes Simplex Type 1<br>Adenovirus Type 5 | Stressed Solution is<br>virucidal | Same |
| Tuberculocidal Activity Test<br>Mycobacterium terrae | Stressed solution is<br>tuberculocidal | Same<br>(M. bovis) |
| Simulated Use Test<br>Mycobacterium terrae | Meets efficacy requirements | Same |
Table 5-2. Comparison Matrix for Resert XL HLD and SPOROX Microbiological Efficacy Testing
{4}------------------------------------------------
| <br>P | A<br>1.5<br>11<br>- | 7 |
|-------|---------------------|---|
| | to | |
| Toxicology Test | Resert XL HLD | SPOROX |
|--------------------------------------------------------------|---------------------------------------------|-------------------------------------------------|
| Primary dermal Irritation | Slight irritant | Moderate irritant |
| Acute Dermal Toxicity | $Dermal LD_{50} >5000$<br>mg/kg body weight | $Dermal LD_{50} >$<br>5000 mg/kg body<br>weight |
| Ocular Irritation | Mild irritant | Severe irritant |
| Acute Oral Toxicity | $OralLD_{50} >5000$ mg/kg<br>body weight | $OralLD_{50} >5000$<br>mg/kg body<br>weight |
| Skin sensitization | Non-sensitizing | Non-sensitizing |
| Bacterial Reverse Mutation Assay<br>(Mutagenicity Ames Test) | Non-mutagenic | Non-mutagenic |
Table 5-3. Comparison Matrix for Resert XL HLD and SPOROX Toxicological Evaluation
#### 8. Device /Material Compatibility Studies
STERIS Corporation demonstrated the material compatibility of Resert XL HLD with medical devices and device materials, as identified in the product labeling, by conducting studies with actual devices, in both Simulated and In Use applications, and by conducting rigorous evaluations of a number of different plastic, metals, and elastomers commonly used in the construction of medical devices. The materials were exposed to Resert XL HLD solution, formulated above its maximum manufacturing specification for hydrogen peroxide, for up to 134 hours at 25+2°C (equivalent up to 1005 disinfectant cycles). A comparison of the compatible and incompatible materials for Resert XL HLD and the predicate device is shown in Table 5-4.
| Materials Compatible | Resert XL HLD | SPOROX |
|----------------------------------------|---------------|--------|
| Plastics: | | |
| Teflon | X | X |
| Polyester | X | X |
| Polystyrene | X | X |
| Polycarbonate | X | X |
| Polypropylene | X | X |
| Acrylic | X | X |
| Polyvinyl Chloride (PVC) | X | X |
| Acrylonitrile- butadiene-styrene (ABS) | X | X |
| Nylon | X | X |
| Metals: | | |
| Stainless Steel (302,316, 410) | X | X |
| Chrome Plate (Steel) | X | X |
Table 5-4. Comparison Matrix for Resert XL HLD and SPOROX Material Compatibility
{5}------------------------------------------------
### 12080420 STERIS Response to 053008 Request for Additional Information - Appendix B K080420 / Resert XL HLD High Level Disinfectant
| Materials Compatible | Resert XL HLD | SPOROX |
|--------------------------------------------|---------------|--------|
| Platinum | P | X |
| Aluminum | X | X |
| Anodized Aluminum (6061) | X | X |
| Gold Plated Steel | X | P |
| Mild Steel | X | P |
| Solder 70/30 | X | P |
| Nickel-Silver Alloy | P | N |
| Monel S | N | X |
| Brass | N | N |
| Copper | N | N |
| Nickel-Plated Steel | N | N |
| Tungsten Carbide | N | P |
| Silver | N | P |
| Elastomers: | | |
| Polyurethane | X | X |
| Neoprene | X | X |
| Silicone rubber | X | X |
| Ethylene Propylene Diene Terpolymer (EPDM) | X | X |
| Viton A | X | P |
| Red Natural Rubber | X | X |
| Black Natural Rubber | P | X |
X= Compatible, N= Not Compatible, P= Not Performed
Material compatibility was confirmed with two Fujinon and Olympus flexible colonoscope sets, which included procedural and cleaning/reprocessing accessories. The devices were immersed in Resert XL HLD for 134 hours (1005 cycles) at 25 ±2℃, and showed no significant cosmetic changes, and were not functionally altered at study completion.
#### 9. Simulated Use Testing
Five clinical endoscope sets comprised of flexible endoscopes, procedural endoscope accessories (i.e. valves), and cleaning accessories were challenged with Mycobacterium terrae ATCC 15755 under worst-case temperatures (18-19°C), and organic (5% serum, final v/v), and inorganic (400 ppm AOAC hard water) conditions. A ≥ 6 log10 reduction of test organism was achieved for all of the endoscope sets after 8 minutes of contact with Resert XL HLD test solution, formulated to below the minimum manufacturing specification for hydrogen peroxide and stressed to the end of its reuse life.
{6}------------------------------------------------
K080420
#### 10. In Use testing
Pre-cleaned, representative patient soiled endoscopes were manually reprocessed in Resert XL HLD for 8 minutes at ≥ 20°C, and shown to be effectively high level disinfected under In Use test conditions.
#### 11. Resert XL HLD Stability
Resert XL HLD will have at least a 1 year shelf life or real time test data to support claimed expiry.
#### 12. Conclusion
The above nonclinical and device /material compatibility studies demonstrate that Resert XL HLD High Level Disinfectant is as safe and effective as the predicate.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 2 2008
Mr. John R. Jack Scoville, Jr. Fellow, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060
Re: K080420
Trade/Device Name: Resert™ XL HLD High Level Disinfectant Regulation Number: 880.6885 Regulation Name: Liquid Chemical Sterilants/ High Level Disinfectants Regulatory Class: II Product Code: MED Dated: August 26, 2008 Received: August 27, 2008
Dear Mr. Scoville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Scoville
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
TK. Samuels, m.d. for 11
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
STERIS Response to 053008 Request for Additional Information -- Appendix A K080420 / Resert XL HLD High Level Disinfectant
## Indications for Use
510(k) Number (if known):
KO80420
Device Name:
Resert™ XL HLD High Level Disinfectant
Indications for Use:
Resert™ XL HLD is a ready to use liquid chemical germicide. The product is a 2.0% nominal hydrogen peroxide solution. The minimum recommended concentration (MRC) is 1.5%. Resert XL HLD is intended to be used by health care practitioners in clinical settings as a ready to use liquid chemical disinfectant for the high level disinfection of semi-critical medical devices for which alternative methods of terminal reprocessing are not suitable or available.
High Level Disinfectant: Resert™ XL HLD High Level Disinfectant is a high level disinfectant when used or reused undiluted for a maximum of 21 days at a minimum temperature of 20℃ (68°F) for a minimum immersion time of 8 minutes.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shula H. Murphy, MD
(Division Sign-Off)
Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of 1
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.