ENCORE SHOULDER PRIMARY HUMERAL STEM, MODEL 500-00, ENCORE SHOULDER HUMERAL HEAD, MODEL 520-38
K080402 · Encore Medical L.P. · KWS · Mar 28, 2008 · Orthopedic
Device Facts
Record ID
K080402
Device Name
ENCORE SHOULDER PRIMARY HUMERAL STEM, MODEL 500-00, ENCORE SHOULDER HUMERAL HEAD, MODEL 520-38
Applicant
Encore Medical L.P.
Product Code
KWS · Orthopedic
Decision Date
Mar 28, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Encore Medical shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the indications for use. While shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
Device Story
Encore Shoulder System is a semi-constrained, cemented shoulder prosthesis for primary or revision arthroplasty. System comprises humeral stem (rectangular proximal body, cylindrical distal stem with flutes, anterior/posterior fins for stability), modular humeral neck, humeral head, and glenoid component. Humeral stem features Morse tapers for modular neck attachment and suture holes for soft tissue/bone fragment reattachment. Humeral heads available in standard and offset configurations to recreate patient anatomy. Glenoid components fabricated from ultra-high molecular weight polyethylene; feature keel or four-peg fixation with annular barbs on central peg. Device used by orthopedic surgeons in clinical settings to restore mobility and reduce pain in patients with shoulder joint disability. Mechanical design allows superior/inferior and anterior/posterior translation.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by mechanical bench testing demonstrating performance under expected clinical conditions.
Technological Characteristics
Semi-constrained, cemented shoulder prosthesis. Components: humeral stem, neck, head, and glenoid. Materials: metal and ultra-high molecular weight polyethylene. Fixation: cemented glenoid; humeral stem features Morse tapers and rotational stability fins. Sterilization: not specified.
Indications for Use
Indicated for patients with disability due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis of humeral head/glenoid), rheumatoid arthritis, functional deformity, or humeral fracture. Also indicated for salvage of failed surgical attempts. Glenoid components indicated for cemented use only.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Bio-Modular Choice Shoulder System - Biomet Orthopedics, Inc.
Global Advantage - DePuy
Submission Summary (Full Text)
{0}------------------------------------------------
#### Summary of Safety and Effectiveness MAR 2 8 2008
Date: March 28, 2008
Manufacturer: Encore Medical, L.P. 9800 Metric Blvd Austin, TX 78758
Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (512) 834-6313 Email: teffany hutto@encoremed.com
| Product | Product Code | Regulation and Classification Name |
|------------------------|--------------|------------------------------------------------------------------------------------------|
| Encore Shoulder System | KWS | Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis per 21 CFR 888.3660 |
### Description:
The Encore Shoulder System consists of a humeral stem, neck and head and a glenoid component. Components are offered for use for either primary or revision surgery applications.
The humeral stem consists of proximal body that is rectangular in cross-sectional geometry and tapers proximal to distal. The distal stem is cylindrical with four flutes. Anterior and posterior fins are located on the proximal body to help provide rotational stability. The anterior and posterior fins have suture holes to allow reattachment of soft tissue and bone fragments in the case of proximal humeral fracture. A suture hole is also placed medially through the proximal body just below the collar. A collar is present on the anterior, posterior and medial faces of the proximal body to resist stem subsidence. A neck stem angle is incorporated. The stem has a female Morse type taper to receive modular humeral necks. Also present is a female locating scallop utilized to orient the modular humeral neck and provide additional rotational stability of the neck and head.
The humeral neck has two male Morse type tapers that differ in size to prevent incorrect installation. The smaller Morse type taper interfaces with the female Morse type taper in the humeral stem, the larger Morse type taper interfaces with the female Morse type taper in the humeral head. A scalloped collar between the two male tapers interfaces with the female locating scallop in the humeral stem to provide orientation when using the angled neck and to provide additional rotational stability of the neck and head
The humeral heads are available in standard and offset configurations. In the offset configuration, the male Morse type taper on the humeral heads is offset from the center that makes it possible to orient the head in asymmetric positions on the symmetric stem, thus allowing the surgeon to intraoperatively select the position of the humeral head to recreate the anatomy of the individual patient.
The glenoid components are fabricated from ultra-high molecular weight polyethylene. The articulating surface has a radius of curvature greater than the corresponding humeral head. This
{1}------------------------------------------------
allows translation in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has either a keel or four pegs for fixation in the glenoid. The central peg has three annular barbs to provide immediate fixation to the patient's glenoid when inserted.
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis of the natural humeral head and/or glenoid;
- . rheumatoid arthritis;
- correction of functional deformity; .
- . humeral fracture.
This device may also be indicated in the salvage of previously failed surgical attempts. The glenoid components are indicated for cemented use only.
Intended Use: Encore Medical shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the indications for use. While shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
### Predicate Devices:
Bio-Modular Choice Shoulder System - Biomet Orthopedics, Inc. Global Advantage - DePuy
Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same indications, materials, sterilization, and indications for use.
Non-Clinical Testing: Mechanical testing has demonstrated the device's ability to perform under expected clinical conditions.
Clinical Testing: None provided.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem with three parallel lines forming the body and wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 8 2008
Encore Medical, L.P. % Ms. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Blvd. Austin, TX 78758
Re: K080402
Trade/Device Name: Encore Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: February 13, 2008 Received: February 14, 2008
Dear Ms. Hutto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Ms. Teffany Hutto
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K080402 510(k) Number (if known): __
Device Name: Shoulder System
Indications for Use:
# Encore Shoulder System Indications for Use
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis of the natural humeral head and/or glenoid;
- . rheumatoid arthritis;
- correction of functional deformity; .
- . humeral fracture.
This device may also be indicated in the salvage of previously failed surgical attempts. The glenoid components are indicated for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Oputu
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K080402
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.