K080290 · American Radiologist Network, Inc. · LLZ · Feb 22, 2008 · Radiology
Device Facts
Record ID
K080290
Device Name
XVIEWNET
Applicant
American Radiologist Network, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 22, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The xViewNet software is intended as a display and analysis tool for the interpretation of diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended to provide access to covered entities for clinical review, throughout the healthcare facility and at the point of image acquisition. This device is not intended for mammography use
Device Story
xViewNet is a web-based PACS software for viewing, editing, manipulating, annotating, analyzing, and storing medical images/data. Inputs include medical images from various imaging sources (CR, CT, MR, NM, US, RF, SC, VL) transmitted via TCP/IP. Images are converted to internal formats, cached locally, and processed for display; original pixel data remains unaltered. Features include 3D multiplanar reconstruction, volume rendering, cine-loop, and measurement tools (distance, angle, grey tone). Used in healthcare facilities by clinicians/radiologists for diagnostic interpretation and clinical review. Output is displayed on standard off-the-shelf PC hardware. Enables efficient image access, visualization, and clinical decision-making support.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of risk analysis, requirements reviews, design reviews, module verification, system integration testing, performance testing, and final acceptance validation.
Technological Characteristics
Supports DICOM 3.0; runs on off-the-shelf PC hardware; communicates via TCP/IP. Features: 3D volume rendering, multiplanar reconstruction (coronal, axial, sagittal), cine-loop, window/level adjustment, pixel interpolation, and annotation tools. Software-based image processing.
Indications for Use
Indicated for trained healthcare professionals (radiologists, physicians, technologists, clinicians, nurses) for the display, analysis, and clinical review of diagnostic medical images. Not for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular seal with text around the perimeter. The text reads "American Radiological Moducet AMRADHET". In the center of the seal is a symbol that looks like a stylized medical or scientific emblem. The emblem consists of a central design with smaller, similar shapes arranged around it.
510(k) Summary (per 21 CFR 807.92)
FEB
22
2000
I. Applicant: American Radiologist Network, Inc. 2298 N.W. 60th Street Boca Raton, FL. 33496, U.S.A
> Neil Iyer, President & CEO Contact Person: Tel: 805-906-1360 Fax: 805-520-0568 e-mail: neil@amradnet.com
#### II. Device Name
Proprietary Name: xViewNet Picture Archiving and Communications System Common/ Usual Name: Classification Name: System, Image Processing, Radiological Regulation Number: 892.2050 LLZ Product Code: 2 Classification:
#### III. Predicate Device
The xViewNet PACS is substantially equivalent to the following predicate devices:
- iSite Radiology (K013630) from Stentor Inc. .
- . iPACS Prism (K030751) from Real Time Image Inc.
## IV. Intended Use of the Device
The xViewNet software is intended as a display and analysis tool for the interpretation of diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended to provide access to covered entities for clinical review, throughout the healthcare facility and at the point of image acquisition.
This device is not intended for mammography use
V. Description of the Device
x ViewNet is a secured imaging system that is used to view, edit, manipulate, annotate, analyze, and store images and data that are stored and managed in the web-based RIS. This software-based product provides capabilities for the
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a circular emblem or seal with text around the perimeter. The text reads "AMERICAN RADIOLOGIST NETWORK" at the top and "ZMRADKEY" at the bottom. In the center of the emblem, there is a symbol that appears to be a stylized image of a medical device or tool, repeated in a pattern. The emblem has a vintage or slightly distressed appearance, with some areas of the image appearing faded or worn.
acceptance, transmission, printing, display, storage, editing and digital processing of medical images and associated data.
Images sent to xViewNet are converted into formats suitable for viewing in its framework, and temporarily stored in a local cache memory. The algorithms used by xViewNet to view JPEG and JPEG 2000 images follow known and acceptable protocols. Changes may be made to the presentation of the images. These changes are saved as display definitions only and do not alter the acquired image pixel data. Any and all display definitions applied to an image can always be reversed to the acquired state.
The xViewNet software includes advanced visualization such as 3-D multiplanar reconstruction of the 2-D images.
x ViewNet uses standard "off-the-shelf" PC hardware and communicates using the standard TCP/IP stack. The network hardware used to support the TCP/IP stack is superfluous to XViewNet.
## VI. Technical Characteristics
- Support for the DICOM 3.0 file specification, including all standard . classes and modalities (CR, CT, MR, NM, US, RF, SC, VL, etc.)
- Overview of study or series with different filter possibilities t
- Interactive scrolling through picture series .
- Thumbnail view of different picture series .
- Tools freely configurable to the different mouse buttons .
- Interactive change of W/L or W/C .
- Picture zoom with pixel interpolation ●
- Magnifying glass .
- Panning of pictures .
- Shows modality-dependent DICOM header information in each picture .
- Placing and show values of grey tones (numeric), distances and angles . (graphic and numeric), text annotations, ellipses, polygons, and arrows
- Several picture series can be shown side by side .
- Allocation of picture by free layout choice .
- Turn picture, reflect picture and invert picture .
- Fast Interactive 3D volume rendering .
- Coronal, Axial and Sagittal views .
- x ViewNet is able to read and display the HUGO dataset .
- Fast continuous looping display of images in a series (cine-loop) at . various speeds: forwards, backwards, and bi-directional
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a circular seal with text around the perimeter. The text reads "American Bildtological Institute Zinradnet". In the center of the seal is a symbol that appears to be a stylized representation of a screw or bolt, with four smaller symbols resembling screws or bolts arranged around it.
- Linking and Unlinking of windows to coordinate actions (window/level, . panning, zooming, etc)
## VII. Testing
The xViewNet program complies with the voluntary standards as detailed in Section 9 of this submission.
The following quality assurance measures were applied to the development of the MedViewNet program:
- Risk Analysis .
- Requirements Reviews .
- Design Reviews .
- Testing on unit level (Module Verification) .
- Integration testing (System Verification) .
- Final Acceptance Testing (Validation) .
- Performance Testing .
- Safety Testing .
American Radiologist Network certifies that the x ViewNet software is designed, developed, tested and validated according to written procedures. These procedures identify individuals within the organization responsible for developing and approving product specifications, coding and testing, validation testing and field maintenance.
For details please see Section 16. Software, item 7. Verification & Validation Testing.
# VIII. Safety and Effectiveness
There are no substantial differences between the xViewNet defined in this 510(k) submission and the stated predicate devices. They are similar to the technologies that are currently used in other similar medical devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
FEB 22 2008
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
American Radiologist Network, Incorporated % Mr. James W. Monroe Staff Engineer -- Medical Devices Intertek Testing Services NA, Incorporated 2307 East Aurora Road Unit B7 TWINSBURG OH 44087
Re: K080290
Trade/Device Name: xViewNet PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 1, 2008 Received: February 4, 2008
Dear Mr. Monroe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo with the letters "FDA" in large, bold font in the center. Above the letters, the numbers "1906-2006" are printed. Below the letters, the word "Centennial" is written in cursive. The logo is surrounded by a circular border with text and stars. The text is difficult to read due to the image quality.
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a circular emblem or seal with text around the perimeter. The text reads "PUNJAB RAJYA BIJLI BOARD AMRADHET". In the center of the emblem is a stylized design that includes geometric shapes, possibly representing electrical components or a power grid. The design is symmetrical and appears to be a logo for an organization related to electricity or power in the Punjab region.
510(k) Number (if known):
K080290
Device Name: xViewNet
Indications for Use:
xViewNet is a secured imaging software system that receives medical images and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
This device is not intended for mammography use
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Agni De Whas
(Division Sign-Off Division of Reproductive, Abdomina and Radiological Devic 510(k) Number
Page 1 of
Page 4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.