← Product Code [GEI](/productcode/GEI) · K080282

# ARTHROCARE CONTROLLERS (SYSTEM 2000,1100,1200), MODEL H2076-00, H2079-00, H3000-00, H4000-00; (K080282)

_Arthrocare Corp. · GEI · Feb 15, 2008 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080282

## Device Facts

- **Applicant:** Arthrocare Corp.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Feb 15, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ArthroCare Topaz ArthroWands are indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: Procedures Body Structure as described below • Fasciotomy Foot • Synovectomy Foot • Tendonotomy Knee, Wrist, Elbow, Ankle, Shoulder, Foot • Rotator Cuff Tendonotomy Shoulder • Capsulotomy Foot

## Device Story

Bipolar, single-use, high-frequency electrosurgical device; used in arthroscopic and orthopedic procedures. Operates via high-frequency energy to perform soft tissue resection, ablation, coagulation, and hemostasis. Used by surgeons in clinical settings. Output is controlled thermal effect on tissue; assists in surgical debridement and tissue management. Benefits include precise tissue removal and bleeding control during orthopedic surgery.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Bipolar electrosurgical device; high-frequency energy source; single-use; sterile. Materials modified from predicate; no specific ASTM standards cited in text.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ArthroCare® Topaz® ArthroWands® ([K053567](/device/K053567.md))
- ArthroCare System 12000 ([K071709](/device/K071709.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# KO 80 282

# 510(k) Summary

## ArthroCare Corporation ArthroCare® Topaz® ArthroWands®

### General Information

FEB 15 . .

| Submitter Name/Address:            | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-3523 |
|------------------------------------|---------------------------------------------------------------------------|
| Establishment Registration Number: | 2951580                                                                   |
| Contact Person:                    | Valerie Defiesta-Ng<br>Director, Regulatory Affairs                       |
| Date Prepared:                     | February 1, 2008                                                          |

**Device Description**

| Trade Name:          | ArthroCare <sup>®</sup> Topaz <sup>®</sup> ArthroWands <sup>®</sup>              |
|----------------------|----------------------------------------------------------------------------------|
| Generic/Common Name: | Electrosurgical Device and Accessories                                           |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) |

**Predicate Devices**

| ArthroCare <sup>®</sup> Topaz <sup>®</sup> ArthroWands <sup>®</sup> | K053567 |
|---------------------------------------------------------------------|---------|
| ArthroCare System 12000                                             | K071709 |

### Product Description

The ArthroCare Topaz ArthroWands are bipolar, single use, high frequency electrosurgical devices designed for specific indications in arthroscopic and orthopedic procedures.

{1}------------------------------------------------

# KO 80282

Page 2 of 2

#### Intended Uses

The ArthroCare Topaz ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

| Procedures                 | Body Structure as<br>described below         |
|----------------------------|----------------------------------------------|
| • Fasciotomy               | Foot                                         |
| • Synovectomy              | Foot                                         |
| • Tendonotomy              | Knee, Wrist, Elbow,<br>Ankle, Shoulder, Foot |
| • Rotator Cuff Tendonotomy | Shoulder                                     |
| • Capsulotomy              | Foot                                         |

#### Substantial Equivalence

This Special 510(k) proposes modifications in the performance specifications, materials, and labeling for the ArthroCare Topaz ArthroWands, which were previously cleared in K053567 (March 6, 2006). The indications for use, technology, principle of operation, and sterilization parameters of the ArthroCare Topaz ArthroWands remain the same as in the predicate cleared 510(k)s.

#### Summary of Safety and Effectiveness

The modified ArthroCare Topaz ArthroWands, as described in this Special 510(k), are substantially equivalent to the predicate device. The proposed modification in the material is a not substantial change or modification, and does not significantly affect the safety or efficacy of the devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of two main elements: the HHS symbol and the department's name. The HHS symbol is a stylized representation of a bird or eagle, with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 15 2008

ArthroCare Corporation % Ms. Valerie DeFiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523

Re: K080282

Trade/Device Name: ArthroCare® Topaz® ArthroWands® Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: February 1, 2008 Received: February 4, 2008

Dear Ms. DeFiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 - Ms. Valerie DeFiesta-Ng

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Mulhausen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement

K_080282 510(k) Number:

ArthroCare® Topaz® ArthroWands® Device Name:

Indications for Use:

The ArthroCare Topaz ArthroWands are indicated for debridement, resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

| Procedures                 | Body Structure as described below            |
|----------------------------|----------------------------------------------|
| • Fasciotomy               | Foot                                         |
| • Synovectomy              | Foot                                         |
| • Tendonotomy              | Knee, Wrist, Elbow,<br>Ankle, Shoulder, Foot |
| • Rotator Cuff Tendonotomy | Shoulder                                     |
| • Capsulotomy              | Foot                                         |

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K080282

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

X

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

x

---

**Source:** [https://fda-staging.innolitics.com/device/K080282](https://fda-staging.innolitics.com/device/K080282)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
