← Product Code [GEX](/productcode/GEX) · K080244

# PALOMAR LUX 1540 HANDPIECE (K080244)

_Palomar Medical Products, Inc. · GEX · Jun 20, 2008 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080244

## Device Facts

- **Applicant:** Palomar Medical Products, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jun 20, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Palomar Lux1540 handpiece is intended for use in dermatological procedures requiring the coagulation of soft tissue, skin resurfacing procedures, mclasma, acne scars, and surgical scars.

## Device Story

Lux1540 handpiece is an accessory for the StarLux Pulsed Light and Laser System; system includes cart, console, chiller, and footswitch. Device delivers laser energy to skin for dermatological procedures. Operated by physicians in clinical settings. Output is laser energy used for coagulation of soft tissue and skin resurfacing. Benefits include treatment of scars and melasma.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to predicate devices.

## Technological Characteristics

Laser surgical instrument (Product Code GEX). Handpiece attaches to StarLux Pulsed Light and Laser System console. Energy source: Laser. Intended for dermatological soft tissue coagulation and resurfacing.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Palomar Lux1540 Handpiece ([K060301](/device/K060301.md))
- Palomar Lux1540 Handpiece ([K061652](/device/K061652.md))
- Reliant Technologies, Inc. Fraxel SR1500 Laser System (Fraxel Re:store™) ([K070284](/device/K070284.md))
- Reliant Technologies, Inc. Fraxel SR1500 Laser System (Fraxel Re:store™) ([K062303](/device/K062303.md))
- Reliant Technologies, Inc. Fraxel SR1500 Laser System (Fraxel Re:store™) ([K042319](/device/K042319.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K080244

pg 1 of 2

# 510(k) Summary Lux1540 Handpiece

JUN 2 0 2008

This 510(k) summary is being submitted in accordance with 21 CFR 807.92

### SUBMITTER'S INFORMATION 1.

| NAME:    | Palomar Medical Technologies, Inc.                                                          |
|----------|---------------------------------------------------------------------------------------------|
| ADDRESS: | 82 Cambridge Street<br>Burlington, MA 01803<br>Phone: (781) 993-2300<br>Fax: (781) 993-2330 |
| CONTACT: | Sharon Timberlake, RAC, CCRA<br>Director of Regulatory Affairs                              |

DATE PREPARED: June 18, 2008

## DEVICE INFORMATION 2.

| Trade/Proprietary Name: | Palomar Lux1540 Handpiece                                                                                     |
|-------------------------|---------------------------------------------------------------------------------------------------------------|
| COMMON/USUAL NAME:      | Lux1540                                                                                                       |
| CLASSIFICATION NAME:    | Laser surgical instrument for use in general and<br>plastic surgery and in dermatology<br>(21 CFR § 878.4810) |
| Product Code:           | GEX                                                                                                           |

### 3. PREDICATE DEVICES

Palomar Medical Technologies, Inc. Lux1540 Handpiece K060301 & K061652

Reliant Technologies, Inc. Fraxel SR1500 Laser System (Fraxel Re:store™) K070284, K062303, K042319

{1}------------------------------------------------

K080244 pg 2 of 2

#### 4. INTENDED USE

The Palomar Lux1540 handpiece is intended for use in dermatological procedures requiring the coagulation of soft tissue, skin resurfacing procedures, mclasma, acne scars, and surgical scars.

#### 5. DEVICE DESCRIPTION

The Lux1540 Handpiece attaches to the StarLux Pulsed Light and Laser Systems. The complete system consists of a cart, system console, chiller, a footswitch, and a handpiece.

#### PERFORMANCE DATA 6.

The specifications and indications for use of the Lux1540 handpiece are substantially equivalent to its predicate devices based on the data provided in the 510(k) premarket notification.

#### 7. SUBSTANTIAL EQUIVALENCE

The Lux1540 Handpiece is substantially equivalent to its predicate devices when used according to its intended use. The information that is provided in this 510(k) premarket notification demonstrates that the Lux1540 Handpiece also shares the same technological characteristics, mechanism of action, intended use and physical properties to its predicates.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Palomar Medical Technologies, Inc. % Ms. Sharon Timberlake, RAC, CCRA Director of Regulatory Affairs 82 Cambridge Street Burlington, MA 01803

JUN 2 0 2008

Re: K080244 Trade/Device Name: Palomar Lux1540 Handpiece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: May 29, 2008 Received: May 30, 2008

Dear Ms. Timberlake:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Ms. Sharon Timberlake, RAC, CCRA

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Milliman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

080244

pg 1 of 1

# Indications for Use

510(k) Number (if known): K080244

Device Name: Palomar Lux 1 540 Handpicce

Indications for Use:

The Palomar Lux1540 handpiece is intended for use in dermatological procedures requiring:

Coagulation of soft tissue, skin resurfacing procedures, melasma, acne scars, and surgical scars.

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil RBQ,L for xkm

Division of General, Restorative, and Neurological Devices

**510(k) Number** K080244

---

**Source:** [https://fda-staging.innolitics.com/device/K080244](https://fda-staging.innolitics.com/device/K080244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com