← Product Code [JAK](/productcode/JAK) · K080211

# MODEL TSX-101A/H/I AQUILION 32/64 SP CT SYSTEM (K080211)

_Toshiba America Medical Systems, In.C · JAK · Mar 14, 2008 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080211

## Device Facts

- **Applicant:** Toshiba America Medical Systems, In.C
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Mar 14, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

This device is designed to produce cross-sectional images of a human body by reconstruction of x-ray transmission data from the same axial plane taken at different angles. These images have been proven to be clinically useful in the diagnosis of spine and head injuries, intracranial tumors, blood clots in the brain, eye trauma, soft tissue lesions in the extremities gastro intestinal lesions , abdominal and levic malignancies, and hepatic metastases. CT is also used to evaluate intestinal obstructions, ass intra-abdominal abnormalities and to examine musculoskeletal degeneration. This device employs no intended uses that are not in cleared devices already found in the market place.

## Device Story

Multi-slice CT system comprising gantry, patient couch, and console. Generates up to 128 axial images per second using selectable slice-thickness multi-row detector. Inputs: x-ray transmission data from multiple angles. Transformation: reconstruction of transmission data into cross-sectional images. Output: axial x-ray images. Used in clinical settings by trained operators. Images assist clinicians in diagnosing head/spine injuries, tumors, blood clots, trauma, lesions, malignancies, and musculoskeletal degeneration. Benefits include non-invasive visualization of internal anatomy for improved diagnostic accuracy.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Multi-slice CT system with gantry, couch, and console. Uses ionizing radiation for imaging. Conforms to 21 CFR 1020.30/1020.33 and IEC 60601-1, 60601-2-32, 60601-44 standards. Multi-row detector architecture.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- TSX-101A/I Aquilion 64 SP CT System ([K051833](/device/K051833.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K080211

## Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification

# 510(k) Summary

MAR 1 4 2008

| Date:                                 | January 25, 2008                                                                  |
|---------------------------------------|-----------------------------------------------------------------------------------|
| Submitter's Name:                     | Toshiba America Medical Systems, Inc.                                             |
| Submitter's Address:                  | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068                      |
| Submitter's Contact:                  | Paul Biggins, Regulatory Affairs Specialist,<br>(714)730-5000                     |
| Establishment Registration<br>Number: | 2020563                                                                           |
| Device Proprietary Name:              | TSX-101A/H, Aquilion 32 SP CT System<br>TSX-101A/I, Aquilion 64 SP CT System      |
| Common Name:                          | Scanner, Computed Tomography, X-Ray<br>[Fed. Reg. No. 892.1750, Pro. Code: 90JAK] |
| Regulatory Class:                     | II (per 21 CFR 892.1750)                                                          |
| Performance Standard:                 | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard               |
| Predicate Device(s):                  | TSX-101A/I Aquilion 64 SP CT System                                               |
| Reason For Submission                 | Modification of cleared device                                                    |

### Description of this Device:

The Aquilion 32/64 SP is a multi-slice CT system, consisting of a gantry, patient couch and console. The system generates uop to 128 axial images ower second using a selectable slice-thickness multi-row detector.

### Summary of Intended Uses:

This device is designed to produce cross-sectional images of a human body by reconstruction of x-ray transmission data from the same axial plane taken at different angles. These images have been proven to be clinically useful in the diagnosis of spine and head injuries, intracranial tumors, blood clots in the brain, eye trauma, soft tissue lesions in the extremities gastro intestinal lesions , abdominal and levic malignancies, and hepatic metastases. CT is also used to evaluate intestinal obstructions, ass intra-abdominal abnormalities and to examine musculoskeletal degeneration. This device employs no intended uses that are not in cleared devices already found in the market place.

{1}------------------------------------------------

## Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification

#### Technological Characteristics:

This device employs similar materials and processes as found in the predicate device. The device produces ionizing radiation that is employed to generate cross sectional images of the anatomy.

#### Safety and Effectiveness Concerns:

This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020.30 and 1020.33, that apply to this upgrade, will be met and reported via a supplement to the initial report for the predicate device. Additionally this system is in conformance with the applicable parts of the IEC 60601-1 {applicable portions}; IEC 60601-2-32, and IEC 60601-2-44. -Medical Device Safety standards.

#### Substantial Equivalence:

This device is similar in materials and processes to that of the predicate device, Toshiba TSX-101A/H, /I Aquilion 32/64 SP CT scanner [k051833].

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized symbol that appears to be a representation of human services.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313

MAR 1 4

Re: K080211

Trade/Device Name: Model TSX-101A/H, /I AQUILION 32/64 SP CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 27, 2008 Received: February 28, 2008

Dear Mr. Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular logo or emblem. The text "1906 - 2006" is at the top of the circle. The letters "FDA" are in the center of the circle, with the word "Centennial" underneath. The logo appears to be a commemorative emblem for the FDA's centennial anniversary.

Protecting and Promoting Public Health.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): 10802

Device Name: TSX-101A/H, // Aquilion 32/64 SP CT System

Indications For Use:

Acquisition and display of axial x-ray images of the whole body to include the head.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of

Sign Off

(Division Sign-OII)
Division of Reproductive, Abdominal and Radiological 510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K080211](https://fda-staging.innolitics.com/device/K080211)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
