FISHER AND PAYKEL HEALTHCARE ZEST NASAL MASK

K080209 · Fisher & Paykel Healthcare, Ltd. · BZD · Sep 12, 2008 · Anesthesiology

Device Facts

Record IDK080209
Device NameFISHER AND PAYKEL HEALTHCARE ZEST NASAL MASK
ApplicantFisher & Paykel Healthcare, Ltd.
Product CodeBZD · Anesthesiology
Decision DateSep 12, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Zest™ Nasal Mask is intended for single patient adult use by individuals who have been diagnosed by a physician as requiring CPAP or Bi-level ventilator treatment in the home, hospital or other clinical setting.

Device Story

Zest™ Nasal Mask is a non-invasive respiratory interface; delivers pressurized gases for CPAP or Bi-level ventilation. Device consists of a silicone seal positioned over the nose and upper lip; secured via headgear straps; connects to breathing tube via swivel adaptor. Integrated exhalation vents (bias holes) flush exhaled gases to room air. Used in home, hospital, or clinical settings; operated by patient or clinician. Output is the delivery of therapeutic pressure to the airway; facilitates ventilation therapy; improves patient comfort via contoured seal design.

Clinical Evidence

Bench testing only. Performance compared to predicate Opus HC482; biocompatibility compared to predicate HC432. No clinical data presented.

Technological Characteristics

Non-invasive nasal mask; silicone seal; headgear straps; swivel adaptor; integrated exhalation vents (bias holes). Single-size design. Non-continuous ventilation (IPPB).

Indications for Use

Indicated for single patient adult use requiring CPAP or Bi-level ventilator treatment in home, hospital, or clinical settings.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 080209 HEALTHCARE Paykel ## SECTION 5 - 510K Summary SEP 1 2 2008 15 Maurice Paykel Place, East Tamaki P O Box 14 348, Panmure Auckland, New Zealand Telephone: +64 9 574 0100 Facsimile: +64 9 574 0196 Website: www.fphcare.com | Contact person | Tina Mason | |------------------------|---------------------------------------------------------------------------------------------------------------| | Date Prepared | 25th January 2008 | | Trade Name | Zest™ Nasal Mask | | Common Name | Nasal Mask | | Classification<br>Name | Non continuous ventilator IPPB<br>(21 CFR § 868.5905, product code BZD) | | Predicate<br>Devices | K063036 Fisher & Paykel Healthcare Opus HC482 Direct Nasal<br>Mask | | | K061236 Fisher & Paykel Healthcare Flexifit HC432 Full Face<br>Mask (predicate for material biocompatibility) | {1}------------------------------------------------ ## 5.1 Description The Zest™ Nasal Mask is a respiratory mask which is non invasive, the Silicone Seal is positioned over the nose, and it seals on the area around the nose, and the upper lip. The mask is held on the face with headgear straps. It connects to a single breathing tube via a swivel adaptor, to receive pressurized gases. On the mask base are exhalation vents (bias holes) that allow exhaled gases to be continually flushed and removed to room air. The silicone seal is contoured for comfort and to reduce leakage. ## 5.2 Intended Use The Zest™ Nasal Mask is intended for single patient adult use by individuals who have been diagnosed by a physician as requiring CPAP or Bi-level ventilator treatment in the home, hospital or other clinical setting. ## 5.3 Technological Characteristics Comparison The Zest™ Nasal Mask is very similar to the predicate Fisher & Paykel Healthcare Opus HC482 Direct Nasal Mask. Both masks are for non continuous ventilation, and deliver CPAP gasses through the nose. The main difference is the way the seal is achieved, the Zest Nasal Mask seals around the nose and on the upper lip, where the Opus HC482 Direct Nasal Mask seals on the entrance of the nostrils. Other differences are that the Zest Nasal Mask has 4 headgear attachment points (as opposed to 2 in the Opus). The Zest Nasal Mask has only one size, where the Opus HC482 has 3 sizes (the Zest Nasal Mask is designed to fit a large proportion on the patient population with only one size) and the Zest Nasal Mask has a larger dead space due the wav it achieves a seal on the face. #### 5.4 Non-clinical Tests Testing of the Zest™ Nasal Mask was compared to the predicate Fisher & Paykel Healthcare Opus HC482 Direct Nasal Mask for performance and the HC432 Full Face Mask for biocompatibility. These tests demonstrate substantial equivalence of the Zest™ Nasal Mask to the predicate mask. Copies of test reports are included in Appendix B. ## 5.5 Conclusion The comparison of features, performance, and intended use demonstrate that the Zest™ Nasal Mask is substantially equivalent to the predicate Fisher & Paykel Healthcare Opus HC482 Direct Nasal Mask for performance and the HC432 Full Face Mask for biocompatibility. The Zest™ Nasal Mask is proven to be safe and effective for CPAP and Bi-level ventilation therapy. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle's head is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # SEP 1 2 2008 Ms. Tina Mason Regulatory Affairs Engineer Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki P.O. Box 14 348, Panmure Auckland NEW ZEALAND Re: K080209 Trade/Device Name: Zest™ Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: July 28, 2008 Received: July 30, 2008 Dear Ms. Mason: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general. controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Mason Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Tikamuli-Lindu-y forey Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {4}------------------------------------------------ ## SECTION 4 - Indications for Use Statement 510(k) Number Zest™ Nasal Mask Device Name Indications for Use: The Zest™ Nasal Mask is intended for single patient adult use by individuals who have been diagnosed by a physician as requiring CPAP or Bi-level ventilator treatment in the home, hospital or other clinical setting. Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K06205
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...