← Product Code [MSX](/productcode/MSX) · K080192

# HYPERVISOR VI CENTRAL MONITORING SYSTEM (INCLUDING TMS) (K080192)

_Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MSX · Apr 3, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080192

## Device Facts

- **Applicant:** Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- **Product Code:** [MSX](/productcode/MSX.md)
- **Decision Date:** Apr 3, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions. Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.

## Device Story

The HYPERVISOR VI Central Monitoring System (CMS) is a PC-based medical information system that aggregates, stores, prints, and reviews physiological data from networked medical devices. The system includes a Telemetry Monitoring System (TMS) sub-system that collects ECG and SpO2 signals via sensors, transmitting them wirelessly over the WMTS band to receivers, which then forward the data to the CMS. The system is operated by medical personnel in hospitals or medical institutions to facilitate remote monitor management and centralized clinical oversight. Key features include ViewOtherBed functionality and real-time printing. By centralizing patient data, the device assists clinicians in monitoring patient status, potentially improving clinical decision-making and patient care efficiency.

## Clinical Evidence

Bench testing only. Laboratory testing was conducted to validate and verify that the CMS and TMS meet all design specifications and comply with applicable standards. No clinical data was provided.

## Technological Characteristics

PC-based platform; network connectivity via hardwired cables and WLAN; wireless transmission via WMTS band; physiological sensors for ECG and SpO2; software-based data processing, storage, and display.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- HYPERVISOR VI Central Monitoring System ([K062194](/device/K062194.md))
- Spacelabs Medical Ultraview Digital Telemetry System ([K983996](/device/K983996.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

APR - 3 2008

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

08019-The assigned 510(k) number is:

#### Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 2658 2888 Fax: +86 755 2658 2680

- Contact Person:
	- Li Dongling

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Date Prepared: .

January 4, 2008

### Name of the Device:

- Trade/Proprietary Name: HYPERVISOR VI Central Monitoring System . (including TMS)
- Common Name: Central Monitoring System (CMS) ●
- Classification: ●

MSX, 21 CFR Part 870.2300 MHX, 21 CFR Part 870.1025

DQA, 21 CFR Part 870.2700

DPZ, 21 CFR Part 870.2710

DRG, 21 CFR Part 870.2910

System, Network and Communication, Physiological Monitors Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Oximeter Oximeter, Ear Transmitters and Receivers, Physiological Signal, Radiofrequency

ge 1 OF 3

{1}------------------------------------------------

#### Legally Marketed Predicate Device:

HYPERVISOR VI Central Monitoring System, Shenzhen Mindray Bio-Medical K062194 Electronics Co., LTD.

Spacelabs Medical Ultraview Digital Telemetry System, Spacelabs Medical, Inc. K983996

#### Description:

Central Monitoring System network is a kind of medical information system, which consists of different networked devices (which have separate 510(k) clearance). Hypervisor VI Central Monitoring System (CMS) is the primary maintainer of communication between other networked devices. It can store, print, review or process information from networked devices. It can also realize remote monitor management function to free doctors from clinical monitoring work and conduct centralized monitoring management.

In this submission, CMS adds the function of supporting communication with Telemetry Monitoring System. Besides this, CMS is improved to include some new functions, such as, add the function of ViewOtherBed and add the function of real-time print.

By using radiofrequency signal, TMS is intended to monitor Electrocardiogram (ECG), Heart Rate (HR), Arrhythmia Detection, ST Segment Analysis, Saturation of Pulse Oxygen (SpO2) and Pulse Rate (PR) for adult and pediatric patients. Physical signals are collected by sensors and wirelessly transmitted by transmitters in WMTS band, receivers get the signals and forward to CMS for processing, displaying, storing, printing, etc. It can be used within a defined coverage area in hospitals or medical institutions.

CMS is based on the PC platform, and the PC can be configured by users according to the requirements on the operator's manual.

#### Statement of Intended Use:

The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions.

Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.

#### Comparison of Technological Characteristics:

The CMS is substantially equivalent to the previously cleared CMS (K062194) in technological characteristics. Both of them apply the same design principles, and the

Page 2 of 3

{2}------------------------------------------------

modifications are concentrated on software consummating and network testing. The TMS is substantially equivalent to Ultraview Digital Telemetry System (K983996). Both of them monitor similar parameters of patients, and apply similar design principles.

## Testing:

Laboratory testing has been conducted to validate and verify that the CMS (including TMS) meet all design specifications. Results of these tests demonstrate compliance to the requirements of all applied standards.

: 上

### Conclusion:

The Central Monitoring System is substantially equivalent to the predicate devices.

Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/12 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol of medicine and health. The caduceus has three snakes winding around a staff.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 3 2008

Shezhen Mindray Bio-Medical Electronics Co., Ltd c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd, Suite 200 Great Neck, NY 11021

Re: K080192

HYPERVISOR VI Central Monitoring System (including Telemetry Monitoring System, Model TMS-6016) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MSX Dated: January 4, 2008 Received: January 25, 2008

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Susan D. Goldstein-Falk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

B.J. Zimmerman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

KO80x92

Device Name: HYPERVISOR VI Central Monitoring System (including TMS)

#### Indications for Use:

The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions.

Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B. Vettumtor

Page

stan Sign-Of scular Devices

---

**Source:** [https://fda-staging.innolitics.com/device/K080192](https://fda-staging.innolitics.com/device/K080192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com