ZENO®PMMA Discs are milling blanks consisting of polymethylmethacrylate. They are designed for making long-term temporary crown and bridge work and have to be machined with the CAD/CAM technique. ZENO PMMA Discs are recommended for manufacturing substructures of single tooth and bridgework with one or two pontics.
Device Story
ZENO PMMA Discs are hot-cured polymethylmethacrylate (PMMA) milling blanks used by dental technicians within the ZENO TEC CAD/CAM system. Process involves scanning a dental model; designing the restoration virtually via CAD software; milling the design from the PMMA disc using CAM technology; and final manual characterization with dental veneering composites and polishing. Used in dental laboratories to create long-term temporary crowns and bridges. The restoration serves a prophylactic function by maintaining abutment tooth position and protecting against bacterial, toxic, and thermal influences; it also allows clinicians to verify occlusion, phonetics, and aesthetics before final restoration. Benefits include improved patient comfort and functional assessment during the interim period before permanent placement.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Hot-cured polymethylmethacrylate (PMMA) milling blanks. Available in two shades (A3, B1) and varying thicknesses. Designed for use with ZENO TEC CAD/CAM milling systems. Material properties include high polymerization grade.
Indications for Use
Indicated for professional dental technicians to fabricate long-term temporary crowns and bridges (single tooth or bridgework with one or two pontics) for patients requiring temporary restorations, including implant-supported cases, to maintain abutment position, protect teeth from bacterial/thermal effects, and evaluate occlusion, phonetics, and aesthetics.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
Predicate Devices
VITA CAD-TEMP FOR INLAB, MODEL, VX59-4002, VX59-4004 (K070991)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "WIELAND" in large, bold, black letters. The letters are sans-serif and appear to be slightly distressed or textured, giving them a somewhat rough appearance. The word is horizontally oriented and fills most of the frame.
K080182
GmbH & Co. KG Schwenninger Straße 13 75179 Pforzheim, Germany Postfach 10 20 40 75120 Pforzheim, Germany Fon +49 72 31/ 37 05 0 lax +49 /2 31/ 35 79 59
WIELAND Dental + Technik
#### 510 (k) Summary 5.
APR - 8 2008
Submitter of 510(k): Wieland Dental + Technik GmbH & Co. KG Schwenninger Str. 13 D-75179 Pforzheim Germany Phone: +49-7231-3705-0
| Contact person: | Dr. Gerhard Polzer |
|-----------------|----------------------------------|
| Phone: | +49-7231-3705-219 |
| Fax: | +49-7231-357959 |
| e-mail: | gerhard.polzer@wieland-dental.de |
Date of Summary: 2008-01-07
Trade name:
ZENO PMMA DISC
Classification name: Temporary crown and bridge resin Product code: EBG C.D.R section: 872.3770 Classification: Class II
| Legally marketed<br>equivalent device: | VITA CAD-TEMP FOR INLAB, MODEL, VX59-4002, VX59-4004 |
|----------------------------------------|------------------------------------------------------|
| 510(k) number: | K 070991 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "WIELAND" in large, bold, black letters against a white background. The font appears to be sans-serif, and the letters are slightly distressed, giving them a textured look. The word is horizontally oriented and fills most of the frame.
WIFLAND Dental + Technik GmbH & Co KG Schwenninger Straße 13 75179 Pforzheim. Germany Postfach 10 20 40 75120 Pforzheim, Germany Fan +49 72 31/ 37 05 - 0 Fax +49 72 31/35 79 59
510 (k) Summarv
## Device description
ZENO PMMA Discs (A3/B1) are milling blanks composed of hot cured polymethylmethacrylate (PMMA). They are intended to be used by professional dental technicians for making long-term temporary crowns and bridgeworks for the sole use of particular patients.
ZENO PMMA Discs can be machined in all machines of the ZENO Tec system. The manufacturing process of the long-term temporary crowns and bridges consists of different steps. At first the model has to be scanned. In the next step, the restoration has to be designed virtually with the help of the CAD technology. Thereafter, the realization of this design has to be carried out by the CAM technology. In a final step the restoration can be individually characterised with dental veneering composites for improving the aesthetics and it can be pre- and high shine polished.
The ZENO PMMA Discs are offered in two different shades (A3, B1) and thicknesses.
# Recommended application
With the introduction of the ZENO PMMA Discs, Wieland Dental+Technik offers to the customer for the ZENO TEC System the possibility of producing long-term temporary crowns and bridgeworks with one or two pontics. These could be used, for example, with patients who should receive an implant -supported restoration, in which the long-term temporary restoration has a prophylactic function to prevent any change of the position of the abutment-teeth. In addition, the teeth are protected by the restoration against the influence of bacterial, toxic as well as thermal effects. Furthermore, it becomes possible to check the occlusion, the phonetics as well as the aesthetics.
### Comparison with the predicate device
ZENO PMMA Discs are substantially equivalent to the dental device VITA CAD-TEMP. Both devices are acrylate polymers and have identical indication for use, comparable physical, biological, and chemical properties and the same prophylactic, diagnostic and aesthetic function.
Due to the high polymerization grade and excellent material properties of the ZENO PMMA Discs, the restorations possess a high-level safety and effectiveness. ZENO PMMA Discs therefore are as safe, as effective, and performs as well as or better than the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 8 2008
Gerhard Polzer, Ph.D. Director Regulatory Affairs Wieland Dental + Technik GmbH & Company KG Schwenninger Straße 13 D-75179 Pforzheim, GERMANY
Re: K080182
Trade/Device Name: ZENO PMMA Disc Regulation Number: 872.3770 Regulation Name: Temporary Crown and Bridge Resin Regulatory Class: II Product Code: EBG Dated: January 18, 2008 Received: January 25, 2008
Dear Dr. Polzer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Polzcr
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Syltte y. Michau Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "WIELAND" in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance, as if they have been stamped or printed with a slightly uneven distribution of ink. The word is presented in all capital letters and is horizontally oriented.
WIELAND Dental + Technik GmbH & Co. KG Schwenninger Straße 13 75179 Pforzheim, Germany Postfach 10 20 40 75120 Pforzheim, Germany lion +49 7231/37 05-0 Fax +49 72 31/ 35 79 59
#### Indication for use statement 11.8.
***Indications for Use***
| 510(k) Number (if known): | k060182 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: ZENO PMMA Disc
Indications for Use:
ZENO®PMMA Discs are milling blanks consisting of polymethylmethacrylate. They are designed for making long-term temporary crown and bridge work and have to be machined with the CAD/CAM technique. ZENO PMMA Discs are recommended for manufacturing substructures of single tooth and bridgework with one or two pontics.
Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Swati Kumar
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K080182 and here is the code to generate that json:```json {
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.