# PASPORT OPTICAL TROCAR, VORTEX SEAL, BLUNT TIP TROCAR, SHIELDED TROCAR, CANNULA ANCHOR (K080161)

_Patton Surgical Corp. · GCJ · May 29, 2008 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K080161

## Device Facts

- **Applicant:** Patton Surgical Corp.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** May 29, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The PassPort® Optical Trocar is an access device that may be used with or without visualization that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures. The PassPort® Vortex Seal functions in combination with Patton Surgical's PassPort® Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures. The PassPort® Blunt Tip Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures. The PassPort® Shielded Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures. The PassPort® Cannula Anchor functions in combination with Patton Surgical's PassPort® Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

## Device Story

PassPort® system consists of single-patient-use surgical access devices (trocars, seals, and anchors) for laparoscopic procedures. Devices create and maintain instrument passageways in abdominal or thoracic cavities. Optical, blunt, and shielded trocar variants provide different entry methods; seals and anchors support trocar function. Used in clinical settings (OR) by surgeons. Devices facilitate instrument insertion, enabling minimally invasive surgery. Benefits include access to internal anatomy for general, gynecologic, thoracic, and urological procedures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Single-patient-use surgical access devices. Includes optical, blunt, and shielded trocar configurations, vortex seals, and cannula anchors. Mechanical design for tissue penetration and instrument port maintenance. Sterilization method not specified.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Endopath® Bladeless Trocar, Endopath® Blunt Tip Trocar, Endopath® Dilating Tip Trocar ([K032676](/device/K032676.md))
- Funnel Trocar ([K992324](/device/K992324.md))
- Shielded Surgical Trocar ([K971475](/device/K971475.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Section 5 510(k) Summary

Image /page/0/Picture/1 description: The image shows the date "MAY 29 2008" in bold, black font. The letters and numbers are closely spaced together. The date is likely extracted from a document or photograph.

This 510(K) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21CFR § 807.92

The assigned 510(K) number is :

K 080161

Submitted by:

Michael Patton 6300 Bridgepoint Parkway Bldg. 2, Ste. 420 Austin, Texas, 78730 Phone: 512 329 0469 Fax: 512 328 9113 Email: mpatton@pattonsurgical.com

Date Prepared:

31 October 2007

Establishment Registration Number: Patton Surgical is located at 6300 Bridgepoint Parkway, Bldg. 2, Suite 420 in Austin, Texas 78730. We are registered with the Food and Drug Administration as Establishment Number 1651380.

Classification Name:

Common/Usual Name:

Proprietary Name:

Disposable Surgical Trocar/Cannula

Laparoscope, General & Plastic Surgery

PassPort® Optical Trocar, PassPort® Vortex Seal, PassPort® Blunt Tip Trocar, PassPort® Shielded Trocar, PassPort® Cannula Anchor

Indication for Use:

The PassPort® Optical Trocar is an access device that may be used with or without visualization that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Vortex Seal functions in combination with Patton Surgical's PassPort® Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Blunt Tip Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

{1}------------------------------------------------

The PassPort® Shielded Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Cannula Anchor functions in combination with Patton Surgical's PassPort® Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

Device Description:

The Candidate Devices are a single patient device to be utilized as an access port.

Substantial Equivalence Claim:

The principles of operation and technology in the candidate device are similar to other devices, which the FDA has found to be substantially equivalent as outlined below:

Product:

Manufacturer: 510(K) Number:

Endopath® Bladeless Trocar, Endopath® Blunt Tip Trocar, Endopath® Dilating Tip Trocar Ethicon Endo-Surgery, Inc. K032676 Substantial Equivalence Date: 10/30/03

Product: Manufacturer: 510(K) Number: Substantial Equivalence Date: 08/19/99

Patton Surgical Corp. K992324

Funnel Trocar

Product: Manufacturer: 510(K) Number: Substantial Equivalence Date: 08/21/97

Shielded Surgical Trocar Ethicon Endo-Surgery, Inc. K971475

-End of summary-

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring an abstract design of an eagle or bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 9 2008

Patton Surgical Corporation % Mr. Michael Patton President 6300 Bridgepoint Parkway Building 2, Suite 420 Austin, Texas 78730

Re: K080161

Trade/Device Name: PassPort® Optical Trocar, PassPort® Vortex Seal, PassPort® Blunt Tip Trocar, Passport® Shielded Trocar, PassPort® Cannula Anchor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: May 22, 2008 Received: May 22, 2008

Dear Mr. Patton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Michael Patton

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N Wilkerson

Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Section 4 Indications for Use Statement K 080161

510(K) Number:

Device Name:

PassPort® Optical Trocar, PassPort® Vortex Seal, PassPort® Blunt Tip Trocar, PassPort® Shielded Trocar, PassPort® Cannula Anchor

Indications for Use:

The PassPort® Optical Trocar is an access device that may be used with or without visualization that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Vortex Seal functions in combination with Patton Surgical's PassPort® Shielded Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Blunt Tip Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Shielded Trocar is an access device that creates and maintains a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

The PassPort® Cannula Anchor functions in combination with Patton Surgical's PassPort® Shielded Trocar to create and maintain a passageway for laparoscopic instruments during a variety of laparoscopic procedures such as general, gynecologic, thoracic, and urological procedures.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) Over-the-Counter (Optional format 1-2-96)

Neil K. Sieler for mxa

sion Sign-Cl Division of General, Restorative, and Neurological Devices

**510(k) Number** K080161

---

**Source:** [https://fda-staging.innolitics.com/device/K080161](https://fda-staging.innolitics.com/device/K080161)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).