APLIO ARTIDA, MODEL SSH-880CV DIAGNOSTIC ULTRASOUND SYSTEM
K080160 · Toshiba America Medical Systems, In.C · IYN · Feb 8, 2008 · Radiology
Device Facts
Record ID
K080160
Device Name
APLIO ARTIDA, MODEL SSH-880CV DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Toshiba America Medical Systems, In.C
Product Code
IYN · Radiology
Decision Date
Feb 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
This device is intended to be used for the following type of studies; cardiac, transesophageal and peripheral vascular.
Device Story
Mobile diagnostic ultrasound system (Track 3); utilizes flat linear and sector array probes (2.5–7.5 MHz). Operates by emitting ultrasonic pulses and processing echo/Doppler signals to generate real-time images. Used by clinicians in clinical settings for cardiac, transesophageal, and peripheral vascular examinations. Output displayed on system monitor for real-time visualization, measurement, and clinical assessment of anatomy and blood flow. Assists physicians in diagnosing cardiovascular conditions and guiding clinical decision-making.
Clinical Evidence
Bench testing only. Device performance verified through compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and AIUM-NEMA UD2/UD3 standards for acoustic output and safety. No clinical data presented.
Technological Characteristics
Mobile ultrasound system; frequency range 2.5–7.5 MHz. Probes: PST-25SX, PST-30SBT, PLT-704SBT, PET-511BTM, PC-20M. Modes: B, THI, M, Color Doppler, Power, Dynamic Flow, TDI, PW, CW, CHI 2D, CHI Dynamic Flow, 4D. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, AIUM-NEMA UD2, and AIUM-NEMA UD3.
Indications for Use
Indicated for cardiac, transesophageal, and peripheral vascular diagnostic ultrasound imaging in adult and pediatric populations.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Toshiba Ultrasound Diagnostic System APLIO ARTIDA MODEL SSH-880A (K072826)
Toshiba Ultrasound Diagnostic System APLIO XG MODEL SSA-790A (K072000)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten text string that appears to be a combination of letters and numbers. The text reads 'KD80160'. The characters are written in a bold, somewhat irregular style, suggesting they were written quickly or with a thick marker.
ﺕ : - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Toshiba America Medical Systems, Inc
# 510(k) Summary of Safety and Effectiveness: 21 CFR 807.92
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
|-------------------|-------------------------------------------------------|
| Address: | PO Box 2068,2441 Michelle Drive Tustin, CA 92781-2068 |
| Contact: | Paul Biggins, Director of Regulatory Affairs |
| Telephone No.: | (714) 730-5000 |
| Device Proprietary Name: | APLIO ARTIDA MODEL SSH-880CV |
|--------------------------|------------------------------|
| Common Name: | Diagnostic Ultrasound System |
Classification:
Regulatory Class: II Review Category: Tier II
Ultrasonic Pulsed Doppler Imaging System - Product Code: 90-IYN [Fed.Reg.No .: 892.1550] Ultrasonic Pulsed Echo Imaging System -- Product Code: 90-IYO [Fed.Reg.No .: 892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [Fed. Reg. No.: 892.1570]
## Identification of Predicate Devices:
Toshiba America Medical Systems believes that this device is substantially equivalent to:
- 1) Toshiba Ultrasound Diagnostic System APLIO ARTIDA MODEL SSH-880A ; 510(k) control number k072826.
- 2) Toshiba Ultrasound Diagnostic System APLIO XG MODEL SSA-790A ; 510(k) control number k072000.
## Device Description:
This device is a mobile system is a Track 3 device that employs a wide range of probes that include flat linear array and sector array with a frequency range of approximately 2.5 MHz to 7.5 MHz.
## Intended Use:
This device is intended to be used for the following type of studies; cardiac, transesophageal and peripheral vascular.
### Safety Considerations:
This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601-1 (applicable portions), IEC 60601-1-2 (applicable portion), IEC60601-2-37 (applicable portions), and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 8 2008
Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K080160
Trade/Device Name: APLIO ARTIDA Model: SSH-880CV Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 22, 2008 Received: January 23, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the APLIO ARTIDA Model: SSH-880CV Diagnostic Ultrasound System, as described in your premarket notification:
### Transducer Model Number
| PST-25SX |
|------------|
| PST-30SBT |
| PLT-704SBT |
| PET-511BTM |
| PC-20M |
{2}------------------------------------------------
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{3}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.
Sincerely yours,
Vou M. Whay
Nancy C. Brogdon दि Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
:
{4}------------------------------------------------
System _X Transducer Model SSH-880CV 510(k) Number(s) 000000
| Clinical<br>Application | B | THI | M | Color<br>Doppler | Power | Dynamic<br>Flow | TDI | PW | CW | CHI<br>2D | CHI<br>Dynamic<br>Flow | Combined<br>(Specify)* | 4D<br>(Realti<br>me 3D) |
|----------------------------------|---|-----|---|------------------|-------|-----------------|-----|----|----|-----------|------------------------|------------------------|-------------------------|
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | N | | | | |
| Small Organ<br>(Specify) | N | N | N | N | N | N | | N | | | | N | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | N | N | N | N | | | N | N | N | N | N | N | N |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | N | N | N | N | N | N | | N | N | | | N | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Additional Comments: Combined Modes: B/M; B/PWD; BDF/PWD; BDF/MDF;
BDF/MDF/PWD; 2D/CWD; BDF/CWD
Small organ = heart Cardiac = adult and pediatric
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (OD
V Prescription Use (Per 21 CFR 801.109)11
lori McWhay
(Divišion Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{5}------------------------------------------------
Toshiba America Medical Systems, Inc
## Diagnostic Ultrasound Indications For Use Form
| System | Transducer X | | | | | | | |
|------------------|--------------|--|--|--|--|--|--|--|
| Model | PST-25SX | | | | | | | |
| 510(k) Number(s) | K080160 | | | | | | | |
| | Mode of Operation | | | | | | | | | | | | |
|----------------------------------|-------------------|-----|---|-----------------------|-------|-----------------|-----|----|----|-----------|------------------------|-----------------------|---------------------|
| Clinical Application | B | THI | M | Color<br>Dopp-<br>ler | Power | Dynamic<br>Flow | TDI | PW | CW | CHI<br>2D | CHI<br>Dynamic<br>Flow | Combined<br>(Specify) | 4D<br>(Realtime 3D) |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify)* | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | P | P | | | | | | | | | | | P |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Additional Comments: Combined Modes: None
Predicate 510(k) K072826
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (OD)
Prescription Use (Per 21 CFR 801.109)
Larry M. Why
(Division Sign Off)
Division of Reproductive, Abdominal and Radiologica 510(k) Number
A-6
{6}------------------------------------------------
Transducer X System PST-30SBT Model 510(k) Number(s) X0100
・)
| | Mode of Operation | | | | | | | | | | | | |
|----------------------------------|-------------------|-----|---|-----------------------|-------|-----------------|-----|----|----|-----------|------------------------|-----------------------|-------------------------|
| Clinical Application | B | THI | M | Color<br>Dopp-<br>ler | Power | Dynamic<br>Flow | TDI | PW | CW | CHI<br>2D | CHI<br>Dynamic<br>Flow | Combined<br>(Specify) | 4D<br>(Realti<br>me 3D) |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify)* | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | P | P | P | P | | | P | P | P | P | P | P | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Additional Comments: Combined Modes: B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/C
Previous 510k of this transducer : k072826
(please do not write below this line - continue on other pages fr ni Concurrence of CDRH, Office of Device Evaluation (9
/ Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
Division of Reproductive, Abdomi
Radiological Devices
510(k) Number K08
{7}------------------------------------------------
System Transducer X Model PLT-704SBT 510(k) Number(s)
KOSOKOO
| | Mode of Operation | | | | | | | | | | | | |
|-------------------------------|-------------------|-----|---|---------------|-------|--------------|-----|----|----|--------|------------------|---------------------|------------------|
| Clinical Application | B | THI | M | Color Doppler | Power | Dynamic Flow | TDI | PW | CW | CHI 2D | CHI Dynamic Flow | Combined (Specify)* | 4D (Realtime 3D) |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify) | P | P | P | P | P | P | | P | | | | P | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | | P | | | | P | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | | |
N= new indication: P = Previously Cleared by FDA: E= Added under Appendix E (LTF)
N=new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Additional Comments: Combined Modes: B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD;
Previous 510(k) of this transducer: K072000
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)11
Lori M. Whay
Division Sign-Off
Division of Reproductive. Abdominal a 510(k) Numbe
{8}------------------------------------------------
Toshiba America Medical Systems, Inc
510(k) Premarket Notification Aplio Artida SSH-880CV Ultrasound System
### Diagnostic Ultrasound Indications For Use Form
System Transducer X PET-511BTM Model 510(k) Number(s) 080160
| | Mode of Operation | | | | | | | | | | | | |
|----------------------------------|-------------------|-----|---|-----------------------|-------|-----------------|-----|----|----|-----------|------------------------|-----------------------|-------------------------|
| Clinical Application | B | THI | M | Color<br>Dopp-<br>ler | Power | Dynamic<br>Flow | TDI | PW | CW | CHI<br>2D | CHI<br>Dynamic<br>Flow | Combined<br>(Specify) | 4D<br>(Realti<br>me 3D) |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify)* | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | | |
| Transesophageal | P | P | P | P | | | | P | P | P | | P | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Combined Modes: B/M; B/PWD; Additional Comments: BDF/PWD; BDF/MDF; BDF/MDF/PWD;B-TDI; M-TDI; 2D/CWD; BDF/CWD;
Previous 510k of this transducer : k072000
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE
Evaluation (GDE)
(Division Sign-Off)
Division Sign-Off) (Division of Reproductive, Abdominal a Radiological Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
A-9
{9}------------------------------------------------
Transducer X System PC-20M Model 510(k) Number(s)
| | Mode of Operation | | | | | | | | | | | | |
|--------------------------|-------------------|-----|---|-----------------------|-------|-----------------|-----|----|----|-----------|------------------------|-----------------------|-------------------------|
| Clinical Application | B | THI | M | Color<br>Dopp-<br>ler | Power | Dynamic<br>Flow | TDI | PW | CW | CHI<br>2D | ેના<br>Dynamic<br>Flow | Combined<br>(Specify) | 4D<br>(Realti<br>me 3D) |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (Specify) | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | P | | | | |
| Small Organ (Specify)* | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac | | | | | | | | | P | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethrai | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | P | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | | |
| Superficial | | | | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | | | | |
| Conventional | | | | | | | | | | | | | |
N= new indication; P = Previously Cleared by FDA; E = Added under Appendix E (LTF)
Additional Comments:
Previous 510k of this transducer : k072000
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON OTHER PAGES IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
v Prescription Use (Per 21 CFR 801.109)
Lori M. Whay
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
A-10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.