K080157 · Ge Medical Systems Information Technologies · MHX · Mar 24, 2008 · Cardiovascular
Device Facts
Record ID
K080157
Device Name
DASH 2500 PATIENT MONITOR
Applicant
Ge Medical Systems Information Technologies
Product Code
MHX · Cardiovascular
Decision Date
Mar 24, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Dash 2500 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of this device is a portable (intrahospital) multiparameter unit designed for monitoring adult, pediatric, and neonate patient vital signs in a hospital subacute care environments, such as same-day surgery, emergency rooms, recovery/PACU, progressive care, interventional radiology, special care units, and GI/endoscopy. The Dash 2500 Patient Monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), heart/pulse rate, respiration rate, ECG, temperature with an reusable electronic thermometer (predictive mode for oral and rectal temperature measurement, monitor mode for axillary temperature measurement), and functional oxygen saturation (SpO2) and pulse rate via spot checking and continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion. The Dash 2500 also detects alarm limit conditions and is capable of recording up to two waveforms. Using this monitor a clinician can view, record and recall clinical data derived from each parameter.
Device Story
Portable, multi-parameter patient monitor for intra-hospital use; monitors adult, pediatric, and neonate vital signs. Inputs: ECG, NIBP, SpO2, respiration, and temperature sensors. Device processes physiological signals to display real-time data, detect alarm limit conditions, and record up to two waveforms. Operated by clinicians in subacute hospital settings; can be shelf, table, or wall-mounted. Provides continuous monitoring and spot-checking capabilities; supports data recall and interface with central stations or servers. Benefits include real-time patient status awareness and timely alarm notification for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, module verification, system integration testing, performance testing, safety testing, and environmental testing.
Technological Characteristics
Portable multi-parameter monitor; powered by AC or battery. Parameters: oscillometric NIBP, ECG, SpO2, respiration, and temperature (predictive/monitor modes). Connectivity: standalone or interface to central station/server via host communications protocol. Mounting options: wall, pole, bed rail, or head/foot board. Complies with voluntary standards for safety and performance.
Indications for Use
Indicated for adult, pediatric, and neonate patients in hospital subacute care environments (e.g., same-day surgery, ER, PACU, progressive care, interventional radiology, special care, GI/endoscopy) for monitoring vital signs including NIBP, heart/pulse rate, respiration rate, ECG, temperature, and SpO2.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K080157
## Section 5: 510(k) Summary or 510(k) Statement
| | MAR 24 2008<br>510(k) Summary of Safety and Effectiveness |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 21, 2008 |
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, Wisconsin 53223 |
| Contact Person: | Sharon Untz<br>Quality and Regulatory Affairs Manager<br>GE Medical Systems Information Technologies<br>Telephone: 414-362-2436<br>Fax: 414-362-2585<br>E-mail: Sharon.Untz@med.ge.com |
| Device: | |
| Trade Name: | Dash 2500 Patient Monitor |
| Common/Usual Name: | Physiological Patient Monitor (Multi-parameter Module) |
| Classification Names: | 21 CFR 870.1025 Physiologic Patient Monitor (with arrhythmia<br>detection or alarms) |
| Predicate Device: | K031376 DINAMAP PRO 1000 V3 MONITOR |
| Device Description: | The Dash 2500 Patient Monitor is a portable (intra-hospital) multi-<br>parameter monitor designed for monitoring adult, pediatric, and neonate<br>patient vital signs.<br>The Dash 2500 Patient Monitor is self-contained and can be powered<br>by batteries or AC. The Monitor has a carrying handle and can be<br>operated on a shelf or table. It can also be mounted in a variety of ways<br>(e.g., wall, pole, bed rail, or head/foot board) using a mounting plate<br>located on the bottom of the Monitor. The Monitor can be used as a<br>stand-alone monitor with the capability to interface to a central station,<br>a server or any other device capable of receiving data using the host<br>communications protocol. |
| Intended Use: | The Dash 2500 Patient Monitor is intended for use under the direct<br>supervision of a licensed healthcare practitioner. The intended use of<br>this device is a portable (intra-hospital) multiparameter unit designed<br>for monitoring adult, pediatric, and neonate patient vital signs in a<br>hospital subacute care environments, such as same-day surgery,<br>emergency rooms, recovery/PACU, progressive care, interventional<br>radiology, special care units, and GI/endoscopy.<br>The Dash 2500 Patient Monitor monitors and displays oscillometric<br>non-invasive blood pressure (systolic, diastolic and mean arterial<br>pressure), heart/pulse rate, respiration rate, ECG, oxygen saturation<br>(SpO2) by non-invasive pulse oximetry, and temperature with an<br>reusable electronic thermometer (predictive mode for oral and rectal<br>temperature measurement, monitor mode for axillary temperature<br>measurement). |
Page 1 or 2
: 上一篇:
15
{1}------------------------------------------------
(08057
The Dash 2500 also detects alarm limit conditions and is capable of recording up to two waveforms. Using this monitor a clinician can view, record and recall clinical data derived from each parameter.
Technology: The Dash 2500 employs the same functional technology as the predicate devices.
Test Summary: The Dash 2500 complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the system:
- � Risk Analysis
- Requirements Reviews �
- Design Reviews �
- � Testing on unit level (Module verification)
- � Integration testing (System verification)
- � Final acceptance testing (Validation)
- Performance testing �
- � Safety testing
- � Environmental testing
Conclusion: The results of these measurements demonstrated that the Dash 2500 is as safe, as effective, and performs as well as the predicate device.
Page 2 of 2
16
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three horizontal lines representing the branches of government. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems Information Technologies c/o Ms. Sharon Untz Quality Assurance and Regulatory Affairs Manager 8200 West Tower Avenue Milwaukee, WI 53223
MAR 2 4 2008
Re: K080157
Dash 2500 Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: January 21, 2008 Received: January 23, 2008
Dear Ms. Untz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou mayy ators provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Sharon Untz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 1080157
## Section 4: Indications for Use Statement
### Indications for Use
510(k) Number (if known): Device Name: Dash 2500 Patient Monitor
Indications for Use:
The Dash 2500 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of this device is a portable (intrahospital) multiparameter unit designed for monitoring adult, pediatric, and neonate patient vital signs in a hospital subacute care environments, such as same-day surgery, emergency rooms, recovery/PACU, progressive care, interventional radiology, special care units, and GI/endoscopy.
The Dash 2500 Patient Monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), heart/pulse rate, respiration rate, ECG, temperature with an reusable electronic thermometer (predictive mode for oral and rectal temperature measurement, monitor mode for axillary temperature measurement), and functional oxygen saturation (SpO2) and pulse rate via spot checking and continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion.
The Dash 2500 also detects alarm limit conditions and is capable of recording up to two waveforms. Using this monitor a clinician can view, record and recall clinical data derived from each parameter.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Div | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number: | K080057 |
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.